K142860 · Lumenis, Ltd. · GEX · Jan 21, 2015 · General, Plastic Surgery
Device Facts
Record ID
K142860
Device Name
Lumenis M22 System
Applicant
Lumenis, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 21, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The subject Lumenis M22 System has connection capability with the following available treatment handpieces, for multi-application treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows: 1. The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200nm (with 8 different filters) is indicated for: · Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, ephelides (freckles) and tattoos · Cutaneous lesions, including warts, scars and striae · Benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations · Removal of unwanted hair from all skin types, and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles · Mild to moderate inflammatory Acne (Acne vulgaris) 2. The Nd: Y AG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd: Y AG) is indicated for: · The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1 - 4.0 mm. diameter) of the leg · The removal of unwanted hair from all skin types, and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles · The non-ablative treatment of facial wrinkles 3. ResurFX module and handpiece, with wavelength of 1565 pm, is indicated for: · Use in dermatological procedures requiring skin resurfacing and coagulation of soft tissue 4. The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064nm is indicated for: - Removal of dark tattoos · Treatment of pigmented lesions
Device Story
Multi-application aesthetic/dermatological platform; utilizes four interchangeable handpieces: IPL (400-1200nm), Multi-Spot Nd:YAG (1064nm), ResurFX (1565nm), and Q-Switched Nd:YAG (1064nm). Inputs: electrical energy converted to light/laser pulses. Operation: physician-controlled delivery of light/laser energy to target tissues (melanin, hemoglobin, water) for coagulation, resurfacing, or hair reduction. Used in clinical settings. Output: thermal/photomechanical tissue effects. Benefits: non-invasive treatment of skin conditions, vascular lesions, and hair reduction.
Clinical Evidence
Bench testing only. Safety and efficacy established via comparison to cleared predicate devices and existing scientific research/clinical studies for identical handpiece technologies.
Indicated for aesthetic and dermatological procedures including treatment of benign epidermal/cutaneous lesions, vascular lesions, unwanted hair reduction (skin types I-V), inflammatory acne, soft tissue coagulation, skin resurfacing, and tattoo removal. Contraindications not specified.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, with three overlapping profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 21, 2015
Lumenis Limited Ms. Elissa Burg Director of Quality Assurance/Quality Systems 6 Hakidma Street Yokneam 20692 Israel
Re: K142860 Trade/Device Name: Lumenis M22 System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX, ONF, ONG Dated: December 22, 2014 Received: December 24, 2014
Dear Ms. Burg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K142860
Device Name
Lumenis M22™ System
#### Indications for Use (Describe)
The subject Lumenis M22 System has connection capability with the following available treatment handpieces, for multiapplication treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows:
1. The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200nm (with 8 different filters) is indicated for:
· Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, ephelides (freckles) and tattoos
· Cutaneous lesions, including warts, scars and striae
· Benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations
• Removal of unwanted hair from all skin types, and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
· Mild to moderate inflammatory Acne (Acne vulgaris)
2. The Nd: Y AG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd: Y AG) is indicated for:
· The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1 - 4.0 mm. diameter) of the leg
· The removal of unwanted hair from all skin types, and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
· The non-ablative treatment of facial wrinkles
3. ResurFX module and handpiece, with wavelength of 1565 pm, is indicated for:
· Use in dermatological procedures requiring skin resurfacing and coagulation of soft tissue
4. The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064nm is indicated for:
- Removal of dark tattoos .
- · Treatment of pigmented lesions
#### *Note
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6. 9. and 12 months after the completion of a treatment regime.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Lumenis logo. The logo consists of a blue circle with a white "L" inside it, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the phrase "Energy to Healthcare" also in blue. The logo is clean and modern, and the colors are bright and inviting.
# Section 2- 510(k) Summary
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Lumenis logo. The logo features a blue circle with a white "L" inside on the left. To the right of the circle is the word "Lumenis" in blue, with the tagline "Energy to Healthcare" underneath in a smaller font. The logo is clean and modern, with a focus on the company name and its mission.
### 510(K) SUMMARY
Lumenis M22 System
## 510(k) Number K
Applicant Name: Lumenis Ltd. 6 Hakidma Street, P.O.B. # 240 Yokneam 2069204 Israel Tel. (972) 4-959-9000 Fax: (972) 4-959-9050
Contact Person: Elissa Burg Director of Regulatory Affairs & Quality Systems Telephone: (972) 4-959-9000 Fax: (972) 4-959-9050 E-mail: Elissa.Burg@lumenis.com
Trade Name: Lumenis M22TM System
Summarv
Preparation Date: October 13, 2014
- Classification: Name: Intense Pulsed Light & Laser System and delivery devices Product Code: GEX, ONF, ONG Regulation No: 21 CFR 878.4810 Class: II Classification Panel: General and Plastic Surgery
{6}------------------------------------------------
#### Device Description:
The subject Lumenis M22 System is a multi-application, multi-technology platform with four (4) available treatment handpieces:
- Universal Intense Pulsed Light (IPL) handpiece (K083733) .
- Multi-Spot Nd:YAG laser handpiece (K083733)
- ResurFX non-ablative laser handpiece (K130028) .
- Q-Switched Nd: Y AG laser handpiece (K043173) ●
#### Intended Use Statement:
The subject Lumenis M22 System has connection capability with the following available treatment handpieces, for multi-application treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows:
#### 1. The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200nm (with 8 different filters) is indicated for:
- Benign epidermal lesions, including dyschromia, hyperpigmentation, ● melasma, ephelides (freckles) and tattoos
- Cutaneous lesions, including warts, scars and striae
- Benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, . truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations
- . Removal of unwanted hair from all skin types, and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
- Mild to moderate inflammatory Acne (Acne vulgaris)
#### 2. The Nd:YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd:YAG) is indicated for:
- · The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1 - 4.0 mm. diameter) of the leg
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Lumenis logo. The logo consists of a blue circle with a white "L" inside it, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the phrase "Energy to Healthcare" in a smaller blue font.
- The removal of unwanted hair from all skin types, and to effect stable long term, or ● permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
- The non-ablative treatment of facial wrinkles
- 3. ResurFX module and handpiece, with wavelength of 1565 nm, is indicated for:
- Use in dermatological procedures requiring skin resurfacing and coagulation of soft tissue
- The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064nm 4. is indicated for:
- Removal of dark tattoos ●
- Treatment of pigmented lesions
#### *Note
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
#### Predicate Devices:
Substantial equivalence to the following predicate devices is claimed:
| Device Name | 510k No | Date of Clearance |
|--------------------------------------|---------|-------------------|
| Lumenis Lum One | K060448 | July 5, 2006 |
| Lumenis LUME 2 | K083733 | May 04, 2009 |
| Lumenis ResurFX laser<br>Module | K130028 | Sep 03, 2013 |
| Lumenis Quantum Q<br>switched Nd:YAG | K043173 | Jan 1, 2005 |
| Palomar Lux V | K040081 | April 1, 2004 |
| Candela Nd:YAG laser<br>System | K063074 | Dec 27, 2006 |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Lumenis logo. On the left is a blue circle with a white "L" inside. To the right of the circle is the word "Lumenis" in blue, with the registered trademark symbol in the upper right corner. Below the word "Lumenis" is the phrase "Energy to Healthcare" in a smaller blue font.
#### Substantial Equivalence to Predicate Devices
The subject Lumenis M22 System's IPL handpiece uses a spectrum of 400-1200 mm, combined by filters used in Lumenis One (K060448), Lumenis LUME 2 (K083733) and Palomar LuxV (K040081). The device contains filters for different parts of this spectrum, same as the predicates. The minor differences in technical specifications do not raise any new questions of safety and efficacy.
Lumenis M22 System also has the Acne filter (Notch filter 400-600 & 800-1.200 nm) for the treatment of inflammatory Acne vulgaris which is similar to the Palomar Lux V (K040081) Acne filter which has a wavelength of 400-700 nm & 870-1200 nm. The minor differences in the wavelengths do not raise any new questions of safety and efficacy.
The subject Lumenis M22 system's Multi-Spot Nd:YAG handpiece uses a 1064nm laser, which is equivalent to the handpiece used by its predicate device: Lumenis Lume 2 (K083733). Both products use a Xenon flash-lamp pumped solid state Nd:YAG, and both utilize the same pulse rate and similar performance characteristics. The Lumenis M22 Nd:YAG laser has an additional tip of 1.5 mm, which has similar characteristics as the predicate device of Candela Nd:YAG laser system (K063074).
The subject Lumenis M22 System's ResurFX handpiece is the same as the ResurFX handpiece that was cleared under K130028 and used with the market cleared Lumenis Lume 2 (K083733), being the case for all their physical parameters as well as their intended uses.
The subject Lumenis M22 System's Q-Switched handpiece is equivalent to the Q-Switched handpiece that has been used with the market cleared Lumenis Quantum (K043173). The only differences are a minor increase in fluence (<10%), caused by the specification of the Nd:YAG rods and an additional 6mm metal tip.
2-5
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Lumenis logo. The logo consists of a blue circle with a white "L" inside it, followed by the word "Lumenis" in blue text. Below the word "Lumenis" is the phrase "Energy to Healthcare" in smaller blue text.
#### Performance Standards:
The subject Lumenis M22 System complies with
- . IEC 60601-1: 2005 Medical Electrical Equipment-Part 1: General Requirements for Safety. Collateral Standard: Safety Requirements for Medical Electrical Systems.
● IEC 60601-1-2:2007 Medical Electrical Equipment Part 1-2: Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
IEC 60601-2-22:2007 - Medical electrical equipment - Part 2-22: ● Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.
. IEC 60825-1:2007 - Safety of laser products - Part 1: Equipment classification, and requirements.
● IEC 60601-2-57:2011 - Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetics.
#### Performance Bench Tests
Bench testing demonstrated that the subject Lumenis M22 System is as safe and effective as the cleared predicate devices.
#### Summary of Pre-Clinical and Clinical Study
The safety and efficacy of all the identified handpieces applied within the subject Lumenis M22 System are well established in scientific research and clinical studies. Multiple studies with these and similar systems have shown safety in aesthetic and dermatological procedures, in addition to the devices cleared for marketing by the FDA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.