← Product Code [HSB](/productcode/HSB) · K143276

# Acumed Small Bone IM Nail System (K143276)

_Acumed, LLC · HSB · Feb 5, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K143276

## Device Facts

- **Applicant:** Acumed, LLC
- **Product Code:** [HSB](/productcode/HSB.md)
- **Decision Date:** Feb 5, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Acumed Small Bone IM Nail System is intended for fixation of fractures and osteotomies of the fibula, radius, and ulna, including fractures where the medullary canal is narrow or flexibility of the implant is paramount.

## Device Story

Acumed Small Bone IM Nail System consists of titanium alloy (Ti-6Al-4V) intramedullary rods/nails; used for fixation of fractures and osteotomies of fibula, radius, and ulna. System includes nails (110mm-270mm lengths; 2.6mm-4.0mm diameters), cortical screws for locking, optional far-end locking (FEL) bushings, and end caps. Device provides mechanical stabilization of bone fractures; implanted by orthopedic surgeons in clinical/surgical settings. Output is physical fixation of bone segments; aids healing by maintaining alignment. Benefits include flexibility for narrow medullary canals.

## Clinical Evidence

Bench testing only. Comparative mechanical testing was performed per ASTM F1264-03, including static four-point bend, static torsion, and bending fatigue tests, demonstrating performance equivalent to the predicate devices.

## Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V). Dimensions: 110mm-270mm length, 2.6mm-4.0mm diameter. Principle: Intramedullary fixation rod with locking screws, FEL bushings, and end caps. Testing standard: ASTM F1264-03. Sterilization: Provided both sterile and non-sterile.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- Synthes Elastic Intramedullary Nail (EIN) ([K971783](/device/K971783.md))
- Synthes EIN End Cap ([K082148](/device/K082148.md))
- Acumed Small Bone Locking Rod System II ([K031438](/device/K031438.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, layered on top of each other to create a sense of depth and connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 5, 2015

Acumed, LLC Mr. Nathan Wolf Regulatory Specialist 5885 North West Cornelius Pass Road Hillsboro, Oregon 97124

Re: K143276

Trade/Device Name: Acumed Small Bone IM Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: November 12, 2014 Received: November 14, 2014

Dear Mr. Wolf:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Mr. Nathan Wolf

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Lori A. Wiggins -S

- for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K143276

#### Device Name

Acumed Small Bone IM Nail System

Indications for Use (Describe)

The Acumed Small Bone IM Nail System is intended for fixation of fractures of the fibula, and ulna, including fractures where the medullary canal is narrow or flexibility of the implant is paramount.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

## Contact Details

| Applicant Name:    | Acumed LLC<br>5885 NW Cornelius Pass Road, Hillsboro, OR 97124-9432                                    |
|--------------------|--------------------------------------------------------------------------------------------------------|
|                    | Nathan Wolf, Regulatory Specialist<br>503-726-6622 (Cell)<br>503-207-1502 (Desk)<br>503-520-9618 (Fax) |
| Date Prepared:     | 12 November 2014                                                                                       |
| <b>Device Name</b> |                                                                                                        |
| Trade Name:        | Acumed Small Bone IM Nail System                                                                       |
| Common Name:       | Intramedullary Fixation Rod/Pin                                                                        |
| Classification:    | 21 CFR 888.3020, Intramedullary Fixation Rod                                                           |
| Class:             | Class II                                                                                               |
| Product Code:      | HSB                                                                                                    |

### Legally Marketed Predicate Device(s)

The Synthes Elastic Intramedullary Nail (EIN), cleared in 1997 (K971783), the Synthes EIN End Cap, cleared in 2008 (K082148), and the Acumed Small Bone Locking Rod System II, cleared in 2003 (K031438) serve as the predicate devices.

### Device Description

The Acumed Small Bone IM Nail is a titanium alloy (Ti-6Al-4V) intramedullary rod/nail manufactured in multiple lengths (110mm to 270mm) and diameters (2.6mm to 4.0mm). The nails have openings used in conjunction with titanium alloy cortical screws, which lock them in place. The nails are compatible with an optional far-end locking (FEL) bushing and set screw that provide a locking option at the distal end of the nail. The nails are also compatible with optional end caps that thread into the proximal portion of the nail to provide additional length if desired. All implants are provided both sterile and non-sterile.

{4}------------------------------------------------

### Intended Use/Indications for Use

The Acumed Small Bone IM Nail System is intended for fixation of fractures and osteotomies of the fibula, radius, and ulna, including fractures where the medullary canal is narrow or flexibility of the implant is paramount.

#### Substantial Equivalence Comparison

In consideration of the comparisons given herein, the Acumed Small Bone IM Nail System has been determined to be substantially equivalent to its predicate devices, the Synthes EIN (K971783), Synthes EIN End Cap (K082148), and Acumed Small Bone Locking Rod System II (K031438). Substantial equivalence was determined due to similarities in materials, technology, function, and dimensions.

#### Non-clinical Testing

Comparative testing between the Acumed Small Bone IM Nailing system and a predicate device was conducted as per ASTM F1264-03. The test data showed the Acumed Small Bone IM Nail was substantially equivalent to the predicate device in a static four-point bend test, static torsion test, and bending fatigue test as described herein.

---

**Source:** [https://fda.innolitics.com/device/K143276](https://fda.innolitics.com/device/K143276)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
