For storage of soft (hydrophilic), hard, and rigid gas permeable contact lenses during chemical disinfection. Use for storage during chemical disinfections only. DO NOT USE WITH HEAT.
Device Story
Contact lens cases (Cat, Sunglasses, 1st, Geeky Eyes models) store soft, hard, and rigid gas permeable contact lenses during chemical disinfection. Device consists of a base with two wells and two screw-on lids to ensure a leak-proof seal. User places lenses and chemical disinfecting solution into wells; L/R markings assist in lens differentiation. Intended for consumer use; not sold sterile; requires rinsing before use. Recommended replacement every 3 months. Benefits include safe storage and maintenance of contact lenses during chemical disinfection process.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Eye Irritation), and ISO 10993-11 (Systemic Toxicity) with passing results.
Technological Characteristics
Materials: Acrylonitrile Butadiene Styrene (ABS) (Polylac PA-707). Colorants: Insoluble pigments (PLASBLAK UN2014, Titanium Dioxide R-350, PV Fast Green GNX, MACROLEX Yellow G, 13575 Orange). Construction: Two-well base with screw-on lids. Dimensions: 60-80mm length, 14-21mm height. Capacity: 2.5-3.0ml per well. Not sold sterile.
Indications for Use
Indicated for storage of soft (hydrophilic), hard, and rigid gas permeable contact lenses during chemical disinfection. Not for use with heat disinfection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 7, 2015
Ni Hau Industrial Co., Ltd. Ms. Lyn Chiu Sales Manager No. 15. Allev 125. Lane 318. Section 2. An Ho Road Tainan City, Taiwan 70967
Re: K142717
> Trade/Device Name: Cat Contact Lens Case, Sunglasses Contact Lens Case, 1st Contact Lens Case, Geeky Eye Contact Lens Case Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LRX Dated: April 10, 2015 Received: April 13, 2015
Dear Ms. Chiu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -A
for Malvina B. Eydelman, MD Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142717
Device Name Cat Contact Lens Case / Sunglasses Contact Lens Case 1st Contact Lens Case Geeky Eyes Contact Lens Case
Indications for Use (Describe)
For storage of soft (hydrophilic), hard, and rigid gas permeable contact lenses during chemical disinfection. Use for storage during chemical disinfections only. DO NOT USE WITH HEAT.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Joseph C. Hut
Digitally signed by Joseph C. Hutter -S DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, 0.9.2342.19200300.100.1.1=1300106195, cn=Joseph C. Hutter -S Date: 2015.05.05 10:56:22 -04'00"
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Nihau" in a bold, sans-serif font. The word is slightly slanted to the right. A circular shape is behind the word, partially obscuring the top and right portions of the letters. The logo has a simple, clean design.
Ni Hau Industrial Co., Ltd. No. 15, Alley 125, Lane 318, Section 2, An Ho Rd., Tainan City, Taiwan TEL:+886- 6- 355- 7000 FAX:+886- 6- 355- 9000 › 355-0297 E-MAIL:lyn@ni-hau.com.tw,
# 510(k) Summary
| (1) 510(k) Proprietary Details | | |
|--------------------------------|------------------------------------------------------------|--|
| Date of Summary: | MAY 05, 2015 | |
| Company: | Ni Hau Industrial Co., Ltd. | |
| Address: | No. 15, Alley 125, Lane 318, Section 2, An Ho Road, Tainan | |
| | City, TAIWAN. Postal Code: 70967 | |
| Telephone: | (886) 6-355 7000 Extension 64 | |
| Fax: | (886) 6 – 355 9000 | |
| Email: | lyn@ni-hau.com.tw | |
| Contact Person: | Ms. Lyn Chiu | |
| (2) 510(k) Device Details: | |
|----------------------------|----------------------------------------------------------------------------------------------------------------|
| Device Name: | Cat Contact Lens Case<br>Sunglasses Contact Lens Case<br>1st Contact Lens Case<br>Geeky Eyes Contact Lens Case |
| Device Model #: | K-489 (Cat)<br>K-488 (Sunglasses)<br>K-506 (1st Contact)<br>K-507 (Geeky Eyes) |
| Device: | Case, Contact Lens |
| Classification | Soft (hydrophilic) contact lens care products. |
| Description: | |
| Review Panel: | Ophthalmic |
| Product Code: | LRX |
| Submission Type: | 510(k) |
| Regulation Number: | 886.5928 |
| Device Class: | II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Nihau" in a stylized font, with the letters slightly overlapping each other. A circular shape is positioned above the word, partially enclosing the top portion of the letters. The overall design appears to be a logo or brand mark.
Ni Hau Industrial Co., Ltd. No. 15, Alley 125, Lane 318, Section 2, An Ho Rd., Tainan City, Taiwan TEL: +886- 6- 355- 7000 FAX:+886- 6- 355- 9000 › 355-0297 E-MAIL:lyn@ni-hau.com.tw,
| (3) Predicate Device: | |
|------------------------------|------------------------------------------------|
| Device Name: | Owl Contact Lens Case<br>Dog Contact Lens Case |
| Device Trade Name: | Owl Contact Lens Case<br>Dog Contact Lens Case |
| Device Model Number: | MG00-A (Owl)<br>MG37-A (Dog) |
| Classification Product Code: | LRX |
| Device Class: | II |
| 510(k) Submitter/ Holder | NINGBO LISHUNDA ELECTRONICS CO., LTD. |
| 510(k) Number | K130930 |
### (4) Device Description:
The contact lens cases in submission are intended to be used to store soft (hydrophilic) and hard contact lenses during chemical disinfection only. These are not to be used in heat disinfection.
There are 2 parts to each contact lens case, the main body/base with two wells, and two screw-on lids. In order to store contact lenses with disinfection solution, the wells for storage need to be leak-proof, which is designed so by the screw construction of the lids.
The plastic that is used in all four contact lens cases is Acrylonitrile Butadene Styrene (Product Name: Polylac ® PA-707, manufactured by Chi Mei Corporation, please refer MSDS attached: Attachment 1), which is a durable plastic that is insoluble in water and can be used for at least two months continuous use.
The cat contact lens case is black, with white printed on the lids. The color pigment used in the case is PLASBLAK® UN2014 (Supplier: Cabot Plastics, please refer MSDS attached: Attachment 2). This pigment is insoluble in water.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Nihau" in a bold, sans-serif font. The word is slightly tilted to the right. A curved line, resembling a crescent or a partial circle, arches over the word, adding a dynamic element to the design. The overall impression is a simple yet recognizable logo or brand mark.
Ni Hau Industrial Co., Ltd. No. 15, Alley 125, Lane 318, Section 2, An Ho Rd., Tainan City, Taiwan TEL:+886- 6- 355- 7000 FAX:+886- 6- 355- 9000 › 355-0297 E-MAIL:lyn@ni-hau.com.tw,
The sunglasses contact lens case is white, with black labels on the lids. The color pigment used in the case is DuPont™ Titanium Dioxide Pigment - Plastics Grade/Type : R-350 (Supplier: DuPont, please refer MSDS attached: Attachment 3). This pigment is insoluble in water.
The 1st contact lens case is green, with clear uncolored lids. There are 2 color pigments used in the green body (1) PV Fast Green GNX (Supplier: Clariant Chemicals (India) Limited, please refer MSDS attached: Attachment 4) and (2) MACROLEX Yellow G Granulate (Supplier: Bayer AG, please refer MSDS attached: Attachment 5). These pigments are insoluble in water.
The Geeky eyes contact lens case is red with black glasses. The color pigment used in the red case is 13575 Orange Color Powder (Supplier: Summit Color Co., Ltd., please refer MSDS attached: Attachment 6). This pigment is insoluble in water.
The individual well capacity of each contact lens are: cat contact lens case 2.8ml, sunglasses contact lens case 2.5ml, 1* contact lens case 3ml, geeky eyes contact lens case 2.8ml. The capacities aforementioned are sufficient for a contact lens to be fully soaked in the disinfection liquid.
At the base of the cat contact lens case and the sunglasses contact lens case, in each well there is L and R embossed to help users differentiate the left and right lenses. The geeky eyes contact lens case also has the L and R markings on the side of the casing outer. The 1d contact lens case does not have L and R markings on it as the two wells are situated one above and one below the disc so it is unlikely to mix up the two wells during use.
Through use, the material of the contact lens case may come into contact with the user's eyes indirectly. The user will need to put contact lenses with chemical disinfecting liquid into the contact lens case for storage. Therefore we have used a plastic material and colorants that are insoluble in water so that it will not come in contact with one's eyes.
These contact lens cases are intended for the storage of soft (hydrophilic) and
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with the word "Nihau" in a stylized font. The word is in black and appears to be slightly italicized. Above and below the word, there is a curved line that forms a partial circle, creating a frame around the text. The logo is simple and monochromatic.
Ni Hau Industrial Co., Ltd. No. 15, Alley 125, Lane 318, Section 2, An Ho Rd., Tainan City, Taiwan TEL:+886- 6- 355- 7000 FAX:+886- 6- 355- 9000 > 355-0297 E-MAIL:lyn@ni-hau.com.tw,
hard contact lenses. The case should ensure the fluids contained do not leak. These cases are for the storage of chemical disinfecting fluids with contact lenses only, they are not to be used in heat disinfection. These are not sold sterile, so the packaging should have instructions written on them to "Rinse the case thoroughly with clean solution before use".
It is recommended to replace the contact lens case at least every 3 months.
# (5) Indications For Use
For storage of soft (hydrophilic), hard, and rigid gas permeable contact lenses during chemical disinfection. Use for storage during chemical disinfections only. DO NOT USE WITH HEAT.
| (6) Comparison of Devices in Application & Substantially Equivalent Predicate<br>Device: | | |
|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| The Device in Submission | Substantially Equivalent<br>Predicate Device | |
| Device Name | Cat Contact Lens Case<br>Sunglasses Contact Lens Case<br>1st Contact Lens Case<br>Geeky Eyes Contact Lens Case | OWL CONTACT LENS CASES<br>DOG CONTACT LENS CASES |
| Device Model #: | K-489 (Cat)<br>K-488 (Sunglasses)<br>K-506 (1st Contact)<br>K-507 (Geeky) | MG00-A (Owl)<br>MG37-A (Dog) |
| Device<br>Appearance | Image: Cat contact lens case and sunglasses contact lens case | Image: Owl contact lens case |
| Image: Green contact lens case | Image: Blue contact lens case | |
| Case Dimension | Cat: 65 x 37 x 14mm | Owl: 71 x 38.5 x 17mm |
| | Sunglasses: 71 x 31 x 14mm | Dog: 69.5 x 38 x 14mm |
| | 1st Contact: 60 x 39.6 mm | |
| | Geeky eyes: 80 x 32 x 21mm | |
| Single Well | Cat: 3ml | Owl: 5.5ml |
| Capacity Volume | Sunglasses: 2.5ml | Dog: 4.8ml |
| | 1st Contact: 3ml | |
| | Geeky eyes: 2.8ml | |
| Device Material | All 4: (ABS) Acrylonitrile<br>Butadene Styrene | Both: (ABS) Acrylonitrile<br>Butadene Styrene |
| Device<br>Construction | All 4: Shaped base with two<br>wells to hold the fluid and two<br>screw lids for the Left and<br>Right eyes of contact lenses. | Both: Shaped base with two<br>wells and two screw lids |
| Device Intended<br>Use | All 4: To hold soft and hard<br>contact lenses. For storing<br>chemical disinfecting fluid<br>only. | Both: To hold soft and hard<br>contact lenses. For storing<br>chemical disinfecting fluid<br>only. |
| Biocompatibility | PASS the following test:<br>Cytotoxicity (ISO 10993-5)<br>Eye Irritation (ISO 10993-10)<br>Systematic Toxicity (ISO<br>10993-11) | PASS the following test:<br>Cytotoxicity (ISO 10993-5)<br>Eye Irritation (ISO 10993-10)<br>Systematic Toxicity (ISO<br>10993-11) |
| Sterilization | Not sold sterile | Not sold sterile |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Nihau. The logo consists of the word "Nihau" in a stylized font, with a circular shape encompassing the top half of the word. The font is bold and slightly slanted, giving it a dynamic appearance. The logo is presented in black and white.
Ni Hau Industrial Co., Ltd. No. 15, Alley 125, Lane 318, Section 2, An Ho Rd., Tainan City, Taiwan TEL:+886- 6- 355- 7000 FAX:+886- 6- 355- 9000 › 355-0297 E-MAIL:lyn@ni-hau.com.tw,
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo with the word "Nihau" in bold, stylized letters. A curved line arches over the word, creating a semi-circular shape above it. The logo has a simple, black-and-white design, with the word "Nihau" being the most prominent element.
Ni Hau Industrial Co., Ltd. No. 15, Alley 125, Lane 318, Section 2, An Ho Rd., Tainan City, Taiwan TEL: +886- 6- 355- 7000 FAX:+886- 6- 355- 9000 \ 355-0297 E-MAIL:lyn@ni-hau.com.tw,
Since the construction, materials and specifications of the devices in submission are very similar to the substantially predicate devices, which have been safely distributed in the USA market, we can conclude that our Cat Contact Lens Case, Sunglasses Contact Lens Case, 1st Contact Lens Case, and Geeky Eye Contact Lens Case are also safe to be distributed in the U.S.A.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.