K142933 · Biomet, Inc. · JWH · Dec 23, 2014 · Orthopedic
Device Facts
Record ID
K142933
Device Name
Biomet Tibial trays
Applicant
Biomet, Inc.
Product Code
JWH · Orthopedic
Decision Date
Dec 23, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Biomet Tibial Trays are modular components that are part of a total knee implant system intended to replace the damaged or diseased knee joint during total knee arthroplasty. Indications for use are as follows: 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved. 2. Correction of varus, valgus, or posttraumatic deformity. 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure. The Regenerex femoral augments are indicated for use with the Vanguard Total Knee System. The Regenerex tibial augments are indicated for use with standard and offset Biomet Tibial Trays. Femoral components and tibial tray components with porous coatings are indicated for cemented and uncemented biological fixation application. Non-coated (Interlok) devices and allpolyethylene patellar components are indicated for cemented application only. Regenerex components are intended only for uncemented biologic fixation application.
Device Story
Biomet Tibial Trays are modular orthopedic implants for total knee arthroplasty; used to replace damaged/diseased knee joints. Configurations include Interlok, porous, stemmed, and offset designs. System integrates with existing Biomet femoral components, tibial bearings, patella components, and augment blocks. Modification involves removing shot peen manufacturing process and introducing 'CementON Tibial Molds' for Interlok tray cementing. Implants provide structural support for knee joint; porous-coated versions allow for biological fixation; non-coated versions require cement. Device used by orthopedic surgeons in clinical/OR settings. Output is a physical prosthetic construct; clinical benefit is restoration of knee function and pain relief.
Clinical Evidence
No clinical data was required to establish substantial equivalence. Evidence consists of non-clinical bench testing, including tibial tray cyclic fatigue testing and pullout force testing for I-beam and finned implants comparing CementON versus manual cement application.
Technological Characteristics
Modular knee prosthesis components (Interlok, porous, stemmed, offset). Materials and design features remain unchanged from predicates. Modifications include removal of shot peen manufacturing step and addition of CementON Tibial Molds for cementing. Fixation via cemented or uncemented biological application depending on coating. Sterilization method unchanged.
Indications for Use
Indicated for patients with painful/disabled knee joints due to osteoarthritis, rheumatoid arthritis, or traumatic arthritis; patients requiring correction of varus, valgus, or posttraumatic deformity; and patients requiring revision of failed osteotomy, arthrodesis, or previous joint replacement.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features the department's emblem, which is a stylized representation of a human figure in profile, repeated three times and arranged in a way that suggests movement or progress. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 23, 2014
Biomet, Incorporated Mr. Jason Heckaman Project Manager, Regulatory Affairs 56 East Bell Drive Warsaw, Indiana 46581
Re: K142933 Trade/Device Name: Biomet Tibial Trays Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis. Regulatory Class: Class II Product Code: JWH, MBH, MBV Dated: October 27, 2014 Received: October 28, 2014
Dear Mr. Heckaman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and appear to be a dark gray or black color. The word is horizontally oriented and centered in the image.
### INDICATIONS FOR USE STATEMENT
510(k) Number (if known): _K142933_
Device Name: Biomet Tibial Trays
### INDICATIONS FOR USE:
- 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.
- 2. Correction of varus, valgus, or posttraumatic deformity.
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.
The Regenerex femoral augments are indicated for use with the Vanguard Total Knee System.
The Regenerex tibial augments are indicated for use with standard and offset Biomet Tibial Trays.
Femoral components and tibial tray components with porous coatings are indicated for cemented and uncemented biological fixation application. Non-coated (Interlok) devices and allpolyethylene patellar components are indicated for cemented application only.
Regenerex components are intended only for uncemented biologic fixation application.
Prescription Use |X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign Off) Division of Orthopedic Devices 510(k) Number: K142933
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and appear to be a dark gray or black color. The word is presented in a straightforward, unadorned manner, emphasizing its legibility and impact.
## 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Biomet Tibial Trays 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | Biomet Inc.<br>56 East Bell Drive<br>PO Box 587<br>Warsaw, IN 46581<br>Establishment Registration Number: 1825034 |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Jason Heckaman<br>Project Manager, Regulatory Affairs<br>Phone: 574-371-3707<br>Fax: 574-372-1683 |
| Date: | October 30, 2014 |
| Subject Device: | Trade Name: Biomet Tibial Trays<br>Common Name: Knee Prosthesis<br><br>Classification Name:<br>• JWH - prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer/metal/polymer (21 CFR §888.3560)<br>• MBH - prosthesis, knee, patello/femorotibial, semi-constrained,<br>uncemented, porous, coated, polymer/metal/polymer (21 CFR §888.3565)<br>• MBV-prosthesis, knee, patellofemorotibial, semi-constrained,<br>UHMWPE, pegged, cemented, polymer/metal/polymer (21 CFR §888.3560) |
Legally marketed devices to which substantial equivalence is claimed:
- K915132 MCK (Maximum Congruent Knee) System
- K010212 Offset Tibial Tray
- K033489 Biomet's Non-Cemented Porous Coated Knee Components
### Device Description
The subject Biomet Tibial Trays are available in four design configurations – Interlok, porous, stemmed, and offset. The trays, as a system construct, are compatible with existing, legally marketed Biomet femoral components, tibial bearings, patella components, stems, and various auxiliary components, including augment blocks and wedges.
This submission includes a modification to remove the shot peen operation from the manufacturing process of the subject Biomet Tibial Trays. In addition, a new cementing
{4}------------------------------------------------
technique option is being introduced for the Interlok tibial trays. The technique is executed via use of new implant-specific instrumentation referred to as CementON Tibial Molds.
### Intended Use and Indications for Use
Biomet Tibial Trays are modular components that are part of a total knee implant system intended to replace the damaged or diseased knee joint during total knee arthroplasty.
Indications for use are as follows:
- 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.
- 2. Correction of varus, valgus, or posttraumatic deformity.
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.
The Regenerex femoral augments are indicated for use with the Vanguard Total Knee System.
The Regenerex tibial augments are indicated for use with standard and offset Biomet Tibial Trays.
Femoral components and tibial tray components with porous coatings are indicated for cemented and uncemented biological fixation application. Non-coated (Interlok) devices and allpolyethylene patellar components are indicated for cemented application only.
Regenerex components are intended only for uncemented biologic fixation application.
### Summary of Technological Characteristics
The subject changes include removal of the shot peen operation from the manufacturing process and the addition of a new cementing technique option for the Interlok trays via use of new implant-specific instrumentation.
The subject modifications do not result in any changes to the intended use, indications for use, materials, design features, or sterilization of the Biomet Tibial Trays.
### Summary of Performance Data (Nonclinical and/or Clinical)
- . Non-Clinical Tests
- o Tibial Tray Cyclic Fatigue Testing
- I-beam Implant Pullout Force from a Cement Mantle after CementON or Manual o Cement Application
- Finned Implant Pullout Force from a Cement Mantle after CementON or Manual o Cement Application
- . Clinical Tests
- Clinical data was not required to establish substantial equivalence between the o subject Biomet Tibial Trays and the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The color of the text is a dark gray, and the background is white.
### Substantial Equivalence Conclusion
Based on the similarities in design, function, intended use and fundamental scientific technology, the devices that are the subject of this submission are similar to the predicate devices and do not introduce any new risks of safety or efficacy. Therefore, Biomet concludes that the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.