Denali Filter System - Femoral Delivery Kit, Denali Filter System - Jugular/Subclavain Delivery Kit
K143208 · C.R. Bard, Inc. · DTK · Dec 9, 2014 · Cardiovascular
Device Facts
Record ID
K143208
Device Name
Denali Filter System - Femoral Delivery Kit, Denali Filter System - Jugular/Subclavain Delivery Kit
Applicant
C.R. Bard, Inc.
Product Code
DTK · Cardiovascular
Decision Date
Dec 9, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3375
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Denali® Filter is indicated for use in the prevent pulmonary embolism via placement in the vena cava in the following situations: -Pulmonary thromboembolism when anticoagulants are contraindicated. -Failure of anticoagulant therapy for thromboembolic disease. -Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced. -Chronic. recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated. The Denali® Filter may be removed according to the instructions supplied under the Section labeled: Optional Procedure for Filter Removal.
Device Story
Denali Filter System is a venous interruption device for permanent or optional retrieval placement in the inferior vena cava (IVC) to prevent pulmonary embolism. Device consists of a laser-cut nickel-titanium filter with twelve shape-memory appendages forming two filtration levels; delivered via femoral or jugular/subclavian approaches using a specialized delivery system. System includes 10F and 8F dilators, introducer sheath, and pusher. Radiopaque marker bands on the dilator assist in measuring IVC diameter (up to 28mm). Physician operates the system under fluoroscopic guidance to deploy the filter. Filter captures emboli while allowing blood flow. Benefits include reduced risk of pulmonary embolism in patients where anticoagulation is ineffective or contraindicated.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was demonstrated via in vitro testing including dimensional and visual inspection, delivery system component tensile testing, simulated use testing, and packaging testing.
Technological Characteristics
Material: Nickel-titanium (Nitinol) shape-memory alloy. Design: 12-appendage filter with two-level filtration. Delivery: 10F and 8F dilators, introducer sheath with hemostasis valve. Connectivity: None. Sterilization: Not specified. Dimensions: Compatible with IVC diameter ≤ 28mm; 15cm and 55cm dilators.
Indications for Use
Indicated for patients requiring pulmonary embolism prevention via vena cava placement due to contraindication to anticoagulants, failure of anticoagulant therapy, or emergency treatment for massive pulmonary embolism.
Regulatory Classification
Identification
A cardiovascular intravascular filter is an implant that is placed in the inferior vena cava for the purpose of preventing pulmonary thromboemboli (blood clots generated in the lower limbs and broken loose into the blood stream) from flowing into the right side of the heart and the pulmonary circulation.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) “Use of International Standards Organization's ISO 10993 ‘Biological Evaluation of Medical Devices Part I: Evaluation and Testing,’ ” and
(2) FDA's:
(i) “510(k) Sterility Review Guidance and Revision of 2/12/90 (K90-1)” and
(ii) “Guidance for Cardiovascular Intravascular Filter 510(k) Submissions.”
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 9, 2014
C.R. Bard, Inc. % Laurie Sang Regulatory Affairs Specialist 1625 W. Third St. Tempe, Arizona 85281
Re: K143208
> Trade/Device Name: Denali Filter System – Femoral. Jugular/Subclavian Delivery Kit Regulation Number: 21 CFR 870.3375 Regulation Name: Cardiovascular Intravascular Filter Regulatory Class: Class II Product Code: DTK Dated: November 7, 2014 Received: November 10, 2014
Dear Laurie Sang,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K143208
Device Name
Denali® Filter System - Femoral and Jugular/Subclavian Delivery Kit
#### Indications for Use (Describe)
The Denali® Filter is indicated for use in the prevent pulmonary embolism via placement in the vena cava in the following situations:
-Pulmonary thromboembolism when anticoagulants are contraindicated.
-Failure of anticoagulant therapy for thromboembolic disease.
-Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced.
-Chronic. recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated.
The Denali® Filter may be removed according to the instructions supplied under the Section labeled: Optional Procedure for Filter Removal.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## DENALI® Filter System 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
## Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>Tempe, Arizona 85281 |
|------------|--------------------------------------------------------------------------------|
| Phone: | 480-350-6069 |
| Fax: | 480-449-2546 |
| Contact: | Laurie Sang, Regulatory Affairs Specialist |
| Date: | November 7, 2014 |
## Subject Device Name:
| Device Trade Name: | Denali® Filter System - Femoral and<br>Jugular/Subclavian Delivery Kit |
|-----------------------|------------------------------------------------------------------------|
| Common or Usual Name: | Filter, Intravascular, Cardiovascular |
| Classification: | Class II |
| Classification Panel: | Cardiovascular Devices |
| Product Code: | DTK |
| Predicate Devices: | (K130366; Clearance May<br>15, 2013) |
# Summary of Change:
The modification from the predicate device, compared to the subject Denalle Filter System - Femoral and Jugular/Subclavian Delivery Kit, is the device will include two introducers, a 10 French dilator and an 8 French dilator and pusher system. The new design will also incorporate a longer handle on the delivery system. Minor changes have been made to the device labeling.
Image /page/3/Picture/12 description: The image shows the word "BARD" in a stylized, bold font. The letters are thick and geometric, with a unique design for the letter "A" that resembles an upside-down "V" with a horizontal line connecting the two sides. The overall impression is a strong and modern logo or wordmark.
{5}------------------------------------------------
#### Device Description:
The Denali® Vena Cava Filter is a venous interruption device designed to prevent pulmonary embolism. The Denali® Filter can be delivered via the femoral and jugular/subclavian approaches. A separate delivery system is available for each approach. The Denali® Filter is designed to act as a permanent filter. When clinically indicated, the Denali® Filter may be percutaneously removed after implantation according to the instructions provided under the "Optional Procedure for Filter Removal" section.
The Denali® Filter of twelve shape-memory laser-cut nickel-titanium appendages. These twelve appendages form two levels of filtration with the legs providing the lower level of filtration and the arms providing the upper level of filtration. The Denal® Filter is intended to be used in the inferior vena cava (IVC) with a diameter less than or equal to 28mm.
The Denali®Vena Cava Filter System consists of a short, 15cm, 10 French dilator, an introducer sheath, a long, 55cm, 8 French dilator, and a preloaded Denall® Filter in a storage tube with a pusher. The 15cm, 10 French dilator accepts a 0.035" quidewire and is intended for pre-dilatation. The long, 55cm, 8 French dilator accepts a 0.035" quidewire and allows for an 800 psi maximum pressure contrast power injection. Radiopaque marker bands on the end of the dilator aid in measuring the maximum indicated IVC diameter. They are spaced at a distance of 28mm (outer-to-outer). The 55cm, 8.4 French I.D. introducer sheath contains a radiopaque marker and hemostasis valve with a side port. The pusher advances the filter through the introducer sheath to the predeployment mark and is then used to fix the filter in place while the filter is unsheathed.
#### Indications for Use of Device:
The Denali® Filter is indicated for use in the prevention of recurrent pulmonary embolism via placement in the vena cava in the following situations:
- . Pulmonary thromboembolism when anticoaqulants are contraindicated
- . Failure of anticoagulant therapy for thromboembolic disease
- Emergency treatment following massive pulmonary embolism where anticipated . benefits of conventional therapy are reduced
Image /page/4/Picture/13 description: The image shows the word "BARD" in a stylized, all-caps font. The letters are bold and outlined in black, giving them a three-dimensional appearance. The "A" in "BARD" is designed with a unique, angular shape, setting it apart from the other letters.
{6}------------------------------------------------
- . Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated.
The Denali® Filter may be removed according to the instructions supplied under the Section labeled: Optional Procedure for Filter Removal.
## Technological Comparison to Predicate Devices:
The technological characteristics of the subject device, the Denali® Filter System -Femoral and Jugular/Subclavian Delivery Kit, are substantially equivalent to those of the predicate devices, in terms of the following:
- Intended use
- Indications for use
- . Target population
- . Delivery system design
- Filter design and material .
- . Fundamental scientific technology
- . Packaging configuration (with addition of the extra dilator)
- . Sterility Assurance and method of Sterilization
# Performance Testing Summary:
To demonstrate substantial equivalence of the subject device to the predicate device, the technological characteristics and performance criteria were evaluated using in vitro testing performed as outlined below:
# In Vitro Delivery System Testing
- . Dimensional and Visual Inspection
- Delivery System Component Tensile Testing .
- . Simulated Use Testing
- . Packaging Testing
Image /page/5/Picture/23 description: The image shows the word "BARD" in a stylized, bold font. The letters are connected, creating a unified design. The overall impression is modern and clean.
{7}------------------------------------------------
The results from these tests demonstrate that the technological characteristics and performance of the Denali® Filter System - Femoral and Jugular/Subclavian Delivery Kit is comparable to the predicate device and that the subject device can perform in a manner substantially equivalent to the predicate device with the same intended use.
## Conclusion:
The subject Denali Filter System met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The Denali® Filter System - Femoral and Jugular/Subclavian Delivery Kit is substantially equivalent to the legally marketed predicate device.
Image /page/6/Picture/6 description: The image shows the word "BARD" in a stylized, bold, sans-serif font. The letters are outlined in black, creating a strong visual impact. The design of the letters is somewhat geometric, with the "A" resembling a triangle and the "D" having a rounded, almost semi-circular shape.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.