K143235 · ArthroCare Corporation · GEI · Oct 23, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143235
Device Name
RF20000a Controller, FLOW 50 Wand
Applicant
ArthroCare Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RF20000a Coblation System, comprised of the FLOW 50 Wand and the RF20000a Controller, is indicated for the resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels in the following arthroscopic and orthopedic procedures: All Joints (Hip, Knee, Shoulder, Wrist, Ankle, Elbow) Ablation/Debridement: Articular Cartilage, Bursectomy, Chondroplasty, Fascia, Ligament, Scar Tissue, Soft Tissue, Synovectomy, Tendon; Excision/Resection: Articular Labrum, Capsule, Cysts, Ligament, Loose Bodies, Plica Removal, Scar Tissue, Soft Tissue, Synovial Membrane, Tendon. Hip Excision/Resection: Acetabular Labrum. Knee Ablation/Debridement: ACL/PCL, Notchplasty; Excision/Resection: Capsular Release, Cartilage Flaps, Discoid Meniscus, Lateral Release, Meniscal Cystectomy, Meniscectomy, Villusectomy. Shoulder Ablation/Debridement: Acromioplasty, Subacromial Decompression; Excision/Resection: Frozen Shoulder Release, Glenoid Labrum. Wrist Excision/Resection: Triangular Fibrocartilage (TFCC).
Device Story
Electrosurgical system for arthroscopic/orthopedic surgery; comprises bipolar RF controller, disposable FLOW 50 wand, and foot control. Wand features integrated suction tubing and electrodes; controller includes integrated peristaltic fluid outflow regulator. System operates in conductive media; delivers RF energy to wand electrodes for tissue ablation/resection and hemostasis. Modes include Hi/Lo (Ablation), Vac (Suction), and COAG (Hemostasis). Surgeon controls modes via wand finger switches or foot pedal. Controller manages RF delivery and fluid outflow rate; provides GUI for settings. Benefits include precise tissue removal with minimal thermal damage to surrounding healthy tissue and consistent hemostasis. Used in clinical/OR settings by orthopedic surgeons.
Clinical Evidence
Bench testing only. Evidence includes design verification, software validation, tissue effect testing (histology, thermal margins), and peak temperature testing. No clinical data presented.
Indicated for resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels in arthroscopic and orthopedic procedures across all major joints (hip, knee, shoulder, wrist, ankle, elbow) for patients requiring such surgical intervention.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
ArthroCare ArthroWands (Ambient Super MultiVac 50) (K083306)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines representing hair or clothing. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 23, 2015
ArthroCare Corporation Ashley J. Dawson, PhD Manager, Regulatory Affairs 7000 West William Cannon Drive Austin, Texas 78735
Re: K143235
Trade/Device Name: RF20000a Coblation System, RF20000a Controller and FLOW 50 Wand Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 21, 2015 Received: September 23, 2015
Dear Dr. Dawson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{2}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
p. 1 of 2
510(k) Number (if known) K143235
Device Name RF20000a Coblation System: RF20000a Controller and FLOW 50 Wand
Indications for Use (Describe) Please see attached.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
The RF20000a Coblation System, comprised of the FLOW 50 Wand and the RF20000a Controller, is indicated for the resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels in the following arthroscopic and orthopedic procedures:
| | Ablation/Debridement | Excision/Resection |
|----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| All Joints (Hip,<br>Knee, Shoulder,<br>Wrist, Ankle,<br>Elbow) | ■ Articular Cartilage<br>■ Bursectomy<br>■ Chondroplasty<br>■ Fascia<br>■ Ligament<br>■ Scar Tissue<br>■ Soft Tissue<br>■ Synovectomy<br>■ Tendon | ■ Articular Labrum<br>■ Capsule<br>■ Cysts<br>■ Ligament<br>■ Loose Bodies<br>■ Plica Removal<br>■ Scar Tissue<br>■ Soft Tissue<br>■ Synovial Membrane<br>■ Tendon |
| Hip | | ■ Acetabular Labrum |
| Knee | ■ ACL/PCL<br>■ Notchplasty | ■ Capsular Release<br>■ Cartilage Flaps<br>■ Discoid Meniscus<br>■ Lateral Release<br>■ Meniscal Cystectomy<br>■ Meniscectomy<br>■ Villusectomy |
| Shoulder | ■ Acromioplasty<br>■ Subacromial Decompression | ■ Frozen Shoulder Release<br>■ Glenoid Labrum |
| Wrist | | ■ Triangular Fibrocartilage<br>(TFCC) |
{5}------------------------------------------------
# 510(k) Summary
# ArthroCare® Corporation
# RF20000a Coblation System: RF20000a Controller with FLOW 50 Wand
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### General Information
| Submitter Name: | ArthroCare Corporation |
|-----------------|--------------------------------------------------------------------------------------------------|
| Address | 7000 West William Cannon Drive<br>Austin, TX 78735 |
| Contact Person: | Ashley J. Dawson, PhD<br>Manager, Regulatory Affairs<br>Phone: 512-358-5773<br>Fax: 512-895-1489 |
| Date Prepared: | January 28, 2015 |
#### Device Name
| Proprietary Name: | RF20000a Coblation System:<br>RF20000a Controller<br>FLOW 50 Wand | |
|--------------------|-------------------------------------------------------------------|--|
| Common Name: | Electrosurgical cutting and Coagulation Device and Accessories | |
| Regulation Name: | Electrosurgical cutting and Coagulation Device and Accessories | |
| Regulatory Class: | II | |
| Product Code: | GEI | |
| Regulation Number: | 21 CFR 878.4400 | |
#### Predicate Devices
ArthroCare® System 12000 (Quantum 2) ArthroCare ArthroWands® (Ambient Super MultiVac 50) K082666 K083306
#### Description
#### RF20000a Coblation System
The RF20000a Coblation System is an electrosurgical system consisting of a bipolar radiofrequency Controller with Integrated Fluid Outflow Regulator; a sterile, disposable, single-use FLOW 50 Wand; and a non-sterile, reusable Foot Control.
{6}------------------------------------------------
This System utilizes bipolar technology specifically designed for the resection, ablation and coagulation of soft tissues and hemostasis of blood vessels in various arthroscopic and orthopedic procedures.
The System offers five distinct Modes of operation: Hi (Ablation), Lo (Ablation), Vac (Vacuum), and COAG (Hemostasis). Each Ablation (Coblation) mode allows for precise ablation with minimal damage to surrounding healthy tissue. The COAG mode allows for consistent and precise hemostasis of blood vessels.
#### FLOW 50 Wand
The FLOW 50 Wand consists of a handle, shaft, integrated cable, and suction tubing. The integrated cable and suction tubing are attached at the proximal end of the handle and connect to the RF20000a Controller and the Fluid Outflow Regulator, respectively. The handle has finger switches that enable Ablation Mode switching (Lo. Med, Hi) as well as activation of the Wand (Vac, Coag, or Ablate). The Foot Control provides an alternate means of controlling these same functions. The Wand is provided sterile and is single-use only.
#### RF20000a Controller
The RF20000a Controller is designed to deliver radiofrequency energy to the electrodes of the FLOW 50 Wand. The Controller is an enclosed unit with incorporated software that runs both the delivery of radiofrequency energy as well as a Graphical User Interface with which the user can control various modes, levels, volume, etc. Ports for connecting the FLOW 50 Wand and the Foot Control Pedal are located on the front panel.
The Controller incorporates a peristaltic integrated Fluid Outflow Regulator, which provides dynamic control of the rate of removal of conductive irrigating solution and/or fine, less dense, free-floating debris.
{7}------------------------------------------------
#### Intended Use/Indications for Use
There are no added procedures to the Indications for Use for the subject devices as compared to the predicate Ambient Super MultiVac Wand (K083306) and the Quantum 2 Controller (K082666). However, previously cleared procedures that utilized the Coagulation setting in order to thermally shrink tissue (e.g., Coagulation of Medial Retinaculum) were removed since the FLOW 50 Wand is not designed to be used in that manner. Removal of previously cleared procedures has no effect on the safety and effectiveness of the device when used as labeled.
The RF2000a Coblation System, comprised of the FLOW 50 Wand and the RF20000a Controller, is indicated for the resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels in the following arthroscopic and orthopedic procedures:
| | Ablation/Debridement | Excision/Resection |
|----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| All Joints<br>(Hip, Knee,<br>Shoulder,<br>Wrist,<br>Ankle,<br>Elbow) | ■ Articular Cartilage<br>■ Bursectomy<br>■ Chondroplasty<br>■ Fascia<br>■ Ligament<br>■ Scar Tissue<br>■ Soft Tissue<br>■ Synovectomy<br>■ Tendon | ■ Articular Labrum<br>■ Capsule<br>■ Cysts<br>■ Ligament<br>■ Loose Bodies<br>■ Plica Removal<br>■ Scar Tissue<br>■ Soft Tissue<br>■ Synovial Membrane<br>■ Tendon |
| Hip | | ■ Acetabular Labrum |
| Knee | ■ ACL/PCL<br>■ Notchplasty | ■ Capsular Release<br>■ Cartilage Flaps<br>■ Discoid Meniscus<br>■ Lateral Release<br>■ Meniscal<br>Cystectomy<br>■ Meniscectomy<br>■ Villusectomy |
| Shoulder | ■ Acromioplasty<br>■ Subacromial<br>Decompression | ■ Frozen Shoulder<br>Release<br>■ Glenoid Labrum |
| Wrist | | ■ Triangular<br>Fibrocartilage<br>(TFCC) |
{8}------------------------------------------------
## Summary of Technological Characteristics
The subject devices have the same technological characteristics (i.e., design, material, chemical composition, and energy source) as the predicate devices with the following exceptions (in bold font):
| | PREDICATE:<br>Controller: Quantum 2 (K082666)<br>Wand: Ambient Super MultiVac<br>(K083306) | SUBJECT:<br>Controller: RF20000a<br>Wand: FLOW 50 |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Intended Use | Resection, Ablation, and Coagulation of<br>Soft Tissue and Hemostasis of Blood<br>Vessels in Arthroscopic and Orthopedic<br>Procedures | Same |
| Electrical Safety/EMC | IEC 60601-1 Compliant<br>IEC 60601-2-2 Compliant | Same |
| Controller Specifications/Features | | |
| Input Power | 100-240V<br>50/60Hz | Same |
| Fuse Rating | 8 A | 15 A |
| Output Frequency<br>(Fundamental) | 100kHz | Same |
| Default Ablation<br>Set Point / Output<br>Voltage (Vrms) | Set Point 7/ 260 | Med-/ 279 |
| Ablation Set Point<br>Range / Output<br>Voltage (Vrms) | Set Points 1 to 9/ 100-314 | Lo⁻ to Hi⁺ / 257-340 |
| Coagulation Set Point<br>Range / Output<br>Voltage (Vrms) | Set Points 1 to 2/65-100 | Coag - Coag Plus/65-85 |
| Outflow Control<br>Mechanism | Hospital Suction with Roller clamp to<br>adjust flow control. Recommended<br>suction:<br>200-400 mmHg | Controller has an integrated low<br>pressure rotational peristaltic pump<br>Range: 0-600 rpm |
| Software Program | Software for<br>Quantum 2, V 2.03 | Graphic User Interface V 0.2<br>RF controller software for RF20000a, V<br>1.0 |
| Weight | <5 kg | 10 kg |
| Controller Input<br>Power | 442 W (217 ohms) | 460 W (290 ohms) |
| Controller Output<br>Power (Ablation, 350<br>ohm load) | 26 – 261 W | 45 – 298 W |
| Controller Output<br>Power (Coagulation,<br>350 ohm load) | 12 - 27 W | 8 – 30 W |
| Controller Crest<br>Factor (350 ohm load) | 1.4 | Same |
| Controller Waveforms | Square | Same |
| | PREDICATE:<br>Controller: Quantum 2 (K082666)<br>Wand: Ambient Super MultiVac<br>(K083306) | SUBJECT:<br>Controller: RF20000a<br>Wand: FLOW 50 |
| Wand Materials | | |
| Electrode | Tungsten | Same |
| Shaft | 304 Stainless Steel | Same |
| Outer Shaft Insulation | Black PET Heat Shrink Tubing | Black Pebax |
| Spacer | Ceramic (Alumina) | Same |
| Adhesive | Epoxy (Loctite 3981) | Epoxy (Loctite 3984) |
| Handle Material | Lexan 104 | Iupilon S3001R |
| Wand Suction Line | PVC | Same |
| Wand Specifications/Features | | |
| Shaft Length | $5.34 \pm 0.08$ inches | $5.31 \pm 0.20$ inches |
| Distal Bend Angle | 50° | 40° |
| Handle Length | 6.03 inches | 6.13 inches |
| Number of Electrodes | 1 active & 1 return | Same |
| Number of Internal<br>Suction Ports | 2 | 1 |
| Suction | Yes | Same |
| Shaft Rigid<br>Construction | Yes | Same |
| Use Limiting Feature | Yes | Yes |
| Temperature Measure | 20 to 60 °C | 10 to 60 °C |
| Finger Switch<br>Activation | Yes | Same |
| Foot Switch<br>Activation | Yes | Same |
| Software in Wand | No | Yes |
| Packaged Sterile | Yes | Same |
| Single Use Disposable | Yes | Same |
| Operates in<br>Conductive Media<br>Environment | Yes | Same |
| Bipolar/<br>Monopolar | Bipolar | Same |
| Sterilization | Radiation | Same |
| Recommended Active<br>Ablation Time | 5 minutes<br>(cumulative ablation) at set point 7 | Lo Mode: 10 minutes<br>Med Mode: 4 minutes<br>Hi Mode: 2 minutes |
| Rated Wand Voltage | 320 Vrms | 340 Vrms |
| Wand Output Power<br>(Ablation, 350 ohm<br>load) | 26 - 255 W | 189 - 298 W |
| | PREDICATE:<br>Controller: Quantum 2 (K082666)<br>Wand: Ambient Super MultiVac<br>(K083306) | SUBJECT:<br>Controller: RF20000a<br>Wand: FLOW 50 |
| Wand Output Power<br>(Coagulation, 350 ohm<br>load) | 12 – 27 W | 12 – 21 W |
{9}------------------------------------------------
{10}------------------------------------------------
## Substantial Equivalence
Non-clinical performance data such as design verification, software validation, tissue effect testing (histology, thermal margins), and peak temperature testing demonstrated that the subject devices are substantially equivalent to the predicate devices and are safe and effective when used as intended.
### Summary
The RF20000a Coblation System is substantially equivalent to the predicate devices. The differences between the ArthroCare RF20000a Coblation System and the predicate devices do not raise any new concerns about the safety or effectiveness of the subject devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.