SonicFill 2 is indicated for direct placement in all cavity classes in anterior and posterior teeth. Additional indications include: base/liner material, repair of enamel defects, repair of provisionals, repair of porcelain restorations, minor occlusal build-ups, pit and fissure sealant, luting of composite/ceramic veneers, core build-ups and incisal abrasions.
Device Story
SonicFill 2 is a light-cured, nano-hybrid, resin-based dental restorative material. Used with SonicFill 2010 Handpiece (K091091) for sonic activation during delivery. Sonic energy reduces material viscosity, allowing close adaptation to cavity walls; deactivation restores higher viscosity for sculpting. Enables bulk filling of cavities up to 5mm deep in single increment. Used by licensed dental professionals in clinical settings. Benefits include reduced shrinkage stress, high depth of cure, and efficient placement.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including mechanical strength, polishability, water sorption/solubility, flexural strength, depth of cure, light sensitivity, radiopacity, volumetric shrinkage, shrinkage stress, color stability, and biocompatibility per ISO 10993 and ISO 4049 standards.
Technological Characteristics
Nano-hybrid, light-cured, resin-based dental restorative. Sonic-activated delivery via SonicFill 2010 Handpiece. Complies with ISO 4049 (dentistry polymer-based filling/restorative/luting materials) and ISO 10993 (biocompatibility).
Indications for Use
Indicated for direct placement in all cavity classes in anterior and posterior teeth; base/liner material; repair of enamel defects, provisionals, and porcelain restorations; minor occlusal build-ups; pit and fissure sealant; luting of composite/ceramic veneers; core build-ups; and incisal abrasions.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a way that they appear to be interconnected. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 5, 2015
Sybron Dental Specialties Ms. Kerri Casino Regulatory Affairs Manager 1717 West Collins Avenue Orange, California 92687
Re: K143209
Trade/Device Name: SonicFill 2 Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: II Product Code: EBF, EBC Dated: November 7, 2014 Received: November 10, 2014
Dear Ms. Casino:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K143209
Device Name SonicFill 2
### Indications for Use (Describe)
SonicFill 2 is indicated for direct placement in all cavity classes in anterior teeth. Additional indications include: base/liner material, repair of enamel defects, repair of povisionals, repair of porcelain restorations, minor occlusal build-ups, pit and fissure sealant, luting of composite/ceramic veneers, core build-ups and incisal abrasions.
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CER 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY for SonicFill 2
Image /page/3/Picture/1 description: The image shows the logo for Sybron Dental Specialties. The logo features the letters 'sds' in a bold, sans-serif font, with a curved line above the letters. Below the letters, the words 'Sybron Dental Specialties' are written in a smaller, sans-serif font. The color of the text and curved line is a dark blue.
### SonicFill 2
- 1. Submitter Information: Sybron Dental Specialties 1717 W. Collins Ave. Orange CA, 92687
| Contact Person: | Kerri Casino |
|-------------------|--------------|
| Telephone Number: | 714-516-7634 |
| Fax Number: | 714-516-7472 |
January 16, 2015 Date Prepared:
- 2. Device Name:
- Proprietary Name: SonicFill 2 . ●
- Tooth Shade Resin Material Classification Name:
- 872.3690 CFR Number: ●
- Device Class: ● II
- . Product Code: EBF, EBC
- 3. Predicate Device:
SonicFill 2 is substantially equivalent to the legally marketed device Metamorphosis (K091023) cleared on May 21, 2009, product code EBF, EBC.
- 4. Description of Device:
SonicFill 2 is a light-cured, low-shrink, resin-based, dental restorative designed for direct placement. SonicFill 2 is used in combination with the SonicFill 2010 Handpiece (K091091), which activates the restorative sonically for delivery. Upon delivery the restorative drops in viscosity, allowing the composite to closely adapt to the cavity walls. When the cavity is filled and the SonicFill 2010 Handpiece (K091091) is deactivated, the restorative resin returns to its original viscosity. The composite in the cavity is then shaped and sculpted to the contours of the tooth. SonicFill 2 is designed to have a high depth of cure and low shrinkage stress allowing a cavity up to 5mm in depth to be filled and cured in a single bulk increment.
- న. Statement of Intended Use:
SonicFill 2 is indicated for direct placement in all cavity classes in anterior and posterior teeth. Additional indications include: base/liner material, repair of enamel defects, repair of provisionals, repair of porcelain restorations, minor occlusal build-ups, pit and fissure sealant, luting of composite/ceramic veneers, core build-ups and incisal abrasions.
{4}------------------------------------------------
#### 6. Identification of Risk Analysis Method
Risk analysis was performed on SonicFill 2 utilizing a process based on ISO 14971:2007. The results of the risk analysis performed on SonicFill 2 concluded that all device design controls and process controls will be able to mitigate known potential failures and effects. In addition, performance testing and biocompatibility testing were performed to mitigate other potential risks.
#### 7. Description of Safety and Substantial Equivalence:
### Technological Characteristics
The technological characteristics of SonicFill 2 are very similar to those of the predicate, Metamorphosis (K091023). Both composites are sonically activated using the SonicFill SonicFill 2010 Handpiece (K091091) during delivery. Both SonicFill 2 and Metamorphosis (K091023) composites behave in the same manner in terms of viscosity dropping during activation while adapting to the cavity, and viscosity increasing after activation ends. Both SonicFill 2 and Metamorphosis (K091023) use the same SonicFill 2010 Handpiece (K091091) and the same unidose delivery system.
## Non-Clinical Performance Data
Non-clinical performance data included testing for mechanical strength, polishability, water sorption and solubility, flexural strength, depth of cure, light sensitivity, radiopacity, volumetric shrinkage, shrinkage stress, and color stability. Working Time testing, stability testing and biocompatibility testing were also performed. The data analyzed from the various tests substantiate that SonicFill 2 is as safe and effective as the predicate Metamorphosis (K091023). The following standards were utilized for the nonclinical performance testing of SonicFill 2:
- . Guidance for Industry and FDA Staff: Dental Composite Resin Devices -Premarket Notification [510(k)] Submissions, November 27, 1998
- ISO 10993-1: 2009 Biological evaluation of medical devices ●
- ISO 10993-3:2003 Biological Evaluation of Medical Devices- Part 3: Tests for . Genotoxicity, Carcinogenicity, and Reproductive Toxicity
- . ISO 10993-5:2009 Biological Evaluation of Medical Devices- Part 5: Tests for in Vitro Cytotoxicity
- . ISO 10993-6:2007 Biological Evaluation of Medical Devices- Part 6: Tests for Local Effects after Implantation
- . ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin Sensitization
- ISO 14971:2007 Risk Management
- ISO 4049:2009 Dentistry- Polymer based Filling, Restorative, and Luting . Materials.
{5}------------------------------------------------
| Element | Predicate Device-<br>Metamorphosis | SonicFill 2 |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | Metamorphosis (K091023) | To be assigned |
| Trade Name | SonicFill, Sonic-Activated Bulk Fill<br>Composite | SonicFill 2, Sonic-Activated Bulk<br>Fill Composite |
| Target Users | Licensed dental professionals | Licensed dental professionals |
| Device Description | Metamorphosis, a nano-hybrid<br>composite, is a light-cured, resin-<br>based, dental restorative designed for<br>direct placement. | SonicFill2, a nano-hybrid<br>composite, is a light cured, resin-<br>based dental restorative designed<br>for direct placement. |
| | Metamorphosis is packaged into tips,<br>which are custom-designed, single-<br>dose and made from plastic. The<br>single-dose tips are dispensed using<br>the SonicFill 2010 Handpiece<br>(K091091). | SonicFill 2 uses the same delivery<br>system as the predicate device<br>Metamorphosis, which are<br>custom-designed, single dose and<br>made from plastic. The single-<br>dose tips are dispensed using the<br>same SonicFill 2010 Handpiece<br>(K091091) as the predicate device. |
| Common Name | Dental Composite Restorative<br>Material | Dental Composite Restorative<br>Material |
| Classification Name | Tooth Shade Resin Material, per<br>CFR § 872.3690 and<br>Pit and Fissure Sealant and<br>Conditioner, per CFR § 872.3765 | Tooth Shade Resin Material, per<br>CFR § 872.3690 |
| Class | II | II |
| Product Code | EBF/ EBC | EBF/EBC |
| Storage | Ambient Temperature | Ambient Temperature |
| Curing Mechanism | Photo initiation | Photo initiation |
| Material<br>Compatibility | Biocompatibility meets requirements | Biocompatibility meets<br>requirements |
| Shelf Life | 24 months based on real time data | 24 months based on accelerated<br>data |
| UV Color Stability<br>per ISO 4049 | Pass | Pass |
| Volumetric Shrinkage | Pass | Equivalent to Predicate |
| Shrinkage Stress | Pass | Equivalent to Predicate |
| Depth of Cure | Pass | Pass |
| Light Sensitivity per<br>ISO 4049 | Pass | Pass |
| Flexural Strength per<br>ISO 4049 | Pass | Pass |
| Radiopacity per ISO<br>4049 | Pass | Pass |
| Translucency | Pass | Pass |
| Working Time | Pass | Pass |
| Water Sorption/<br>Solubility per ISO<br>4049 | Pass | Pass |
| Polishability | N/A | Pass |
# Table 5.1: Predicate and Proposed Device Comparison Table
{6}------------------------------------------------
## Clinical Performance Data
Clinical performance testing has not been performed for SonicFill 2.
## Conclusion as to Substantial Equivalence
SonicFill 2 has been tested for its strength, depth of cure, shrinkage, polishability, color It also has been verified as being substantially equivalent in stability, and safety. performance, safety, and effectiveness to the predicate Metamorphosis (K091023).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.