K142782 · Kimberly-Clark Health Care · FRG · Mar 9, 2015 · General Hospital
Device Facts
Record ID
K142782
Device Name
KIMGUARD ONE-STEP Sterilization Wrap
Applicant
Kimberly-Clark Health Care
Product Code
FRG · General Hospital
Decision Date
Mar 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
KIMGUARD ONE-STEP® Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using: - V-PRO® 60 Low Temperature Sterilization System that include: - Lumen Cycle - Non Lumen Cycle - Flexible Cycle KIMGUARD* ONE-STEP® Sterilization Wrap (KC100, KC200, KC300, KC400, KC500 and KC600) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.
Device Story
KIMGUARD ONE-STEP® Sterilization Wrap consists of two ultrasonically seamed sheets of nonwoven polypropylene fabric. Used in clinical settings by healthcare providers to wrap medical devices prior to sterilization in V-PRO® 60 Low Temperature Sterilization System. The wrap allows sterilant penetration and aeration during the sterilization cycle while maintaining a microbial barrier to preserve sterility of the enclosed instruments until point of use. The device is single-use and provided non-sterile.
Clinical Evidence
Bench testing only. Validation studies confirmed sterilant penetration, effective aeration, and maintenance of package integrity for 180 days post-sterilization in V-PRO® 60 Low Temperature Sterilization System (Lumen, Non-Lumen, and Flexible cycles). Biocompatibility testing also performed.
Indicated for use by healthcare providers to enclose medical devices (metal/non-metal, including rigid/semi-rigid/flexible endoscopes and instruments with diffusion-restricted spaces) for sterilization in V-PRO® 60 Low Temperature Sterilization System (Lumen, Non-Lumen, and Flexible cycles) and to maintain sterility until use.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 9, 2015
Halyard Health (formerly known as Kimberly-Clark Health Care) Thomas Kozma, Ph.D. Director of Regulatory Affairs 5450 Windward Parkway Alpharetta, GA 30004
Re: K142782
> Trade/Device Name: KIMGUARD* ONE-STEP® Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600) Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: February 12, 2015 Received: February 13, 2015
Dear Dr. Kozma,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Post market Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K142782
#### Device Name
.
KIMGUARD* ONE-STEP® Sterilization Wrap (Models KC100, KC200, KC300, KC500, and KC600)
#### Indications for Use (Describe)
KIMGUARD ONE-STEP® Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using:
- V-PRO® 60 Low Temperature Sterilization System that include:
- o Lumen Cycle
- Non Lumen Cycle o
- Flexible Cycle O
KIMGUARD ONE-STEP® Sterilization Wrap (KC100, KC200, KC300 and KC600) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.
Test results validated that KIMGUARD ONE-STEP® Sterilization Wraps (KC100, KC300, KC500, and KC600) allowed sterilization of the enclosed devices by the V-PRO® 60 Low Temperature Sterilization System (i.e., Lumen, Non Lumen, and Flexible Cycles). Additionally, the KIMGUARD ONE-STEP® Sterilization Wrap was validated to allow effective aeration under the pre-programmed sterilization cycles.
All models of the KIMGUARD ONE-STEP® Sterilization Wrap (KC100, KC300, KC300, KC500, and KC600) have been validated for use with the V-PRO® 60 Low Temperature Sterilization System cycles in Table 1.
KIMGUARD ONE-STEP® Sterilization Wraps (KC100, KC200, KC300, and KC600) Recommendations for Use with the V-PRO® 60 Low Temperature Sterilization System are provided in Table 2.
| TABLE 1: Validated V-PRO® 60 Low Temperature Sterilizer System Cycles | |
|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| V-PRO® 60 Low<br>Temperature<br>Sterilizer<br>Cycles | Intended Load |
| Lumen Cycle | Reusable metal and non-metal medical devices including instruments with diffusion-restricted spaces (such as the hinged<br>portion of forceps or scissors) and single, dual or triple channeled rigid/semi-rigid endoscopes, with the following<br>configurations:<br>Single or dual channeled devices with stainless steel lumens with An inside diameter of 0.77 mm or larger and a length of 410 mm or shorter Triple channeled devices with stainless steel lumens with An inside diameter of 1.2 mm or larger and a length of 257 mm or shorter. An inside diameter of 1.8 mm or larger and a length of 310 mm or shorter or An inside diameter of 2.8 mm or larger and a length of 317 mm or shorter |
| Non Lumen<br>Cycle | Reusable metal and non-metal non-lumened medical devices including non-lumened rigid, semi-rigid and flexible<br>endoscopes and medical devices with stainless steel or titanium diffusion-restricted spaces such as the hinged portion of<br>forceps or scissors. |
| Flexible Cycle | Single or dual channeled Flexible Surgical Endoscopes or Bronchoscopes with lumens that have:<br>An inside diameter of 1 mm or larger and a length of 990 mm or shorter. |
{3}------------------------------------------------
| TABLE 2: Recommended Loads for KIMGUARD ONE-STEP® Sterilization Wrap (KC100, KC200, KC300, KC400, KC500, & KC600) for use with V-PRO® 60 Low Temperature Sterilizer System | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| KIMGUARD ONE-STEP® Sterilization Wrap Models | Intended Loads1 | Maximum Wrapped Package Content Weights Used in Sterility Maintenance Validation Study2 | Description of Loads Used in Sterility Maintenance Validation Study2 |
| KC100 | Very Light Weight Package (for example batteries) | 3 lbs. | 3 lb of metal mass<br>6 surgical forceps<br>1 self-contained biological indicator (SCBI) (no tray was included) |
| KC200 | Light Weight Package (for example telescope with light cord) | 6.5 lbs. | • 2.5 lbs metal mass<br>• 6 surgical forceps<br>• 1 SCBI<br>• V-PRO tray (17" x 10" x 3½") at 4 lbs |
| KC300 | Light to Moderate Weight Package (for example: general use medical instruments) | 9 lbs. | • 5 lbs metal mass<br>• 6 surgical forceps<br>• 1 SCBI<br>• V-PRO tray (17" x 10" x 3½") at 4 lbs |
| KC400 | Moderate to Heavy Weight Package (for example: general use medical instruments) | 12 lbs. | • 8 lbs metal mass<br>• 6 surgical forceps<br>• 1 SCBI<br>• V-PRO tray (17" x 10" x 3½") at 4 lbs |
| KC500 | Heavyweight Package (for example: general use medical instruments) | 12 lbs. | • 8 lbs metal mass<br>• 6 surgical forceps<br>• 1 SCBI<br>• V-PRO tray (17" x 10" x 3½") at 4 lbs |
| KC600 | Very Heavy Weight Package (for example: general use medical instruments) | 12 lbs. | • 8 lbs metal mass<br>• 6 surgical forceps<br>• 1 SCBI<br>• V-PRO tray (17" x 10" x 3½") at 4 lbs |
| 1 Individual results may differ due to factors such as variations in handling practices, wrapping techniques, and folding methods. Results may also differ due to the use of irregularly shaped contents, which may put added stress on the wrap. Each healthcare facility should determine for itself which wrap model is most appropriate for each intended use.<br>2 It is recommended to not exceed the maximum wrapped package content weights indicated for each wrap model. Furthermore, it is recommended to not exceed the number, weight, and size of individual content types that were validated for KIMGUARD ONE-STEP® Sterilization Wraps. | | | |
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D)
区 | Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
### Applicant's Name, Address, Telephone, FAX, Contact Person:
Halyard Health (formerly known as Kimberly-Clark Health Care) 5450 Windward Parkwav. Alpharetta. GA 30004. USA
Establishment Registration Number: 1033422
| Contact Name: | Thomas Kozma, Director of Regulatory Affairs |
|------------------------|--------------------------------------------------------------------------------------------------------------------|
| E-mail: | thomas.kozma@hyh.com |
| Phone: | 470.448.5681 |
| FAX: | 678.254.0099 |
| DATE PREPARED: | March 6, 2015 |
| TRADE NAME: | KIMGUARD* ONE-STEP® Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600) |
| CLASSIFICATION NAME: | Sterilization Wrap |
| COMMON/USUAL NAME: | Sterilization Wrap |
| PRODUCT CODE: | FRG |
| DEVICE CLASSIFICATION: | Class II per 21 CFR §880.6850 |
| PREDICATE DEVICES: | K092167 and K112805 - KIMGUARD ONE-STEP® Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600). |
### INDICATIONS FOR USE
KIMGUARD ONE-STEP® Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using:
- V-PRO® 60 Low Temperature Sterilization System that include: ●
- o Lumen Cycle
- Non Lumen Cycle o
- Flexible Cvcle O
KIMGUARD* ONE-STEP® Sterilization Wrap (KC100, KC200, KC300, KC400, KC500 and KC600) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.
Test results validated that KIMGUARD ONE-STEP® Sterilization Wraps (KC100, KC200, KC300, KC400, KC500, and KC600) allowed sterilization of the enclosed devices by the V-PRO® 60 Low Temperature Sterilization System (i.e., Lumen, Non Lumen, and Flexible Cycles). Additionally, the KIMGUARD ONE-STEP® Sterilization Wrap was validated to allow effective aeration under the pre-programmed sterilization cycles. All models of the KIMGUARD ONE-STEP® Sterilization Wrap (KC100, KC200, KC300, KC400, KC500, and KC600) have been validated for use with the V-PRO® 60 Low Temperature Sterilization System cycles in TABLE 1.
{5}------------------------------------------------
| TABLE 1: Validated V-PRO® 60 Low Temperature Sterilizer System Cycles | |
|------------------------------------------------------------------------|--|
|------------------------------------------------------------------------|--|
| Cycles | Intended Load |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Lumen<br>Cycle1 | Reusable metal and non-metal medical devices including instruments with diffusion-restricted spaces<br>(such as the hinged portion of forceps or scissors) and single, dual or triple channeled rigid/semi-rigid<br>endoscopes, with the following configurations:<br>Single or dual channeled devices with stainless steel lumens with An inside diameter of 0.77 mm or larger and a length of 410 mm or shorter Triple channeled devices with stainless steel lumens with An inside diameter of 1.2 mm or larger and a length of 257 mm or shorter. An inside diameter of 1.8 mm or larger and a length of 310 mm or shorter or An inside diameter of 2.8 mm or larger and a length of 317 mm or shorter |
| Flexible<br>Cycle1 | Single or dual channeled Flexible Surgical Endoscopes or Bronchoscopes with lumens that have:<br>An inside diameter of 1 mm or larger and a length of 990 mm or shorter. |
| Non-Lumen<br>Cycle1 | Reusable metal and non-metal non-lumened medical devices including non-lumened rigid, semi-rigid<br>and flexible endoscopes and medical devices with stainless steel or titanium diffusion-restricted<br>spaces such as the hinged portion of forceps or scissors. |
KIMGUARD ONE-STEP® Sterilization Wraps (KC100, KC200, KC300, KC400, KC500, and KC600) Recommendations for Use with the V-PRO® 60 Low Temperature Sterilization System are provided in TABLE 2.
| Wrap Models | Intended Loads1 | Maximum Wrapped<br>Package Content Weights<br>MPI Study2 | Description of Loads Used in Sterility<br>Maintenance Validation Study2 |
|-------------|------------------------------------------------------------------------------------------|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| KC100 | Very Light Weight Package<br>(for example batteries) | 3 lbs. | 3 lb of metal mass<br>6 surgical forceps<br>1 self-contained biological indicator (SCBI)<br>(no tray was included) |
| KC200 | Light Weight Package (for<br>example telescope with light<br>cord) | 6.5 lbs. | • 2.5 lbs metal mass<br>• 6 surgical forceps<br>• 1 SCBI<br>• V-PRO® tray (17" x 10" x 3½") at 4 |
| KC300 | Light to Moderate Weight<br>Package (for example:<br>general use medical<br>instruments) | 9 lbs. | • 5 lbs metal mass<br>• 6 surgical forceps<br>• 1 SCBI<br>• V-PRO® tray (17" x 10" x 3½") at 4 |
| KC400 | Moderate to Heavy Weight<br>Package (for example:<br>general use medical<br>instruments) | 12 lbs. | • 8 lbs metal mass<br>• 6 surgical forceps<br>• 1 SCBI<br>• V-PRO® tray (17" x 10" x 3½") at 4 |
| KC500 | Heavyweight Package (for<br>example: general use<br>medical instruments) | 12 lbs. | • 8 lbs metal mass<br>• 6 surgical forceps<br>• 1 SCBI<br>• V-PRO® tray (17" x 10" x 3½") at 4 |
| KC600 | Very Heavy Weight Package<br>(for example: general use<br>medical instruments) | 12 lbs. | • 8 lbs metal mass<br>• 6 surgical forceps<br>• 1 SCBI<br>• V-PRO® tray (17" x 10" x 3½") at 4 |
number, weight, and size of individual content types that were validated for KIMGUARD ONE-STEP® Sterilization Wraps.
{6}------------------------------------------------
# Description of Device
KIMGUARD* ONE-STEP® Sterilization Wrap is comprised of two sheets of KIMGUARD* Sterilization Wrap that is ultrasonically seamed on two edges. This seamed configuration allows for convenient wrapping of an article using two sheets simultaneously.
The blue or white sheets of sterilization wrap are square or rectangular fabric produced using a three-layer SMS (spunbound-meltblown-spunbound) process. The blue wrap fabric is composed of polypropylene with the addition of less than 2% by weight of phthalocyanine blue pigment, less than 1% by weight titanium dioxide pigment, and less than 0.008% by weight of anti-static treatment. The white sheet has the same material composition but contains no blue pigment. The wrap allows a sterilized package to be opened aseptically.
## Substantial Equivalence to Predicate Device
KIMGUARD ONE-STEP® Sterilization Wrap (i.e., subject of this Premarket Notification) is substantially equivalent to the predicate Kimberly-Clark KIMGUARD* ONE-STEP® Sterilization Wraps (K092167 & K112805) in technology, design, and materials.
The following table compares the subject KIMGUARD* ONE-STEP® Sterilization Wrap to the predicate KIMGUARD* ONE-STEP® Sterilization Wrap.
{7}------------------------------------------------
# Device Comparison Table (Technological, Design, & Materials)
| Characteristics | Predicate Devices:<br>KIMGUARD* ONE-STEP® Sterilization<br>Wrap (KC100, KC200, KC300, KC400,<br>KC500, & KC600) (K092167 & 112805) | Proposed Device:<br>KIMGUARD* ONE-STEP® Sterilization<br>Wrap (KC100, KC200, KC300, KC400,<br>KC500, & KC600) |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Kimberly-Clark Corporation | Kimberly Clark Corporation |
| Regulation/Product<br>Code | Sterilization Wrap: 880.6850 / FRG | Sterilization Wrap: 880.6850 / FRG |
| Indications for Use | The device is intended to be used to enclose<br>another medical device that is to be sterilized by a<br>healthcare provider by the Amsco® V-PRO™ 1<br>Plus Low Temperature Sterilization System's<br>Lumen (identical to the V-PRO™ 1 Cycle) and<br>Non-Lumen Cycles, and the V-PRO™ Low<br>Temperature Sterilization System's Flexible Cycle.<br>The wrap is intended to allow sterilization of the<br>enclosed medical device(s) and also to maintain<br>sterility of the enclosed device(s) until opened<br>within the period of time for which performance<br>data demonstrating maintenance of sterility has<br>been provided. The KIMIGUARD ONE-STEP*<br>Sterilization Wrap was validated to be effectively<br>aerated during the pre-programmed V-PRO™, V-<br>PRO™ 1 Plus, and V-PRO™ Flexible Cycles. | KIMGUARD ONE-STEP® Sterilization Wraps are<br>intended to enclose another medical device that<br>is to be sterilized by a healthcare provider using:<br>· V-PRO® 60 Low Temperature Sterilization<br>System that include:<br>o Lumen Cycle<br>o Non Lumen Cycle<br>o Flexible Cycle<br>KIMGUARD* ONE-STEP® Sterilization Wrap<br>(KC100, KC200, KC300, KC400, KC500 and<br>KC600) are intended to allow sterilization of the<br>enclosed medical device(s) and also maintain<br>sterility of the enclosed device(s) until used. |
| Sterilization Cycles | Amsco® V-PRO™ 1 Plus Low Temperature<br>Sterilization System's Lumen (identical to the<br>V-PRO™ 1 Cycle) and Non-Lumen Cycles,<br>and the V-PRO max Low Temperature<br>Sterilization System's Flexible Cycle | V-PRO® 60 Low Temperature<br>Sterilization System:<br>o Lumen Cycle<br>o Non Lumen Cycle<br>o Flexible Cycle |
| Maintenance of<br>Package Sterility | For models KC100, KC200, KC300, K400,<br>KC500, and KC600 for at least 30 days. | Real-time testing following sterilization using<br>the V-PRO® 60 Low Temperature<br>Sterilization System supports maintenance<br>of package sterility for 180 days for all<br>models of KIMGUARD* ONE-STEP® Sterilization<br>Wrap. |
| Technology | Tortuous sheet material used to enclose<br>medical devices that are to be sterilized by a<br>healthcare provider to allow sterilization of<br>the enclosed medical device(s) and maintain<br>sterility of the enclosed device(s) until used. | Tortuous sheet material used to enclose<br>medical devices that are to be sterilized by a<br>healthcare provider to allow sterilization of<br>the enclosed medical device(s) and maintain<br>sterility of the enclosed device(s) until used. |
| Device Design | Two sheets of nonwoven polypropylene<br>fabric. Each sheet is composed of three<br>thermally- bonded layers consisting of a<br>Meltblown polypropylene layer surrounded by<br>Spunbound polypropylene layers (SMS) | Two sheets of nonwoven polypropylene<br>fabric. Each sheet is composed of three<br>thermally- bonded layers consisting of a<br>Meltblown polypropylene layer surrounded<br>by Spunbound polypropylene layers (SMS) |
| Method for bonding<br>SMS layers | Thermal bonding with round pin, hexagonal,<br>triangle bond pattern ("daisy" pattern) | Thermal bonding with round pin, hexagonal,<br>triangle bond pattern ("daisy" pattern) |
| Materials | Polypropylene with blue and white<br>pigments | Polypropylene with blue and white<br>pigments |
| Distribution | Non-Sterile and<br>Over-the-Counter | Non-Sterile and<br>Over-the-Counter |
| Single Use Device | Yes | Yes |
{8}------------------------------------------------
# SUMMARY OF NONCLINICAL TESTS
Performance of KIMGUARD* ONE-STEP® Sterilization Wrap (KC100, KC200, KC300, KC400, KC500, KC600) has been tested in accordance with the applicable requirements. All results of testing met acceptance criteria demonstrating that the KIMGUARD* ONE-STEP® Sterilization Wrap allows sterilization of contents by the V-PRO® 60 Low Temperature Sterilization System (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle) and maintains sterility of contents until used.
| Summary of Testing Performed | Results |
|-------------------------------------------------------------------------------------------------------------------|---------|
| V-PRO® 60 Low Temperature Sterilization System<br>Sterilant Penetration | Passed |
| Post V-PRO® 60 Low Temperature Sterilization System<br>Sterilization - Maintenance of Package Integrity (30 Days) | Passed |
| Post V-PRO® 60 Low Temperature Sterilization System<br>Sterilization - Performance Testing | Passed |
| Post V-PRO® 60 Low Temperature Sterilization System<br>Sterilization - Material Biocompatibility | Passed |
## Overall Performance Conclusions
The nonclinical studies demonstrate that the KIMGUARD* ONE-STEP® Sterilization Wrap performs as intended as a sterilization packaging system of medical devices when terminally sterilized in the V-PRO® 60 Low Temperature Sterilization System (Lumen Cycle, Non Lumen Cycle, and Flexible Cycle). These studies demonstrate that the KIMGUARD* ONE-STEP® Sterilization Wrap met the same criteria as the predicate devices and are substantially equivalent.
## BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
Based on the intended use, technological characteristics, performance data and nonclinical tests performed, the subject KIMGUARD* ONE-STEP® Sterilization Wrap is substantially equivalent and is as safe and as effective as the legally marketed predicate devices, K092167 and K112805.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.