K143216 · Xhale, Inc. · DQA · Mar 17, 2015 · Cardiovascular
Device Facts
Record ID
K143216
Device Name
Assurance Alar / Nasal SpO2 Sensor
Applicant
Xhale, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Mar 17, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Assurance™ Nasal / Alar SpO2 Sensor is indicated for single patient use for continuously noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients (weighing > 30kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision.
Device Story
Disposable pulse oximetry sensor; attaches to nasal ala using soft silicone rubber cushions. Inputs: red (660 nm) and infrared (880 nm) light transmitted through tissue; detected by silicon photodiode. Principle: spectrophotometric measurement of light absorption by oxygenated vs. deoxygenated blood. Output: SpO2 and pulse rate data transmitted via DB-9 connector to compatible pulse oximetry monitors. Used in clinical healthcare environments under professional supervision. Healthcare providers view output on monitor screen to assess patient oxygenation status and heart rate; aids in clinical decision-making regarding respiratory support or patient stability. Benefits: provides continuous, noninvasive monitoring for patients where traditional finger sensors may be unsuitable or poorly perfused.
Clinical Evidence
Controlled desaturation study conducted on healthy volunteers per ISO 80601-2-61. Performance evaluated across 8 different oximetry platforms. Primary endpoint: SpO2 accuracy (ARMS) in the 70-100% range under steady-state/non-motion conditions. Results showed ARMS values ranging from 1.5 to 2.2 across platforms, meeting the specification of ≤ 3%.
Technological Characteristics
Disposable nasal/alar sensor; silicone rubber cushions; red (660 nm) and IR (880 nm) LEDs; silicon photodiode detector. Connects via DB-9 connector. Complies with IEC 60601-1, IEC 60601-1-2 (EMC), and ISO 80601-2-61 (mechanical strength, fluid ingress, temperature). Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for single patient use for continuous noninvasive monitoring of functional arterial hemoglobin oxygen saturation (SpO2) and pulse rate in adult and pediatric patients weighing > 30kg, including well or poorly perfused patients, in healthcare environments under professional supervision.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 17, 2015
Xhale Inc. c/o Mr. Paul Dryden Regulatory Consultant 3630 SW 47th Avenue, Suite 100 Gainesville, FL 32608
Re: K143216
Trade/Device Name: Assurance™ Alar / Nasal SpO2 Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: February 12, 2015 Received: February 13, 2015
### Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K143216
Device Name
# Assurance™ Nasal / Alar SpO2 Sensor
#### Indications for Use (Describe)
The Assurance™ Nasal / Alar SpO2 Sensor is indicated for single patient use for continuously noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients (weighing > 30kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (9/13)
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{3}------------------------------------------------
### 510(k) Summary Page 1 of 6
| Date Prepared: | 16-Mar-15 |
|----------------------------|---------------------------------------------|
| | Xhale, Inc. |
| | 3630 SW 47th Ave, Suite 100 |
| | Gainesville, FL 32608 |
| | Phone: (352) 371-8488 |
| | Fax: (352) 375-3133 |
| Official Contact: | Jeffrey Hoebelheinrich |
| | Director of Quality and Regulatory Affairs |
| Proprietary or Trade Name: | Assurance™ Alar / Nasal SpO2 Sensor |
| Common/Usual Name: | Oximeter (Accessory – sensor) |
| Classification Name: | Oximeter |
| | Product Classification – DQA |
| | 21 CFR 870.2700 |
| | Class II |
| Predicate Devices: | K122996 – Xhale Assurance™ Aral SpO2 sensor |
| | K060576- Nellcor OxiMax sensors |
### Device Description:
The Alar / Nasal SpO2 Sensor is a disposable, single patient use Pulse Oximetry sensor designed to attach to the patient's nasal alar region - the fleshy region at the side of the nose. Skin contact and adhesive free sensor retention is via soft silicone rubber cushions encapsulating the optical components. The Alar / Nasal SpO2 Sensor with its associated patient cable, terminates in a DB-9 connector compatible with monitors employing Nellcor SpO2 technology.
The sensor utilizes red and IR LED light sources of 660 nm and 880 nm respectively along with a silicon photodiode detector to detect changes in oxygen saturation in the blood. Since oxygen saturated blood absorbs different amounts of light at each wavelength (red and infrared) as compared with unsaturated blood, the amount of light absorbed at each wavelength by the blood in each pulse can be used to calculate oxygen saturation.
### Indications for Use:
The Assurance™ Nasal / Alar SpO2 Sensor is indicated for single patient use for continuously noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO-) and pulse rate from the nasal ala of adult and pediatric patients (weighing > 30kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision.
### Comparison to Predicates
The following table compares the predicates.
{4}------------------------------------------------
# 510(k) Summary
| Attribute | Xhale<br>Alar SpO2 sensor<br>K122996 | Nellcor<br>OxiMax Sensors<br>K060576 | Proposed Xhale<br>Alar / Nasal SpO2 Sensor |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Assurance™ Alar SpO2 Sensor is<br>indicated for single patient use for<br>continuously noninvasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate from<br>the nasal ala of adult and pediatric<br>patients (weighing > 30kg), who are well<br>or poorly perfused. The sensor can be<br>used in a variety of healthcare<br>environments where compatible pulse<br>oximetry monitors are indicated for use,<br>under professional supervision. | Indicated for single patient use for<br>continuously noninvasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate.<br><br>Neonate, pediatric, and adult who are<br>well or poorly perfused, in hospitals,<br>hospital-type facilities, intra-hospital<br>transport, and home environments. | The Assurance™ Nasal / Alar SpO2 Sensor<br>is indicated for single patient use for<br>continuously noninvasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate from the<br>nasal ala of adult and pediatric patients<br>(weighing > 30kg), who are well<br>or poorly perfused. The sensor can be<br>used in a variety of healthcare<br>environments where compatible pulse<br>oximetry monitors are indicated for use,<br>under professional supervision. |
| Environments of use | variety of healthcare environments<br>where compatible pulse oximetry<br>monitors are indicated for use, under<br>professional supervision | hospitals, hospital-type facilities,<br>intra-hospital transport, and home<br>environments | variety of healthcare environments where<br>compatible pulse oximetry monitors are<br>indicated for use, under professional<br>supervision |
| Patient population | adult and pediatric patients (weighing ><br>30kg) | Neonate, pediatric, and adult | adult and pediatric patients (weighing ><br>30kg) |
| Single patient use | Yes | Yes | Yes |
| Technology, Features and Specifications | | | |
| Principle of Operation | Spectrophotometric measurement of<br>functional arterial oxygen saturation by<br>transmissive mode pulse oximetry | Spectrophotometric measurement of<br>functional arterial oxygen saturation by<br>transmissive mode pulse oximetry | Spectrophotometric measurement of<br>functional arterial oxygen saturation by<br>transmissive mode pulse oximetry |
| LEDS | Red (~660 nm) and IR (~880 nm) | Red (~660 nm) and IR (~880 nm) | Red (~660 nm) and IR (~880 nm) |
| Detector | Photodiode | Photodiode | Photodiode |
| Connector | 9 pin DB-9 style | 9 pin DB-9 style | 9 pin DB-9 style |
{5}------------------------------------------------
# 510(k) Summary
| Attribute | Xhale<br>Alar SpO2 sensor<br>K122996 | Nellcor<br>OxiMax Sensors<br>K060576 | Proposed Xhale<br>Alar / Nasal SpO2 Sensor |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Testing - Clinical and Non-clinical | | | |
| Patient Contact Materials<br>Classification | Materials in the gas pathway<br>External Communicating<br>Tissue<br>Prolonged duration<br><br>Materials in direct patient contact<br>Surface Contact<br>Mucosal<br>Prolonged duration | Materials in the gas pathway<br>External Communicating<br>Tissue<br>Prolonged duration<br><br>Materials in direct patient contact<br>Surface Contact<br>Mucosal<br>Prolonged duration | Materials in the gas pathway<br>External Communicating<br>Tissue<br>Prolonged duration<br><br>Materials in direct patient contact<br>Surface Contact<br>Mucosal<br>Prolonged duration |
| SpO2 Accuracy ( $A_{RMS}$ )* | 70-100% ± 2% | 70-100% ± 2% | 70-100% ± 3% |
| BPM | 30-250 bpm ± 3 bpm | 20-250 bpm ± 3 bpm | 30-240 bpm ± 3 bpm |
| IEC 60601-1 | Yes | - | Yes |
| IEC 60601-1-2 | EMC | - | EMC |
| ISO 80601-2-61 | Mechanical strength<br>Storage and Operating Temperature and<br>humidity<br>Fluid ingress | - | Mechanical strength<br>Storage and Operating Temperature<br>and humidity<br>Fluid ingress |
| Pulse rate accuracy | Yes | - | Yes |
| Inter-device reliability and<br>accuracy | Yes | - | Yes |
{6}------------------------------------------------
## 510(k) Summary Page 4 of 6 16-Mar-15
## Substantial Equivalence Discussion
The table above compares the key features of the proposed Assurance™ Alar / Nasal SpO2 Sensor as compared to the predicate Xhale Assurance™ Alar SpO2 Sensor (K122996) and the predicate Nellcor OxiMax Sensor (K060576).
## Discussion of Differences:
The differences between the proposed Alar / Nasal SpO- Sensor and the predicate Alar Sensor (K122996):
- · The Alar / Nasal SpO2 Sensor has 1 (one) wire connecting the sensor to the DB-9 connector.
- The predicate device had a separate AB-N Adaptor Cable that was necessary to o connect to a DB-9 port.
- The proposed Alar / Nasal SpO2 Sensor removed the need for the additional o adaptor by direct connection of the sensor assembly to the cable with a DB-9 connector.
- The proposed Alar / Nasal SpO2 Sensor has a PCB with embedded software that is ● directly connected to the DB-9 connector.
- This allows the sensor to communicate with the compatible pulse oximetry o equipment and display the SpO2 and heart rate data on the screen.
- Pulse Rate Accuracy - Testing was performed for the accuracy of pulse rate over the range from 30-240 bpm using a SpO2 simulator the results were identical pulse rate accuracy within + 3 bpm over the tested range. We tested 30-240 bpm when the predicate Alar (K122996) was tested at 30-250 bpm. This test method difference is not significant as long as it is disclosed in the labeling and is equivalent to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor.
Discussion - These differences can be viewed as substantially equivalent and do not raise any new safety or effectiveness concerns when compared to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor.
In summary one can conclude that substantial equivalence is met based upon the following:
Indications for Use – The indications for use are identical for the proposed device when compared to the predicate - K122996 - Xhale Assurance™ Alar SpO2 sensor. Discussion - Each device is indicated for use as a pulse oximeter sensor.
Technology and construction - The technology is identical to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor. The basic design, fabrication, constructions and materials are identical to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor. Discussion - The basic design, fabrication, constructions and materials are identical to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor.
Patient Population - The patient population is identical to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor.
Discussion - The patient population is identical - predicate and adults > 30kg.
{7}------------------------------------------------
## 510(k) Summary Page 5 of 6 16-Mar-15
Environment of Use - The environments of use are identical to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor. Discussion - Both devices are used in the identical settings.
### Non-Clinical Testing Summary -
Pulse Rate Accuracy - Testing was performed for the accuracy of pulse rate over the range from 30-240 bpm using a SpO2 simulator the results were identical pulse rate accuracy within + 3 bpm over the tested range.
Discussion - The Pulse Rate Accuracy is equivalent to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor.
Inter-device reliability and accuracy - Testing of the sensor across a range of test conditions and the testing showed the pulse rate to be within + 3 bpm over the range.
Discussion - The inter-device reliability and accuracy is identical to the predicates K122996 -Xhale Assurance™ Alar SpO2 sensor and K060576 - Nellcor OxiMax sensor.
Skin Temperature - Per ISO 80601-2-61 the maximum temperature limit is 41°C. Testing showed that corrected for 35°C the proposed sensor did not exceed this limit. Results ranged from 34.3 to 39.1℃.
Discussion - The results are substantially equivalent to the predicates s K122996 – Xhale Assurance™ Alar SpO2 sensor and K060576 - Nellcor OxiMax sensor.
Materials - The materials were tested per ISO 10993-1 for the submission K122996 and we have not changed the materials.
Discussion - The materials are identical to the predicate K122996 - Xhale Assurance™M Alar SpO2.
### Clinical Testing
We performed a controlled desaturation study with healthy volunteers per ISO 80601-2-61 and the results across the 8 oximetry platforms showed that the SpO2 was within ARMS specification of 3 under steady state / non-motion conditions for the range of 70-100%
Discussion - The clinical testing demonstrated that the performance is substantially equivalent to the predicates K122996 - Xhale Assurance™ Alar SpO2 sensor and K060576 - Nellcor OxiMax sensor. ISO 80601-2-61 allows for an ARMS range of up to + 3% across the range of 70-100%.
| Xhale Assurance A1x sensor with<br>model tested | Arms SpO2<br>70-100% |
|-------------------------------------------------|----------------------|
| Nellcor N-600x | 1.8 (257 pts) |
| Philips Intellivue A04 | 1.5 (257 pts) |
| Nellcor N-595 | 2.1 (258 pts) |
| Philips Intellivue A02 | 2.2 (258 pts) |
| Nellcor N-395 | 1.7 (266 pts) |
| Philips Intellivue A01 | 1.7 (264 pts) |
| DataScope Passport 2 | 1.7 (279 pts) |
| GE Dash 3000 | 2.2 (279 pts) |
{8}------------------------------------------------
# 510(k) Summary Page 6 of 6 16-Mar-15
# Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.