The Sensititre® MIC and Breakpoint Susceptibility system is an in vitro diagnostic product for clinical susceptibility testing of non-fastidious Gram negative isolates, comprising of Enterobacteriaceae, Pseudomonas aeruginosa, and other non-Enterobacteriaceae and of non-fastidious Gram positive isolates, comprising of Staphylococcus sp., Enterococcus sp., and Beta haemolytic Streptococci other than S. pneumoniae. The Sensititre® ESBL confirmatory test plate is an in vitro diagnostic product for detection of ESBLs in clinical isolates of Klebsiella pneumoniae, Klebsiella oxytoca and Escherichia coli. MIC and ESBL plates can either be read manually or automatically on the Sensititre Autoreader® / OptiRead® and/or ARIS®. TREK Diagnostic Systems products have only been validated with TREK manufactured broths.
Device Story
Sensititre susceptibility plates are 96-well microtiter plates pre-dosed with antimicrobial agents (Oritavancin 0.0005-8 µg/mL) and fluorogenic substrates. Clinical laboratory workflow: bacterial isolates are inoculated into plates using the Sensititre AIM Autoinoculator; plates are incubated for 18-24 hours. Principle of operation: bacterial surface enzymes cleave fluorogenic substrates, releasing a fluorophore; fluorescence intensity correlates with bacterial growth. Results are read automatically via OptiRead/ARIS or manually via Vizion. Output: Minimum Inhibitory Concentration (MIC) values. Healthcare providers use MIC results to determine antimicrobial susceptibility, guiding antibiotic therapy decisions for patients with Gram-positive infections. The device provides quantitative susceptibility data to assist in clinical management of non-fastidious Gram-positive organisms.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing of the antimicrobial susceptibility system.
Technological Characteristics
In vitro diagnostic susceptibility test system; utilizes MIC panels with Oritavancin (0.0005-8 ug/mL). Designed for non-fastidious Gram-positive organisms. Platform-based testing.
Indications for Use
Indicated for clinical susceptibility testing of non-fastidious Gram-positive isolates, specifically Staphylococcus aureus (including MRSA and MSSA) and Enterococcus faecalis (vancomycin-susceptible only).
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Related Devices
K143526 — The Sensititre HP MIC Susceptibility plate with Oritavancin (0.0005-8mcg/ml) · Thermo Fisher Scientific · Mar 5, 2015
K150236 — Sensititre 18 - 24 hour MIC or Breakpoint Susceptibility System with Dalbavancin (0.0005 - 2ug/mL) · Thermo Fisher Scientific · Jul 27, 2015
K183324 — Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Omadacycline in the dilution range of 0.008 - 32 ug/mL · Thermo Fisher Scientific · Feb 15, 2019
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is an emblem featuring a stylized image of three human profiles facing right, with a design that resembles a caduceus or flowing fabric beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
THERMO FISHER SCIENTIFIC CYNTHIA KNAPP DIRECTOR OF CLINICAL OPERATIONS 1 THERMO FISHER WAY OAKWOOD VILLAGE OH 44146
April 1, 2015
Re: K143288
Trade/Device Name: The Sensititre 18-24 hour MIC or Breakpoint Susceptibility System with Oritavancin (0.0005-8ug/mL) Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder
Regulatory Class: II Product Code: JWY, LRG, LTT, LTW Dated: March 16, 2015 Received: March 17, 2015
Dear Ms. Knapp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Uwe Scherf - S for
Sally Hojvat, M.Sc., PhD Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) k143288
Device Name
The Sensititre 18 - 24 hour MIC or Breakpoint System with Oritavancin in the dilution range of 0,000-8 us/ml
## Indications for Use (Describe)
The Sensititre 18 - 24 hour MIC or Breakpoint Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of non fastidious isolates.
This 510(k) is for the newly approved Oritavancin in the dilution range of 0.0005-8 ug/ml for testing non-fastidious gram positive organisms on the Sensititre 18 - 24 hour MIC panel.
The approved primary "Indications for Use" and clinical significance for non-fastidious Gram positive isolates:
Staphylococcus aureus (including methicillin-resistant (MRSA) and methicillin susceptible (MSSA) isolates)
Enterococcus faecalis(vancomycin-susceptible isolates only)
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.