Clinical study of 99 subjects (18-60 years) comparing Zmachine algorithm to consensus human scores
>1 (certified polysomnographic technologists)
Indications for Use
The CRI Zmachine is a single-channel, EEG acquisition and analysis system, designed for use in the home or clinical environments. This device is intended to be used by qualified healthcare practitioners to monitor the wake and sleep states of adult patients and as an adjunct to their diagnosis of sleep disorders.
Device Story
Battery-operated, single-channel EEG acquisition/analysis system; uses differential-mastoid EEG channel input; processes data to determine wake/sleep states (wake, light sleep, deep sleep, REM) in 30-second epochs; used in home or clinical settings by healthcare practitioners; provides summary sleep statistics to assist in evaluating treatment efficacy; aids clinicians in diagnosing sleep disorders.
Clinical Evidence
Clinical study of 99 subjects (18-60 years, 52F/47M) comparing Zmachine algorithm output against consensus human scoring (R&K guidelines) of PSG data. Overall kappa agreement was 0.716 across 85,206 epochs. P1 (sensitivity) and P2 (positive predictive value) metrics for Wake, Light Sleep, Deep Sleep, and REM were calculated and compared favorably to the Sleep Profiler predicate.
Technological Characteristics
Single-channel EEG acquisition system; differential-mastoid electrode placement; battery-powered; proprietary algorithm using time and frequency domain features; real-time EEG analysis; provides summary sleep statistics.
Indications for Use
Indicated for adult patients to monitor wake and sleep states as an adjunct to the diagnosis of sleep disorders. No specific contraindications listed.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure with three overlapping profiles, symbolizing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 20, 2015
Consolidated Research Of Richmond, Inc % Richard Kaplan President 26250 Euclid Ave Suite 709 Euclid. Ohio 44132
K142825 Re:
> Trade/Device Name: Zmachine DT-200 Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLZ, OLV, OMC Dated: December 15, 2014 Received: December 19, 2014
Dear Dr. Kaplan,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
Carlos L. Peña. PhD. MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K142825
Device Name Zmachine DT-200
Indications for Use (Describe)
The CRI Zmachine is a single-channel, EEG acquisition and analysis system, designed for use in the home or clinical environments. This device is intended to be used by qualified healthcare practitioners to monitor the wake and sleep states of adult patients and as an adjunct to their diagnosis of sleep disorders.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
C)
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "CRI Leaders in Advanced EEG Technologies". The letters "CRI" are in large, red font, with a shadow effect. The text "Leaders in Advanced EEG Technologies" is in a smaller, gray font and is located below and to the right of the letters "CRI".
## 510(k) Summary
## Zmachine® 510(k) Notification Consolidated Research of Richmond, Inc.
| 510(k) Owner | Consolidated Research of Richmond, Inc. |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | 26250 Euclid Avenue, Suite 709<br>Euclid, Ohio 44132<br>Phone: (216) 289-2331<br>Fax: (216) 393-0079 |
| Contact Person | Richard F. Kaplan, Ph.D., President<br>Phone: (216) 289-2331 Extension 1001<br>E-mail: kaplan@cri-systems.com |
| Date Prepared | December 16, 2014 |
| Trade Name | Zmachine® DT-200 |
| Common Name | Sleep monitoring system |
| Classification Name | Electroencephalograph<br>21 CFR 882.1400 |
| Product Codes | OLZ, OLV and OMC |
| Indications for Use | The CRI Zmachine is a single-channel, EEG acquisition and analysis<br>system, designed for use in the home or clinical environments. This<br>device is intended to be used by qualified healthcare practitioners to<br>monitor the wake and sleep states of adult patients and as an adjunct to<br>their diagnosis of sleep disorders. |
{5}------------------------------------------------
| Device<br>Description | The CRI Zmachine is a battery-operated, single-channel, EEG<br>acquisition and analysis system. The Zmachine system includes the<br>Zmachine device, disposable EEG sensors, sensor cable, and a wall<br>charger. The device operates on data from the differential-mastoid<br>EEG channel to determine the wake and sleep states (wake, light sleep,<br>deep sleep and REM) of the patient every 30 seconds. |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial<br>Equivalence | The Zmachine DT-200 is substantially equivalent to the Zmachine DT-<br>100 (K101830) by Consolidated Research of Richmond, Inc., sharing<br>the same indications for use, the same primary wake/sleep detection<br>algorithm, and the same hardware platform. The firmware of<br>Zmachine DT-200 has been upgraded to further subdivide epochs<br>detected as sleep by the primary wake/sleep detection algorithm into<br>light sleep (stages N1 and N2), deep sleep (N3) and REM. The further<br>subdivision of sleep stages are substantially equivalent to the Sleep<br>Profiler algorithm (K120450) by Advanced Brain Monitoring, Inc. |
The table below summarizes the technological characteristics of the Zmachine DT-200 in comparison to the predicate devices.
| | Zmachine DT-200<br>(New) | Zmachine DT-100<br>(K101830) | Sleep Profiler<br>(K120450) |
|------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>Name | Consolidated<br>Research of<br>Richmond, Inc. | Consolidated<br>Research of<br>Richmond, Inc. | Advanced Brain<br>Monitoring, Inc. |
| 510(k) Number | This 510(k) | K101830 | K120450 |
| Classification<br>Regulation | 882.1400 | 882.1400 | 882.1400 |
| Product Code | OLZ, OLV and<br>OMC | OLV and OMC | OLZ |
| | Zmachine DT-200 | Zmachine DT-100 | Sleep Profiler |
| Indications for Use | (New)<br>The CRI Zmachine<br>is a single-channel,<br>EEG acquisition and<br>analysis system,<br>designed for use in<br>the home or clinical<br>environments. This<br>device is intended to<br>be used by qualified<br>healthcare<br>practitioners to<br>monitor the wake<br>and sleep states of<br>adult patients and as<br>an adjunct to their<br>diagnosis of sleep<br>disorders. | (K101830)<br>The CRI Zmachine<br>is a single-channel,<br>EEG acquisition and<br>analysis system,<br>designed for use in<br>the home or clinical<br>environments. This<br>device is intended to<br>be used by qualified<br>healthcare<br>practitioners to<br>monitor the wake<br>and sleep states of<br>adult patients and as<br>an adjunct to their<br>diagnosis of sleep<br>disorders. | (K120450)<br>Sleep Profiler is<br>intended for the<br>diagnostic<br>evaluation by a<br>physician to assess<br>sleep quality in<br>adults only. The<br>Sleep Profiler is a<br>software-only<br>device to be used<br>under the<br>supervision of a<br>clinician to analyze<br>physiological<br>signals and<br>automatically score<br>sleep study results,<br>including the<br>staging of sleep,<br>detection of<br>arousals and<br>snoring. |
| Patient Population | Adults | Adults | Adults |
| Type of Device | EEG-based sleep<br>monitor | EEG-based sleep<br>monitor | EEG-based sleep<br>staging software |
| No. of EEG<br>Channels | 1 | 1 | 1 |
| Electrode<br>Placement | Mastoid | Mastoid | Frontal |
| Analyzes EEG<br>Data in Real Time | Yes | Yes | Yes |
| Sleep<br>Classification | Wake and N1+N2<br>(light sleep). N3<br>(deep sleep), and<br>REM | Wake and Sleep | Wake and N1, N2,<br>N3, and REM |
| EEG Analysis<br>Methodology | Proprietary<br>algorithm using<br>time and frequency<br>domain features | Proprietary<br>algorithm using<br>time and frequency<br>domain features | Proprietary<br>algorithm using<br>time and frequency<br>domain features |
| | Zmachine DT-200<br>(New) | Zmachine DT-100<br>(K101830) | Sleep Profiler<br>(K120450) |
| Provides<br>Information to<br>Assist the<br>Healthcare<br>Provider in<br>Evaluation of<br>Treatment<br>Efficacy? | Yes | Yes | Yes |
| Calculates<br>Summary Sleep<br>Statistics? | Yes | Yes | Yes |
| Battery Powered? | Yes | Yes | Yes |
{6}------------------------------------------------
{7}------------------------------------------------
### Performance Testing - Clinical
A clinical study was conducted in which overnight laboratory polysomnographic (PSG) data, and data from the differential-mastoids (A1-A2), were acquired from 99 subjects (52F/47M, 18-60 years, median age 32.7 years), including those reporting normal sleep and those reporting complaints consistent with various sleep disorders.
The standard PSG channels were scored independently by two to four certified polysomnographic technologists, using the Rechtschaffen and Kales (R&K) visual sleep staging guidelines. The individual score files were then combined, on a 30-second epoch basis, using a majority voting rule, to generate a single score file per subject ("Human Scores"). EEG data acquired from A1-A2 was processed by the Zmachine Algorithm which determines wake, light sleep (N1&N2), deep sleep (N3) and REM on a 30-second epoch basis.
The performance of the Zmachine Algorithm was evaluated by comparing against the Human Scores for stages wake, light sleep (N1&N2), deep sleep (N3) and REM. Overall, kappa agreement between Zmachine Algorithm and Human Scores for 85,206 epochs is 0.716, with P1 (the probability that Zmachine will correctly assign an epoch when the PSG Consensus assigns the epoch to a particular stage) and P2 (when the Zmachine assigns an epoch to a particular stage, the probability of such assignment is correct) for each stage summarized in the table below.
| Zmachine DT-200 | | | | |
|-----------------|-------|-------------|------------|-------|
| | Wake | Light Sleep | Deep Sleep | REM |
| P1 | 0.908 | 0.835 | 0.738 | 0.721 |
| P2 | 0.843 | 0.848 | 0.782 | 0.732 |
Using the performance data published in K120450 (44 subject study) of the predicate Sleep Profiler algorithm, stages N1 and N2 were combined into Light Sleep for both the
{8}------------------------------------------------
algorithm and scorers (to permit a direct comparison with the Zmachine Algorithm) with a resulting P1 and P2 for each stage summarized in the table below.
| Sleep Profiler | | | | |
|----------------|-------|-------------|------------|-------|
| | Wake | Light Sleep | Deep Sleep | REM |
| P1 | 0.789 | 0.855 | 0.757 | 0.719 |
| P2 | 0.831 | 0.812 | 0.787 | 0.782 |
The P1 and P2 values obtained during clinical validation of the Zmachine Algorithm are similar to those obtained by the predicate Sleep Profiler algorithm, which was validated using its own data set.
### Conclusion
This submission demonstrates that the Zmachine is substantially equivalent to the predicate devices based on descriptive information and clinical performance testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.