← Product Code [KRD](/productcode/KRD) · K143038

# Gel-Block 10X embolization pledgets (K143038)

_Vascular Solutions, Inc. · KRD · Nov 20, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K143038

## Device Facts

- **Applicant:** Vascular Solutions, Inc.
- **Product Code:** [KRD](/productcode/KRD.md)
- **Decision Date:** Nov 20, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.3300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Gel-Block 10x embolization pledgets is intended for use in the embolization of hypervascular tumors and arteriovenous malformations (AVMs).

## Device Story

Gel-Block 10x embolization pledgets are porcine-derived gelatin pledgets designed for vascular embolization. The device consists of 10 radially compressed pledgets packaged in a glass vial. In clinical use, a physician removes pledgets from the vial, places them into a delivery syringe, and introduces them into a delivery catheter (not included). Upon exiting the catheter into a blood vessel, the pledgets expand to occlude blood flow. The device is intended for single use. By physically obstructing blood flow to hypervascular tumors or arteriovenous malformations, the device facilitates therapeutic embolization. The device is used in a clinical setting by physicians.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including deliverability, package validation, sterilization validation, and biocompatibility testing (cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, and hematology). All tests met acceptance criteria.

## Technological Characteristics

Porcine-derived gelatin pledgets; radially compressed; three sizes compatible with catheters ≥ 0.021", 0.027", or 0.038" ID. Sterilized via electron-beam irradiation. Packaging includes glass vial with screw cap and delivery syringe. Biocompatibility testing performed per ISO 10993-1.

## Regulatory Identification

A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.

## Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).

## Predicate Devices

- Gel-Block embolization pledgets ([K113266](/device/K113266.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is depicted in a simple, black and white design, with the snakes forming a continuous line. The text is written in a sans-serif font and is also in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 20, 2014

Vascular Solutions, Inc. Adam Ettl Sr. Regulatory Product Specialist 6464 Sycamore Court North Minneapolis, Minnesota 55369

Re: K143038

> Trade/Device Name: Gel-block 10x Embolization Pledgets Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: October 21, 2014 Received: October 23, 2014

Dear Adam Ettl,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Melissa A. Torres -S

For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K143038

Device Name Gel-Block 10x embolization pledgets

Indications for Use (Describe)

The Gel-Block 10x embolization pledgets is intended for use in the embolization of hypervascular tumors and arteriovenous malformations (AVMs).

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 2 510(k) Summary

[As required by 21 CFR 807.92] Date Prepared: October 21, 2014 510(k) Number: ____K143038

#### Submitter's Name / Contact Person

Manufacturer Vascular Solutions, Inc. 6464 Sycamore Court North Minneapolis, MN 55369 USA Establishment Registration # 2134812 Contact Person Adam Ettl Sr. Regulatory Product Specialist Tel: 763-656-4300 Fax: 763-656-4253

#### General Information

| Trade Name          | Gel-Block 10x embolization pledgets                                  |
|---------------------|----------------------------------------------------------------------|
| Common / Usual Name | Device, vascular, for promoting embolization                         |
| Classification Name | 870.3300, KRD - vascular embolization device; Class II               |
| Predicate Device    | K113266 - Gel-Block embolization pledgets (Vascular Solutions, Inc.) |

#### Device Description

The Gel-Block 10x embolization pledgets is an embolic device consisting of 10 radially compressed porcine derived gelatin pledgets packaged in one transparent glass vial. A delivery syringe is provided for introduction of the pledgets into the delivery catheter (not included). The pledgets are available in three sizes compatible with a delivery catheter having a minimum inner diameter of ≥ 0.038", ≥ 0.027", or ≥ 0.021". When unconstrained in a blood vessel, each pledget will occlude blood flow by swelling to a specified swell size. The finished product is sterilized by electron-beam irradiation and is intended for single use only.

#### Intended Use / Indications

The Gel-Block 10x embolization pledgets is intended for use in embolization of hypervascular tumors and arteriovenous malformations (AVMs).

#### Technological Characteristics

The porcine derived gelatin pledgets in the Gel-Block 10x embolization pledgets and the Gel-Block embolization pledgets are identical. The only differences between the subject and predicate device are the package materials and configuration. The subject device contains 10 pledgets in a glass vial with a screw cap. while the predicate device contains two delivery assemblies each containing one pledget. The subject device does not include or require the use of the delivery assemblies as the pledgets are removed from the glass vial and placed directly in the tip of the delivery syringe. The subject and predicate devices each contain one delivery syringe, either 1cc or 3cc depending on the model. The

> Vascular Solutions, Inc. 6464 Sycamore Court North ◆ Minneapolis, MN ◆ 55369 Phone: 763-656-4300 ♦ Fax: 763-656-4253

{4}------------------------------------------------

subject device components (glass vial with screw cap, 10 pledgets, and one delivery syringe) are provided on a packing card and placed in a Tyvek pouch. The predicate device components (two individual pledgets packaged in separate delivery assemblies and one delivery syringe) are provided in a thermoformed tray with Tyvek lid that is placed in a Tyvek pouch. The subject and predicate devices are packaged in a retail box each containing five Tyvek pouches.

### Substantial Equivalence and Summary of Studies

The technological differences between the subject and predicate devices have been evaluated through biocompatibility and bench tests to provide evidence of substantial equivalence. The Gel-Block 10x embolization pledgets is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics, and indications for use. Therefore the only new design verification test necessary was deliverability. Package validation tests and sterilization validation tests were completed to support the new package configuration. Biocompatibility tests were required to assess the Gel-Block 10x embolization pledgets package materials.

The following biocompatibility tests were performed as recommended by ISO 10993-1:

- . Cytotoxicity
- Sensitization ●
- Irritation/Intracutaneous Reactivity ●
- Acute Systemic Toxicity o
- Hematology

The results of the deliverability, package, sterilization and biocompatibility tests met the specified acceptance criteria and did not raise new safety or performance issues. Therefore, the Gel-Block 10x embolization pledgets is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K143038](https://fda.innolitics.com/device/K143038)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
