The VersaJet II Hydrosurgery System

K143115 · Smith & Nephew, Inc. · FQH · Jul 15, 2015 · General Hospital

Device Facts

Record IDK143115
Device NameThe VersaJet II Hydrosurgery System
ApplicantSmith & Nephew, Inc.
Product CodeFQH · General Hospital
Decision DateJul 15, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5475
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Versalet II Hydrosurgery System is intended for wound debridement (acute and chronic wounds, burns), soft tissue debridement, and cleansing of the surgical site in applications in which, in the physicians judgment, require sharp debridement and pulsed lavage irrigation.

Device Story

VersaJet II Hydrosurgery System is a surgical tool for wound debridement and cleansing. System consists of a reusable power console, footswitch, and single-use sterile handpiece/tubing assembly. Operation involves a high-velocity pressurized stream of sterile saline to tangentially ablate tissue, cut, and remove foreign matter; simultaneously, the device vacuums effluent and debris through the handpiece into a collection system. Pressure is adjustable (1-10 settings) via console or footswitch. Used in clinical/surgical settings by healthcare providers. Output is the physical removal of necrotic tissue and contaminants, facilitating wound healing. Benefits include precise, simultaneous irrigation and evacuation, reducing manual labor for clinicians during debridement.

Clinical Evidence

Bench testing only. Testing included environmental testing, reliability testing, aerosol monitoring, and performance testing (pressure range, priming, cut-off function, and insertion/removal forces). Biocompatibility testing (cytotoxicity, skin irritation, sensitization) performed per ISO 10993. Electrical safety verified per IEC 60601-1 3rd Edition.

Technological Characteristics

Jet lavage system; pressurized saline stream for ablation/irrigation. Components: reusable power console, single-use sterile handpieces (VERSAJET EXACT/PLUS). Materials: nylon resin, Rulon™ bushings, metal components. Sterilization: Ethylene Oxide (handpieces). Connectivity: standalone console with footswitch. Standards: IEC 60601-1 (3rd ed), IEC 60601-1-2, ISO 10993, ISO 11135, ISO 13485, ISO 14971.

Indications for Use

Indicated for wound debridement (acute and chronic wounds, burns), soft tissue debridement, and surgical site cleansing requiring sharp debridement and pulsed lavage irrigation.

Regulatory Classification

Identification

A jet lavage is a device used to clean a wound by a pulsatile jet of sterile fluid. The device consists of the pulsing head, tubing to connect to a container of sterile fluid, and a means of propelling the fluid through the tubing, such as an electric roller pump.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing right, with flowing lines representing hair or clothing. The profiles are arranged in a staggered formation, creating a sense of depth. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Smith and Nephew Incorporated Ms. Laura Reynolds Director Regulatory Affairs 970 Lake Carillon Drive, Suite 110 Saint Petersburg, Florida 33716 Re: K143115 Trade/Device Name: The VersaJet II Hydrosurgery System Regulation Number: 21 CFR 880.5475 Regulation Name: Jet lavage Regulatory Class: Class II Product Code: FQH Dated: June 11, 2015 Received: June 12, 2015 Dear Ms. Reynolds: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. July 15, 2015 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerelv vours. Joshua C. Nipper -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K143115 Device Name The VersaJet II Hydrosurgery System #### Indications for Use (Describe) The Versalet II Hydrosurgery System is intended for wound debridement (acute and chronic wounds, burns), soft tissue debridement, and cleansing of the surgical site in applications in which, in the physicians judgment, require sharp debridement and pulsed lavage irrigation. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|----------------------------------------------------------------------| | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the text "K143115" in black font at the top. Below that is the Smith & Nephew logo, with the words "We are" in gray and "smith&nephew" in orange. The logo is a stylized flower with orange petals. Wound Management Smith & Nephew, Inc. 970 Lake Carillon Drive Suite 110 St. Petersburg, FL 33716 F 727 392-1261 F 727 392-6914 or 727 392-0797 Customer Care Center: 1 800 876-1261 www.smith-nephew.com # 510(k) Summary | General Information | | |------------------------------------|------------------------------------------------------------------------------------------| | Submitters Name/Address: | Smith & Nephew, Inc.<br>970 Lake Carillon Drive<br>Suite 110<br>St. Petersburg, FL 33716 | | Establishment Registration Number: | 3006760724 | | Contact Person: | Laura Reynolds<br>Director Regulatory Affairs | | Phone Number: | (727) 329-7702 | | Date Prepared: | July 8, 2015 | | Device Description | | | Trade Name: | VERSAJET™ II Hydrosurgery System | | Generic/Common Name: | Jet Lavage | | Classification Name: | Jet Lavage, 21 CFR 880.5475 | | | Product Code: FQH | ## Predicate Device Information | Current Device | Predicate Device | 510k# | Clearance Date | |-------------------------------------|-------------------------------------|---------|----------------| | VERSAJET™ II Hydrosurgery<br>System | VERSAJET™ II Hydrosurgery<br>System | K110958 | 08/01/2011 | #### Device Description The VERSAJET™ II Hydrosurgery System consists of two basic system components: the reusable power console and the single-use, sterile (Ethylene Oxide) handpiece and tubing assembly. A footswitch is also available for operating the system. The VERSAJET™ II Hydrosurgery System uses a pressurized stream of sterile fluid to cut, ablate and remove tissue and foreign matter from wounds and to resect and remove material in a variety of surgical applications. The device provides cutting, irrigation and evacuation in the same tool. The high velocity stream of saline simultaneously washes the tissue surface and vacuums away foreign material from the wound. The fluid acts to tangentially ablate the surface of the tissue and {4}------------------------------------------------ propel excised tissue and debris out of the wound. The debris and fluid are directed into the handpiece into a flexible tube, which carries the effluent to the drain or collection canister. The pressure can be adjusted using either the footswitch or a sealed membrane switch on the front panel of the power console. Pressure settings range from 1-10 in factory pre-set increments, with the pressure increasing with each higher setting number, depending on user preference and the needs of a particular application. Associated accessories include: - VERSAJET EXACT Handpieces, single-use, sterile, in three configurations . - . VERSAJET PLUS Handpieces, single-use, sterile, in three configurations - Footswitch #### Indications for Use The VERSAJET II Hydrosurgery System is intended for wound debridement (acute and chronic wounds, burns), soft tissue debridement, and cleansing of the surgical site in applications in which, in the physicians judgment, require sharp debridement and pulsed lavage irrigation. | | New Device: | Predicate Device: | |--------------------------|-------------------------------------|-------------------------------------| | | | 510(k)# K110958 | | Trade Name: | VERSAJET™ II<br>Hydrosurgery System | VERSAJET™ II<br>Hydrosurgery System | | Indications for Use: | Identical | Identical | | Environment of use | Identical | Identical | | VersaJet II Handpiece | Substantially Equivalent | Substantially Equivalent | | Console case | Identical | Identical | | Power Supply | Substantially Equivalent | Substantially Equivalent | | Motor | Identical | Identical | | Transmission | Substantially Equivalent | Substantially Equivalent | | User Interface PCB | Substantially Equivalent | Substantially Equivalent | | User Interface Housing | Substantially Equivalent | Substantially Equivalent | | User Interface Chamber | Substantially Equivalent | Substantially Equivalent | | Method of Sterilization: | Identical | Identical | | Biocompatibility | Identical | Identical | | Software | Identical | Identical | #### Summary Comparison between New and Predicate Devices {5}------------------------------------------------ # Table of Modifications | Modification | Reason For Change | Verification Testing<br>Performed | |----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------| | Saline Bag Spike resin | Obsolete material | Comprehensive verification<br>was completed which<br>demonstrated acceptable<br>device performance;<br>retested biocompatibility | | Retainer Fitting-changed<br>from a metal injection<br>molded part to a<br>machined part; no<br>specification changes | Improve manufacturability of the<br>part | Comprehensive verification<br>was completed which<br>demonstrated acceptable<br>device performance | | Body Pump with Chamfer<br>- increase length of dowel<br>pin component | Length of the dowel pin was<br>increased to assure full<br>engagement between the pump<br>body and retaining fitting;<br>improve manufacturability | Comprehensive verification<br>was completed which<br>demonstrated acceptable<br>device performance | | High Pressure Hose -<br>nylon material resin<br>change | Obsolete material | Comprehensive verification<br>was completed which<br>demonstrated acceptable<br>device performance;<br>retested biocompatibility | | Change to power supply<br>Printed Circuit Board<br>(PCB) layout | Address electrical issues of not<br>turning on | Comprehensive verification<br>and reliability was<br>completed which<br>demonstrated acceptable<br>device performance | | Changes to the oil seals,<br>gasket and compression<br>spring in the transmission | Increase reliability and address<br>oil leakage | Comprehensive verification<br>and reliability was<br>completed which<br>demonstrated acceptable<br>device performance | | Replacing passivation<br>process with electro-<br>polishing of the user<br>interface components | Increase their corrosion<br>resistance | Comprehensive verification<br>and reliability was<br>completed which<br>demonstrated acceptable<br>device performance | | Replace a bronze bushing<br>in the user interface<br>chamber, with a Rulon™<br>bushing | Eliminate galvanic corrosion | Comprehensive verification<br>and reliability was<br>completed which<br>demonstrated acceptable<br>device performance | | Minor modifications to<br>comply with IEC 60601-1<br>3rd Edition safety<br>standards | Compliance with safety standard | Device complies with the<br>IEC 60601-1 3rd Edition<br>series of electrical safety<br>standards | {6}------------------------------------------------ ## Non-Clinical Tests (Bench) Testing has been conducted to verify the modifications to the VERSAJET™ II Hydrosurgery System meet design specifications and demonstrate substantial equivalence to the predicate device. The list below summarizes the bench testing undertaken and successfully completed for the VERSAJET™ II Hydrosurgery System device: - Environmental testing of the device including the extremes of the operating range and under general operating room conditions. - Reliability testing demonstrating that the modified components improved the reliability of ● the device. - . Monitoring of aerosols generated through the use of the device outside of the operating room in conditions such as out-patient surgery. - . Performance testing demonstrating that the effective cutting pressure range per console power setting, system priming, console cut off pressure function, and handset insertion, removal, engagement and disengagement forces are substantially equivalent to the predicate. Biocompatibility on the VersaJet Handpieces has been successfully completed in accordance with applicable parts of ISO 10993 as follows: - Cytotoxicity - Skin Irritation - Sensitization Device complies with the following standards: - . ISO13485:2003, Medical Devices - Quality Management Systems - ISO 14971:2012 Medical Devices Application of Risk Management to Medical Devices ● - ISO 15223-1:2012 Medical devices Symbols to be used with medical device labels, . labelling, and information to be supplied Part 1: General requirements.(General) - . ISO 15223-2:2010 Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied-Part 2: Symbol development, selection and validation. (General) - BS EN 980:2008 Graphical Symbols for use in the labeling of Medical Devices - BS EN 1041:2008 +A1:2013 Information Supplied by the Manufacturer with Medical Devices - IEC 60601-1-2:2007(3rd edition) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests, Interpretation Sheet - . IEC 60601-1:2005 (3rd edition). Medical Electrical Equipment - Part 1: General Requirements for Safety - ISO 11135: 2014 Medical Devices - Validation and routine control of ethylene oxide sterilization - ISO 11607-1:2006 Packaging for terminally sterilized medical devices Part 1: ● Requirements for materials, sterile barrier systems and packaging systems - . ISO 10993-1:2009 Biological evaluation of medical devices Part 1: Evaluation and testing with a risk management process {7}------------------------------------------------ - ISO 10993-10:2010 Biological evaluation of medical devices- Part 10:Tests for irritation and skin sensitization #### Conclusions In establishing substantial equivalence to the currently marketed predicate devices, Smith & Nephew, Inc. evaluated the indications for use, materials, technology, product specifications and energy requirements of the device. Performance testing, biocompatibility testing and electrical safety testing has been successfully completed to demonstrate that the VERSAJET™ II Hydrosurgery System is substantially equivalent to the predicate device for the intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...