EOS System

K142773 · Eos Imaging · KPR · Jan 22, 2015 · Radiology

Device Facts

Record IDK142773
Device NameEOS System
ApplicantEos Imaging
Product CodeKPR · Radiology
Decision DateJan 22, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2
AttributesPediatric

Indications for Use

EOS is intended for use in general radiographic examinations and applications, excluding the evaluation of lung nodules and examinations involving fluoroscopy, angiography, and mammography. EOS allows the radiographic acquisition of either one or two orthogonal X-ray images for diagnostic purposes, in one single scan, of the whole body or a reduced area of investigation of a patient in the upright or seated position. The Micro Dose feature is indicated for imaging with a patient entrance dose of 10 to 90 µGy for assessing global skeletal deformities in follow-up pediatric examinations. Micro Dose is not indicated for focal skeletal abnormalities and/or other pediatric abnormalities. Micro Dose is not indicated for use in patients with a Body Mass Index over 30.

Device Story

Digital radiography system; uses two orthogonal X-ray tubes and xenon gas-filled detectors to acquire simultaneous frontal and lateral images in one scan. Operates in upright or seated positions. Includes 'Micro Dose' feature for low-dose (10-90 µGy) pediatric follow-up imaging. Images stored in local database; displayed on high-resolution medical monitors for physician diagnosis. Transmits via DICOM 3.0 for archiving/printing. Benefits pediatric patients by reducing radiation exposure during skeletal deformity follow-ups.

Clinical Evidence

Bench and clinical testing performed. Bench testing included safety (IEC standards) and image quality assessment. Clinical testing evaluated visibility of anatomical landmarks for scoliosis follow-up and measured reproducibility/accuracy of scoliosis angles. Results confirmed Micro Dose feature generates images sufficient for global skeletal deformity assessment in follow-up exams where fine bone structure analysis is not required.

Technological Characteristics

Digital radiography system; two sets of xenon gas-filled detectors and X-ray tubes in orthogonal configuration. Features 'Micro Dose' acquisition mode (10-90 µGy). Connectivity via DICOM 3.0. Images displayed on high-resolution medical-quality monitors.

Indications for Use

Indicated for general radiographic examinations (excluding lung nodules, fluoroscopy, angiography, mammography) in upright or seated patients. Micro Dose feature indicated for follow-up pediatric skeletal deformity assessment (10-90 µGy entrance dose); contraindicated for focal skeletal abnormalities, other pediatric abnormalities, and patients with BMI > 30.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 22, 2015 EOS Imaging % Mr. John J. Smith Regulatory Counsel Hogan Lovells US L.L.P. 555 Thirteenth Street, NW WASHINGTON DC 20004 Re: K142773 Trade/Device Name: EOS System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR, MOB Dated: December 12, 2014 Received: December 12, 2014 Dear Mr. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert A. Ochs Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K142773 Device Name EOS System #### Indications for Use (Describe) EOS is intended for use in general radiographic examinations and applications, excluding the evaluation of lung nodules and examinations involving fluoroscopy, angiography, and mammography. EOS allows the radiographic acquisition of either one or two orthogonal X-ray images for diagnostic purposes, in one single scan, of the whole body or a reduced area of investigation of a patient in the upright or seated position. The Micro Dose feature is indicated for imaging with a patient entrance dose of 10 to 90 µGy for assessing global skeletal deformities in follow-up pediatric examinations. Micro Dose is not indicated for focal skeletal abnormalities and/or other pediatric abnormalities. Micro Dose is not indicated for use in patients with a Body Mass Index over 30. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) SUMMARY ### EOS imaging's EOS System | EOS imaging | |-----------------| | 10 rue Mercoeur | | PARIS F-75011 | | FRANCE | | Phone: | + 33 1 55 25 60 60 | |------------|--------------------| | Facsimile: | + 33 1 55 25 60 61 | | Contact Person: | Karine Chevrie, Quality and Regulatory Affairs Officer | |-----------------|--------------------------------------------------------| | Date Prepared: | January 22, 2015 | Name of Device and Name/Address of Sponsor: | EOS System | |-----------------| | EOS imaging | | 10 rue Mercoeur | | PARIS F-75011 | | FRANCE | | Common or Usual Name: | Digital Radiography System | |----------------------------|-----------------------------------------------------------------------------------------------| | Classification Regulation: | 21 C.F.R. § 892.1680; Stationary X-Ray System<br>Product Code KPR – System, X-Ray, Stationary | Predicate Devices: EOS imaging's EOS (K123740) ## Intended Use/ indication for use EOS is intended for use in general radiographic examinations and applications, excluding the evaluation of lung nodules and examinations involving fluoroscopy, angiography and mammography. EOS allows the radiographic acquisition of either one or two orthogonal X ray images for diagnostic purposes, in one single scan, of the whole body or a reduced area of investigation of a patient in the upright or seated position. The Micro Dose feature is indicated for imaging with a patient entrance dose of 10 to 90 µGy for assessing global skeletal deformities in follow-up pediatric examinations. Micro Dose is not indicated for focal skeletal abnormalities and/or other pediatric abnormalities. Micro Dose is not indicated for use in patients with a Body Mass Index over 30. ## Technological Characteristics EOS is a digital radiography system in which two sets of xenon gas filled digital detectors and Xray tubes are positioned orthogonally to generate frontal and lateral images simultaneously by scanning the patient over the area of interest. A new acquisition feature named Micro Dose allows image acquisition with a patient entrance dose of 10 to 90 µGy for assessing global skeletal deformities in follow-up pediatric exams. The diagnostic images are stored in a local {4}------------------------------------------------ database and are displayed on a high-resolution, medical-quality monitor, where the diagnosis is performed. The diagnostic image can be transmitted through a DICOM 3.0 compatible digital network for printing and archiving. ## Performance Data The company's performance testing was designed to confirm performance of the EOS System at a dose level used by the Micro Dose feature. Performance testing included safety testing according to IEC, image quality bench testing, and clinical testing. The objective of these performance tests was to confirm the ability of the EOS System's Micro Dose acquisition feature to generate very low X-ray dose images allowing for the follow-up assessment of global skeletal deformities in pediatric patients. The Micro Dose image quality was assessed by: (1) rating the visibility of the anatomical landmarks used for the measurement of the clinical parameters for spine scoliosis follow up; and (2) assessing the reproducibility and the accuracy of the measurements of clinically relevant scoliosis angles. Based on the results of these performance tests it was concluded that the Micro Dose feature enables the analysis of global skeletal deformity in the context of radiographic examinations which do not require fine bone structure analysis, such as follow-up exams of the spine or lower limb. In such exams, the patient entrance dose of 10 to 90 µGy associated with the Micro Dose feature is an advantage, especially in pediatric patients who are more susceptible to radiation. ## Substantial Equivalence The EOS is essentially the same device as its predicate device with the exception of the revised indications for use that includes the indication for use of the Micro Dose acquisition protocol, which is more specific compared to the indication for use of higher X-ray dose acquisition settings. The Micro Dose feature is indicated for imaging with a patient entrance dose of 10 to 90 µGy for assessing global skeletal deformities in follow-up pediatric examinations. This change does not alter the device's diagnostic effect, and performance testing demonstrates safety and performance comparable to the predicate. Thus, the EOS is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...