BP652N (HEM-6300-Z)

K142917 · Omron Healthcare, Inc. · DXN · Dec 1, 2014 · Cardiovascular

Device Facts

Record IDK142917
Device NameBP652N (HEM-6300-Z)
ApplicantOmron Healthcare, Inc.
Product CodeDXN · Cardiovascular
Decision DateDec 1, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 1/4 inches (13.5 cm to 21.5 cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

Device Story

Automatic non-invasive wrist blood pressure monitor; battery-powered (2 AAA). Device inflates wrist cuff via integral pump; deflates via electronic valve. Monitors cuff pressure; extracts pulse waveform data. Software analyzes waveform to determine systolic/diastolic pressure and pulse rate. Includes memory function (100 measurements); calculates average of last three. Features Advanced Positioning Sensor (APS) using an accelerometer to provide visual feedback on cuff height relative to heart. Used by home users. Output displayed on LCD. Helps users monitor hypertension; irregular heartbeat detection provides warning signal.

Clinical Evidence

Clinical accuracy validated in accordance with ANSI/AAMI/ISO 81060-2:2013. Bench testing included verification testing, IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and AAMI SP10 compliance. Patient-contacting materials tested per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity).

Technological Characteristics

Oscillometric measurement; piezo resistance pressure sensor; piezo-electric pump; electronic control valve. Powered by 2 AAA batteries. Wrist cuff. APS uses accelerometer for positioning feedback. Complies with IEC 60601-1, IEC 60601-1-2, and AAMI SP10. Patient-contacting materials tested per ISO 10993-1.

Indications for Use

Indicated for adult patients with wrist circumference 13.5 cm to 21.5 cm for noninvasive blood pressure and pulse rate measurement in home settings. Detects irregular heartbeats.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three abstract human profiles facing to the right, stacked on top of each other. The profiles are depicted in a flowing, wave-like design. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 1, 2014 Omron Healthcare, Inc. % Mr. Paul Dryden Consultant ProMedic, Inc. 24301 Woodsage Drive Bonita Springs, Florida 34134-2958 Re: K142917 > Trade/Device Name: Omron BP652N (HEM-6300-Z) with APS Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: October 2, 2014 Received: October 7, 2014 Dear Mr. Paul Dryden, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {2}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Melissa A. Torres -S For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ K142917 Page __1__ of__1_ 510(k) Number (if known): __ K142917 Device Name : __ Omron BP652N (HEM-6300-Z) with APS_____ # Indications For Use: The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 1/4 inches (13.5 cm to 21.5 cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. Environments of Use: Home Patient Population: Adult # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The-Counter Use {4}------------------------------------------------ ## 510(k) Summary Page 1 of 7 10/2/2014 | Omron Healthcare, Inc. | | |-----------------------------------|-------------------------------------------------------------------------| | 1925 West Field Court | Tel - 847-247-5626 | | Lake Forest, IL 60045 USA | Fax- 847-680-5626 | | Official Contact: | Renee Thornborough – Director QA/RA | | Proprietary or Trade Name: | Model BP652N(HEM-6300-Z) with APS | | Common/Usual Name: | Noninvasive blood pressure measurement system | | Classification Name/Code: system. | DXN - Noninvasive blood pressure measurement | | | 21CFR 870.1130 | | | Class II | | Device: | Model BP652N(HEM-6300-Z) with APS | | Predicate Device: | Model BP652N (HEM-6300-Z) K123498<br>Braun Series BP 200 Series K003732 | | Device Description: | | This device originally received clearance under 510(k) K123498. The device is an automatic non-invasive blood pressure system. The device is battery powered by 2 "AAA" batteries, there is no connection to external power. The device inflates a wrist cuff with an integral pump, then deflates the cuff via an electronically controllable valve. During inflation the cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic pressure. The device has a memory function that automatically stores up to 100 of the latest measurements. It can also display an average of the last three values The device also detects the appearance of irregular heartbeats during measurement. As originally submitted under 510(k) K123498, the device contained the APS feature. The feature was removed prior to clearance. This submission is for the addition of the Advanced Positioning Sensor (APS) into the device. The APS is an aid to help the user determine if the cuff is at the correct height in relationship to the heart. The addition of APS does not change the intended use of the device. The addition of APS does not alter the fundamental scientific technology. There are no changes in hardware. There are no changes in the software algorithm which {5}------------------------------------------------ # 510(k) Summary Page 2 of 7 10/2/2014 determines blood pressure or pulse rate. The addition of the APS feature is an optional feature which the user can elect to use. The APS feature does not alter the accuracy of the device, but serve to assist the user in positioning the device at the level of the heart. Though there is no change in algorithm a clinical study in accordance with ANSI/AAMI/ISO 81060-2:2013 was performed to validate the clinical accuracy of the device when the arm is positioned by use of the APS, see Section 20. # Description of APS: The Advanced Positioning Sensor is an aid to help the user determine if the cuff is at the correct height in relation to the heart. It makes this determination based on the reading of an accelerometer (to measure the angle of the arm in relation to the table) integral to the device. The APS has no interface to the software algorithm which determines blood pressure or pulse rate, it is a visual feedback to the user indicating whether the device is positioned appropriately in accordance with the APS angle measurement criteria as below. Image /page/4/Picture/7 description: The image shows three Omron blood pressure monitors on wrists, demonstrating correct and incorrect positioning. The first monitor, labeled "Too Low," displays a reading with the indicator light positioned below the heart symbol. The second monitor, labeled "Correct," shows the indicator light aligned with the heart symbol. The third monitor, labeled "Too High," shows the indicator light above the heart symbol. As above, there are no changes to the hardware, blood pressure or pulse rate algorithms. Intended User Home user # Patient Population This device is intended for use on adults. {6}------------------------------------------------ #### 510(k) Summary Page 3 of 7 10/2/2014 # Indications for Use: The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5 1/4 inches to 8 1/2 inches (13.5 cm to 21.5 cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. Environment of Use: Home # Contraindications There are no known contraindications. ## Predicate Device Comparison The BP652N(HEM-6300-Z) with APS was compared to the predicates Model BP652N (HEM-6300-Z) K123498 and Braun Series BP 200 Series K003732 as in the device comparison table below. {7}------------------------------------------------ #### 510(k) Summary Page 4 of 7 10/2/2014 # Device Comparison | | PREDICATE | PREDICATE | Device Under Review | Comment | |-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------| | | Omron BP652N (HEM-6300-Z)<br>K123498 | Braun BP2000 series K003732 | Omron BP652N (HEM-6300-Z) with<br>APS | | | Indications for Use | The device is a digital monitor<br>intended for use in measuring blood<br>pressure and pulse rate in adult<br>patient population with wrist<br>circumference ranging from 5 1/4<br>inches to 8 1/2 inches (13.5 cm to<br>21.5 cm).<br>The device detects the appearance of<br>irregular heartbeats during<br>measurement and gives a warning<br>signal with readings. | The Braun PrecisionSensor™ (BP2000<br>series) wrist blood pressure monitor is<br>indicated for use for the noninvasive<br>measurement of blood pressure (systolic<br>and diastolic} and pulse rate in adults, in<br>a home use setting. Use may be initiated<br>by the individual or as part of a<br>hypertension monitoring and<br>management program supervised by a<br>health care provider. | The device is a digital monitor intended<br>for use in measuring blood pressure and<br>pulse rate in adult patient population<br>with wrist circumference ranging from 5<br>1/4 inches to 8 1/2 inches (13.5 cm to<br>21.5 cm).<br>The device detects the appearance of<br>irregular heartbeats during measurement<br>and gives a warning signal with<br>readings. | Identical K123498 | | Patient Population | Adult | Adult | Adult | Identical K123498 and K003732 | | Environment of<br>Use | Home | Home | Home | Identical K123498 and K003732 | | Prescriptive | OTC | OTC | OTC | Identical K123498 and K003732 | | Patient Connection | Yes via cuff | Yes via cuff | Yes via cuff | Identical K123498 and K003732 | | Technology | Oscillometric | Oscillometric | Oscillometric | Identical K123498 and K003732 | | Advanced<br>Positioning Sensor | - | Device has position Sensor | Added* | Identical in function to K003732<br>Braun BP2000 series | | Measurement<br>range | Pressure: 0-299 mmHg<br>Pulse rate: 40 to 180 bpm | Pressure: 0-300 mmHg<br>Pulse rate: 40 to 160 bpm | Pressure: 0-299 mmHg<br>Pulse rate: 40 to 180 bpm | Identical K123498 | | Accuracy of<br>pressure indicator | +/- 3 mmHg or 2% of reading | +/- 3 mmHg | +/- 3 mmHg or 2% of reading | Identical K123498 | | Pressure sensor | Piezo resistance sensor | Not specified | Piezo resistance sensor | Identical K123498 | | Accuracy Pulse<br>Rate | +/-5% | +/-5% | +/-5% | Identical K123498 and K003732 | | Inflation Method | Piezo-electric pump | Not specified | Piezo-electric pump | Identical K123498 | | Deflation Method | Internal valve | Electronic control valve | Internal valve | Identical K123498 and K003732 | | Display Type | LCD | LCD | LCD | Identical K123498 and K003732 | | Irregular pulse<br>detection | Yes | No | Yes | Identical K123498 | | Power Source | AAA batteries | AAA batteries | AAA batteries | Identical K123498 and K003732 | | Operating<br>Conditions | Temperature: 10° to 40° C<br>Humidity: 15 to 85% RH | Temperature: 10° to 40° C<br>Humidity: <85% | Temperature: 10° to 40° C<br>Humidity: 15 to 85% RH | Identical K123498 | | Storage Conditions | Temperature: -20° to 60° C<br>Humidity: 10 to 95% RH | Temperature: -20° to 60° C<br>Humidity: <85% | Temperature: -20° to 60° C<br>Humidity: 10 to 95% RH | Identical K123498 | | Dimensions | 89(W) x 61(D) x 13(H) mm | Not specified | 89(W) x 61(D) x 13(H) mm | Identical K123498 | | Weight | 80g | Not specified | 80g | Identical K123498 | {8}------------------------------------------------ #### 510(k) Summary Page 5 of 7 10/2/2014 By "added" we mean that the feature was re-enabled thus restoring it to the state it was in when originally submitted under 510(k) K123498 . {9}------------------------------------------------ ## 510(k) Summary Page 6 of 7 10/2/2014 # Differences Between Other Legally Marketed Predicate Devices The Omron BP652N (HEM-6300-Z) with APS is viewed as substantially equivalent to the predicate devices because: The BP652N (HEM-6300-Z) with APS uses the exact same technology and has identical indications for use. The differences that exist between the devices are insignificant in the terms of safety or effectiveness. # Indications - The indications for use are identical. Prescriptive - The BP652N (HEM-6300-Z) with APS and predicates are OTC. Design and Technology - The BP652N (HEM-6300-Z) with APS has equivalent design and technology when compared to the predicates. Performance and Specifications - The BP652N (HEM-6300-Z) with APS has equivalent specifications of performance compared to the predicates. Compliance with standards - The BP652N (HEM-6300-Z) with APS and predicate device (Omron BP652N (HEM-6300-Z) K123498) declare compliance with the identical standards: SP10, IEC 60601-1 and IEC 60601-1-2. # Materials - The patient contacting materials of the cuffs has been tested in accordance with ISO 10993-1 and FDA Guidance. The tests included Cytotoxicity, Sensitization, and Intracutaneous Reactivity and are identical to Omron BP652N (HEM-6300-Z) K123498 . Section 15 contains a Materials Certification attesting to this. ## Patient Population - The BP652N (HEM-6300-Z) with APS and predicates are indicated for adults ## Non-Clinical Testing Summary: We have performed bench tests and found that the BP652N (HEM-6300-Z) with APS met all requirements specifications and standards requirements and were found to be equivalent in comparison to the predicate. Testing includes the following: - Verification Testing - Testing for compliance to IEC 60601-1 ● - Testing for compliance to IEC 60601-1-2 - Testing for compliance to AAMI SP10 ● {10}------------------------------------------------ ## 510(k) Summary Page 7 of 7 10/2/2014 Testing to insure clinical accuracy of the device in accordance with ANSI/AAMI/ISO 81060-2: 2013 is documented in Section 20. # Substantial Equivalence Conclusion Omron maintains that the BP652N (HEM-6300-Z) with APS is substantially equivalent to the predicate devices Model BP652N (HEM-6300-Z) K123498 and Braun Series BP 200 Series K003732 in indications for use, patient population, and environment for use, technology characteristics, specifications / performance and compliance with international standards
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...