← Product Code [DQK](/productcode/DQK) · K142901

# RhythmView Workstation (K142901)

_Topera, Inc. · DQK · Dec 19, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K142901

## Device Facts

- **Applicant:** Topera, Inc.
- **Product Code:** [DQK](/productcode/DQK.md)
- **Decision Date:** Dec 19, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1425
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Rhythm View Workstation is a computerized system that assists in the diagnosis of complex cardiac arrhythmias. The Rhythm View Workstation is used to analyze electrogram and electrocardiogram signals and display them in a visual format.

## Device Story

RhythmView Workstation is a computerized diagnostic system for complex cardiac arrhythmias; used by physicians during electrophysiological studies. Input: electrical signals from multi-polar electrophysiology catheters. Operation: proprietary algorithms compute spatial organization of heart rhythm disorders; identifies electrical rotors or focal beat sources. Output: interactive graphic displays of electrical activity, including panoramic animations of atrial and ventricular maps. Hardware includes cart, monitor, keyboard, mouse, RFID reader/writer, and USB switch. Physician reviews animated graphic representations of signal potentials (depolarization/repolarization) to aid diagnosis. Benefits: assists in identifying sources of arrhythmias to guide clinical decision-making.

## Clinical Evidence

Bench testing only. Performance data included software verification, hardware qualification (RFID, USB switch, SSD), and EMC testing. No clinical trial data presented.

## Technological Characteristics

Software-driven workstation; C++ programming language. Components: cart, monitor, keyboard, mouse, RFID reader/writer, two-port USB switch, solid-state hard drive. Signal processing: electrogram/ECG analysis. Connectivity: USB data transfer. Display modes: Electrical Activity, Contours, DContours, Rotational Activity, Profile. Sterilization: N/A.

## Regulatory Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

## Predicate Devices

- RhythmView ([K130827](/device/K130827.md))
- RhythmView ([K123295](/device/K123295.md))
- RhythmView ([K110878](/device/K110878.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines beneath them. The profiles are black, and the text is also in black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 19, 2014

Topera, Inc. Dennis Pozzo Senior Regulatory Affairs Master Specialist 3668 S. Gever Road. Suite 365 St. Louis, Missouri 63127

K142901 Trade/Device Name: Rhythmview Workstation Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: November 24, 2014 Received: November 25, 2014

Dear Dennis Pozzo,

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Melissa A. Torres -S

For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

K142901

Device Name

RhythmView Workstation

Indications for Use (Describe)

The Rhythm View Workstation is a computerized system that assists in the diagnosis of complex cardiac arrhythmias. The Rhythm View Workstation is used to analyze electrogram and electrocardiogram signals and display them in a visual format.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Appendix 1: 510(k) Summary per 21CFR §807.92

| Submitter's information    | Topera, Inc.<br>3668 S. Geyer Road, Suite 365<br>St. Louis, MO 63127<br>Contact: Dennis Pozzo<br>Phone 314-300-6580<br>Date: May XX, 2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |  |  |  |  |  |  |  |  |  |  |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|
| Device/classification name | Device Name: RhythmView™ Workstation Classification/Common name: Programmable diagnostic computer The marketed device(s) to which substantial equivalence is claimed:  K130827, cleared Dec. 16, 2013 K123295, cleared April 24, 2013 K110878, cleared Sept. 23, 2011                                                                                                                                                                                                                                                                                                                                                                     |  |  |  |  |  |  |  |  |  |  |
| Device description         | The RhythmView™ Workstation is comprised of the following components:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |  |  |  |  |  |  |  |  |  |  |
|                            | Cart Keyboard   Monitor/Display Mouse   Computer Software   Radio-Frequency Identification (RFID) Reader/Writer (new component) Two Port USB Switch (new component)   Solid State Hard Drive (new optional component)                                                                                                                                                                                                                                                                                                                                                                                                                     |  |  |  |  |  |  |  |  |  |  |
|                            | RhythmView™ takes electrical signals collected from multi-polar electrophysiology catheters and outputs a graphic display that assists in the diagnosis of cardiac arrhythmias.<br>The RhythmView™ computes and displays electrical rotors or focal beat sources responsible for maintaining human heart rhythm disorders including focal AT, AFL, other SVT, AF, VT and VF in a given patient. The product takes as input electrical signals recorded during the heart rhythm disorder under consideration, typically from multiple specified locations within the heart during an electrophysiological study. The RhythmView™ then uses |  |  |  |  |  |  |  |  |  |  |

proprietary patented algorithms and methods to compute spatial organization during the heart rhythm disorder. These computed elements are displayed graphically in interactive form for review to aid diagnosis by the physician during an electrophysiology study.

Continued on next page

{4}------------------------------------------------

# Appendix 1: 510(k) Summary per 21CFR §807.92, Continued

| Indications for use              | The RhythmView™ Workstation is a computerized system that assists in the<br>diagnosis of complex cardiac arrhythmias. The RhythmView™ Workstation<br>is used to analyze electrogram and electrocardiogram signals and display<br>them in a visual format.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>characteristics | Both the proposed and predicate RhythmView™ Workstations are software<br>driven devices that translate electrical signals within the heart into graphic<br>representations in order to assist the physician in diagnosis. They also allow<br>the user to:<br>Review and select a time sequence of electrical signals from various<br>electrodes; Analyze the signals; View a graphic display (Electrical Activity) of the signal potentials<br>showing progressive depolarization and repolarization in grayscale for<br>the particular arrhythmia; Play/Replay the animated graphic representation of electrical signals.              Furthermore,  The ventricular maps appear in the panoramic animation display in the<br>same manner as atrial maps. All of the same display options available in atrial maps may be used<br>for ventricular maps. |

Continued on next page

{5}------------------------------------------------

### Appendix 1: 510(k) Summary per 21CFR §807.92, Continued

| Device Characteristic                                                  | Predicate Device<br>RhythmView™ 4.1                                                         | Proposed Device<br>RhythmView™ 4.2                                                           |
|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Signal processing                                                      | Yes                                                                                         | Yes                                                                                          |
| Post-processing display                                                | Yes                                                                                         | Yes                                                                                          |
| Grid display of electrode signals                                      | Yes                                                                                         | Yes                                                                                          |
| Programming Language                                                   | C++                                                                                         | C++                                                                                          |
| Export of processed file into video<br>format                          | Yes                                                                                         | Yes                                                                                          |
| Manual tagging by user of<br>electrograms                              | No                                                                                          | No                                                                                           |
| OTS Software requirements                                              | Same                                                                                        | Same                                                                                         |
| Display options for review of<br>processed signals                     | • Electrical Activity<br>• Contours Only<br>• DContours<br>• Rotational Activity<br>Profile | • Electrical Activity<br>• Contours Only<br>• DContours<br>• Rotational Activity<br>Profile" |
| RFID Reader/Writer Function                                            | No                                                                                          | Yes                                                                                          |
| Data transfer via Two Port Switch                                      | No                                                                                          | Yes                                                                                          |
| Direct data transfer via USB cable<br>to RV Workstation from EP system | No                                                                                          | Option available                                                                             |
| Atrial Function                                                        | Yes                                                                                         | Yes                                                                                          |
| Ventricular Function                                                   | No                                                                                          | Option available                                                                             |

The Software, Hardware (RFID, Two Port USB Switch and SSD Performance data Qualification Testing and EMC Testing) and User testing that have been performed provide reasonable assurance that the proposed device has been tested to verify conformance to requirements for its intended use. Therefore, it has been demonstrated that the Rhythm View™ Workstation is safe and effective for its intended use.

---

**Source:** [https://fda.innolitics.com/device/K142901](https://fda.innolitics.com/device/K142901)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
