K143143 · Custom Spine, Inc. · MAX · Jul 22, 2015 · Orthopedic
Device Facts
Record ID
K143143
Device Name
PATHWAY ELIF
Applicant
Custom Spine, Inc.
Product Code
MAX · Orthopedic
Decision Date
Jul 22, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PATHWAY ELIF device(s) is intended for spinal fusion procedure at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative Disc Diseases (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spiral surgery at the involved spinal level(s). The PATHWAY ELIF device is intended to be used with supplemental spinal fixation systems that have been cleared for lumbosacral spine (i.e. posterior pedicle screws and rod systems, and anterior screw and rod systems. The device(s) is intended to be used with autogenous bone graft. Patients must have undergone a regiment of at least (6) months of non-operative treatment prior to being treated with the PATHWAY ELIF device(s). The PATHWAY ELIF device can be used in one of two methods: Transforaminal Lumbar Interbody Fusion (TLIF) Used as a TLIF, a single device is implanted in the appropriate location to provide support for a transforaminal approached surgery. Posterior Lumbar Interbody Fusion (PLIF) Used as a PLIF, two devices are implanted in the appropriate locations to provide support to the spine for a posterior surgery.
Device Story
PATHWAY ELIF is an expandable intervertebral body fusion device for lumbar spinal fusion. Implant features three lordotic angle/height positions (0°, 8°, 16°) and two lengths; provides height restoration and lordotic adaptability. Comprised of upper/lower expandable surfaces with posterior height maintenance component; superior/inferior surfaces coated with plasma-sprayed titanium for roughened finish. Used in TLIF (single device) or PLIF (two devices) procedures. Implanted by surgeons in clinical settings to support spine during fusion; used with supplemental fixation (pedicle screws/rods) and autogenous bone graft. Benefits include structural support and stabilization of spinal segments in patients with DDD.
Clinical Evidence
No clinical data. Evidence consists of bench testing (static/dynamic axial compression, shear, subsidence, and expulsion) per ASTM standards and a cadaver study to demonstrate usability and functionality.
Technological Characteristics
Titanium alloy (Ti-6Al-4V ELI per ISO 5832-3 and ASTM F136) with plasma-sprayed titanium coating. Expandable design with fixed configurations (0°, 8°, 16°). Tested per ASTM F2077, F2267, F1580, F1854, F1044, F1147, F1160, and F1978.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD defined as discogenic back pain with degeneration confirmed by history/radiography; may include up to Grade I spondylolisthesis or retrolisthesis. Includes patients with prior non-fusion spinal surgery at involved levels. Requires 6 months of failed non-operative treatment. Contraindications not explicitly listed.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three faces in profile, stacked on top of each other. The faces are facing to the right. The image is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 22, 2015
Custom Spine, Incorporated Mr. Mahmoud Abdelgany Chief Executive Officer 9 Campus Drive Parsippany, New Jersey 07054
Re: K143143
Trade/Device Name: PATHWAY ELIF Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: June 15, 2015 Received: June 18, 2015
Dear Mr. Abdelgany:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Mr. Mahmoud Abdelgany
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mark N. Melkerson -S
Mark N. Melkerson Director, Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
K143143
Page 1 of 1
510(k) Number (if known) K143143
Device Name PATHWAY ELIF
#### Indications for Use (Describe)
The PATHWAY ELIF device(s) is intended for spinal fusion procedure at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative Disc Diseases (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spiral surgery at the involved spinal level(s).
The PATHWAY ELIF device is intended to be used with supplemental spinal fixation systems that have been cleared for lumbosacral spine (i.e. posterior pedicle screws and rod systems, and anterior screw and rod systems. The device(s) is intended to be used with autogenous bone graft.
Patients must have undergone a regiment of at least (6) months of non-operative treatment prior to being treated with the PATHWAY ELIF device(s).
The PATHWAY ELIF device can be used in one of two methods:
Transforaminal Lumbar Interbody Fusion (TLIF)
Used as a TLIF, a single device is implanted in the appropriate location to provide support for a transforaminal approached surgery.
Posterior Lumbar Interbody Fusion (PLIF)
Used as a PLIF, two devices are implanted in the appropriate locations to provide support to the spine for a posterior surgery.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section VIII. 510(K) SUMMARY
#### Name of Firm
Custom Spine, Incorporated 9 Campus Dr. Parsippany, NJ 07054 Phone: (973) 808-0019 Fax: (877) 770-7746
### Date Prepared
July 17, 2015
#### Official Correspondent
Mahmoud Abdelgany Chief Executive Officer 9 Campus Dr. Parsippany, NJ 07054 Phone: (973) 808-0019 Fax: (877) 770-7746 E-mail: mahmoud(@customspine.com
#### Establishment Number
3005129649
#### Device Name
Legally Marketed Trade Name: PATHWAY ELIF Common Name: INTERVERTEBRAL BODY FUSION DEVICE Device Classification: Class II Regulation Number: 21 CFR 888.3080 Device Product Codes: MAX
#### Predicate Devices
Custom Spine PATHWAY (K111774)
#### Device Description
The PATHWAY ELIF implant features three prescribed lordotic angle / height positions and is available in two lengths for posterior lumbar fusion. It allows lordotic adaptability and height restoration to meet patient anatomy. It is comprised of upper and lower expandable surfaces preassembled with a posterior height maintenance component. The device can be expanded to fixed configurations of 0°, 8°, and 16° positions. Each position after 0° provides approximately 2mm of additional mid-body height increase. The superior and inferior surfaces of all components are coated with plasma sprayed titanium to provide a roughened finish.
{4}------------------------------------------------
# Indications for Use
The PATHWAY ELIF device(s) is intended for spinal fusion procedure at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative Disc Diseases (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD mav also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).
The PATHWAY ELIF device is intended to be used with supplemental spinal fixation systems that have been cleared for lumbosacral spine (i.e. posterior pedicle screws and rod systems, anterior plate systems, and anterior screw and rod systems. The device(s) is intended to be used with autogenous bone graft.
Patients must have undergone a regiment of at least (6) months of non-operative treatment prior to being treated with the PATHWAY ELIF device(s).
The PATHWAY ELIF device can be used in one of two methods:
Transforaminal Lumbar Interbody Fusion (TLIF) Used as a TLIF, a single device is implanted in the appropriate location to provide support for a transforaminal approached surgery.
Posterior Lumbar Interbody Fusion (PLIF)
Used as a PLIF, two devices are implanted in the appropriate locations to provide support to the spine for a posterior surgery.
## Materials
Materials used in the PATHWAY ELIF device(s) are:
- Titanium allov: Ti-6AI-4V ELI according to ISO 5832-3 and ASTM F136. and are coated with ● plasma spray titanium.
## Performance Data
Testing was performed in accordance with the following standard(s):
- ASTM F2077-11, "Test Methods for Intervertebral Body Fusion Devices" ●
- o ASTM F2267-04(2011), "Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression"
- . ASTM F04.25.02.02, Draft standard for implant expulsion
- o ASTM F1580-07 Standard Specification for Titanium 6 Aluminum-4 Vanadium Alloy Powders for Coatings of Surgical Implants
- o ASTM F1854-09, Standard Test Method for Stereological Evaluation of Porous Coatings on Medical Implants
- o ASTM F1044-05, Standard Test Method for Shear Testing of Calcium Phosphate Coatings and Metallic Coatings
- . ASTM F1147-05, Standard Test Method for Tension Testing of Calcium Phosphate and Metallic Coatings
- ASTM F1160-05, Standard Test Method for Shear and Bending Fatigue Testing of Calcium Phosphate and Metallic Medical and Composite Calcium Phosphate/Metallic
- o ASTM F1978-00, Standard Test Method for Measuring Abrasion Resistance of Metallic Thermal Spray Coatings by Using the TaberTM Abraser
{5}------------------------------------------------
# Non-Clinical Testing
The PATHWAY ELIF System demonstrated equivalent performance to the predicate through static and dynamic axial compression shear testing per ASTM F2077, subsidence testing per ASTM F2267, and expulsion testing per ASTM Draft F04.25.02.02. In addition, a cadaver study was conducted to demonstrate usability and functionality.
# Technological Characteristics
The PATHWAY ELIF System consists of a series of Titanium alloy: Ti-6Al-4V ELI according to ISO 5832-3 and ASTM F136, and are coated with plasma spray titanium. The proposed devices are the same as the current predicate devices already on the market with minor differences in shapes and sizes.
Test data has shown that the proposed devices are equivalent to the predicate devices and the minor differences will not impact the devices safety and effectiveness.
## Substantial Equivalence Statement
Documentation is provided to demonstrate that the PATHWAY ELIF System is substantially equivalent to its predicate devices in terms of its material, design, indications for use, and performance characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.