K143191 · Mcginley Innovations · HWE · Feb 20, 2015 · General, Plastic Surgery
Device Facts
Record ID
K143191
Device Name
IntelliSense Drill
Applicant
Mcginley Innovations
Product Code
HWE · General, Plastic Surgery
Decision Date
Feb 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4820
Device Class
Class 1
Indications for Use
The IntelliSense Drill is intended to bore a hole into bone for insertion of a screw, wire, cable, plate, pin, bolt, etc. In Bicortical mode, the IntelliSense Drill is indicated to be used in bicortical long bone, such as the femur, tibia, fibula, humerus, ulna, and radius.
Device Story
IntelliSense Drill is a surgical power tool for boring holes in bone. System includes drill unit with cable, control console, and proprietary disposable drill bits/bushings. Operates by monitoring load on drilling mechanism to determine drill bit location relative to bone; provides automated depth measurement for screw sizing. Used in surgical settings by clinicians. Console features 7-inch touch screen for parameter adjustment and status display. Benefits include combining drilling and depth measurement into a single procedure, potentially improving surgical efficiency and accuracy in screw selection.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and technological comparison.
Technological Characteristics
Components: Drill unit, control console, disposable drill bits/bushings. Materials: Drill (AISI 316 stainless steel), drill bits (AISI 17-4 stainless steel), bushing (polycarbonate), tray (5052-H32 aluminum, silicone, 302/304/430 stainless steel, Radel). Power: AC. Motor: Bidirectional, max 1,200 RPM, 1.9Nm peak torque. Connectivity: Internal USB port for parameter updates. Software: Microprocessor-based control system.
Indications for Use
Indicated for patients requiring bone drilling for fixation device insertion (screws, wires, cables, plates, pins, bolts). Bicortical mode indicated for long bones: femur, tibia, fibula, humerus, ulna, radius.
Regulatory Classification
Identification
Surgical instrument motors and accessories are AC-powered, battery-powered, or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. Accessories or attachments may include a bur, chisel (osteotome), dermabrasion brush, dermatome, drill bit, hammerhead, pin driver, and saw blade.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines connecting them. The profiles are arranged in a way that suggests unity and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 20, 2015
McGinley Orthopaedic Innovations % Mr. Justin Eggleton MCRA, LLC 1311 H Street North West, 12th Floor Washington, District of Columbia 20005
Re: K143191
Trade/Device Name: McGinley Innovations IntelliSense Drill Regulation Number: 21 CFR 878.4820 Regulation Name: Surgical instrument motors and accessories/attachments Regulatory Class: Class I Product Code: HWE, GEY, GFG, HTW Dated: January 30, 2015 Received: February 3, 2015
Dear Mr. Eggleton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K143191
Device Name McGinley Innovations IntelliSense Drill
#### Indications for Use (Describe)
The IntelliSense Drill is intended to bore a hole into bone for insertion of a screw, wire, cable, plate, pin, bolt, etc. In Bicortical mode, the IntelliSense Drill is indicated to be used in bicortical long bone, such are the fibri, fiber, fibel, f.but, f.but, humerus, ulna, and radius.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other spect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for McGinley Innovations. The logo features a gold emblem with the letter "M" in the center, surrounded by a decorative gold border. To the right of the emblem, the words "MCGINLEY INNOVATIONS" are written in gold, with the tagline "REDEFINING MEDICINE" underneath in a smaller font.
| Section 5: 510(k) Summary | |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DATE OF PREPARATION: | November 5, 2014 |
| COMPANY/OWNER: | Dr. Joseph McGinley<br>McGinley Orthopaedic Innovations<br>2435 King Blvd, Ste 230<br>Casper, WY 82604 |
| CONTACT: | Dr. Joseph McGinley<br>McGinley Orthopaedic Innovations<br>2435 King Blvd, Ste 230<br>Casper, WY 82604 |
| TELEPHONE | 215-847-1462 |
| FAX: | 435-304-1170 |
| EMAIL: | mcgjoe26@yahoo.com> |
| DEVICE TRADE NAME: | IntelliSense Drill |
| COMMON NAME: | Surgical Drill and accessories |
| CLASSIFICATION NAME: | Surgical instrument motors and accessories/attachments |
| REGULATION NUMBER: | 21 CFR §878.4820 |
| PRODUCT CODE: | HWE |
| SUBSEQUENT PRODUCT<br>CODE: | GEY, GFG, HTW |
| DEVICE CLASS: | Class I |
| PREDICATE DEVICES: | Stryker Total Performance (TPS) System, K943589 |
| DEVICE DESCRIPTION: | The IntelliSense Drill device will contain a drill unit (with<br>attached cable), Control Console, and Sterilization Tray.<br>Proprietary Drill Bits and Drill Bit Bushing will be sold<br>separately as accessories. The drill bits and companion bushings<br>are sold as a pair and are disposable (single-procedure). |
| | The IntelliSense Drill's design combines the drilling and depth<br>measurement processes into one procedure, which determines<br>drill bit location. By monitoring the load placed upon the drilling<br>mechanism, the precise location of the drill bit may be<br>determined relative to the bone. Secondary depth monitoring will<br>provide the measurement of the drill bit depth and appropriate<br>screw size. |
| INTENDED USE: | The IntelliSense Drill is intended to bore a hole into bone for |
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for McGinley Innovations. The logo features a gold emblem with the letter "M" in the center, surrounded by a decorative border. To the right of the emblem, the words "MCGINLEY INNOVATIONS" are written in a serif font, with the tagline "REDEFINING MEDICINE" underneath in a smaller font.
insertion of a screw, wire, cable, plate, pin, bolt, etc. In Bicortical mode, the IntelliSense Drill is indicated to be used in bicortical long bone, such as the femur, tibia, fibula, humerus, ulna, and radius.
### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The IntelliSense Drill demonstrates substantial equivalence to the Stryker Total Performance (TPS) System, K943589. Table 5-1 summarizes the key technological characteristics and features of both the predicates and the new device.
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for McGinley Innovations. The logo features a gold emblem with the letter "M" in the center, followed by the company name "MCGINLEY INNOVATIONS" in gold lettering. Below the company name is the tagline "REDEFINING MEDICINE" in a smaller font.
# Table 5-1: Comparison of Proposed Device with Predicate Devices
| | | Predicate |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | IntelliSense Drill | Stryker Total Performance<br>System |
| 510(k) Number | | K943589 |
| Intended Use | The IntelliSense Drill is<br>intended to bore a hole into<br>bone for insertion of a screw,<br>wire, cable, plate, pin, bolt,<br>etc. In Bicortical mode, the<br>IntelliSense Drill is indicated<br>to be used in bicortical long<br>bone, such as the femur, tibia,<br>fibula, humerus, ulna, and<br>radius. | The Stryker Total<br>Performance (TPS) System is<br>intended for use in the cutting,<br>drilling, reaming,<br>decorticating, and smoothing<br>of bone, teeth and other bone<br>related tissue in a variety of<br>surgical procedures, including<br>but not limited to Dental,<br>ENT, Neuro, and Endoscopic.<br>It is also usable in the<br>placement of cutting of<br>screws, wires, pins, and other<br>fixation devices. It can also be<br>used to cut metal |
| Outer profile: Console | Control Console has a 7 inch<br>diagonal graphical touch<br>screen for the user interface. | Bench-top style console with<br>touch screen display and<br>optional irrigation pump. |
| Operating Parameters: | User adjusts operating<br>parameters via touch screen | User adjusts operating<br>parameters via touch screen |
| System Status Display | Displayed on Touch Screen<br>interface | Displayed on Touch Screen<br>interface |
| Software | Microprocessor | Microprocessor |
| Non-volatile memory | System setup parameters such<br>as touch screen calibration<br>data, cortex breakthrough<br>detection parameters and drill<br>bit offsets. The setup<br>parameters can be viewed,<br>updated, and set to their<br>default values, using the<br>internal USB port on the SBC. | All current setting are stored<br>when the power is down. |
| Energy (input)/Power<br>Source | AC | AC |
| Console Controller<br>Operation: | commercial bidirectional<br>motor controller | Bidirectional |
| Weight | 1.5 Kg (3.3 lbs) | 1.36 Kg (3.0 lbs) [TPS<br>Universal Driver] |
| Dimensions: Drill | Drill to fit in envelope of size | Not stated |
| Feature | IntelliSense Drill | Predicate<br>Stryker Total Performance<br>System |
| Materials: Drill<br>Accessories | Stainless Steel AISI 316 | Stainless Steel |
| Materials:<br>Sterilization Tray | 5052-H32 Aluminum,<br>Silicone, 302 Stainless Steel,<br>304 Stainless Steel Radel, 430<br>Stainless Steel | N/A |
| Materials: Drill Bit | Stainless Steel: AISI 17-4 | N/A |
| Materials: Drill Arm<br>Bushing | Polycarbonate | N/A |
| Technological Characteristics | | |
| Design: | The IntelliSense Drill is<br>comprised of three primary<br>subsystems: the Drill Unit<br>(with attached cable), the<br>Control Console, and the<br>proprietary Drill Bits. | The device description of the<br>Stryker System includes<br>drills, shavers, shields,<br>guards, motors, attachments,<br>saws, wire drivers, collets,<br>console, irrigation pump,<br>cords, footswitch, handswitch,<br>clips, tubing, cutting<br>accessories, and sterilization<br>cases. |
| Bidirectional Motor | Forward / Reverse Trigger | Forward / Reverse Trigger |
| Drill RPM | speeds up to 1,200 rpm | 1,500 rpm [TPS Universal<br>Driver] |
| Torque | Peak torque 1.9Nm with 0.3<br>Nm continuous torque at<br>speeds up to 1,200 RPM | 3.3 Nm [Stryker System<br>2000] |
| Drill Bit Depth<br>Measurement | By monitoring the load placed<br>upon the drilling mechanism,<br>the precise location of the drill<br>bit may be determined relative<br>to the bone. Secondary depth<br>monitoring will provide the<br>measurement of the drill bit<br>depth and appropriate screw<br>size | Manual depth gauge to<br>determine proper screw size |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for McGinley Innovations. The logo features a gold emblem with the letter "M" in the center. To the right of the emblem, the words "MCGINLEY INNOVATIONS" are written in gold, with the tagline "REDEFINING MEDICINE" underneath in a smaller font. The overall design is elegant and professional.
## Conclusions:
McGinley Innovations provided sufficient information to demonstrate the IntelliSense Drill is substantially equivalent to predicate Stryker Total Performance System (K943589).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.