S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System
K142714 · Sonoscape Company Limited · IYN · Jan 29, 2015 · Radiology
Device Facts
Record ID
K142714
Device Name
S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System
Applicant
Sonoscape Company Limited
Product Code
IYN · Radiology
Decision Date
Jan 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SonoScape S8 Exp/S9 Pro system is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.
Device Story
Portable digital color Doppler ultrasound system; utilizes ultrasonic energy transmission and echo reception to generate real-time images of internal anatomy and fluid flow. Operates via console with keyboard, LCD monitor, and various probes (linear, convex, phased arrays; 2.0-15.0 MHz). Supports B-Mode, M-Mode, Tissue Harmonic Imaging, Color-Flow/Pulsed/Continuous Wave/Power Doppler, Elastography, and 3D/4D imaging. Used by physicians in clinical settings for diagnostic evaluation. Output displayed on monitor for clinical assessment of anatomic structures and hemodynamics; aids in diagnosis and clinical decision-making across multiple specialties.
Clinical Evidence
No clinical testing was required. Bench testing only, including electrical, mechanical, thermal, and electromagnetic compatibility safety, biocompatibility, and acoustic output. Phantom testing verified Elastography function, accuracy, and repeatability.
Technological Characteristics
Portable console with color LCD; probes include linear, convex, and phased arrays (2.0-15.0 MHz). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-5, ISO 10993-10, NEMA UD2, NEMA UD3. Software-controlled imaging modes: B-Mode, M-Mode, Doppler (Color, PW, CW, Power), Elastography, 3D/4D.
Indications for Use
Indicated for use by qualified physicians for diagnostic ultrasound imaging of fetal, abdominal, pediatric, small organ, cephalic, trans-rectal, trans-vaginal, peripheral/cerebral vascular, musculo-skeletal, cardiac, trans-esophageal, laparoscopic, OB/Gyn, and urological structures.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
SonoScape S8 Exp Portable Digital Color Doppler Ultrasound System (K132768)
SonoScape S30 Digital Color Doppler Ultrasound System (K132527)
Philips EPIQ Diagnostic Ultrasound System (K132304)
Philips CX50 3.0 Diagnostic Ultrasound System (K123754)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 29, 2015
Sonoscape Company Limited % Mrs. Toki Wu Regulatory Affairs Manager Yizhe Building, Yuquan Road, Nanshan Shenzhen 518051 P.R. CHINA
Re: K142714
Trade/Device Name: S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: September 18, 2014 Received: January 09, 2015
Dear Mrs. Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K142714
### Device Name
S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System
Indications for Use (Describe)
The SonoScape S8 Exp/S9 Pro system is a general-purpose ultrasonic intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Muscular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <div><span style="font-size:12px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
# 510(k) Summary
# 1. Submitter [21 CFR807.92 (a) (1)]
| Submitter: | SonoScape Company Limited |
|-----------------|---------------------------------------------------------------------|
| Address: | Yizhe Building, Yuquan Road, Nanshan, Shenzhen<br>518051, P.R.China |
| Contact Person: | Toki Wu |
| Tel: | +86 755 26722890 |
| Fax: | +86 755 26722850 |
| Email: | wusq@sonoscape.net |
| Date Prepared | September 18, 2014 |
# 2. Device [21 CFR807.92 (a)(2)]
| Trade Name: | S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System |
|--------------|----------------------------------------------------------------|
| Common Name: | Diagnostic Ultrasound System and Transducers |
Classification Regulatory:
| | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
Classification Panel: Radiology Device Class: ll
# 3. Predicate Device(s) [21 CFR 807.92(a) (3)]
| The identified predicate devices within this submission are as follows: | |
|-------------------------------------------------------------------------|---------|
| SonoScape S8 Exp Portable Digital Color Doppler Ultrasound System | K132768 |
| SonoScape S30 Digital Color Doppler Ultrasound System | K132527 |
| Philips EPIQ Diagnostic Ultrasound System | K132304 |
| Philips CX50 3.0 Diagnostic Ultrasound System | K123754 |
# 4. Device Description [21 CFR 807.92(a) (4)]
This SonoScape S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.
{5}------------------------------------------------
The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body.
This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array with a frequency range of 2.0 MHz to 15.0 MHz. This system consists of a portable console with keyboard control panel, power supply module, color LCD monitor and optional probes.
This system is a portable, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or the combination of these modes, Elastography, 3D/4D.
The subject of this submission is the addition of new indications, new probes and special function to original SonoScape S8 Exp and the addition of SonoScape S9 Pro.
| New indications | Trans-esoph.(Cardiac), Cerebral Vascular and Laparoscopic. |
|------------------|--------------------------------------------------------------------------------------------------|
| New probes | Add C613, MPTEE, MPTEE mini, CWD2.0, CWD5.0, PWD2.0 and LAP7 probes. |
| Special function | Elastography. |
| New Model | Add S9 Pro model which is identical to S8 Exp except for the difference of probes configuration. |
### 5. Intended Use [21 CFR 807.92(a)(5)]
The SonoScape S8 Exp/S9 Pro system is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.
### 6. Comparison with the Predicate Devices [21 CFR 807.92(a) (6)]
S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System is comparable with and substantially equivalent to the predicate devices:
| SonoScape S8 Exp Portable Digital Color Doppler Ultrasound System | K132768 |
|-------------------------------------------------------------------|---------|
| SonoScape S30 Digital Color Doppler Ultrasound System | K132527 |
| Philips EPIQ Diagnostic Ultrasound System | K132304 |
{6}------------------------------------------------
#### Philips CX50 3.0 Diagnostic Ultrasound System K123754
# Intended Use Comparison:
Compared with SonoScape S30 (K132527), the Subject Device S8 Exp/S9 Pro has the same intended uses except for Laparoscopic indication which is comparable and equivalent to the Predicate Device Philips CX50 3.0 (K123754) and Cerebral Vascular indication which is comparable and equivalent to the Predicate Device Philips EPIQ (K132304). The detailed analysis can be found in the probes comparison.
# Technical Characteristics Comparison:
The basic and main technical features of the Subject Device S8 Exp/S9 Pro are the same as the original SonoScape S8 Exp (K132768), including Design, Operation Controls, Display Modes, Measurement Items, Cine Loop, Power Supply, Operating and Storage Condition and Screen Size. For Operation Modes, Elastography is a special Operation Mode for the Subject Device S8 Exp/S9 Pro, but already employed by many marketed devices and considered Substantially Equivalent to the Predicate Device Philips EPIQ (K132304) and detailed comparison analysis can be found in Substantial Equivalence Comparison. The detailed technical features can be found in General Device Descriptions of the submission.
### Probes Comparison:
Subject device S8 Exp/S9 Pro has the similar probes as the predicated device original SonoScape S8 Exp (K132768), SonoScape S30 (K132527), Philips EPIQ (K132304) and Philips CX50 3.0 (K123754).
| Subject device | Predicate Device | Remark |
|-------------------------|-------------------------|-------------|
| SonoScape S8 Exp/S9 Pro | SonoScape S8 Exp | |
| C322 Micro-curved Array | C322 Micro-curved Array | Same |
| C344 Curved Array | C344 Curved Array | |
| C353 Curved Array | C353 Curved Array | |
| C542 Curved Array | C354 Curved Array | |
| | C362 Curved Array | |
| | C542 Curved Array | |
| VC6-2 Curved Array | VC6-2 Curved Array | Same |
| C613 Micro-curved Array | C611 Micro-curved Array | SE |
| | C311 Micro-curved Array | Analysis1a) |
| 2P1 Phased Array | 2P1 Phased Array | Same |
| 2P2 Phased Array | 2P2 Phased Array | |
| 3P1 Phased Array | 3P1 Phased Array | |
### Table 1 Probes Comparison
{7}------------------------------------------------
| 5P1 Phased Array | 5P1 Phased Array | Same |
|------------------------------------------|------------------------------------------|-------------------|
| 5P2 Phased Array | 5P2 Phased Array | |
| 8P1 Phased Array | 8P1 Phased Array | |
| L741 Linear Array | L741 Linear Array | |
| L742 Linear Array | L742 Linear Array | |
| L743 Linear Array | L743 Linear Array | |
| L752 Linear Array | L752 Linear Array | |
| 10L1 Linear Array | 10L1 Linear Array | |
| 6V1 Micro-curved Array | 6V1 Micro-curved Array | |
| 6V3 Micro-curved Array | 6V3 Micro-curved Array | Same |
| EC9-5 Micro-curved Array | EC9-5 Micro-curved Array | |
| BCC9-5 Micro-curved Array | BCC9-5 Micro-curved Array | |
| BCL10-5 Biplane (Curved + Linear Array) | BCL10-5 Biplane (Curved + Linear Array) | |
| Subject device | Predicate Device | Remark |
| SonoScape S8 Exp/S9 Pro | SonoScape S30 | |
| MPTEE Phased Array<br>(Multi-plane) | MPTEE Phased Array<br>(Multi-plane) | Same |
| MPTEE mini Phased Array<br>(Multi-plane) | MPTEE mini Phased Array<br>(Multi-plane) | |
| Subject device | Predicate Device | Remark |
| SonoScape S8 Exp/S9 Pro | Philips CX50 3.0 | |
| CWD2.0, 2.0 MHz | D2cwc, 2.0 MHz | SE<br>Analysis1b) |
| CWD5.0, 5.0 MHz | D5cwc, 5.0 MHz | |
| Continuous Wave Doppler | Continuous Wave Doppler | |
| LAP7, Linear Array, 10.0-5.0 MHz | L10-4 lap, Linear Array, 10.0-4.0 MHz | |
| Subject device | Predicate Device | Remark |
| SonoScape S8 Exp/S9 Pro | Philips EPIQ | |
| PWD2.0, 2.0 MHz | D2tcd, 2.0 MHz | SE |
Note: C344, 2P1, 2P2, 5P1, L741 and 6V1 only for S8 Exp; and 3P1, 5P2 and 8P1 only for S9 Pro.
# SE Analysis 1:
- a) Compared with the predicate device original SonoScape S8 Exp (K132768), there is a new probe (C613) for the subject device. And the frequency, performance and clinical application of C613 probe are the same as C611 for predicate device.
- b) Compared with the predicate device Philips CX50 3.0 (K123754), the frequency and clinical application of CWD2.0 and CWD5.0 probe is the same as D2cwc and D5cwc cleared for use with predicate device; the clinical application of LAP7 probe is the same
{8}------------------------------------------------
as L10-4 lap cleared for use with predicate device, and though the frequency of them is different, they both meet the clinical use.
- c) Compared with the predicate device Philips EPIQ (K132304), the frequency and clinical application of PWD2.0 probe is the same as D2tcd cleared for use with predicate device.
Note: detailed description information about the new probes can be found in Substantial Equivalence Comparison.
Moreover, compared with predicated devices, the subject device (S8 Exp/S9 Pro) complies with the same regulation and safety standards and has the consistent acoustic output levels. Therefore they can be considered Substantially Equivalent in safety and effectiveness, and no new risk is raised, so the SE is not affected.
## 7. Non-Clinical Tests [21 CFR 807.92(b) (1)]
The S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System has been evaluated for electrical, mechanical, thermal and electromagnetic compatibility safety, biocompatibility and acoustic output.
Laboratory tests were conducted to verify that the S8 Exp/S9 Pro system met all design specifications and the S8 Exp/S9 Pro system conformed to applicable medical device standards. Phantom test was conducted to verify that the strain Elastography function was effective and Elastography performance met design specifications, including accuracy and repeatability of strainratio measurement and etc.
The S8 Exp/S9 Pro system has been designed and manufactured to meet the following standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-5, ISO10993-10, UD2, and UD3.
### 8. Clinical Test [21 CFR 807.92(b) (2)]
No clinical testing was required.
# 9. Substantially Equivalent Conclusions [21 CFR 807.92(b) (3)]
In accordance with the 21 CFR Part 807 and based on the information provided in this premarket notification, SonoScape Company Limited concludes that S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System is substantially equivalent to the predicate devices with regard to safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.