The CVUE ADVANCED (hioxifilcon D) Single Vision Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with nondiseased eyes. The lens may be worn by persons who require up to 3.00 Diopters of add and who exhibit astigmatism of up to 0.75 Diopters that does not interfere with visual acuity. The CVUE ADVANCED (hioxifilcon D) Toric and Multifocal Toric Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia), presbyopia and astigmatism up to 4.00 diopters in aphakic and/or notaphakic persons with non-diseased eyes. Eye care practitioners may prescribe the lens for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system.
Device Story
C-VUE Advanced (hioxifilcon D) soft contact lenses are semi-scleral flexible shells covering the cornea and partial sclera; designed as aspheric single vision, multifocal, or toric multifocal. Toric designs utilize prism ballast or dynamic axis stabilization (double slab off). Multifocal designs feature center-plus power progression. Lenses are prescribed by eye care practitioners for daily wear; patients perform cleaning and chemical disinfection. Lenses correct refractive errors and presbyopia, improving visual acuity for patients with non-diseased eyes.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on established safety profile, pre-clinical toxicology, and manufacturing/chemistry data consistent with previously cleared devices.
Technological Characteristics
Material: hioxifilcon D (non-ionic copolymer of 2-HEMA and GMA, crosslinked with EGDMA). Water content: 54%. Oxygen permeability (Dk): 23 x 10^-11 Fatt units (ANSI Z80.20:2004). Refractive index: 1.408. Dimensions: 12.5-17.0mm diameter, 7.0-10.5mm base curve. Stabilization: prism ballast or double slab off. Tint: phthalocyanato (2)-(copper).
Indications for Use
Indicated for daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism) and presbyopia in aphakic or non-aphakic persons with non-diseased eyes. Single vision lenses support up to 3.00D add and 0.75D astigmatism. Toric/Multifocal Toric lenses support up to 4.00D astigmatism.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing to the right, with flowing lines suggesting movement or connection. The profiles are arranged in a way that they appear to be overlapping. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 30, 2014
Unilens Corp., USA Mr. Alan J. Frazer Director of Quality Assurance 10431 72nd Street, North Largo, FL 33777
Re: K142851
Trade/Device Name: C-VUE® Advanced™ (hioxifilcon D) Soft (hydrophilic) Regulation Number: 21 CFR 886.5925 Regulation Name: Lenses, soft contact, daily wear Regulatory Class: Class II Product Code: LPL Dated: September 29, 2014 Received: September 30, 2014
Dear Mr. Frazer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K142851
Device Name
CVUE® Advanced™ (hioxifilcon D) Soft (hydrophilic) Contact Lens for Daily Wear
#### Indications for Use (Describe)
The CVUE ADVANCED (hioxifilcon D) Single Vision Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with nondiseased eyes. The lens may be worn by persons who require up to 3.00 Diopters of add and who exhibit astigmatism of up to 0.75 Diopters that does not interfere with visual acuity.
The CVUE ADVANCED (hioxifilcon D) Toric and Multifocal Toric Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia), presbyopia and astigmatism up to 4.00 diopters in aphakic and/or notaphakic persons with non-diseased eyes.
Eye care practitioners may prescribe the lens for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:14px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:14px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### Applicant Information
| Date prepared: | 17 Sep 2014 |
|-------------------|--------------------------------------------------------------------|
| Name:<br>Address: | Unilens Corp., USA<br>10431 72nd Street, North<br>Largo, FL 33777 |
| Contact person: | Alan J. Frazer<br>Director of Quality Assurance |
| Phone number: | (727) 544-2531 |
| Fax number: | (727) 545-1883 |
#### Device Information
| Device classification: | Class II |
|------------------------|------------------------------------------------------------------|
| Classification number: | LPL |
| Classification name: | Lenses, Soft Contact, Daily Wear |
| Trade name: | C-VUE® Advanced™ (hioxifilcon D) Soft (hydrophilic) Contact Lens |
### Equivalent device
The C-VUE® Advanced™ (hioxifilcon D) Soft (hydrophilic) Contact Lens for Daily Wear is substantially equivalent to the predicate device identified below in terms of intended use.
Predicate devices:
LifeStyle MV2 Toric (polymacon) Soft (Hydrophilic) Multifocal Contact Lens for Daily Wear
C-VUE® Advanced™ (hioxifilcon D) Soft (hydrophilic) Contact Lens for Daily Wear
### Device description
CVUE ADVANCED (hioxifilcon D) soft contact lenses are semi-scleral flexible shells which cover the cornea and may cover a portion of the adjacent sclera and are available as aspheric single vision, multifocal, toric multifocal designs. Torics have a toroidal posterior optic zone, and multifocals have the most plus power is in the center of the lens, with the power progressively becoming more minus towards the periphery. Torics have the option of dynamic axis stabilization (double slab off) or prism ballast stabilization. All lenses share a base curve with a flattened peripheral curve which approximates the curvature of the sclera. The lens material, (hioxifilcon D), is a non-ionic copolymer of 2-hydroxyethyl methacrylate (2-HEMA) and 2, 3-dihydroxypropyl methacrylate (Glycerol Methacrylate, GMA) and crosslinked with ethylene glycol dimethacrylate (EGDMA). It consists of 46% hioxifilicon D
{4}------------------------------------------------
and 54% water by weight when immersed in normal buffered saline solution. The lens is available with or without a blue visibility handling tint, phthalocyanato (2) -(copper).
The CVUE ADVANCED (hioxifilcon D) soft contact lens is a hemispherical shell of the following dimensions:
| Chord diameter: | 12.5 to 17.0mm |
|-----------------------------|------------------------------------------------|
| Center thickness: | 0.13 to 0.73; varies with power |
| Base curve: | 7.0 to 10.5mm |
| Powers: | -20.00 to +20.00 diopters |
| ADD powers (multifocal): | Up to +3.00 diopters |
| Cylinder (toric): | Up to 4.00 diopters |
| Axis (toric): | 0° to 180° in 1° steps |
| Axis stabilization (toric): | prism ballast; or<br>dynamic (double slab off) |
| Optical zone diameters: | 5.0 to 10.0mm |
| The physical/optical properties of the lens are: | |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Refractive Index | 1.408 (hydrated) |
| Light Transmission – tinted | greater than 90% |
| Water Content | 54% |
| Specific Gravity | 1.299 (dry) |
| Oxygen Permeability (Dk Value) | 23 x 10-11 Fatt Units (cm2/sec) (ml O2/ml x mm |
| | Hg @ 35°C), ANSI Z80.20:2004 upgraded polarographic method<br>corrected for boundary-layer and edge effects |
## Intended Use (Indications)
The CVUE ADVANCED (hioxifilcon D) Single Vision Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) and presbyopia in aphakic and/or not-aphakic persons with nondiseased eyes. The lens may be worn by persons who require up to 3.00 Diopters of add and who exhibit astigmatism of up to 0.75 Diopters that does not interfere with visual acuity.
The CVUE ADVANCED (hioxifilcon D) Toric and Multifocal Toric Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia), presbyopia and astigmatism up to 4.00 diopters in aphakic and/or notaphakic persons with non-diseased eyes.
Eye care practitioners may prescribe the lens for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system.
{5}------------------------------------------------
Substantial equivalence
The CVUE ADVANCED (hioxifilcon D) Single Vision Contact Lenses will be manufactured according using the specified process controls and the quality management system currently in place. The device will underqo manufacturing, packaging and sterilization procedures that are the same as devices currently manufactured, marketed and distributed by Unilens Corp., USA.
The established safety profile (pre-clinical toxicology and manufacturing/chemistry data) of the device the same to the CVUE ADVANCED (hioxifilcon D), 510(k) K082853.
The CVUE ADVANCED (hioxifilcon D) Single Vision Contact Lenses indications for use do not change
Physical construction and safety and effectiveness do not change from the predicate devices.
This meets the requirements per section 510(k) of the act regarding substantial equivalence and does not raise any new or different questions of safety and effectiveness than the predicate devices identified above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.