← Product Code [FTM](/productcode/FTM) · K142887

# SIS Inguinal Hernia Repair Graft (K142887)

_Cook Biotech Incorprated · FTM · Oct 1, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K142887

## Device Facts

- **Applicant:** Cook Biotech Incorprated
- **Product Code:** [FTM](/productcode/FTM.md)
- **Decision Date:** Oct 1, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The SIS Inguinal Hernia Repair Graft is intended for implantation to reinforce soft tissues where weakness exists, including the repair of inguinal hernias. The graft is supplied sterile and is intended for one-time use.

## Device Story

SIS Inguinal Hernia Repair Graft is a flat, bioabsorbable sheet derived from porcine Small Intestinal Submucosa (SIS); contains collagens I, III, IV, and VI. Used by surgeons in clinical settings for open, tensionless repair of inguinal hernias; implanted to reinforce soft tissue defects. Device provides mechanical support during healing; resorbs over time. Clinical outcomes include hernia recurrence rates and post-operative pain assessment compared to synthetic polypropylene mesh. Benefits include potential for reduced chronic pain compared to synthetic alternatives. Packaged in dried state, sterile, single-use.

## Clinical Evidence

Clinical evidence includes 3 peer-reviewed studies (95 patients total). One US randomized trial (n=100) compared SIS graft to polypropylene (PP) mesh in Lichtenstein repair; 1-year recurrence: 6.7% (SIS) vs 0% (PP); 3-year recurrence: 15.6% (SIS) vs 4% (PP). Two OUS randomized trials showed 0% recurrence for SIS at 1 and 3 years. Studies indicate SIS graft is effective for tensionless repair, with some evidence of lower post-operative pain compared to PP mesh. Recurrence rates may be higher in obese patients (BMI >30).

## Technological Characteristics

Material: Porcine Small Intestinal Submucosa (SIS) extracellular matrix containing collagens I, III, IV, VI. Form factor: Flat sheet, rectangular or pre-shaped (various sizes 6x10 cm to 13x18 cm). Thickness: 0.1-1.0 mm. Sterilization: Ethylene Oxide (EtO). Biocompatibility: ISO 10993-1 compliant. Mechanical properties: Suture retention, burst, and tensile strength tested.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- SurgiMend Collagen Matrix ([K083898](/device/K083898.md))
- SIS Hernia Repair Device ([K062697](/device/K062697.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other. The symbol is black and has a stylized, flowing design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 1, 2015

Cook Biotech Incorporated Mr. Perry W. Guinn Vice President/Quality Assurance and Regulatory Affairs 1425 Innovation Place West Lafayette, Indiana 47906

Re: K142887

Trade/Device Name: SIS Inguinal Hernia Repair Graft Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM Dated: September 10, 2015 Received: September 11, 2015

Dear Mr. Guinn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR [SELECT ONE: Part 801 [or, for IVDs only] Parts 801 and

{1}------------------------------------------------

809]); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

| 04/9-£ÞÞ (10€) ຂອງທ່ານຈຽ (301) 443-974<br>てーレ<br>ਜੂੜ                                      | ા ૧૦ । રાજ્યન                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | משלה (1888 משה במקום המועד המודד המודד המורד של המוזיקה במשמעות המושב המושב המושב של המו |
|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
|                                                                                           | to uoilsellos e  ot buodser of beninber ton si nosted e business to toubuco jou year your your your your you<br>" чәдини ЯМО рікел үрдә ипә е гүеldsib זו ггелип чоңгеттолгі                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                          |
|                                                                                           | of ,nebud sint guisuber 107 suoiteeppus guibuloni ,noitselloo noitemnonni sint to<br>əsindən bir bəhri bir nəhrindən bura və ya və məni çəkildir. Bir mənilər və mənisələr mənisələr mənisələr mən<br>əd birinci ilə sənəqsən təq suna çəkin və bir növünün in noitəlif və növü. İstinadlar və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və<br>. MOJES SSARIOA JIAMS FAATS ASS SHT OT MAOS GETTINGS AUOY CHESS TON OCT<br>. See to to to is noitoubes your to sinemeriupes of your seilade notices sinT<br><br><br>Designed of Health and Human Services<br>A month and to a noitoube Research (A) September |                                                                                          |
|                                                                                           | (ອານາຣູເນນັງ) (HACO) ເຖີໂຣອH ໄຂວ່ເງິດໄດ້ເນືອນມີຄົມ ການ ຂອງເຂົາມອົງ ໄດ້ ເອີ້ນອາການປ່ຽ<br>Provinsi USB OULE O                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                          |
| аараан аг                                                                                 | STAR FRASARIE A NO SUMITHOO – SHIJ SIHT WOJIBE STIRW TON OG 32A3J9                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                          |
| AFC (S) ອຣປ າອານາດວົ-ອປັT-19VO ---<br>00 Super 108                                        | (0 heqdns 108 Яन2 12 ກr2 12 ກr2 12 ກrest 12                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | aller and the same to the supers and to advises and                                      |
|                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                          |
|                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                          |
|                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                          |
|                                                                                           | əsi əmir ədə 101 bəblərini bax əlirəsi bəildqus si ilk saidlər 10 ninqsai 10 niedər əh<br>əribələri əsərindən ərədi vəzir ilə əsərlərini ol nərindən ini ilk bəhnəldi il ilk və il istifadə kiməli istifi                                                                                                                                                                                                                                                                                                                                                                                                                                           | Indications for Use (Describe                                                            |
|                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 12 Institutional Herminal Hermingul SIS<br>Device Name                                   |
|                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | (umoux yı) vəqimləri (xidmətlər və bir sə<br>882442887                                   |
| יינו<br>woled themetic Statements of the sec<br>54 million Dote: January 2018 minuten and | гээл хэвэл иймин ама НТТАЭН ЭО ТИЭМТЯАЧЭЭ<br>Indications for Use<br>uoitestsinimbA purQ bus booti                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                          |
|                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                          |

{3}------------------------------------------------

# 510(k) Summary

### Sep. 30, 2015

### Cook Biotech Incorporated

### SIS Inguinal Hernia Repair Graft

Manufacturer Name: Cook Biotech Incorporated 1425 Innovation Place West Lafayette, Indiana 47906 Telephone: +1 (765) 497-3355 FAX: +1 (765) 807-7709 Official Contact: Perry W. Guinn

# DEVICE NAME AND CLASSIFICATION:

| Trade/Proprietary Name:     | SIS Inguinal Hernia Repair Graft |
|-----------------------------|----------------------------------|
| Common Name:                | Surgical graft                   |
| Classification Regulations: | Class II, 21 CFR §878.3300 (FTM) |

### INTENDED USE:

The SIS Inguinal Hernia Repair Graft is intended for implantation to reinforce soft tissues where weakness exists, including the repair of inguinal hernias. The graft is supplied sterile and is intended for one-time use.

### DEVICE DESCRIPTION:

The SIS Inguinal Hernia Repair Graft is a flat sheet constructed of an animal sourced bioabsorbable, extracellular matrix collagen membrane derived from porcine Small Intestinal Submucosa, (SIS). It is intended for soft tissue repair with sizes and shapes appropriate for the repair of inguinal hernias. SIS Inguinal Hernia Repair Graft contains collagens I, III, IV and VI. The device is packaged in a dried state and supplied sterile in a sealed double pouch system.

### EQUIVALENCE TO MARKETED DEVICES:

The SIS Inguinal Hernia Repair Graft is similar with respect to intended use, materials and technological characteristics to its predicate devices in terms of section 510(k) Substantial Equivalence, as shown in pre-clinical (biocompatibility, conducted in accordance to ISO 10993-1 standards), mechanical, and clinical testing.

{4}------------------------------------------------

## Biocompatibility Testing

The following biocompatibility tests were performed on sterilized SIS devices, which are identical in composition to the SIS Inguinal Hernia Repair Graft (according to the ISO 10993-1 standard):

- Genotoxicity
- Direct contact in vitro hemolysis
- Cytotoxicity ●
- Muscle implantation
- Acute intracutaneous reactivity
- ISO Sensitization
- Acute systemic toxicity
- . Pyrogenicity
- LAL endotoxins ●
- . Subchronic systemic toxicity

The results of these tests provide evidence that the SIS Inguinal Hernia Repair Graft meets the biocompatibility requirements of the ISO standard.

### Mechanical Testing

The SIS Inguinal Hernia Repair Graft material was tested for the following:

- Suture retention strength
- Burst strength ●
- . Tensile strength

The results of the mechanical tests provide evidence that the SIS Inguinal Hernia Repair Graft provides adequate mechanical strength for its application.

### Clinical Testing

The SIS Inguinal Hernia Repair Graft (also known as Surgisis Inguinal Hernia Matrix SIHM) was implanted in 95 patients in 3 separate clinical studies and the results were published in 3 peer reviewed articles. The results address the durability of repair with resorbable SIS Inguinal Hernia graft as used for open, tensionless repairs of inguinal hernias. One of the studies was a recent US randomized clinical trial comparing the use of SIS Inguinal Hernia Repair Graft (SIHRG) to polypropylene mesh (PP) using the Lichtenstein hernia repair procedure. This repair included bridging the defect. Results showed hernia recurrence rates of 6.7% (3/45 patients) and 0% (0/50 patients) at 1 year follow-up in the treatment (SIHRG) versus control groups. A difference in hernia recurrence rate was also seen at 3 years with 15.6% (7/45 patients) and 4.0% (2/50 patients) for the SIHRG and PP groups, respectively* . In addition, post-operative pain was assessed with a 1 year follow-up. Persistent pain trended higher in the PP group (6% vs. 4%). Authors of the U.S. study note that factors other than device performance could influence hernia recurrence outcomes, e.g., surgical approach, surgical experience, etc.

{5}------------------------------------------------

Two OUS prospectively randomized clinical investigations were conducted comparing SIHRG to PP using the Lichtenstein Repair. These studies found no hernia recurrences for SIHRG-treated patients at 1 and 3 years, (0/15, 1 year)2 and 0% (0/35, 3 years) 3, respectively. Recurrence rates for the PP-treated group found hernia recurrences of 0% (0/15, 1 year)2 and 2.9% (1/35, 3 years) 3. Both studies found lower post-operative pain and discomfort in the SIHRG patients at 30 days 2. The reported differences regarding hernia recurrence between the OUS and US clinical investigations may be reflective of many factors, chief among these being the probable difference in surgical procedures.

In clinical studies in which SIHRG was used to repair an inguinal hernia, the mean BMI in patients undergoing inguinal herniorraphy was 26 kg/m². Recurrence rates may be higher in obese patients (BMI>30) when using SIHRG in a tensionless inguinal hernia repair. A synopsis of each study is presented in the following tables.

*Three year follow-up data not yet validated or published.

{6}------------------------------------------------

| Number of centers                                                       | One center; Baltimore VA hospital; 7 investigators including 4 surgeons                                                                                                                                                                                                                                                                               |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Number of patients                                                      | 100 male patients randomized in a 1:1 fashion to open<br>Lichtenstein repair of the test groups, Surgisis Inguinal<br>Hernia Matrix (SIHM) and control polypropylene mesh:<br>50:50 patients<br>Note: In SIHM group 5 patients were withdrawn prior to<br>surgery due to emergency surgery or traumatic event<br>resulting in 45:50 patients          |
| Study inclusion/exclusion<br>criteria                                   | Exclusion Criteria: life expectancy<3 years, ASA class IV<br>and V, bowel obstruction, strangulation, peritonitis, bowel<br>perforation, local or systemic infection, history of inguinal<br>hernia repair with mesh<br>Inclusion Criteria: 18 years of age or older, unilateral<br>hernia, able to provide informed consent.                         |
| Patient Age, BMI and Hernia<br>Types                                    | All male patients<br>Mean Age: 64(24-85) SIHM and 59 (25-97) Polypropylene<br>BMI: 36(18-39) SIHM; 25(19-37) Polypropylene<br>Hernia type: Direct 20 (44% SIHM) 21(42%<br>Polypropylene)<br>Indirect 26(58% SIHM) 29 (58%<br>Polypropylene)<br>Sliding 24 (53% SIHM) 19 (38%<br>Polypropylene)<br>Non-sliding 26 (58% SIHM) 31 (62%<br>Polypropylene) |
| Operative procedure specifics -<br>anesthesia, duration of<br>procedure | Lichtenstein open repair<br>Anesthesia: Spinal: 4 (9% SIHM) 6 (12% Polypropylene)<br>General: 42 (93% SIHM) 44 (88%<br>Polypropylene)<br>Procedure time (Minutes):<br>134 (SIHM) 115 (Polypropylene)                                                                                                                                                  |
| Patient duration of follow-up<br>outcomes/adverse events                | All patients were followed up for 12 months                                                                                                                                                                                                                                                                                                           |
|                                                                         | Recurrences:                                                                                                                                                                                                                                                                                                                                          |
|                                                                         | 3 hernia recurrences all in the SIHM group incidence<br>6.7% vs 0% in Polypropylene group                                                                                                                                                                                                                                                             |
|                                                                         | All recurrences occurred in patients who originally had<br>direct inguinal hernias (recurrence rate in subset of<br>patients with direct hernias: 3/20 or 15% at 1 year)                                                                                                                                                                              |
|                                                                         | Unpublished 3 year data recurrence rates: 15.6%<br>(7/45) SIHM vs 4% (2/50) Polypropylene                                                                                                                                                                                                                                                             |
|                                                                         | Post-operative pain:                                                                                                                                                                                                                                                                                                                                  |
|                                                                         | At 2 weeks : 9 (20% SIHM) vs 8 (16% Polypropylene)                                                                                                                                                                                                                                                                                                    |
|                                                                         | At 1 year : 2 (4% SIHM) vs 3 (6% Polypropylene)                                                                                                                                                                                                                                                                                                       |
|                                                                         | Adverse events:                                                                                                                                                                                                                                                                                                                                       |
|                                                                         | Hematoma: 6 (13% SIHM) vs 1 (2% Polypropylene)                                                                                                                                                                                                                                                                                                        |
|                                                                         | Incisional pain: 2 (4% SIHM) vs 4 (8% Polypropylene)                                                                                                                                                                                                                                                                                                  |
|                                                                         | Surgical site reaction : 3 (7% SIHM) vs 0 ( 0%<br>Polypropylene)                                                                                                                                                                                                                                                                                      |
|                                                                         | Seroma : 5 (11% SIHM) vs 0 (0% Polypropylene)                                                                                                                                                                                                                                                                                                         |
|                                                                         | Neuralgia: 4 ( 9% SIHM) vs 6 ( 12% Polypropylene )                                                                                                                                                                                                                                                                                                    |
|                                                                         | Infection: 0 ( 0% SIHM) vs 0 (0% Polypropylene)                                                                                                                                                                                                                                                                                                       |
|                                                                         | Testicular Problems: 5 (11% SIHM) vs 4 (8%<br>Polypropylene)                                                                                                                                                                                                                                                                                          |
|                                                                         | Urinary retention: 6 (13% SIHM) vs 3 (6% Polypropylene)                                                                                                                                                                                                                                                                                               |
|                                                                         | Spermatic cord injury: 0 (0% SIHM) vs 1 (2%<br>Polypropylene)                                                                                                                                                                                                                                                                                         |
|                                                                         | *One death in SIHM group due to myocardial infarction.                                                                                                                                                                                                                                                                                                |

Table 1: Bochicchio GV, et al., Biologic vs Synthetic Inguinal Hernia Repair: 1-Year Results of a Randomized Double-Blinded Trial. J Am Coll Surg 2014; 218:751-759.

{7}------------------------------------------------

{8}------------------------------------------------

Table 2: Ansaloni L, et al. Inguinal hernia repair with porcine small intestine submucosa: 3year follow-up results of a randomized controlled trial of Lichtenstein's repair with

| Number of centers                                                    | 1 OUS Center / 2 investigators                                                                                                                                                                                                                                                                      |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Number of patients                                                   | 70 patients randomized to:<br>• Polypropylene<br>• Porcine small intestinal submucosa (Surgisis)                                                                                                                                                                                                    |
| Study inclusion/exclusion criteria                                   | Excluded patients with recurrent hernia, any condition<br>preventing a correct evaluation of pain, hypersensitivity to<br>drugs used in study, intraoperative findings of pathology<br>other than inguinal hernia                                                                                   |
| Patient Age, BMI and Hernia<br>Types                                 | Mean age: 61.3 years for polypropylene (SD 17.7 years)<br>56.2 years for SIHM (SD 18.0 years)<br>Mean BMI: 26<br>Mix of direct and indirect inguinal hernias in each group                                                                                                                          |
| Operative procedure specifics -<br>anesthesia, duration of procedure | Operative time: SHIM 68.6 Minutes / PP 66.0 Minutes<br>Preoperative antibiotics<br>General or Spinal anesthesia (patient's choice/<br>anesthetist's preference)                                                                                                                                     |
| Patient duration of follow-up<br>outcomes/adverse events             | 36 month follow-up<br>• Hernia recurrence:<br>0% SIHM / 2.9% PP<br>• Chronic pain:<br>6 months 11% SIHM / 31% PP<br>12 months 8% SIHM / 23% PP<br>36 months 3% SIHM / 14% PP<br>• Surgical site occurrence at 1 week post-surgery:<br>Hematoma: 5.7% SIHM / 5.7% PP<br>Seroma: 5.7% SIHM / 17.1% PP |

polypropylene mesh versus Surgisis Inguinal Hernia Matrix. Am J Surg 2009; 198:303-312.

{9}------------------------------------------------

| Number of centers                                                 | 1 OUS Center / 5 investigators                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Number of patients                                                | 45 patients with unilateral primary inguinal hernia receiving Lichtenstein repair randomized to:<br>• Polypropylene<br>• Polyglactin and polypropylene<br>• Porcine small intestinal submucosa (Surgisis)                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Study inclusion/exclusion criteria                                | Excluded patients with history of major surgery in lower abdomen other than cancer or immune deficiency                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Patient Age, BMI and Hernia Types                                 | Mean age 54 (range 26-74 years)<br>Mean BMI 26<br>Mix of direct and indirect inguinal hernias in each group                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Operative procedure specifics - anesthesia, duration of procedure | Operative time for all patients 45 minutes (range 35-80 min)<br>Preoperative antibiotics<br>Local anesthesia                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Patient duration of follow-up outcomes/adverse events             | 12 mo (1-16 mo)<br>All patients received 3 month follow-up, using ultrasound U/S - no evidence of prosthesis in SIS group, prosthesis visible in other groups<br>• Polypropylene – early complications (< 30 days): 1 hematoma, 1 seroma, 1 delayed wound healing, 8 discomfort; late complication (> 30 days) 1 hyperesthesia; hernia recurrence 0<br>• Polyglactin and polypropylene – early complications: 2 hematoma, 1 prolonged pain, 1 sensory loss, 7 discomfort; long term: 1 hyperesthesia, 1 prolonged pain, 1 sensory loss, hernia recurrence 0<br>• Porcine small intestinal submucosa (Surgisis) – early: 1 seroma, 2 discomfort; late – none, hernia recurrence 0 |

Table 3: Puccio F, et al.. Comparison of three different mesh materials in tension-free hernia repair: Prolene versus Vypro versus Surgisis. Int Surg 2005:90:S21-S23.

{10}------------------------------------------------

### Substantial Equivalence

See Table 4 for a comparison of the subject device and its predicates.

| Device         | SIS Inguinal Hernia<br>Repair Graft                                                                                                               | SurgiMend Collagen<br>Matrix                                                                                                                                                                                                                                                                                                                                                                                    | SIS Hernia Repair<br>Device                                                                                                                          |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer   | Cook Biotech<br>Incorporated                                                                                                                      | TEI Biosciences Inc.                                                                                                                                                                                                                                                                                                                                                                                            | Cook Biotech<br>Incorporated                                                                                                                         |
| 510(k) Number  | K142887                                                                                                                                           | K083898                                                                                                                                                                                                                                                                                                                                                                                                         | K062697                                                                                                                                              |
| Intended Use   | For implantation to<br>reinforce soft tissues where<br>weakness exists, including<br>the repair of inguinal<br>hernias.                           | For implantation to<br>reinforce soft tissue<br>where weakness exists<br>and for the surgical repair<br>of damaged or ruptured<br>soft tissue membranes.<br>SurgiMend is specifically<br>indicated for: plastic and<br>reconstructive surgery,<br>muscle flap<br>reinforcement, hernia<br>repair including<br>abdominal, inguinal,<br>femoral, diaphragmatic,<br>scrotal, umbilical, and<br>incisional hernias. | For implantation to<br>reinforce soft tissues<br>where weakness exists.<br>Indications for use<br>include repair of a hernia<br>or body wall defect. |
| Product code   | FTM                                                                                                                                               | FTM                                                                                                                                                                                                                                                                                                                                                                                                             | FTL                                                                                                                                                  |
| 21 CFR         | 878.3300                                                                                                                                          | 878.3300                                                                                                                                                                                                                                                                                                                                                                                                        | 878.3300                                                                                                                                             |
| Material       | Porcine small intestinal<br>submucosa (porcine)<br>Primarily Types I, III, IV<br>and VI collagen<br>(constituents of the<br>extracellular matrix) | Collagen based derived<br>from fetal bovine dermis                                                                                                                                                                                                                                                                                                                                                              | Porcine small intestinal<br>submucosa (porcine)<br>Primarily Types I, III, IV<br>and VI collagen<br>(constituents of the<br>extracellular matrix)    |
| Dimensions     | Rectangular or pre-<br>shaped<br>6 x 10 cm<br>8 x 15 cm<br>10 x 15 cm<br>12 x 15 cm<br>13 x 18 cm                                                 | 3 x 3 cm to 20 x 20 cm<br>square, 0.3 x 25 cm to 25<br>x 40 cm rectangle, 7 x 17<br>cm to 10 x 20 rectangle,<br>8 x 16 cm to 15 x 15 cm<br>semi-oval                                                                                                                                                                                                                                                            | 5 x 8 cm to 20 x 30 cm                                                                                                                               |
| Thickness      | 0.1 - 1.0 mm                                                                                                                                      | 0.5 to 4.4 mm                                                                                                                                                                                                                                                                                                                                                                                                   | 0.1 - 1.5 mm                                                                                                                                         |
| EtO Sterilized | Yes                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                  |

# Table 4 – Substantial Equivalence Comparison

### CONCLUSION:

The mechanical, pre-clinical, and clinical tests performed on the SIS Inguinal Hernia Repair Graft show that the device is substantially equivalent to its predicates.

# REFERENCES:

1.Bochicchio, GV, et al., J Am Coll Surg, 2014. 218(4): p. 751-7.

- 2.Puccio, F, et al., Int Surg, 2005. 90(3 Suppl): p. S21-3.
- 3.Ansaloni, L, et al., Am J Surg, 2009. 198(3): p. 303-12.

---

**Source:** [https://fda.innolitics.com/device/K142887](https://fda.innolitics.com/device/K142887)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
