K142912 · Merits Health Products Co., Ltd. · INI · Jun 23, 2015 · Physical Medicine
Device Facts
Record ID
K142912
Device Name
Merits Yoga Power Scooter, S542
Applicant
Merits Health Products Co., Ltd.
Product Code
INI · Physical Medicine
Decision Date
Jun 23, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 890.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Merits Yoga Power Scooter, S542, is to provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls.
Device Story
Battery-powered, three-wheeled motorized scooter; rear transaxle drive; controlled via PG S-Drive 45amp controller. User interface consists of wig-wag control lever and speed knob on tiller. Input: user manipulation of wig-wag lever. Transformation: controller releases solenoid electromechanical brakes (normally closed) to allow movement; releasing lever engages brakes for stop. Output: physical mobility. Used in indoor/outdoor environments on dry, level surfaces (concrete, asphalt). Operated by patient. Benefits: provides personal mobility for individuals with limited physical capacity; foldable frame for storage/transport.
Clinical Evidence
Bench testing only. No clinical data. Device performance validated against ISO 7176 series standards, including static/dynamic stability, braking effectiveness, obstacle climbing, electromagnetic compatibility, and fatigue strength. Lithium battery safety verified via UN 38.3 testing.
Technological Characteristics
Welded aluminum frame; 24V/11.5Ah Li-ion battery; PG S-Drive 45amp controller; solenoid electromechanical brakes (normally closed); front leaf spring suspension; PU tires. Complies with ISO 7176 series (stability, braking, dimensions, fatigue, EMC) and ISO 14971 (risk management). Upholstery meets ISO 7176-16:2012 ignition resistance.
Indications for Use
Indicated for adult persons with limited mobility who are restricted to a sitting position and possess the capability to operate simple controls.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
Predicate Devices
Travel Ease Electric Scooter, Model Pioneer 5 (K013788)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 23, 2015
Merits Health Products Co., Ltd. Vincent Chen, Manager No.18 Jingke Rd. Nantun District Taichung, 40852 Taiwan R.O.C.
Re: K142912
> Trade/Device Name: Merits Yoga Power Scooter, S542 Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: May 26, 2015 Received: June 1, 2015
Dear Mr. Chen,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Felipe Aquel -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142912
Device Name Merits Yoga Power Scooter, S542
Indications for Use (Describe)
The Merits Yoga Power Scooter, S542, is to provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
Merits Health Products Co., LTD. 510(k) Premarket Notification
#### Submitter:
Merits Health Products Co., LTD. No.18, Jingke Rd., Nantun Dist. Taichung City 40852, Taiwan (R.O.C.)
#### Contact Person:
Vincent Chen Merits Health Products Co., LTD. No.18, Jingke Rd., Nantun Dist., Taichung City 40852, Taiwan (R.O.C.) Phone: +886-4-23594985 ext.200 Fax: +886-4-23594992
#### Date Prepared:
Dec. 18, 2014
#### Proprietary Name:
Merits Yoga Power Scooter, S542
#### Common name:
Three Wheel Power Scooter
#### Classification name:
Physical Medicine / Motorized Three - Wheeled Vehicle
#### Product Code:
INI
# Regulation number:
890.3800
#### Comparison to Predicate Devices:
This submission indicates the Substantial Equivalence of the Merits Yoga Power Scooter, S542, with the predicate Travel Ease Electric Scooter, Model Pioneer 5 (K013788). S542 has the same intended uses and similar indications, technological characteristics and principles of operation with predicate device.
{4}------------------------------------------------
#### Device Description
The Merits Yoga Power Scooter, S542, is battery powered, rear transaxle driven and is controlled by the PG S-Drive 45amp controller. The user interface is a wig-wag control lever and a speed knob on the control panel. S542 is powered by one 12 VDC 11.5Ah battery. The batteries are charged by 4.3A off-board charger connect with 3-pin Microphone Connector to charging socket on the rear shroud. The approximate driving range on fully charged batteries is up to 18km (11.3mi). The scooter frame is a welded aluminum construction and includes 2 rear wheels with drive units (including motor, gear box, and brake), a removable battery pack, a front wheel set which consists of 2 wheels for better stability and a tiller to control its turning direction. When the user activates the wig-wag control lever, the controller receives a signal to release the brakes. With the brakes released, the scooter is allowed to move forwards or backwards. When the user releases the control lever, the scooter slows to a stop and the brakes are automatically re-engaged. The solenoid electromechanical brakes allow the user stop by letting go of the control lever. The frame is designed to be foldable for easy storage and transportation.
For safety reason, this solenoid electromechanical brake is designed as normal close type. In case of the brake or electrical failure, the brake shoe will re-engage automatically to prevent the movement of the scooter. The upholstery of the device complies with ISO 7176-16:2012 Resistance to ignition of postural support devices.
The device can be operated on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions.
The Merits Yoga Power Scooter. S542, is substantially equivalent to the Merits Travel Ease Electric Scooter, Model Pioneer 5 (K013788). Both products are battery power, motorized vehicles designed for use as personal power mobility aids. Performance characteristics and drive mechanisms are similar and all have the same intended function and use which is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered vehicle. Additional, they are all constructed from the same basic materials, have the same basic operational principles and all use DC batteries as their source of power.
Although there are some minor differences between S542 and its predicate device. But they raise no new issues of safety or effectiveness. Performance data demonstrate that S542 is safe. The non-clinical testing and the predicate comparisons demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness.
#### Intended Use
The Merits Yoga Power Scooter, S542, is to provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls.
{5}------------------------------------------------
#### Discussion of Non-clinical Tests Performed for Determinations of Substantial equivalence are as follows:
- ISO 7176-1:1999 Determination of Static Stability
- ISO 7176-2:2001 Determination of Dynamic Stability of electric wheelchairs
- ISO 7176-3:2012 Determination of effectiveness of brakes
- Energy consumption of electric wheelchairs and scooters for ISO 7176-4:2008 determination of theoretical distance range
- Determination of overall dimensions, mass and maneuvering ISO 7176-5:2008 space
- Determination of maximum speed, acceleration and ISO 7176-6:2001 deceleration of electric wheelchairs
- ISO 7176-7:1998 Method of Measurement of Seating and Wheel Dimensions
- Requirements and test methods for static, impact and fatigue ISO 7176-8:1998 strengths
- ISO 7176-9:2009 Climatic tests for wheelchairs
- Determination of obstacle-climbing ability of electrically power ISO 7176-10:2008 wheelchairs
- ISO 7176-11:2012 Test dummies
- ISO 7176-13:1989 Determination of coefficient of friction of test surfaces
- Power and control systems for electrically powered ISO 7176-14:2008 wheelchairs and scooter- Requirements and test methods.
- Requirements for Information Disclosure, Documentation and ISO 7176-15:1996 Labeling
- ISO 7176-16:2012 Resistance to ignition of postural support devices
- requirements and test methods for electromagnetic ISO 7176-21:2009 compatibility of electrically powered wheelchairs and scooters, and battery chargers.
- Medical devices -- Application of risk management to medical ISO 14971:2007 devices
The results of the testing confirm that the device meets specifications and is substantially equivalent to the predicate device.
{6}------------------------------------------------
### Substantial Equivalence Discussion
The Merits S542 Scooter is substantially equivalent to the Merits Travel Ease Electric Scooter, Model Pioneer 5 (K013788). Both products are battery power, motorized, seating systems designed for use with electric scooters. Performance characteristics and drive mechanisms are similar and all have the same intended function and use which is to provide mobility to adult persons limited to a sitting position and capable to operate a few simple controls. Additional, they are all constructed from the same basic materials, have the same basic operational principles and all use DC batteries as their source of power. The comparison table is as follow:
| Characteristics | | Merits Yoga Power Scooter,<br>S542 | Merits Travel Ease Electric<br>Scooter, Model Pioneer 5 |
|----------------------------|---------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Intended Use | | To provide mobility to adult<br>persons limited to a sitting<br>position and capable to operate<br>a few simple controls. | To provide mobility to adult<br>persons limited to a sitting<br>position and capable to operate<br>a few simple controls. |
| Dimension | Length | 39" (±0.6") / 20.2" (±0.6") | 38.5" |
| | Width | 17.2" (±0.6") | 19.7" |
| | Height | 39.6" (±0.6") / 24.2" (±0.6") | 48" |
| Weight | Total | 56.7 lb | 92 lb |
| | Base | 53 lb (Seat Not Removable) | 50 lb |
| | Seat | | 19 lb |
| | Battery | 3.7lb | 23lb |
| Seat Height (From<br>Deck) | | 15.6" | 16" |
| Seat Width | | 14.2" | 14" |
| Seat Depth | | 12.4" | 12.5" |
| Armrest Type | | Removable & Flip Backward | Removable & Flip Backward |
| Weight Capacity | | 250 lb | 250 lb |
| Gradient | | 6° | 6° |
| Turning Radius | | 37.8" | 33.8" |
| Ground Clearance | | 3.7" | 2.4" |
| Tires | | Front: 6"/ Rear: 7" (PU Tires) | Front: 8"/ Rear: 8" (PU Tires) |
| Suspension | | Front Wheel Leaf Spring | No Suspension |
| Max. Speed | | 4 mph | 4.5 mph |
| Range up to | | 11.25 miles | 10 miles |
| Power | | 24V/11.5Ah Li-ion Battery | 24V/12Ah Lead Acid Battery |
There are some difference between S542 and its predicate device raise:
- a. Turning Radius: The device passed ISO 7176-2:2001(Determination of Dynamic Stability of electric wheelchairs): Item 10.3 and 10.4. So there is no deleterious affection of safety and effectiveness about the difference on Turning Radius with predicated device.
- b. Dimension: The device passed ISO 7176-1:1999 (Determination of Static Stability) and ISO 7176-2:2001(Determination of Dynamic Stability of electric wheelchairs). So there is no deleterious affection of safety and effectiveness about the difference on Dimension with predicated device.
- c. Range up to: The travel range of S542 is better than the predicated device.
- d. Max. Speed: The device has passed ISO 7176-2:2001(Determination of Dynamic Stability of electric wheelchairs) & ISO 7176-8 (Requirements and test methods for static, impact and fatigue strengths) test (Please see Appendix IV: Test reports). So there is no deleterious affection of safety and
{7}------------------------------------------------
effectiveness about the difference on Max. Speed with predicated device.
- e. Weight: Although the total weight of S542 is lighter than Merits Travel Ease Electric Scooter, Model Pioneer 5 (K013788). The device has passed ISO 7176-8 (Requirements and test methods for static, impact and fatigue strengths) test (Please see Appendix IV: Test reports). So there is no deleterious affection of safety and effectiveness about the difference on Weight with predicated device. As the S542 is lighter than the predicate device, it is more suitable for user to carry and transport by vehicles.
- f. Battery: The S542 is using Li-ion Battery. It is lighter than the conventional Lead Acid Batteries. Although the capacity of the new battery is slightly smaller than the predicated device. Its travel rang is longer than the predicated device. The device has passed ISO 7176-8 (Requirements and test methods for static, impact and fatigue strengths) test (Please see Appendix IV: Test reports). And the Li-ion Battery has passed the UN 38.3 Transportation Testing Required for Lithium Battery Safety during Shipping to ensure its safety. (Please see Appendix X: Part I Test reports). So there is no deleterious affection of safety and effectiveness about the difference on Battery with predicated device.
- g. The seat height, width and depth for both devices are almost the same. They are all equipped with Removable and Flip Backward Armrest for users to leave the seats easily and more compact for storage.
The front wheel set of S542 is equipped with a leaf spring suspension. This makes the S542 more comfortable than the predicated device. The device has passed ISO 7176-2:2001(Determination of Dynamic Stability of electric wheelchairs) & ISO 7176-8 (Requirements and test methods for static, impact and fatigue strengths) test (Please see Appendix IV: Test reports). So there is no deleterious affection of safety and effectiveness about the difference on this suspension with the predicated device.
#### Conclusions
According to comparison table and discussions above, the differences on total weight, speed, battery size, suspension type and dimensions do not deleteriously affect the safety and effectiveness of the device.
So based on the design, performance specifications and testing and intended use, the Merits Yoga Power Scooter, S542, is substantially equivalent to the Merits Travel Ease Electric Scooter, Model Pioneer 5 (K013788).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.