The device is a spirometer which measures patient respiratory parameters including FVC, FEV1, FEV6, MVV and VC. The Vitalograph Pneumotrac is a desktop spirometer designed for luse on adults and pediatics, 5 years and older, in a variety of environments such as hospital wards, health centers and private homes under the supervision of a healtheare provider.
Device Story
Desktop spirometer; measures dynamic lung volumes/flow rates (FVC, FEV1, FEV6, PEF, MVV, VC) via Fleisch pneumotachograph; connects via USB to Vitalograph Model 7000 Spirotrac software for data management; used in hospitals, health centers, and homes under healthcare provider supervision; provides audible incentives during testing; outputs patient demographic data and spirometry reports; aids physicians in detection, diagnosis, and monitoring of chest diseases.
Clinical Evidence
Bench testing only. Performance validated against ISO 26782 (spirometers), ISO 23747 (peak flow meters), and ATS/ERS 2005 standards. Electrical safety and EMC testing per IEC 60601-1 and 60601-1-2. Connectivity/data integrity validated with Model 7000 Spirotrac software. No clinical data required.
Technological Characteristics
Fleisch pneumotachograph sensing; USB-powered; ABS plastic, silicone rubber, stainless steel, aluminum, TPX plastic construction; dimensions 183x105x70mm; weight 0.45kg; connectivity via USB to PC-based Spirotrac software; complies with IEC 60601-1, IEC 60601-1-2, ISO 23747, ISO 26782, ATS/ERS 2005.
Indications for Use
Indicated for adults and pediatrics (5+ years) for lung function testing (FVC, FEV1, FEV6, PEF, MVV, VC) in clinical or home settings under healthcare provider supervision.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health and Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 29, 2015
Vitalograph Ireland Ltd. Tom J. Healy Regulatory Affairs/QA Manager Gort Road Business Park, Ennis, Co Clare IRELAND
Re: K142812
Trade/Device Name: Vitalograph Model 6800 Pneumotrac Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic spirometer Regulatory Class: II Product Code: BZG Dated: May 27, 2015 Received: May 29, 2015
Dear Mr. Healy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Tom J. Healy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K142812
#### Device Name
Vitalograph Model 6800 Pneumotrac
### Indications for Use (Describe)
The device is a spirometer which measures patient respiratory parameters including FVC, FEV1, FEV6, MVV and VC. The Vitalograph Pneumotrac is a desktop spirometer designed for luse on adults and pediatics, 5 years and older, in a variety of environments such as hospital wards, health centers and private homes under the supervision of a healtheare provider.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510K Summary
## as required by 21 CFR 807.92
- 1. Company Information: Name: Vitalograph (Ireland) Ltd Address: Gort Road Business Park, Ennis, Co Clare, Ireland.
Tel: +353656864100 Fax: +353656829289.
- 2. Contact Person / Official Correspondent: Mr. Tom J Healy Regulatory Affairs / Quality Assurance Manager
- 3. Date prepared: 19th June 2015.
- 4. Device Trade Name: Vitalograph Model 6800 Pneumotrac
- 5. Common / Usual name: Vitalograph Pneumotrac,
- 6. Classification number:
Classified per 21 CFR 868.1840. Part 868 - Anesthesiology Devices, Subpart B--Diagnostic Devices Class 2 Diagnostic Spirometer Product Code BZG.
- 7. Predicate Device:
| Manufacturer | : | Vitalograph |
|--------------|---|-------------------------------------|
| Device Name | : | Model 2120 |
| 510(k) No | : | K100687, Class 2, Product Code BZG. |
## 8. Description of Device:
The intended use of the Vitalograph Pneumotrac desktop spirometer, which connects to Vitalograph Model 7000 Spirotrac software, ref 510(k) K141546, shall provide a USB-powered desktop spirometer for creating, adding and recalling subjects and performing Spirometry testing on those subjects to aid in the measuring the effect of lung disease on pulmonary function.
Spirometry is in the simple assessment of respiratory function through the measurement of dynamic lung volumes.
{5}------------------------------------------------
Its primary functions are;
- 1. Standalone spirometric measures using single breath or multiple-breath testing techniques, display and record lung volumes and flow rates (including FVC, MVV, VC) and it's subdivisions,
- Display a single subject's demographic data entered by the user. 2.
- ന്ന് Produce reports
In a clinical setting, the measurements obtained from a lung function test form part of the various findings of a physician in the detection, diagnosis and monitoring of chest diseases.
## 9. Indications for Use:
The device is a spirometer which measures patient respiratory parameters including FVC, FEV1, FEV6, PEF, MVV and VC. The Vitalograph Pneumotrac is a desktop spirometer designed for lung function testing for use on adults and pediatrics, 5 years and older, in a variety of environments such as hospital wards, health centers and private homes under the supervision of a healthcare provider.
## 10. Technological Characteristics
The characteristics of the Pneumotrac are similar to those of the predicate device listed in comparison table below. The similarities are:
- · Identical operating principle
- Identical spirometry parameters measured
- Both the subject and predicate devices comply with international performance standards including electrical safety {IEC 60601-1}, EMC {IEC 60601-1-2} and performance bench tests {ISO 23747, ISO 26782, ATS/ERS}
- Sharing of Identical patient interface accessories
The primary differences between the Vitalograph Model 6800 and the predicate devices is that the predicate operates as a portable spirometer and the Model 6800 is a desktop spirometer.
The indications for use for the Vitalograph Pneumotrac now include pediatric population in line with the updated FDA guidance. The Model 6800 has the same indications for use, including pediatric population, as the Model 7000 as cleared under K141546. No new testing was required for this revised indication for use. No new risks have been introduced as a result of pediatric population inclusion. The device complies with the existing international performance standards to cater for all population groups.
All of the device materials were previously cleared in the predicate (K100687) and create no new risk profiles. The Model 7000 Spirotrac which connects to the Pneumotrac device was previously cleared. Materials used in the Vitalograph 6800 Pneumotrac are identical to those used in previously cleared devices. Thus separate biocompatibility testing was not performed for the subject device.
{6}------------------------------------------------
Risks have been evaluated and the connectivity / communication by model 6800 Pneumotrac has been validated with the Vitalograph Model 7000 Spirotrac software. This validation is on file.
The operating principle, measuring range, application and use are unchanged. Packaging and labelling has been updated to conform to current regulatory requirements but otherwise remains substantially unchanged. Precautions, warnings, contra-indications and software functionality remain unchanged.
| Parameter | Vitalograph model 6800<br>Pneumotrac | Predicate Device – Vitalograph<br>Model 2120 (K100687) |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Volume Range | 0-10 Liters | 0-10Litres |
| Max Flow Range | 0-16 Liters / second | 0-16 Liters / second |
| Back Pressure | Less than 0.1kPa/L/s | 0.1kPa/L/s |
| Accuracy FEV1 | +/- 3% | +/- 3% |
| Accuracy FEV6 | +/- 3% | +/- 3% |
| Accuracy PEF | +/-5% | +/-5% |
| Calibration | Using calibration syringe | Using calibration syringe |
| Technology | Fleisch<br>Pneumotachograph | Fleisch Pneumotachograph |
| Set Predicted /<br>reference values | Yes | Yes |
| Memory Type /<br>Storage: | Uses Spirotrac database.<br>{K141546} | On-board or connection to<br>Spirotrac database {K141546} |
| Sounds | Audible beeps emitted<br>whilst performing a test,<br>for incentives. | Audible beeps emitted whilst<br>performing a test, for incentives. |
| Communication | USB | USB |
| Download to P.C | Yes. Downloads to<br>Spirotrac {K141546} | Yes. Downloads to Spirotrac<br>{K141546} |
| To be serviced | Yes, Has Service Manual. | Yes. Has Service Manual. |
| Connection to<br>external printer | Yes, via Spirotrac<br>{K141546} | Yes. Direct or via Spirotrac<br>{{K141546}} |
| Internal printer | No | No |
| Battery Type | USB Power | USB Power and 3.7V Li-ion |
| power | | Battery. |
| Dimensions | 183x105x70mm | 160x100x45mm |
| Weight | 0.45Kg | 0.23Kg |
| Material Type | ABS plastic Body, Silicone<br>Rubber, Stainless Steel,<br>Aluminum, TPX plastic | ABS plastic Body (identical),<br>Silicone Rubber(identical),<br>Stainless Steel(identical), |
| Biocompatibility | No new testing required.<br>All materials previously<br>cleared in K100687 &<br>K925085. | |
| Operating Temp: | 10-40°C.<br>{At least 17-37°C required<br>per ATS 2005} | 10 to 40°C<br>{At least 17-37°C required per<br>ATS 2005} |
| Storage Temp | 0 - 50 °C | 0 - 50°C |
| Humidity | 10 - 95% relative<br>humidity | 10 - 95% relative humidity |
| Performance<br>Standards<br>{bench tests} | ATS ERS 2005, ISO<br>23747:2009 for PEF<br>{formerly EN13826:2003}.<br>EN ISO 26782:2009<br>ISO 10993-1 Fourth<br>Edition | ATS ERS 2005, ISO 23747:2009<br>for PEF {formerly<br>EN13826:2003}. EN ISO<br>26782:2009<br>ISO 10993-1 Fourth Edition |
| Compliance: | IEC / EN 60601 {EN<br>60601-1-1 and EN 60601-<br>1-2} | IEC / EN 60601 {EN 60601-1-1<br>and EN 60601-1-2} |
| Interface with<br>user | Spirotrac {K141546} | On-board Button Keypad /<br>touchscreen display or via<br>Spirotrac {{K141546}} |
| Regulatory | FDA - 510(k)<br>CE {0086} Class 2a | FDA - 510(k)<br>CE {0086} Class 2a |
| Product Code,<br>Class, CFR | BZG, Class 2, 868.1840 | BZG, Class 2, 868.1840 |
| Warranty | 1 Year | 1 Year |
| Patient interface<br>accessories | Model 2820 BVF<br>Mouthpiece (K942779)<br>{Class 2 with an active<br>device listing}. | Model 2820 BVF Mouthpiece<br>(K942779) {Class 2 with an active<br>device listing}.<br>Model 2020 SafeTway<br>Mouthpiece {Class 1 with an<br>active device listing}. |
| Indications for<br>Use | The device is a<br>spirometer which<br>measures patient<br>respiratory parameters<br>including FVC, FEV1,<br>FEV6, PEF, MVV and<br>VC. The Vitalograph<br>Pneumotrac is a<br>desktop spirometer<br>designed for lung<br>function testing for<br>use on adults and<br>pediatrics, 5 years and<br>older, in a variety of<br>environments such as<br>hospital wards, health<br>centers and private<br>homes under the<br>supervision of a<br>healthcare provider. | The device is a battery operated<br>spirometer which measures<br>three basic patient respiratory<br>parameters {FVC, MVV and VC}.<br>The model 2120 is a hand held<br>spirometer designed for lung<br>function testing in a variety of<br>environments such as hospital<br>wards, health centers and<br>private homes. The model 2120<br>can be configured as a stand-<br>alone spirometer or connected<br>to a printer. |
{7}------------------------------------------------
{8}------------------------------------------------
The Vitalograph Model 6800 Spirotrac underwent validation bench testing to ensure performance according to its specifications against current standards. These tests included performance testing against international standards such as
- ISO 26782{Anesthetic and respiratory equipment -- Spirometers intended ● for the measurement of time forced expired volumes in humans},
- ATS/ERS 2005 {ATS/ERS Task Force: Standardization of Lung Function Testing} and
- . ISO 23747 {Anesthetic and respiratory equipment -- Peak expiratory flow meters for the assessment of pulmonary function in spontaneously breathing humans}.
- IEC 60601-1 ●
Biocompatibility was reviewed during the risk management process. Separate biocompatibility testing was not carried out as all materials we previously cleared in earlier submissions including K100687 and K925085.
Validation of the interfaces with model 7000 Spirotrac {K141546} was performed to ensure that the integrity of the information is maintained and that the information may be successfully downloaded and stored within the Spirotrac database.
All tests and validations demonstrated satisfactory results. Evidence of successful completion of tests and validations has been provided with this submission.
# 11. Conclusion:
Based on the above, including the successful completion of all device testing Vitalograph conclude that Vitalograph model 6800 Pneumotrac is as safe and as effective as the predicate devices. No new issues of safety or effectiveness have been introduced as a result.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.