K143114 · Airstrip Technologies, Inc. · LQK · Mar 10, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K143114
Device Name
Sense4Baby System Model B+ (MSA)
Applicant
Airstrip Technologies, Inc.
Product Code
LQK · Obstetrics/Gynecology
Decision Date
Mar 10, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2730
Device Class
Class 2
Indications for Use
The Sense4Baby System Model B+ is indicated for conventional antepartum fetal monitoring applications in pregnancies greater than or equal to 24 weeks gestation. It may be used for antenatal monitoring (e.g., non-stress testing and/or uterine activity monitoring) in a health care setting or home. It is to be used by health care professionals and patients on the order of a physician. Before the Sense4Baby System Model B+ is prescribed for home use, the user (patient) must be instructed/trained in proper use of the equipment. Home uterine activity monitoring has not been shown to prevent the onset of preterm labor nor will it prevent the occurrence of preterm birth.
Device Story
Non-invasive, wireless, external monitoring system; measures fetal heart rate (FHR), maternal heart rate, and uterine contractions during antepartum testing. Inputs: pulsed Doppler ultrasound (1 MHz) via piezo-electric crystals for FHR; relative pressure sensor (tocodynamometer) for uterine activity. Data transmitted to gateway device, then to HIPAA-compliant web portal for storage and physician review. Used in clinical settings or home; operated by health care professionals or patients (following training). Output: graphical overlay of FHR and uterine activity on screen or paper. Benefits: enables remote antenatal monitoring, facilitating physician review of fetal/maternal status outside traditional clinical settings.
Clinical Evidence
Bench testing only. Verification and validation included simulated use, hardware, software, environmental, mechanical, water ingress (IEC 60529), and shipping (ASTM D4169-09) testing. Formative and summative usability studies conducted to evaluate lay user operation in home environments; results demonstrated a success rate statistically equivalent to health care provider environments.
Technological Characteristics
Pulsed Doppler ultrasound (1 MHz) for FHR; relative pressure sensor (tocodynamometer) for uterine activity. Wireless (Bluetooth) connection for toco sensor. Rechargeable batteries. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1/5/10. Biocompatible materials. Non-sterile. Dimensions: 25mm x 95mm x 75mm; weight <0.2 kg.
Indications for Use
Indicated for conventional antepartum fetal monitoring in pregnancies >= 24 weeks gestation. Used for antenatal monitoring (non-stress testing, uterine activity) in health care settings or homes. Prescribed by physicians for use by health care professionals and patients.
Regulatory Classification
Identification
A home uterine activity monitor (HUAM) is an electronic system for at home antepartum measurement of uterine contractions. The HUAM system comprises a tocotransducer and an at-home recorder. This device is intended for use in women with a previous preterm delivery to aid in the detection of preterm labor.
Special Controls
*Classification.* Class II (special controls); guidance document (Class II Special Controls Guidance for Home Uterine Activity Monitors).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the seal is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2015
Airstrip Technologies, Inc. % Curtis Egan Sr. Ouality, Regulatory & Design Engineer Certified Compliance Solutions Inc. 11665 Avena Place, Suite 203 San Diego, California 92128
Re: K143114
> Trade/Device Name: Sense4baby System Model B+ (MSA) Regulation Number: 21 CFR 884.2730 Regulation Name: Home Uterine Activity Monitor Regulatory Class: Class II Product Code: LOK, MOH, HGM Dated: December 1, 2014 Received: December 10, 2014
Dear Curtis Egan,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for Benjamin Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K143114
Device Name Sense4Baby System, Model B+ (MSA)
#### Indications for Use (Describe)
The Sense4Baby System Model B+ is indicated for conventional antepartum fetal monitoring applications in pregnancies greater than or equal to 24 weeks gestation. It may be used for antenatal monitoring (e.g., non-stress testing and/or uterine activity monitoring) in a health care setting or home.
It is to be used by health care professionals and patients on the order of a physician.
Before the Sense4Baby System Model B+ is prescribed for home use, the user (patient) must be instructed/trained in proper use of the equipment.
Home uterine activity monitoring has not been shown to prevent the onset of preterm labor nor will it prevent the occurrence of preterm birth.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 1. 510(k) Summary
| Submitter: | Airstrip Technologies, Inc. |
|----------------------|-----------------------------------|
| Address: | 565 Pearl St suite 209 |
| | La Jolla, CA 92037 |
| Phone number: | (201) 847-4496 |
| Fax number: | (201) 847-4845 |
| Contact person: | Curtis M. Egan |
| Phone number: | (951) 723-7261 |
| Fax number: | (858) 675-8201 |
| Date prepared: | October 21, 2014 |
| Trade name: | Sense4Baby System, Model B+ (MSA) |
| Common Name: | Fetal Monitor |
| Classification Name: | Home Uterine Activity Monitor |
| Product Code: | LQK, MOH, HGM |
| Regulation: | 21 CFR 884.2730 |
Substantial equivalence claimed to: Huntliegh Healthcare Limited Fetal Assist, K020390
#### 1.1. Description:
The Sense4Baby System, Model B+ (MSA) is a non-invasive, wireless, external monitoring system used to measure fetal heart rate, maternal heart rate, and uterine contractions during antepartum (non-stress) testing. The system transmits the measured data to a gateway device, and then a HIPAA Compliant web based portal for storage and physician review.
{4}------------------------------------------------
## Sense4Babv System B+ 510(k) Submission
### 1.2. Indications for Use:
The Sense4Baby System Model B+ is indicated for conventional antepartum fetal monitoring applications in pregnancies greater than or equal to 24 weeks gestation. It may be used for antenatal monitoring (e.g., non-stress testing and/or uterine activity monitoring) in a health care setting or home.
It is to be used by health care professionals and patients on the order of a physician.
Before the Sense4Baby System Model B+ is prescribed for home use, the user (patient) must be instructed/trained in proper use of the equipment.
Home uterine activity monitoring has not been shown to prevent the onset of preterm labor nor will it prevent the occurrence of preterm birth.
### 1.3. Technological Characteristics
Sense4Baby, Inc. Model B+ (MSA) is substantially equivalent to the predicate device the Huntliegh Healthcare Limited Fetal Assist (K020390).
The Sense4 Baby System Model B+ (MSA) complies with applicable voluntary safety standards, in the areas of:
- · Electrical Safety
- · Electro-Magnetic Compatibility (EMC)
- · Material safety
- Risk Management
Safety testing performed in conformance to these voluntary standards was conducted at accredited independent test facilities to demonstrate that the Sense4Baby System Model B+(MSA) is as safe and effective as the predicate device. The specific standards employed were:
- IEC 60601-1:2005 + CORR.2 (2007) + A1 (2012), Medical electrical equipment Part 1: General requirements for safety.
- · National deviations of IEC 60601-1: 2005, such as EN 60601-1, Ed. 3 (European Harmonized Standard), AAMI ES 60601-1, Ed. 1 (US National Standard), CSA - C22.2 No 60601-1:2008 (Canadian National Standard).
- · IEC 60601-1-2:2007, Medical electrical equipment Part 1-2: General requirements for safety -Collateral standard: Electromagnetic compatibility - Requirements and tests.
- IEC 60601-2-37:2007 Medical Electrical Equipment Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.
- ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process.
- · ISO 10993-5:2009: Biological evaluation of medical devices, Part 5- Tests for in vitro cytotoxicity.
- ISO 10993-10:2010 Biological evaluation of medical devices, Part 10- Tests for irritation and skin sensitization.
- · ASTM D4169:2009 Standard Practice for Performance Testing of Shipping Containers and Systems.
- · IEC 60529:1989 + A1 (1999) Degrees of protection provided by enclosures (IP Code).
The possibility of hazards arising from hardware and software errors was further minimized in compliance with ISO 14971:2007 and ISO 13485:2003
| Sense4Baby System Model B+ |
|-------------------------------|
| Premarket Notification 510(k) |
{5}------------------------------------------------
# Sense4Baby System B+ 510(k) Submission
| General Device | Sense4Baby | Huntliegh Healthcare Limited |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | System B+ (MSA) | Fetal Assist (K020390) |
| Indication for Use | The Sense4Baby System Model B+ is<br>indicated for conventional antepartum<br>fetal monitoring applications in<br>pregnancies greater than or equal to 24<br>weeks gestation. It may be used for<br>antenatal monitoring (e.g., non-stress<br>testing and/or uterine activity<br>monitoring) in a health care setting or<br>home.<br>It is to be used by health care<br>professionals and patients on the order<br>of a physician.<br>Before the Sense4Baby<br>-<br>System Model B+ is<br>prescribed for home use, the<br>user (patient) must be<br>instructed/trained in proper<br>use of the equipment.<br>Home uterine activity<br>monitoring has not been<br>shown to prevent the onset<br>of preterm labor nor will it<br>prevent the occurrence of<br>preterm birth. | The FETAL ASSIST is suitable for use in all conventional antepartum<br>fetal monitoring application from a gestation age of approximately 26<br>weeks. It is particularly intended for use in the following specific areas:<br>Antenatal monitoring in the hospital, doctors office, health clinic,<br>home or community.<br>External Labor monitoring.<br>Waterbith monitoring using optional waterproof transducers.<br>Home Uterine Activity Monitoring (HUAM) in conjunction with<br>standard high-risk care, for the daily at-home measurement of uterine<br>activity for women with history of preterm birth.<br>Domiciliary Non Stress Testing Applications.<br>NB: before being used in domiciliary HUAM and/or Non Stress Testing<br>applications the mother would be instructed in the use of the equipment.<br>The FETAL ASSIST is a prescription device. |
| Product Code | LQK, MOH, HGM | LQK, MOH |
| Classification | 884.2730, 884.2740 | 884.2730 |
| Indication | Antenatal monitoring | Antenatal monitoring |
| Target Patient | Gravid patient | Gravid patient |
| Anatomical Site | Maternal Abdomen | Maternal Abdomen |
| Use/Reuse | Reusable | Reusable |
| Sterility | Non-sterile | Non-sterile |
| Fetal Heart Rate | Sense4Baby<br>System B+ (MSA) | Huntliegh Healthcare Limited<br>Fetal Assist (K020390) |
| | | |
| Technology | Pulsed Doppler Ultrasound | Doppler Ultrasound |
| Sensors | Piezo-electric crystals | Piezo-ceramic transmitter crystals |
| Ultrasound frequency | 1 MHz | 1.5 Mhz |
| Ultrasound burst rep rate | 3125 Hz | 3200 Hz |
| Signal Processing | Analog to digital conversion | Analog to digital conversion |
| FHR sensor power source | Rechargeable battery | Rechargeable battery |
| FHR audio signal | Speakers (phone) | Speaker |
| Fetal Motion Indicator | Yes (3 levels) | Yes |
| Signal Quality Indicator | Yes (3 levels) | Yes |
| External Uterine Pressure | Sense4Baby<br>System B+ (MSA) | Huntliegh Healthcare Limited<br>Fetal Assist (K020390) |
| Technology | Relative pressure sensor<br>(tocodynamometer) | Relative pressure sensor (tocodynamometer) |
| Sensor | Differential pressure sensor | Differential pressure sensor |
| Signal Processing | Analog to digital conversion | Analog to digital conversion |
| Toco sensor power source | Rechargeable battery | Rechargeable battery |
| Connection of toco sensor | Wireless (Bluetooth) | Cable |
| Physical | Sense4Baby<br>System B+ | Huntliegh Healthcare Limited<br>Fetal Assist (K020390) |
| Size | 25mm(H) x 95mm(W) x 75MM(D) | 2" (50mm) x 10" (250mm) x 6 (150mm) |
| Weight | <0.2 kg | 3.5 lb (1.6Kg) (including battery) |
| Safety | Sense4Baby<br>System B+ (MSA) | Huntliegh Healthcare Limited<br>Fetal Assist (K020390) |
| Hardware | IEC 60601 compliant | EN60601-1: 1990 compliant |
| Ultrasound | IEC 60601-2-37 compliant | -- |
| EMC | IEC 60601-1-2 compliant | EN60601-1-2 : 1993 compliant |
| Applied Parts Type | BF | -- |
| Biocompatible materials | Yes | -- |
Sense4Baby System Model B+ Premarket Notification 510(k)
Confidential
Page 3 of 7 Volume 026, Document 002
{6}------------------------------------------------
# Sense4Baby System B+ 510(k) Submission
{7}------------------------------------------------
### 1.4. Substantial Equivalence Discussion
The S4B System B+ (MSA) is substantially equivalent to the Huntliegh Healthcare Limited Fetal Assist (K020390) Systems based on indications for use and comparison of the functional capabilities. All of the devices are intended to provide fetal heart rate monitoring. Features common to all systems include:
- Doppler Ultrasound Technology
- Analog to digital signal conversion
- . Rechargeable Batteries
- . Relative Pressure Sensors
### Huntliegh Healthcare Limited Fetal Assist (K020390)
The S4B System B+ (MSA) is substantially equivalent to the Huntliegh Healthcare Limited Fetal Assist System (K020390). This is based on the comparison of the indications for use and comparison of the functional capabilities. All of the devices are intended to provide fetal heart rate monitoring.
| DEVICE | | Comparison |
|----------------|------------------------------------------------------|--------------|
| Model B+ (MSA) | The Model B+ (MSA) uses Pulsed Doppler | Equivalent |
| | ultrasound technology to determine the fetal heart | to Predicate |
| | rate, and obtain indications of fetal motion. | |
| Fetal Assist | The Fetal Assist uses the Doppler principle for non- | |
| | invasive monitoring of Fetal Heart Rate and Fetal | |
| | movement. | |
For this primary function, the Model B+ (MSA) utilizes scientific technology / operating principles which are identical or equivalent to the Fetal Assist. These are detailed below:
| Device | Technology | Comparison |
|------------------------|-------------------------------------------------------------|--------------|
| Model B+ (MSA) | The fetal heart rate is commonly measured using | Equivalent |
| | Doppler ultrasound devices and uterine contractions are | to Predicate |
| | commonly measured with a strain gauge based | |
| | tocodynamometer. | |
| | The Sense4Baby System Model B+ uses Piezo-electric | |
| | crystals to generate the ultrasound signals and | |
| | conventional technologies to measure the fetal heart rate | |
| | and uterine contraction patterns. The result is a graphical | |
| | overlay of both measurements that can be viewed either | |
| | on a screen or on paper. | |
| Fetal Assist (K020390) | The Fetal Assist ultrasound transducer contains a | |
| | transmitter and receiver. In use, the transducer sends out | |
| | a pulsed ultrasonic signal, generated by the piezo- | |
| | ceramic transmitter crystals. | |
The minor differences between the S4B System B+ (MSA and the Huntliegh Healthcare Limited Fetal Assist System (K020390) do not raise new questions of safety or effectiveness. These differences will be discussed in turn below:
{8}------------------------------------------------
### Sense4Babv System B+ 510(k) Submission
### Ultrasound Frequency:
The Model B+(MSA), use a frequency of 1 MHz, while the Fetal Assist (K020390) uses a frequency of 1.5 MHz. While these frequencies are similar, the lower the ultrasound frequency the better the depth pf penetration of ultrasound. The Model B+ (MSA) has a slightly deeper penetration than the Fetal Assist (K020390).
### Ultrasound Burst Rate:
The Model B+ (MSA), use a burst rate of 3125Hz, while the Fetal Assist (K020390) uses a burst rate of 3200 Hz. The 3125Hz burst rate sets the rate at which 1MHz pulses are transmitted and the echoes are sampled and received. The 3200Hz sampling rate may result in a negligible increase in the resolution of the fetal heart rate (FHR) at the high end of the FHR range. This difference between the two systems does not raise any safety and effectiveness concerns.
A detailed comparison of the technological characteristics of the Sense4Baby System and the referenced predicate device is in section 5.3. No applicable performance standards have been issued under Section 514 of the Food, Drug, and Cosmetic Act for this device type. FDA Form 3654 is included in this submission to identify national and international standards used in the development and testing of the Sense4Baby System Model B+ including the following voluntary safety standards:
- . IEC 60601-1:2005 + CORR.2 (2007) + A1(2012), Medical electrical equipment = Part 1: General requirements for safety.
- National deviations of IEC 60601-1: 2005, such as EN 60601-1, Ed. 3 (European Harmonized Standard), AAMI ES 60601-1, Ed. 1 (US National Standard), CSA - C22.2 No 60601-1:2008 (Canadian National Standard).
- . IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for safety -Collateral standard: Electromagnetic compatibility - Requirements and tests.
- IEC 60601-1-11:2010. Medical electrical equipment Part 1-11: General requirements for basic ● safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- . IEC 60601-2-37:2007 Medical Electrical Equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.
- ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process.
- . ISO 10993-5:2009: Biological evaluation of medical devices, Part 5- Tests for in vitro cytotoxicity.
- . ISO 10993-10:2010 Biological evaluation of medical devices. Part 10- Tests for irritation and skin sensitization.
- . ASTM D4169:2009 Standard Practice for Performance Testing of Shipping Containers and Systems.
- . IEC 60529:1989 + A1(1999) Degrees of protection provided by enclosures (IP Code). Additional standards used in the development and risk analyses of the Sense4Baby System Model B+ are listed below:
- . ISO14971:2007 Medical Devices- Application of risk management to medical devices.
- ISO13485:2003 Medical Devices- Quality management systems- Requirements for regulatory purposes.
{9}------------------------------------------------
### Sense4Babv System B+ 510(k) Submission
### 1.5. Test Summary
Verification and validation activities established the performance, functionality, and reliability characteristics of the device with respect to the predicate. Testing involved sub-system, as well as system level tests. Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results demonstrated substantial equivalence. Executed testing included:
- Simulated Use testing
- . Hardware testing
- . Predicate comparison testing
- . Software testing
- Environmental testing (operating and storage)
- Mechanical testing (drop, cleaning, etc.)
- Water ingress testing (to IEC 60529) ●
- Shipping testing (to ASTM D4169-09)
- Formative Usability Testing
- . Summative Usability Testing
#### 1.6. Conclusion:
Therefore AirStrip Technologies, Inc. (S4B) is seeking pre-market clearance to introduce the Sense4Baby System Model B+ (MSA) maternal/fetal monitoring system into commercial distribution, with patient use as part of the indications for use. This document inclusive of its attachments contains a complete description of the Sense4Baby System Model B+ (MSA) Fetal monitoring system, a side by side comparison of the Sense4Baby System Model B+ (MSA) to the Huntliegh Healthcare Limited Fetal Assist (K020390) predicate device, a summary of all testing performed and the test results obtained. This information, when considered in its totality provides ample objective evidence that the Sense4Baby System Model B+ (MSA) has been demonstrated to be substantially equivalent to the predicate devices, and raises no new concerns related to safety or efficacy; allowing FDA to make a determination of substantial equivalence and clear the device as described for commercial distribution.
S4B conducted two separate studies to determine the ability of the user to operate the device in the home environment. These studies focused on key aspects of…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.