K142940 · Titan Spine, LLC · OVE · Jan 28, 2015 · Orthopedic
Device Facts
Record ID
K142940
Device Name
Endoskeleton TCS System
Applicant
Titan Spine, LLC
Product Code
OVE · Orthopedic
Decision Date
Jan 28, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Endoskeleton® TCS System is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as neck pain with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment prior to treatment with the device is indicated to be used with autograft bone. The device is a stand-alone system when used with Endoskeleton® TCS integrated screws and when used without the integrated screws it requires additional supplemental fixation cleared in the cervical spine.
Device Story
Anterior cervical intervertebral body fusion device; titanium alloy (Ti-6Al-4V-ELD) implant; hollow center for autograft bone; acid-etched surfaces for bone fixation; integrated screw option for stand-alone use; requires supplemental fixation if screws not used. Implanted by surgeons in cervical spine; provides structural support and facilitates biologic fusion; windows allow radiographic visualization of graft and bone formation.
Clinical Evidence
Bench testing only. Mechanical testing performed per ASTM F2077 and ASTM F2267, including static/dynamic compression, compression-shear, torsion, subsidence, and expulsion testing.
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one level from C2 to T1, following 6 weeks of failed non-operative treatment. Used with autograft bone.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 28, 2015
Titan Spine, LLC % Ms. Christine Scifert Managing Partner Memphis Regulatory Consulting, LLC 3416 Roxee Run Cove Bartlett, Tennessee 38133
Re: K142940
Trade/Device Name: Endoskeleton® TCS System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE, ODP Dated: December 30, 2014 Received: December 31, 2014
Dear Ms. Scifert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
Page 2 - Ms. Christine Scifert
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Page 1 of 2
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K142940
Device Name Endoskeleton(R) TCS System
### Indications for Use (Describe)
The Endoskeleton® TCS System is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as neck pain with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment prior to treatment with the device is indicated to be used with autograft bone. The device is a stand-alone system when used with Endoskeleton® TCS integrated screws and when used without the integrated screws it requires additional supplemental fixation cleared in the cervical spine.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
Endoskeleton® System January 27, 2015
| Company: | Titan Spine, LLC<br>6140 West Executive Drive, Suite A<br>Mequon, WI 53092, USA |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration: | 3006340236 |
| Primary Contact: | Christine Scifert<br>Phone: 901-831-8053 |
| Company Contact: | Jane Rodd<br>Phone: 866-822-7800<br>Fax: 262-242-7802 |
| Trade Name: | Endoskeleton® TCS System |
| Common Name: | Intervertebral Body Fusion Device with Integrated Fusion Cervical<br>Intervertebral Fusion Device with Bone Graft, Cervical |
| Classification: | Class II |
| Regulation Number: | 21 CFR 888.3080 (Intervertebral body fusion device) |
| Panel: | 87- Orthopedic |
| Product Code: | OVE ODP |
| Predicate Devices: | Primary Predicate<br>Titan Spine Endoskeleton® TC (K100889 S.E. July 29, 2010) |
| | Additional Predicates<br>K2M Chesapeake Spinal System (K133494 S.E. June 10, 2014)<br>Surgicraft STALIFTM C (K072415 S.E. January 25, 2006)<br>Centinel Spine STALIF® (K120819 S.E. August 31, 2012)<br>LDR Spine Cervical Interbody Fusion System (K091088 S.E. July 14,<br>2009)<br>LDR Spine Cervical Interbody Fusion System – ROI-C Lordotic Implants<br>(K113559 S.E. December 29, 2011) |
{5}------------------------------------------------
## January 17, 2012) Medtronic PEEK PREVAIL™ Cervical Interbody Device (K073285 S.E. May 15, 2008)
### Device Description:
The Endoskeleton® TCS Interbody Fusion Device (IBD) implants are available in a variety of sizes with a variety of lordotic angles, to accommodate patient anatomy. Endoskeleton® TCS IBD implants are intended for treatment in Cervical Interbody Fusion procedures and are used as single placement devices in the disc space, and are designed with a large hollow region in the center to house autograft bone. The new bone formation through the implant is intended to provide long-term structural support and biologic fusion at the implanted disc space. The design incorporates "windows" through the implant to permit visualization of the graft material and over time formation of new bone. The superior and inferior surfaces are acid etched to improve fixation to the adjacent bone.
The device is a stand-alone system when used with Endoskeleton® TCS integrated screws and when used without the integrated screws it requires additional supplemental fixation cleared in the cervical spine.
An implant holding feature has been incorporated into the anterior surface of the implant to mate with the implant holder, and to facilitate placement of the interbody space. All implantable components are manufactured from medical grade titanium alloy (Ti-6Al-4V-ELD).
## Indications for Use:
The Endoskeleton® TCS System is an anterior cervical intervertebral body fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) at one disc level from C2 to T1. Patients should have received 6 weeks of non-operative treatment prior to treatment with the device is indicated to be used with autograft bone. The device is a stand-alone system when used with Endoskeleton® TCS integrated screws and when used without the integrated screws it requires additional supplemental fixation cleared in the cervical spine.
### Substantial Equivalence:
The subject Endoskeleton® TCS IBD is substantially equivalent to Titan Spine's previously cleared Endoskeleton® TC (K100889 S.E. July 29, 2010), K2M Chesapeake Spinal System (K133494 S.E. June 10, 2014), Surgicraft STALIF™ C (K072415 S.E. January 25, 2006), Centinel Spine STALIF® (K120819 S.E. August 31, 2012), LDR Spine Cervical Interbody Fusion System (K091088 S.E. July 14, 2009). LDR Spine Cervical Interbody Fusion System -ROI-C Lordotic Implants, (K113559 S.E. December 29, 2011), Medtronic PEEK PREVAIL® Cervical Interbody Device (K113252 S.E. January 17, 2012) and Medtronic PEEK PREVAIL™ Cervical Interbody Device (K073285 S.E. May 15, 2008).
{6}------------------------------------------------
## Performance Testing:
Mechanical testing, including static compression, static compression-shear, static torsion, subsidence, expulsion, dynamic compression, dynamic compression-shear, and dynamic torsion have been performed per ASTM F2077 and ASTM F2267 on the subject Endoskeleton® TCS IBD and predicate devices. The resulting data from testing demonstrated that the Endoskeleton® TCS IBD was substantially equivalent to the predicate interbody devices identified in the substantial equivalence section.
## Conclusion:
There are insignificant differences between the subject Endoskeleton® TCS IBD and the predicates. The Indications for Use, Materials, and Geometry for predicate devices are all inclusive of the subject device. The difference between the subject Endoskeleton® TCS IBD and the Endoskeleton® TC IBD is that the subject allows for use of screws for fixation and various dimensions. The difference between the subject and the additional predicate devices include dimensions. Testing shows that the Endoskeleton® TCS IBD performs equivalent to the predicate devices. Thus, it can be concluded that the subject does not raise new questions about safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.