← Product Code [LYZ](/productcode/LYZ) · K142771

# Xuanye Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) (K142771)

_Xuanhua Metallurgical Industry Co., Ltd. · LYZ · Dec 22, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K142771

## Device Facts

- **Applicant:** Xuanhua Metallurgical Industry Co., Ltd.
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** Dec 22, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Device consists of non-sterile, powder-free vinyl (PVC) patient examination gloves; intended for single-patient use. PVC film acts as a physical barrier against body fluids and bloodborne pathogens. Tensile properties allow the material to conform to the hand for procedural movement. Used in clinical settings by healthcare providers to prevent contamination between patient and examiner. No electronic components or software involved.

## Clinical Evidence

Bench testing only. No clinical data required or provided. Performance verified via ASTM D5250-06 (dimensions, physical properties), ASTM D5151-06 (freedom from pinholes), ASTM D6124-06 (powder residual), and ISO 10993-10 (biocompatibility).

## Technological Characteristics

Material: Poly (vinyl chloride) (PVC). Surface coating: PU. Standards: ASTM D5250-06 (physical properties/dimensions), ASTM D5151-06 (pinholes), ASTM D6124-06 (powder residual), ISO 10993-10 (biocompatibility). Non-sterile. Disposable. Form factor: Hand-worn glove in sizes Small to X-Large.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Powder-Free Vinyl Patient Examination Glove (Non-colored) ([K091663](/device/K091663.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health and Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 22, 2014

Xuanhua Metallurgical Industry Co., Ltd C/O Mr. Chu Xiaoan Rm. F302 Bldg., 41. Jing Cheng Ya Ju. Courtyard 6 of Southern Dou Ge Zhuang Beijing, 100121 CN Chaoyang District, Beijing 100121 CHINA

Re: K142771

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYZ Dated: December 1, 2014 Received: December 11, 2014

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Susan Runno, DDS, MA

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K142771

### Device Name

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)

Indications for Use (Describe)

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### Section C 510(k) Summary

# 510(K) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: K142771

### Premarket Notification [510(k)] Summary

#### 1.0 Submitter:

| Submitter's name :      | Xuanhua Metallurgical Industry Co., Ltd.                                                                    |
|-------------------------|-------------------------------------------------------------------------------------------------------------|
| Submitter's address :   | Bridge North, Weisan Rd, East Ring Rd, High-Tec District, Zhangjiakou City, Hebei Province.075000,P.R.China |
| Phone number :          | 86-313-6216205                                                                                              |
| Fax number :            | 86-313-5809316                                                                                              |
| Name of contact person: | Mr.Dong Yanlong                                                                                             |
| Date of preparation :   | 2014-09-14                                                                                                  |

#### 2.0 Name of the Device

| Device Name:            | Powder Free Vinyl Patient Examination<br>Gloves, Clear (non-colored)        |
|-------------------------|-----------------------------------------------------------------------------|
| Proprietary/Trade name: | XUANYE Powder Free Vinyl Patient<br>Examination Gloves, Clear (non-colored) |
| Common Name:            | Exam gloves                                                                 |
| Classification Name:    | Patient examination glove                                                   |
| Device Classification:  | I                                                                           |
| Regulation Number:      | 21 CFR 880.6250                                                             |
| Panel:                  | General Hospital (80)                                                       |
| Product Code:           | LYZ                                                                         |

### 3.0 Predicate device

| Device Name:   | Powder-Free Vinyl Patient Examination Glove<br>(Non-colored) |
|----------------|--------------------------------------------------------------|
| Company name:  | Zhang Jia Gang Fengyuan Plastic Product Co.Ltd.              |
| 510(K) Number: | K091663.                                                     |

### 4.0 Device Description:

#### 4.1 How the device functions:

PVC films form a barrier to body fluids and bloodborne Pathogens

#### 4.2 Scientific concepts that form the basis for the device

The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a 

{4}------------------------------------------------

medical procedure.

### Physical and performance characteristics such as design, materials and 4.3 physical properties:

Poly (vinyl chloride) glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.

## 5.0 Device Intended Use (Indication for use):

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## 6.0 Summary of the Technological Characteristics of the Device:

The Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

| Characteristics          | Standard                                                                         | Device performance                         |
|--------------------------|----------------------------------------------------------------------------------|--------------------------------------------|
| Dimension                | ASTM standard D<br>5250-06(Reapproved 2011).                                     | Meets                                      |
| Physical Properties      | ASTM standard D<br>5250-06(Reapproved 2011).                                     | Meets                                      |
| Freedom from<br>pinholes | 21 CFR 800.20                                                                    | Meets                                      |
| Powder Residual          | ASTM standard D 5250-06<br>(Reapproved 2011).and<br>D6124-06(Reaffirmation 2011) | Meets<br><2mg/glove                        |
| Biocompatibility         | Primary Skin Irritation in rabbits<br>ISO 10993-10: 2010-08-01                   | Passes<br>Not a Primary Skin<br>Irritation |
|                          | Dermal sensitization in the<br>guinea pig<br>ISO 10993-10: 2010-08-01            | Passes<br>Not a Dermal<br>sensitization    |

## 7.0 Substantial Equivalent Based on Assessment of Nan-Clinical Performance Data:

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), meet requirements per ASTM D5250-06 (Reapproved 2011), per ASTM D6124-06 Reaffirmation 2011), per 21 CFR 800.20 and ISO 10993-10:2010-08-01.

The performance test data of the non clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.

#### 8.0 Substantial Equivalent Based on Assessment of Clinical Performance Data:

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

#### Substantial Equivalence Comparison: 9.0

{5}------------------------------------------------

| Features<br>&<br>Description                                                                         | Predicate Device                                                                                                                                                                                                                                  | Subject Device                                                                                                                                                                                                                                      | Result<br>of<br>Comparison |
|------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Company                                                                                              | Zhang Jia Gang Fengyuan<br>Plastic Product Co.Ltd.                                                                                                                                                                                                | Xuanhua<br>Metallurgical<br>Industry Co., Ltd.                                                                                                                                                                                                      | --                         |
| 510(K) Number                                                                                        | K091663                                                                                                                                                                                                                                           | K142771                                                                                                                                                                                                                                             |                            |
| Product name                                                                                         | Powder Free Vinyl Patient<br>Examination Gloves, Clear<br>(Non-colored)                                                                                                                                                                           | Powder Free Vinyl Patient<br>Examination Gloves, Clear<br>(non-colored)                                                                                                                                                                             | same                       |
| Product Code                                                                                         | LYZ                                                                                                                                                                                                                                               | LYZ                                                                                                                                                                                                                                                 | same                       |
| Size                                                                                                 | Small/ Medium/<br>Large/X large                                                                                                                                                                                                                   | Small/ Medium/<br>Large/X large                                                                                                                                                                                                                     | same                       |
| Intend for use                                                                                       | Powder free Vinyl Patient<br>Examination Gloves,<br>Clear(Non-colored)is a<br>disposable device intended<br>for medical purposes that is<br>worn on the examiner's hand<br>or finger to prevent<br>contamination between<br>patient and examiner. | Powder free Vinyl Patient<br>Examination Gloves, Clear<br>(Non-colored) is a<br>disposable device intended<br>for medical purposes that is<br>worn on the examiner's hand<br>or finger to prevent<br>contamination between<br>patient and examiner. | same                       |
| Device Description<br>and Specifications                                                             | Meets ASTM D5250-06<br>(Reapproved 2011)                                                                                                                                                                                                          | Meets ASTM D5250-06<br>(Reapproved 2011)                                                                                                                                                                                                            | same                       |
| Dimensions<br>--Length                                                                               | Meets ASTM<br>D5250 -06<br>(Reapproved 2011)<br>≥230mm min                                                                                                                                                                                        | 230mm min for all sizes                                                                                                                                                                                                                             | same                       |
| Dimensions<br>-- Width                                                                               | Meets ASTM D5250-06<br>(Reapproved 2011)<br>Small<br>80-90 mm<br>Medium<br>90-100mm<br>Large<br>100-110mm<br>Xlarge<br>110-120 mm                                                                                                                 | Small<br>80-85 mm<br>Medium<br>94-100mm<br>Large<br>102-108mm<br>X large<br>113-116 mm                                                                                                                                                              | same                       |
| Dimensions<br>--Thickness                                                                            | Meets ASTM D5250-06<br>(Reapproved 2011)<br>Finger 0.05mm min.<br>Palm 0.08mm min.                                                                                                                                                                | Thickness (mm) min.<br>Finger 0.09-0.11<br>Palm 0.09-0.12                                                                                                                                                                                           | same                       |
| Physical Properties                                                                                  | Meets ASTM D 5250-06<br>(Reapproved 2011)<br>Before aging/after aging<br>Elongation ≥300%<br>Tensile Strength≥ 14MPa                                                                                                                              | Before aging/after aging<br>Elongation :390-420%<br>Tensile Strength:15-20 MPa                                                                                                                                                                      | same                       |
| Freedom<br>from<br>Pinholes                                                                          | Meets<br>21 CFR 800.20<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D 5151-06<br>(Reapproved 2011)                                                                                                                                               | Meets ASTM<br>D5151-06<br>(Reapproved 2011)<br>Holes at<br>Inspection Level I AQL2.5                                                                                                                                                                | same                       |
| Residual Powder                                                                                      | Meets ASTM<br>D 6124-06<br>(Reapproved 2011)<br>below 2mg of residual<br>powder                                                                                                                                                                   | Meets ASTM<br>D 6124-06<br>(Reapproved 2011)<br>Results generated values<br>below 2mg of residual<br>powder                                                                                                                                         | same                       |
| Materials used to<br>fabricate the<br>devices                                                        | PVC                                                                                                                                                                                                                                               | PVC                                                                                                                                                                                                                                                 | same                       |
| Dusting or<br>Donning Powder:                                                                        | PU                                                                                                                                                                                                                                                | PU                                                                                                                                                                                                                                                  | same                       |
| Dusting or<br>Donning Powder:<br>name                                                                | PU                                                                                                                                                                                                                                                | Surface Coating Agent                                                                                                                                                                                                                               | same                       |
| Compare<br>performance data<br>supporting<br>substantial<br>equivalence                              | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reaffirmation 2011)                                                                                                                        | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reaffirmation 2011)                                                                                                                          | same                       |
| Single Patient Use                                                                                   | Single Patient Use                                                                                                                                                                                                                                | Single Patient Use                                                                                                                                                                                                                                  | same                       |
| Biocompatibility                                                                                     | Under the conditions of this<br>study, the test article was a<br>non-irritant or non-<br>sensitizer.<br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION STUDIES<br>Meets ISO<br>10993-10:2002/Amd.1:2006                                           | Under the conditions of this<br>study, the test article was a<br>non-irritant or non-<br>sensitizer.<br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10 Third Edition<br>2010-08-01                                 | same                       |
| Labeling for the<br>legally marketed<br>device to which<br>substantial<br>equivalence is<br>claimed. | -Powder Free<br>-Patient Examination Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot                                                                                                                                                    | -Powder Free<br>-Patient Examination Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot                                                                                                                                                      | same                       |

{6}------------------------------------------------

#### Substantial Equivalence Comparison: 10.0

It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL meet labeling claims.

It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear is as safe, as effective, and performs as well as the predicate device, (non-colored) Powder-Free Vinyl Patient Examination Glove (Non-colored) Zhang Jia Gang Fengyuan Plastic Product Co., Ltd. K091663.

---

**Source:** [https://fda.innolitics.com/device/K142771](https://fda.innolitics.com/device/K142771)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com