K143147 · Medline Industries, Inc. · FRG · May 7, 2015 · General Hospital
Device Facts
Record ID
K143147
Device Name
Gemini Bonded Sterilization Wrap
Applicant
Medline Industries, Inc.
Product Code
FRG · General Hospital
Decision Date
May 7, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Gemini Bonded Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of such content for 180 days following sterilization. Gemini Bonded Sterilization Wrap is validated for use in steam or STERRAD® sterilization processes in the following sterilization modes and cycles. - Pre-vacuum steam cycles - * Minimum 4 Minutes Exposure at 270°F/132°C with minimum 20 minutes dry time - Gravity steam cycles - * Maximum 30 Minutes Exposure at 250°F/121°C with minimum 20 minutes dry time - STERRAD® Sterilization - * STERRAD® 100NX, Flex cycle - * STERRAD® 100NX, Standard cycle - * STERRAD® 100NX, Express cycle - * STERRAD® 100S. Standard cycle - * STERRAD® NX, Advanced cycle - * STERRAD® NX, Standard cycle
Device Story
Gemini Bonded Sterilization Wrap consists of two-ply polypropylene sheets (SSMMS process) ultrasonically bonded along two sides in a dotted pattern; used by healthcare providers to wrap medical devices prior to sterilization. Device acts as a microbial barrier; allows sterilization via pre-vacuum steam, gravity steam, or STERRAD cycles; maintains sterility of contents for 180 days. Output is a sterile-wrapped package; facilitates aseptic opening. Benefits include reduced wrapping/unwrapping time compared to traditional double-wrapping methods.
Clinical Evidence
No clinical data. Bench testing only: Primary skin irritation (ISO 10993-10) and microbial barrier properties (ISO 11607-1, ASTM F1980) met all requirements.
Technological Characteristics
Two-ply polypropylene sheets (SSMMS: spunbond-meltblown-meltblown-spunbond); ultrasonically bonded in three-dotted line pattern. Dimensions: 12"x12" to 54"x72". Sterilization compatibility: Pre-vacuum steam, gravity steam, STERRAD (100NX, 100S, NX).
Indications for Use
Indicated for healthcare providers to enclose medical devices for sterilization via steam or STERRAD processes and maintain sterility for 180 days post-sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads, each head facing a different direction. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 7, 2015
Medline Industries, Inc. Mr. Matt Clausen Senior Regulatory Affairs Specialist One Medline Place Mundelein, IL 60060
Re: K143147
Trade/Device Name: Gemini Bonded Sterilization Wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: April 15, 2015 Received: April 17, 2015
Dear Mr. Clausen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Clausen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143147
Device Name Gemini Bonded Sterilization Wrap
#### Indications for Use (Describe)
Gemini Bonded Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of such content for 180 days following sterilization. Gemini Bonded Sterilization Wrap is validated for use in steam or STERRAD® sterilization processes in the following sterilization modes and cycles.
- Pre-vacuum steam cycles
- * Minimum 4 Minutes Exposure at 270°F/132°C with minimum 20 minutes dry time
#### · Gravity steam cycles
- * Maximum 30 Minutes Exposure at 250°F/121°C with minimum 20 minutes dry time
#### · STERRAD® Sterilization
- * STERRAD® 100NX, Flex cycle
- * STERRAD® 100NX, Standard cycle
- * STERRAD® 100NX, Express cycle
- * STERRAD® 100S. Standard cycle
- * STERRAD® NX, Advanced cycle
- * STERRAD® NX, Standard cycle
The Gemini Bonded Sterilization Wrap intended loads with pre-vacuum steam sterilization, gravity steam sterilization, and the Advanced Sterilization Products (ASP) STERRAD® 100S, and STERRAD® 100S, and STERRAD® NX cycles are provided in Table 1.
The Gemini Bonded Sterilization Wrap Recommendations for Use with Pre-vacuum steam cycles. and the Advanced Sterilization Products STERRAD® Sterilization Systems are provide in Table 2.
(tables 1 and 2 are on a separate attachment. SEE ATTACHMENT)
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
V | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Number: K143147
Device Name: Gemini Bonded Sterilization Wrap
#### Table 1 – Validated Advanced Sterilization Products STERRAD® 100S, and STERRAD® NX Cycles
| Cycle | Maximum recommended<br>Chamber Load | Intended Load |
|-----------------------------------------------|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pre-Vacuum Steam<br>& Gravity Steam<br>Cycles | Lightweight 6 lbs. | 2 huck towels (17 in. x 29 in.), 3 fluid-resistant drapes (108 in. x 72 in.) |
| | Regular<br>Weight 9 lbs. | 16 huck towels (17 in. x 29 in.), 1 fluid resistant table cover (90 in. x 60 in.). 5 lbs. of metal mass. |
| | Medium<br>Weight 13 lbs. | 4 tray liners (20 in. x 25 in.) stacked 23 in. x 11 in. x 3.5 in. tray containing 11 lbs. of metal mass. |
| | Heavyweight 17 lbs. | 4 tray liners (20 in. x 25 in.) stacked 23 in. x 11 in. x 3.5 in. tray containing 15 lbs. of metal mass. |
| | Super<br>Heavyweight 25 lbs. | 4 tray liners (20 in. x 25 in.) stacked 23in. x 11in. x 3.5 in. tray containing 20 lbs. of metal mass. |
| STERRAD® 100NX,<br>Flex Cycle | 10.7 lbs. | One or two single-channel Flexible Endoscope with or without a silicone mat and no additional load. The flexible endoscope may contain:<br>• A single-channel- Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter. (A maximum of two flexible endoscope, one per tray per sterilization cycle.) |
| STERRAD® 100NX,<br>Standard Cycle | 10.7 lbs. | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:<br>• An inside diameter of 0.7 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens. (A maximum of two flexible endoscopes, one per tray per sterilization cycle.) |
| STERRAD® 100NX,<br>EXPRESS Cycle | 10.7 lbs. | Non-lumened reusable metal and non-metal devices requiring surface sterilization, and sterilization of diffusion-restricted spaces such as the hinged portions of forceps and scissors, and rigid or semi-rigid endoscopes without lumens. |
{5}------------------------------------------------
| STERRAD® 100S,<br>Standard Cycle | Light Weight | 6<br>lbs. | Reusable metal and non-metal medical devices, including<br>up to 10 lumens of the following dimensions per chamber |
|----------------------------------|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| | Regular Weight | | load: |
| | Medium Weight | | • An inside diameter of 1mm or larger and a length of |
| | Heavyweight | | 125 mm or shorter of single-channel stainless steel |
| | Super Heavyweight | | lumens. |
| | | 10.7<br>lbs. | • An inside diameter OF 2 mm or larger and length of |
| | | | 250 mm or shorter of single-channel stainless steel<br>lumens. |
| | | | • An inside diameter of 3 mm or larger and a length of |
| | | | 400 mm or shorter of single-channel stainless lumens. |
| | | | • An inside diameter of 6 mm or larger and a length of |
| | | | 310 mm or shorter of single-channel<br>Teflon/Polyethylene lumens. |
| | | | |
| STERRAD® NX,<br>Advanced Cycle | 10.7 lbs. | Reusable metal and non-metal medical devices, including up to 10<br>lumens of the following dimensions per chamber load:<br>• An inside diameter of 1 mm of larger and a length of 500 mm | |
| | | or shorter of single-channel stainless steel lumens | |
| | | Or | |
| | | One single-channel Flexible Endoscope with or without a silicone | |
| | | mat and no additional load. The flexible endoscope may contain: | |
| | | • A single-channel Teflon®/Polyethylene lumen with an inside | |
| | | diameter of 1 mm or larger and a length of 850 mm or | |
| | | shorter. | |
| STERRAD® NX,<br>Standard Cycle | 10.7 lbs. | Reusable metal and non-metal medical devices, including up to 10 | |
| | | lumens of the following dimensions per chamber load: | |
| | | • An inside diameter of 1 mm of larger and a length of 150 mm | |
| | | or shorter of single-channel stainless steel lumens. | |
| | | • An inside diameter of 2 mm or larger and a length of 400 mm | |
| | | or shorter of single-channel stainless steel lumens. | |
| | | | |
Table 2 – Wrap Model Recommendations for Use with Pre-vacuum steam cycles, Gravity steam cycles, and the Advanced Sterilization Products STERRAD® Sterilization Systems
| | | | Maximum Wrapped Package<br>Content Weights Based on Validated Sterility<br>Cycle | | |
|-----------------------|-----------------------|------------------------------------------------------|----------------------------------------------------------------------------------|--------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Gemini Wrap<br>Weight | Item Number<br>Series | Intended Load | Pre-Vacuum<br>Steam &<br>Gravity<br>Steam | Advanced<br>Sterilization<br>Products<br>(ASP)<br>STERRAD®<br>100S | Advanced<br>Sterilization<br>Products<br>(ASP)<br>STERRAD®<br>NX and<br>100NX |
| Light Weight | GEM11XX<br>GEM11XXS | Light weight package (for<br>example: standard linen | 6 lbs | 6 lbs | 10.7 lbs |
{6}------------------------------------------------
| | GEM11XXT | packs) | | | |
|----------------------|---------------------------------|------------------------------------------------------------------------------------------|--------|---------|----------|
| Regular Weight | GEM21XX<br>GEM21XXS<br>GEM21XXT | Light to moderate weight<br>package (for example:<br>general use medical<br>instruments) | 9 lbs | 9.7 lbs | 10.7 lbs |
| Medium Weight | GEM31XX<br>GEM31XXS<br>GEM31XXT | Moderate to heavy<br>weight package (for<br>example: general use<br>medical instruments) | 13 lbs | 9.7 lbs | 10.7 lbs |
| Heavyweight | GEM41XX<br>GEM41XXS<br>GEM41XXT | Heavyweight package (for<br>example: general use<br>medical instruments) | 17 lbs | 9.7 lbs | 10.7 lbs |
| Super<br>Heavyweight | GEM51XX<br>GEM51XXS<br>GEM51XXT | Very heavyweight<br>package (for example:<br>general use medical<br>instruments) | 25 lbs | 9.7 lbs | 10.7 lbs |
The following loads were used in the pre-vacuum steam and gravity steam Sterility Validation Studies:
- Light Weight: 2 huck towels (17 in. x 29 in.), 3 fluid-resistant drapes (108 in. x 72 in.)
- . Regular Weight: 16 huck towels (17 in. x 29 in.), 1 fluid resistant table cover (90 in. x 60 in.). 5 lbs of metal mass.
- Medium Weight: 4 tray liners (20 in. x 25 in.) stacked 23 in. x 11 in. x 3.5 in. tray containing 11 Ibs of metal mass.
- o Heavyweight: 4 tray liners (20 in. x 25 in.) stacked 23 in. x 11 in. x 3.5 in. tray containing 15 lbs of metal mass.
- . Super Heavyweight: 4 tray liners (20 in. x 25 in.) stacked 23in. x 11in. x 3.5 in. tray containing 20 lbs of metal mass.
The following loads were used in the Advanced Sterilization Products (ASP) STERRAD 100S Sterility Validation Studies:
- Light Weight: 15 in. x 10 in. x 1.2 in. tray containing metal instruments.
- Regular Weight: 15 in. x 10 in. x 1.2 in. tray containing metal instruments.
- . Medium Weight: 15 in. x 10 in. x 1.2 in. tray containing metal instruments.
- Heavyweight: 15 in. x 10 in. x 1.2 in. tray containing metal instruments.
- Super Heavyweight: 15 in. x 10 in. x 1.2 in. tray containing metal instruments.
{7}------------------------------------------------
The following loads were used in the Advanced Sterilization Products (ASP) STERRAD NX and STERRAD 100NX Sterility Validation Studies:
- Light Weight: 23 in. x 11 in. x 4 in tray containing metal instruments. ●
- Regular Weight: 23 in. x 11 in. x 4 in tray containing metal instruments.
- Medium Weight: 23 in. x 11 in. x 4 in tray containing metal instruments.
- Heavyweight: 23 in. x 11 in. x 4 in tray containing metal instruments. .
- . Super Heavyweight: 23 in. x 11 in. x 4 in tray containing metal instruments.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Medline. The word "MEDLINE" is written in blue, stylized, sans-serif font. Below the text is a blue, four-pointed star-like shape, with the bottom point extending further down than the other three. The logo is simple and clean, using only the color blue.
# 510(k) Summary (as required per 21 CFR 807.92)
# Summary Preparation Date
May 7, 2015
## Submitter / 510(k) Sponsor
Medline Industries, Inc. One Medline Place Mundelein, IL 60060
# Contact Person
Matt Clausen Sr. Regulatory Affairs Specialist Phone: 847-643-4785 Fax: 847-643-4482
## Device Name / Classification
Device Name: Gemini Bonded Sterilization Wrap Proprietary Name: Medline's Gemini Bonded Sterilization Wrap Common Name: Sterilization Wrap Classification Name: Sterilization Wrap (21 CRF 880.6850, product code - FRG)
## Predicate Device
Gemini Sterilization Wrap - K113353
## Device Description
Gemini Bonded Sterilization Wrap is comprised of two ply sheets ultrasonically bonded together for use by customers in accordance with standard hospital practices which require that two sheets are used each time a medical device or collections of medical devices are wrapped. The Gemini Bonded Sterilization Wrap provides the protection of double-wrapping in an efficient manner; reducing wrapping/unwrapping time significantly compared to traditional double-wrapping methods. Gemini Bonded Sterilization Wrap items are square or rectangular sheets of fabric produced using a five-layer SSMMS (spunbondspunbond-meltblown-meltblown-spunbond) process. The standard blue wrap fabric is made of polypropylene with the addition of blue and white pigmentation. The wrap allows for aseptic opening of the sterilized package. The two-tone blue/pink wrap fabric is made of polypropylene with the addition of blue, white and red pigmentation. The wrap allows for aseptic opening of the sterilized package. Gemini sterilization wrap is available in sizes ranging from 12"x12" to 54"x72" across five different material weights/models.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Medline. The logo is blue and consists of the word "MEDLINE" in a sans-serif font, with a stylized starburst symbol below it. The starburst has four points and a curved line connecting the bottom two points.
# Indications for Use
Gemini Bonded Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until opened. Gemini Bonded Sterilization Wrap is validated for use in steam or STERRAD® sterilization processes in the following sterilization modes and cycles.
- Pre-vacuum steam cycles
- Minimum 4 Minutes Exposure at 270°F/132°C with minimum 20 minutes dry time o
- Gravity steam cycles
- Maximum 30 Minutes Exposure at 250°F/121°C with minimum 20 minutes dry time o
- STERRAD® Sterilization ●
- STERRAD® 100NX, Flex cycle o
- STERRAD® 100NX, Standard cycle o
- STERRAD® 100NX, Express cycle o
- STERRAD® 100S, Standard cycle O
- STERRAD® NX, Advanced cycle o
- STERRAD®NX, Standard cycle O
The Gemini Bonded Sterilization Wrap intended loads with pre-vacuum steam sterilization, gravity steam sterilization, and the Advanced Sterilization Products (ASP) STERRAD® 100NX, STERRAD® 100S, and STERRAD® NX cycles are provided in Table 1.
The Gemini Bonded Sterilization Wrap Recommendations for Use with Pre-vacuum steam cycles, Gravity steam cycles, and the Advanced Sterilization Products STERRAD® Sterilization Systems are provided in Table 2.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "MEDLINE" in blue, with a stylized star symbol to the right of the word. The word is in a bold, italicized font, and the star symbol has four points. The star symbol is also in blue and appears to be connected to the word "MEDLINE".
# Table 1 – Validated Advanced Sterilization Products STERRAD® 100NX, STERRAD® 100S, and STERRAD® NX Cycles
| Cycle | Maximum recommended Chamber Load | Intended Load |
|-----------------------------------------|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pre-Vacuum Steam & Gravity Steam Cycles | Lightweight 6 lbs. | 2 huck towels (17 in. x 29 in.), 3 fluid-resistant drapes (108 in. x 72 in.) |
| | Regular Weight 9 lbs. | 16 huck towels (17 in. x 29 in.), 1 fluid resistant table cover (90 in. x 60 in.). 5 lbs. of metal mass. |
| | Medium Weight 13 lbs. | 4 tray liners (20 in. x 25 in.) stacked 23 in. x 11 in. x 3.5 in. tray containing 11 lbs. of metal mass. |
| | Heavyweight 17 lbs. | 4 tray liners (20 in. x 25 in.) stacked 23 in. x 11 in. x 3.5 in. tray containing 15 lbs. of metal mass. |
| | Super Heavyweight 25 lbs. | 4 tray liners (20 in. x 25 in.) stacked 23in. x 11in. x 3.5 in. tray containing 20 lbs. of metal mass. |
| STERRAD® 100NX, Flex Cycle | 10.7 lbs. | One or two single-channel Flexible Endoscope with or without a silicone mat and no additional load. The flexible endoscope may contain:<br>• A single-channel- Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter. (A maximum of two flexible endoscope, one per tray per sterilization cycle.) |
| STERRAD® 100NX, Standard Cycle | 10.7 lbs. | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:<br>• An inside diameter of 0.7 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens. (A maximum of two flexible endoscopes, one per tray per sterilization cycle.) |
| STERRAD® 100NX, EXPRESS Cycle | 10.7 lbs. | Non-lumened reusable metal and non-metal devices requiring surface sterilization, and sterilization of diffusion-restricted spaces such as the hinged portions of forceps and scissors, and rigid or semi-rigid endoscopes without lumens. |
| STERRAD® 100S, Standard Cycle | Light Weight 6 lbs. | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load: |
| | Regular Weight | 10.7 |
| | Medium Weight<br>Heavyweight<br>Super Heavyweight | lbs. |
| | | An inside diameter of 1mm or larger and a length of 125 mm or shorter of single-channel stainless steel lumens. An inside diameter OF 2 mm or larger and length of 250 mm or shorter of single-channel stainless steel lumens. An inside diameter of 3 mm or larger and a length of 400 mm or shorter of single-channel stainless lumens. An inside diameter of 6 mm or larger and a length of 310 mm or shorter of single-channel Teflon/Polyethylene lumens. |
| STERRAD®<br>NX, Advanced<br>Cycle | 10.7 lbs. | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load: An inside diameter of 1 mm of larger and a length of 500 mm or shorter of single-channel stainless steel lumens Or<br>One single-channel Flexible Endoscope with or without a silicone mat and no additional load. The flexible endoscope may contain: A single-channel Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter. |
| STERRAD®<br>NX, Standard<br>Cycle | 10.7 lbs. | Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load: An inside diameter of 1 mm of larger and a length of 150 mm or shorter of single-channel stainless steel lumens. An inside diameter of 2 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. |
| | | |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Medline. The logo consists of the word "MEDLINE" in blue, with a stylized blue starburst symbol below it. The starburst has four points, with the bottom point extending downwards.
Medline Industries, Inc. One Medline Place Mundelein, IL 60060-4486
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "MEDLINE" in blue, with a star-like symbol below it. The word is written in a bold, italicized font. The star symbol has four points, with the bottom point being longer than the other three. The image is set against a white background.
Table 2 – Wrap Model Recommendations for Use with Pre-vacuum steam cycles, Gravity steam cycles, and the Advanced Sterilization Products STERRAD® Sterilization Systems
| | Item Number<br>Series | | Maximum Wrapped Package<br>Content Weights Based on Validated<br>Sterility Cycle | | |
|-----------------------|---------------------------------|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Gemini Wrap<br>Weight | | Intended Load | Pre-<br>Vacuum<br>Steam &<br>Gravity<br>Steam | Advanced<br>Sterilizatio<br>n Products<br>(ASP)<br>STERRAD<br>® 100S | Advanced<br>Sterilizatio<br>n Products<br>(ASP)<br>STERRAD<br>® NX and<br>100NX |
| Light Weight | GEM11XX<br>GEM11XXS<br>GEM11XXT | Light weight package<br>(for example: standard<br>linen packs) | 6 lbs | 6 lbs | 10.7 lbs |
| Regular<br>Weight | GEM21XX<br>GEM21XXS<br>GEM21XXT | Light to moderate<br>weight package (for<br>example: general use<br>medical instruments) | 9 lbs | 9.7 lbs | 10.7 lbs |
| Medium<br>Weight | GEM31XX<br>GEM31XXS<br>GEM31XXT | Moderate to heavy<br>weight package (for<br>example: general use<br>medical instruments) | 13 lbs | 9.7 lbs | 10.7 lbs |
| Heavyweight | GEM41XX<br>GEM41XXS<br>GEM41XXT | Heavyweight package<br>(for example: general<br>use medical<br>instruments) | 17 lbs | 9.7 lbs | 10.7 lbs |
| Super<br>Heavyweight | GEM51XX<br>GEM51XXS<br>GEM51XXT | Very heavyweight<br>package (for example:<br>general use medical<br>instruments) | 25 lbs | 9.7 lbs | 10.7 lbs |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Medline. The word "MEDLINE" is in blue, block letters. Below the word is a blue star-like shape with four points. The bottom point of the star is elongated.
# Summary of Technological Characteristics
# Comparison of Subject and Predicate Devices
| Characteristic | Subject Device | Predicate Device | Compariso<br>n |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| 510(k) | TBD | K113353 | n/a |
| Product Code | FRG | FRG | Same |
| Regulation No. | 21 CRF 880.6850 | 21 CRF 880.6850 | Same |
| Class | II | II | Same |
| Intended Use | Gemini Bonded Sterilization<br>Wrap is intended to be used to<br>enclose another medical device<br>that is to be sterilized by a health<br>care provider. It is intended to<br>allow sterilization of the enclosed<br>medical device and also to<br>maintain sterility of such content<br>for 180 days following<br>sterilization. Gemini Bonded<br>Sterilization Wrap is validated for<br>use in steam or STERRAD®<br>sterilization processes. | Gemini Sterilization Wrap is<br>intended to be used to enclose<br>another medical device that is to<br>be sterilized by a health care<br>provider. It is intended to allow<br>sterilization of the enclosed<br>medical device and also to<br>maintain sterility of such content<br>for 180 days following<br>sterilization. Gemini Sterilization<br>Wrap is validated for use in steam<br>or STERRAD® sterilization<br>processes. | Same |
| OTC / single<br>use | Yes | Yes | Same |
| Device<br>materials | Polypropylene w/ blue and pink<br>pigmentation | Polypropylene w/ blue and pink<br>pigmentation | Same |
| Device features | Square or rectangular sheets mfd.<br>by spunbond-meltblown process | Square or rectangular sheets mfd.<br>by spunbond-meltblown process | Same |
| Sizes | 12"x12" to 54"x72" | 12"x12" to 54"x72" | Same |
| Wrapping<br>technique | Sequential/simultaneous double<br>wrapping | Sequential/simultaneous double<br>wrapping | Same |
| Bonding<br>method | Ultrasonically seamed in a dotted<br>line pattern along two sides | Unbonded | Different |
| Sterilization | Pre-Vacuum Steam Cycle: 4<br>-<br>Minutes Exposure at<br>270°F/132C° with minimum<br>20 minutes dry time<br>Gravity Steam Cycle: 30 | Pre-Vacuum Steam Cycle: 4<br>-<br>Minutes Exposure at<br>270°F/132C° with minimum<br>20 minutes dry time<br>Gravity Steam Cycle: 30 | Same |
| | Minutes Exposure at<br>250°F/121°C with minimum<br>20 minutes dry time | | Minutes Exposure at<br>250°F/121°C with minimum<br>20 minutes dry time |
| - | STERRAD® 100NX (Flex,<br>Standard, and Express<br>Cycles) | - | STERRAD® 100NX (Flex,<br>Standard, and Express<br>Cycles) |
| - | STERRAD® 100S,<br>Standard Cycle | - | STERRAD® 100S,<br>Standard Cycle |
| - | STERRAD® NX (Advance<br>and Standard Cycles) | - | STERRAD® NX (Advance<br>and Standard Cycles) |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the Medline logo. The logo is blue and consists of the word "MEDLINE" in a stylized font, with a star-like symbol to the right of the word. The star symbol has four points, and there is a curved shape below the star.
Medline Industries. Inc. One Medline Place Mundelein, IL 60060-4486
This table illustrates the identical nature of the subject/predicate devices across all but one of the comparative features. The singular difference of ultrasonic bonding is the reason for this Special 510(k) submission.
### Discussion of similarities and differences:
The proposed Gemini Bonded Sterilization Wrap is substantially equivalent in intended use, materials, device features / specifications and function in comparison to the predicate Gemini Sterilization Wrap. As noted in the table above, these characteristic comparisons render the devices identical. The difference between subject/predicate devices is related to:
- . Bonding Method: The process involves the ultrasonic bonding on two single ply sheets of Gemini Sterilization Wrap. High-frequency ultrasonic acoustic vibrations and pressure fuse bond the layers together in a three-dotted line pattern along two open sides. There is no adhesive or other materials introduced to facilitate the bonding operation - strictly ultrasonic acoustic vibration and pressure.
Once the bonding process is completed, the predicate Gemini Wrap is converted into the proposed Gemini Bonded Wrap.
The following responses to FDA's 510(k) "Substantial Equivalence" Decision-Making Process (Overview) flow chart demonstrate substantial equivalence in terms of indications for use and technological characteristics.
### Does the New Device Have the Same Intended Use?
Yes, the subject and predicate (K113353) have the same intended uses.
## Does the New Device Have Technological Characteristics That Raise New Types of Safety and Effectiveness Ouestions?
There are no new technology characteristics. The Gemini Bonded Sterilization Wrap incorporates the same design features and principles of operation found in currently marketed product.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the Medline logo and address. The logo is a blue star with the word "MEDLINE" in blue letters. The address is "Medline Industries, Inc., One Medline Place, Mundelein, IL 60060-4486".
This product modification involving the subject and predicate devices does not significantly alter the Gemini Wrap or raise questions regarding safety or effectiveness.
Information within this Premarket Notification demonstrates that there are no significant differences in technological characteristics between Medline's Gemini Bonded Sterilization Wrap and the cited predicate device.
### Summary of Non-Clinical Testing
#### Table 1 - Summary of Performance Testing
| Test Objective | Testing Standards | Performance Results |
|---------------------------------|---------------------------|------------------------------------|
| Primary Skin Irritation Testing | ISO 10993-10 | Meets ISO Requirements |
| Microbial Barrier Properties | ISO 11607-1<br>ASTM F1980 | Meets ISO and ASTM<br>Requirements |
The safety and effectiveness of Medline's Gemini Bonded Sterilization Wrap is adequately supported by the substantial equivalence information, materials information, and Design Control activities referenced within this Premarket Notification.
### Summary of Clinical Testing
Not applicable.
#### Conclusion
In accordance with 21 CFR Part 807, and based on the information provided in this premarket notification, Medline Industries, Inc. concludes that the Gemini Bonded Sterilization Wrap is as safe, as effective, and performs as well as the predicate device [Gemini Sterilization Wrap (K113353)].
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.