← Product Code [GCJ](/productcode/GCJ) · K143091

# Surgisleeve Wound Protector (K143091)

_Covidien, LLC · GCJ · Nov 26, 2014 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K143091

## Device Facts

- **Applicant:** Covidien, LLC
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Nov 26, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Surgisleeve™ Wound Protector is indicated for use to access the abdominal cavity during surgery through an atraumatically retracted incision, deliver maximum exposure of the abdominal cavity with minimum incision size, and protect against wound contamination during laparoscopic and open surgery. Additionally, the small size Wound Protector is indicated for use to access the thoracic cavity during cardiac and general surgical procedures through an atraumatically retracted incision. The extra-small Wound Protector is also indicated for use to access the thoracic cavity and other soft tissue retraction during cardiac and general surgical procedures through an atraumatically retracted incision.

## Device Story

Wound retraction device providing thoracic and abdominal access; protects against wound contamination during surgery. Consists of cylindrical film with flexible interior and exterior rings for insertion, retraction, and removal. Large and extra-large models include a rigid circular retraction ring accessory to maintain maximum retraction. Used in OR by surgeons; provides atraumatic incision retraction and exposure. Benefits include reduced incision size and minimized wound contamination risk.

## Clinical Evidence

Bench testing only. In-vitro testing included proximal ring rolling, film-to-ring attachment strength, film weld strength, and retraction ring compression force. In-vivo testing included digital proximal ring insertion, proximal ring rolling, retraction ring insertion/removal, and digital distal ring removal. Results demonstrate performance equivalent to the predicate.

## Technological Characteristics

Cylindrical film with flexible interior/exterior rings; rigid circular retraction ring accessory for large/extra-large models. Materials comply with ISO 10993-1. Mechanical retraction principle.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Surgisleeve™ Wound Protector ([K140064](/device/K140064.md))

## Reference Devices

- Surgisleeve™ ([K140014](/device/K140014.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name, "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA," arranged in a circular pattern around the edge. At the center of the seal is a stylized emblem consisting of three human profiles facing to the right, stacked on top of each other, representing the department's focus on health and human well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 26, 2014

Covidien LLC Mr. Michael Koczocik Product Specialist, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06473

Re: K143091

Trade/Device Name: Surgisleeve " Wound Protector Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ, KKX Dated: October 27, 2014 Received: October 28, 2014

Dear Mr. Koczocik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely yours,

# David-Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### 5. Indications for Use

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration |  | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------|--|---------------------------------------------------------------------------------------------------|
| Indications for Use                                                     |  |                                                                                                   |
| 510(k) Number (if known)                                                |  |                                                                                                   |
| K143091                                                                 |  |                                                                                                   |
| Device Name                                                             |  |                                                                                                   |

Surgisleeve™ Wound Protector

#### Indications for Use (Describe)

The Surgisleeve™ Wound Protector is indicated for use to access the abdominal cavity during surgery through an atraumatically retracted incision, deliver maximum exposure of the abdominal cavity with minimum incision size, and protect against wound contamination during laparoscopic and open surgery. Additionally, the small size Wound Protector is indicated for use to access the thoracic cavity during cardiac and general surgical procedures through an atraumatically retracted incision. The extra-small Wound Protector is also indicated for use to access the thoracic cavity and other soft tissue retraction during cardiac and general surgical procedures through an atraumatically retracted incision.

Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Covidien. The logo consists of a blue square with a white square inside, followed by the word "COVIDIEN" in blue. The word "COVIDIEN" is in a sans-serif font and is slightly larger than the square logo.

## 510(k) Summary

This 510(k) summary of data used to demonstrate substantial equivalence is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR క్తి807.98

| NAME:                      | COVIDIEN                                                                                                                    |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| ADDRESS:                   | 60 Middletown Avenue<br>North Haven, Connecticut 06473 USA                                                                  |
| CONTACT PERSON:            | Michael Koczocik<br>Product Specialist, Regulatory Affairs                                                                  |
| PHONE NUMBER:              | (203) 492-6312                                                                                                              |
| FAX NUMBER:                | (203) 492-5029                                                                                                              |
| DATE PREPARRED:            | October 27, 2014                                                                                                            |
| TRADE/PROPRIETARY NAME:    | Surgisleeve™                                                                                                                |
| COMMON/USUAL NAME:         | Wound Protector                                                                                                             |
| CLASSIFICATION NAME:       | Endoscope and Accessories per 21 CFR §<br>876.1500<br>Surgical Drape and Drape Accessories per 21<br>CFR § 878.4370         |
| PRODUCT CODE:              | GCJ, KKX                                                                                                                    |
| CLASSIFICATION PANEL NAME: | Gastroenterology and Urology<br>General and Plastic Surgery,                                                                |
| FDA PANEL NUMBER:          | 78, 79                                                                                                                      |
| DEVICE CLASS:              | Pursuant to 21 CFR § 876.1500  and 21 CFR §<br>878.4370 an endoscope and<br>accessories/surgical drape is a Class II device |
| PREDICATE DEVICE(S):       | Surgisleeve™ Wound Protector K140064                                                                                        |
| DEVICE DESCRIPTION:        | Wound retraction device providing thoracic and<br>abdominal access and protection from wound<br>contamination               |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Covidien. The logo consists of a blue square with a white cross in the center, followed by the word "COVIDIEN" in blue, sans-serif font. The square is positioned to the left of the word.

## INTENDED USE:

The Surgisleeve™ Wound Protector is indicated for use to access the abdominal cavity during surgery through an atraumatically retracted incision, deliver maximum exposure of the abdominal cavity with minimum incision size, and protect against wound contamination during laparoscopic and open surgery. Additionally, the small size Wound Protector is indicated for use to access the thoracic cavity during cardiac and general surgical procedures through an atraumatically retracted incision. The extra-small Wound Protector is also indicated for use to access the thoracic cavity and other soft tissue retraction during cardiac and general surgical procedures through an atraumatically retracted incision.

SUMMARY COMPARING THE TECHNOLOGICAL CHARACTERISTICS OF THE PROPOSED AND PREDICATE DEVICE(S)

Cylindrical Film is designed to retract an incision and provide protection from wound contamination. The interior and exterior rings are flexible to aid insertion, film retraction, and removal. The technological characteristics of the extra-large Surgisleeve™ remain the same as the 510(k) cleared Surgisleeve™ device (K140014). The retraction ring of the large and extra-large Surgisleeve™ Wound Protector is a rigid circular ring accessory that attaches to the exterior ring that aids in maintaining maximum retraction for both the large and extra-large device when necessary

The Surgisleeve™ Wound Protector's

MATERIALS:

All components of the Surgisleeve™ Wound Protector are comprised of materials which are in accordance with ISO 10993-1

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Covidien logo. The logo consists of a blue square with a white square inside, and the word "COVIDIEN" in blue text to the right of the square. The logo is simple and modern, and the colors are clean and professional.

## PERFORMANCE DATA:

In-vitro and in-vivo tests were performed to verify that the performance of the Surgisleeve™ Wound Protector with Retraction Ring is substantially equivalent to the predicate devices. To validate that the proposed device performs as intended to provide access into the body cavity and maintain maximum retraction, the following tests were performed:

- In-vitro proximal ring rolling .
- . In-vitro strength of attachment between film and proximal (exterior) and distal (interior) rings
- In-vitro film weld strength .
- In-vitro retraction ring compression ● force
- . In-vivo digital proximal ring insertion
- In-vivo proximal ring rolling
- In-vivo retraction ring insertion and ● removal
- . In-vivo digital distal ring removal

## CONCLUSION:

The results of the performance evaluation demonstrate that the Surgisleeve™ Wound Protector extra-large with Retraction ring performed substantially equivalent to the predicate device, Surgisleeve™ Wound Protector K140064.

---

**Source:** [https://fda.innolitics.com/device/K143091](https://fda.innolitics.com/device/K143091)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).