entré Model PD08-U

K143185 · Tactile Systems Technology, Inc. · JOW · May 7, 2015 · Cardiovascular

Device Facts

Record IDK143185
Device Nameentré Model PD08-U
ApplicantTactile Systems Technology, Inc.
Product CodeJOW · Cardiovascular
Decision DateMay 7, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Intended Use

The entré System is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions: - Chronic edema - Lymphedema - Venous insufficiency - Wound healing

Device Story

The entré Model PD08-U is a pneumatic compression system consisting of a controller and an 8-chambered garment. The garment wraps around an affected extremity; the controller inflates the chambers via an 8-tube harness assembly to provide sequential compression. The device is used in clinical or home settings by medical professionals or patients under supervision. The controller features an internal valve manifold and a front-panel 8-port connector for garment attachment. By applying controlled pneumatic pressure to the limb, the device assists in managing edema and promoting wound healing. Healthcare providers use the system to deliver prescribed compression therapy, which helps reduce fluid accumulation and improve circulation in patients with lymphatic or venous conditions.

Clinical Evidence

Bench testing only. Performance testing included garment chamber pressure testing and cycle time testing to verify equivalence to predicate devices. Additional non-clinical testing included electromagnetic safety (IEC 60601-1-2:2007) and electrical safety (IEC 60601-1:2005).

Technological Characteristics

Compressible limb sleeve system; 8-chamber garment; pneumatic controller with internal valve manifold; 8-port connector. Compliant with ANSI/AAMI ES60601-1, IEC 60601-1 Ed. 3, CAN/CSA C22.2 No. 60601-1, EN 60601-1-2, IEC 60601-1-6, and IEC 60601-1-11.

Indications for Use

Indicated for medical professionals and patients under medical supervision for the treatment of chronic edema, lymphedema, venous insufficiency, and wound healing.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 7, 2015 Tactile Systems Technology Inc Daniel Chase V.P. Engineering & Operations 1331 Tyler St NE Minneapolis, Minnesota 55413 Re: K143185 > Trade/Device Name: entré Model PD08-U Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: March 27, 2015 Received: March 30, 2015 Dear Daniel Chase. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, M.A. Hilleman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K143185 Device Name entré Model PD08-U Indications for Use (Describe) The entré System is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions: - · Chronic edema - · Lymphedema - · Venous insufficiency - · Wound healing | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Tactile Medical. The logo features a stylized letter "T" in blue and gray, with the top of the "T" forming a triangle. To the right of the "T" is the word "Tactile" in blue, with the word "MEDICAL" in gray underneath. The logo is clean and modern, with a focus on the company's name. # 510(k) Summary: K143185 | Submission Date: | April 30, 2015 | |--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | Daniel G. Chase<br>VP, Engineering and Operations<br>Tactile Medical<br>1331 Tyler Street NE, Suite 200<br>Minneapolis, MN 55413 USA<br>Telephone: (612) 355-5133 | | Contact Person: | Daniel G. Chase | | Name of Device: | entré Model PD08-U | | Classification: | Compressible Limb Sleeve (21 CFR 870.5800) | | Predicate Devices: | Petite Basic System 701 ELT (K131420)<br>CircuFlow 5150 (K123959)<br>CircuFlow 5208 (K123647) | ## Device Description The entré model PD08-U system consists of two main components: A controller and a garment. The garment is to be wrapped around the affected extremity providing a snug yet comfortable fit. The garment will have eight (8) chambers that are filled with air by the controller to provide compression on the extremity. A harness assembly consisting of eight (8) individual tubes will connect individual chambers with the controller. The controller has an internal valve manifold with a directly connected single eight (8) port connector extension that is accessible to the user on the front panel allowing easy garment connection. ### Intended Use The entré System is intended for use by medical professionals and patients who are under medical supervision, for the treatment of the following conditions: - . Chronic edema - Lymphedema ● - Venous insufficiency - Wound healing ● {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Tactile Medical. The logo consists of a blue stylized letter "T" that also resembles a house, with the word "Tactile" in blue next to it. Below "Tactile" is the word "MEDICAL" in gray, with a blue line underneath. ## Comparison of Technical Characteristics The entré model PD08-U system has the same technological characteristics to the predicate devices with respect to intended use, design, materials used, and construction. The device has been certified to be compliant with ANSI/AAMI ES60601-1, IEC 60601-1 Ed. 3, CAN/CSA C22.2 No. 60601-1 (Medical Electrical Equipment- General require for Safety), EN 60601-1-2 (Electromagnetic Compatibility), IEC 60601-1-6 (Usability), and IEC 60601-1-11 (In-Home Medical Equipment). ### Test Summary The following performance testing and analysis was completed to verify the substantial equivalence between the entré model PD08-U, the subject of this submission, and the predicate devices. #### Garment Chamber Pressure Testing . Both the entré model PD08-U and the predicate devices operate within similar pressure ranges. Pressure testing was performed to verify that the target pressures were met as stated in the device labeling. #### . Cycle Time Testing Predicated devices and the entré model PD08-U are pressure based. Comparative cycle time data was recorded and evaluated. Both devices performed similarly and as stated in the device labeling. #### . Additional Testing Additional non-clinical testing performed for the entré model PD08-U included: - Electromagnetic Safety (IEC 60601-1-2:2007) o - o Electrical Safety (IEC 60601-1:2005) ### Substantial Equivalence Conclusion The results from nonclinical device testing demonstrates that the entré model PD08-U system raises no new safety or effectiveness concerns and is substantially equivalent to the predicate devices for its intended use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...