The ActiveCare+S.F.T. and ActiveCare+DTx Systems are portable, ambulatory, sequential, intermittent pneumatic compression devices (IPCDs) prescribed by health care professionals. The systems include a rechargeable battery powered option allowing patient mobility and ease of use. These devices simulate muscle contractions in order to treat or enhance blood flow velocity in individuals experiencing venous impairment or reduced pulsatility (dysfunction of the muscle pump) when blood flow may become challenged or compromised such as during and after major orthopedic surgery procedures e.g. total joint (hip and knee) arthroplasty. They are intended for use in the clinical setting or home environment. These devices are indicated for use in: - Preventing Deep Vein Thrombosis (DVT) - Diminishing post-operative pain and swelling - Reducing wound healing time - Patients at risk for deep vein thrombosis (DVT) and related pulmonary embolism (PE) (Venous Thromboembolism (VTE)) - Treatment of venous stasis - Treatment and assistance in healing: Stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers - Enhancing blood circulation - Treatment of chronic venous insufficiency - Reducing edema. The ActiveCare+S.F.T. and ActiveCare+DTx Systems are intended to provide external compression in synchrony with the specific patient's natural venous blood flow return profile in order to achieve a high pulsatile venous blood flow. In addition, the ActiveCare+DTx System can detect hemodynamic changes in venous blood flow.
Device Story
Portable, ambulatory, sequential intermittent pneumatic compression devices (IPCDs); used in clinical or home settings by patients or healthcare professionals. Systems apply sequential pressure to lower limbs (foot, calf, thigh) via single or multi-celled sleeves. Devices monitor venous phasic blood flow in real-time to synchronize compression with patient's natural venous return profile, maximizing pulsatile blood flow. ActiveCare+DTx variant includes capability to detect hemodynamic changes potentially indicative of venous obstruction; not a diagnostic tool. Output provides external compression to simulate muscle contractions, enhancing circulation, reducing edema, and preventing DVT. Battery-powered for mobility.
Clinical Evidence
Bench testing including electrical safety, EMC, usability, software validation, and risk analysis. Clinical evidence supported by published literature. No new clinical trials conducted for this specific submission.
Technological Characteristics
Portable, battery-powered or AC/DC sequential intermittent pneumatic compression system. Features single-celled foot sleeves and three-celled calf/thigh cuffs. Real-time venous blood flow monitoring for synchronization. Software-controlled compression cycles.
Indications for Use
Indicated for patients at risk of DVT/PE (VTE), venous impairment, or reduced muscle pump pulsatility; post-operative patients (e.g., total joint arthroplasty); patients requiring treatment for venous stasis, chronic venous insufficiency, edema, or healing assistance for stasis dermatitis, venous stasis ulcers, and arterial/diabetic leg ulcers.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
Predicate Devices
ActiveCare+S.F.T. and ActiveCare+DTx Systems (K140755)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 4, 2014
Medical Compressions System (DBN) Ltd. Adely Levy RA & QA General Manager 12 Ha'ilan Street, PO Box 75 Or Akiva, Israel 30600
Re: K142728
> Trade/Device Name: Activecare+SFT; Activecare+SFT Homecare; Activecare+DTx ; Activecare+DTx Homecare Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: September 19, 2014 Received: September 23, 2014
Dear Adely Levy,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hilleman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known):K142728
Device Name: ActiveCare+S.F.T. and ActiveCare+DTx Systems
The ActiveCare+S.F.T. and ActiveCare+DTx Systems are portable, ambulatory, sequential. intermittent pneumatic compression devices (IPCDs) prescribed by health care professionals. The systems include a rechargeable battery powered option allowing patient mobility and ease of use. These devices simulate muscle contractions in order to treat or enhance blood flow velocity in individuals experiencing venous impairment or reduced pulsatility (dysfunction of the muscle pump) when blood flow may become challenged or compromised, such as during and after major orthopedic surgery procedures e.g total joint (hip and knee) arthroplasty. They are intended for use in the clinical setting or home environment.
These devices are indicated for use in:
- Preventing Deep Vein Thrombosis (DVT) ●
- Diminishing post-operative pain and swelling ●
- Reducing wound healing time ●
- Patients at risk for deep vein thrombosis (DVT) and related pulmonary embolism (PE) ● (Venous Thromboembolism (VTE))
- Treatment of venous stasis
- Treatment and assistance in healing: Stasis dermatitis, venous stasis ulcers, arterial and ● diabetic leg ulcers
- Enhancing blood circulation ●
- Treatment of chronic venous insufficiency
- Reducing edema
The ActiveCare+S.F.T. and ActiveCare+DTx Systems are intended to provide external compression in synchrony with the specific patient's natural venous blood flow return profile in order to achieve a high pulsatile venous blood flow.
In addition, the ActiveCare+DTx System can detect hemodynamic changes in venous blood flow.
Prescription Use __X__________________ AND/OR Over-The-Counter Use (Per 21 C.F.R. 801 Subpart D) (Per 21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
# 510(K) SUMMARY: K142728
Medical Compression Systems (DBN) Ltd's ActiveCare+S.F.T and ActiveCare+DTx
Date Prepared: November 21, 2014
| Submitter and Manufacturer | Contact Person |
|-------------------------------------------|--------------------------------|
| Medical Compression Systems (DBN) Ltd. | Adely Levy |
| 12 Ha'ilan Street, PO Box 75 | 12 Ha'ilan Street, P.O. Box 75 |
| Or Akiva 30600, Israel | Or Akiva 30600, Israel |
| Tel:+972(4) 6266630 | Telephone:+972 (4) 6266630 |
| Fax: +972 (4) 6266640 | Fax: +972 (4) 6266640 |
| E-mail: adely@mcsmed.com | E-mail: adely@mcsmed.com |
| Manufacturer Registration Number: 9616558 | |
#### Name of Device
| Trade Names: | ActiveCare+S.F.T. and ActiveCare+DTx Systems |
|----------------------|----------------------------------------------|
| Common Names: | Pneumatic Compression System |
| Classification Name: | Compressible Limb Sleeve |
## Device Classification/FDA Reviewing Branch
The Division of Cardiovascular Devices has classified Compressible Limb Sleeves as Class II devices pursuant to 21 C.F.R. § 870.5800 (JOW).
#### Predicate Device Information
| Predicate Device Name | 510(k) | Manufacturer |
|-------------------------------------------------------|---------|-----------------------------|
| ActiveCare+S.F.T. and<br>ActiveCare+DTx Systems | K140755 | Medical Compression Systems |
| VENAPRO | K133274 | Innovamed Health |
| Kendall SCD 700 Sequential<br>Compression Controller, | K102737 | Covidien |
| Pulstar Logix Pump | K141609 | AlbaHealth LLC |
{4}------------------------------------------------
#### Intended Use
The ActiveCare+S.F.T. and ActiveCare+DTx Systems are portable, ambulatory, sequential, intermittent pneumatic compression devices (IPCDs) prescribed by health care professionals. The systems include a rechargeable battery powered option allowing patient mobility and ease of use. These devices simulate muscle contractions in order to treat or enhance blood flow velocity in individuals experiencing venous impairment or reduced pulsatility (dysfunction of the muscle pump) when blood flow may become challenged or compromised such as during and after major orthopedic surgery procedures e.g. total joint (hip and knee) arthroplasty. They are intended for use in the clinical setting or home environment.
These devices are indicated for use in:
- . Preventing Deep Vein Thrombosis (DVT)
- . Diminishing post-operative pain and swelling
- . Reducing wound healing time
- Patients at risk for deep vein thrombosis (DVT) and related pulmonary ● embolism (PE) (Venous Thromboembolism (VTE))
- . Treatment of venous stasis
- Treatment and assistance in healing: Stasis dermatitis, venous stasis ulcers, . arterial and diabetic leg ulcers
- . Enhancing blood circulation
- Treatment of chronic venous insufficiency ●
- Reducing edema .
The ActiveCare+S.F.T. and ActiveCare+DTx Systems are intended to provide external compression in synchrony with the specific patient's natural venous blood flow return profile in order to achieve a high pulsatile venous blood flow.
In addition, the ActiveCare+DTx System can detect hemodynamic changes in venous blood flow.
#### Device Description/Technological Characteristics
The ActiveCare+DTx and ActiveCare+S.F.T. Systems are prescriptive, portable, sequential, intermittent pneumatic compression devices designed to apply sequential compression to the lower limb. The systems include a rechargeable battery powered option (in addition to an AC/DC adapter) allowing patient mobility and ease of use. The control units of the Systems provide the user with several treatment options: compression of the foot - single or double, compression of the calf – single or double, compression of the thigh – single or double, and combined compression of any combination of two sleeves. The foot compression program is an intermittent pressure pulse application to a single celled foot sleeve. The calf and thigh compression program is a sequential intermittent application of a pressure to a three-celled cuff sleeve.
{5}------------------------------------------------
The ActiveCare+S.F.T. and ActiveCare+DTx Systems are intended to provide continuous monitoring of each patient's venous phasic blood flow on a real time basis so that the timing of external compression is in synchrony with the specific patient's natural venous blood flow return profile in order to achieve a high pulsatile venous blood flow.
In addition, the ActiveCare+DTx System can detect hemodynamic changes that may be indicative of the development of a venous blood flow obstruction event in the treated patient's limbs. The ActiveCare+DTx is not a diagnostic tool.
# Performance Data
Testing referenced in support of this submission includes: Electrical Safety, EMC, Usability, Software Validation, Risk Analysis, internal testing and published Clinical Literature. The data provided demonstrates that the modified ActiveCare+S.F.T. and ActiveCare+DTx Systems are substantially equivalent to its predicates, and raises no new safety or effectiveness issues.
### Substantial Equivalence
The ActiveCare+S.F.T. and ActiveCare+DTx Systems are substantially equivalent in intended use and technological characteristics to the predicate devices. There are no new questions of safety and efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.