← Product Code [DOF](/productcode/DOF) · K142705

# LeadCare Plus Blood Lead Testing System (K142705)

_Magellan Diagnostics · DOF · Jul 7, 2015 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K142705

## Device Facts

- **Applicant:** Magellan Diagnostics
- **Product Code:** [DOF](/productcode/DOF.md)
- **Decision Date:** Jul 7, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3550
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K142705 · Jul 7, 2015 | LeadCare Plus Blood Lead Testing System | Magellan Diagnostics | Clinical whole blood samples (native patient samples) | Clinical samples were used to evaluate the accuracy of the LeadCare Plus system by comparing results against the predicate device (LeadCare Ultra) and a reference method (GFAAS). | Method comparison; Clinical samples; Accuracy validation |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Method comparison study | Clinical whole blood samples; Sample Size: 284 total results (169 within reportable range); Number of Sites: 1 | LeadCare Ultra (predicate) and GFAAS (reference method) | Accuracy (linear regression analysis) |
| Matrix comparison study | Native patient samples and contrived blood samples; Sample Size: 50 (23 native patient samples, 27 contrived) | GFAAS | Linear regression |

## Indications for Use

The LeadCare® Plus™ Blood Lead Testing System is intended for the quantitative measurement of lead in a whole blood sample. The LeadCare Plus Blood Lead Testing System is intended for in vitro (external) use only. The test kit components are for use with both the LeadCare Plus and LeadCare Ultra® Blood Lead Testing Systems. This test system is for prescription use only. This system is not intended for point of care use.

## Device Story

LeadCare® Plus™ Blood Lead Testing System is a single-channel analyzer for quantitative lead measurement in whole blood. System utilizes Anodic Stripping Voltammetry (ASV); whole blood sample mixed with hydrochloric acid-based treatment reagent to lyse cells; lead ions collected on sensor via applied electrical potential; analyzer measures lead concentration. Used in clinical settings by trained personnel; not for point-of-care. Calibration performed via lot-specific electronic button. Output displayed in μg/dL. Benefits include rapid, quantitative lead assessment to support clinical decision-making regarding lead exposure.

## Clinical Evidence

Clinical method comparison study conducted at one site with 284 samples (169 within 1.9–65 µg/dL range). LeadCare Plus results compared against predicate (LeadCare Ultra) and reference method (GFAAS). Regression analysis vs. LeadCare Ultra: slope 0.985, intercept -0.10, R²=0.994. Regression analysis vs. GFAAS: slope 1.029, intercept -0.98, R=0.994. Non-clinical testing included precision (CLSI EP05-A2), linearity (CLSI EP06-A), and LoB/LoD/LoQ (CLSI EP17-A2).

## Technological Characteristics

Single-channel potentiostat; Anodic Stripping Voltammetry (ASV) principle. Uses screen-printed sensors with conductive inks, plastic spacer/lid, and colloidal gold electrode. Dimensions: 9"x6.5"x3.5". Power: AC or battery. Connectivity: Calibration button reader for lot-specific parameters. Software: Firmware-based, alphanumeric LCD interface. Sterilization: Not specified.

## Regulatory Identification

A lead test system is a device intended to measure lead, a heavy metal, in blood and urine. Measurements obtained by this device are used in the diagnosis and treatment of lead poisoning.

## Predicate Devices

- LeadCare Ultra® Blood Lead Testing System ([K123563](/device/K123563.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE

A. 510(k) Number:
K142705

B. Purpose for Submission:
New Device

C. Measurand:
Lead

D. Type of Test:
Quantitative

E. Applicant:
Magellan Diagnostics

F. Proprietary and Established Names:
LeadCare® Plus™ Blood Lead Testing System

G. Regulatory Information:
1. Regulation section:
21 CFR§ 862.3550, Lead test system
2. Classification:
Class II
3. Product code:
DOF
4. Panel:
Toxicology (91)

{1}

2

H. Intended Use:

1. Intended use(s):

See Indications for Use below.

2. Indication(s) for use:

The LeadCare® Plus™ Blood Lead Testing System is intended for the quantitative measurement of lead in a whole blood sample. The LeadCare Plus Blood Lead Testing System is intended for in vitro (external) use only. The test kit components are for use with both the LeadCare Plus and LeadCare Ultra® Blood Lead Testing Systems.

This test system is for prescription use only. This system is not intended for point of care use.

3. Special conditions for use statement(s):

- Prescription use only
- This system is not intended for point of care use

4. Special instrument requirements:

LeadCare Plus Blood Lead Analyzer

I. Device Description:

The LeadCare® Plus™ Blood Lead Testing System is a one-channel analyzer system version of the predicate, LeadCare Ultra®, that is a 6-channel analyzer. The new LeadCare Plus Analyzer is used with the same Blood Lead Sensors, Treatment Reagent Tubes, and LeadCare Plus Blood Lead Controls (Level 1 and Level 2) as previously cleared with the predicate device (k123563; LeadCare Ultra® Blood Lead Testing System).

J. Substantial Equivalence Information:

1. Predicate device name(s):

LeadCare Ultra® Blood Lead Testing System

2. Predicate 510(k) number(s):

K123563

3. Comparison with predicate:

{2}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device LeadCare® Plus™ | Predicate LeadCare Ultra® (K123563)  |
|  Indication For Use | The LeadCare® Plus™ Blood Lead Testing System is intended for the quantitative measurement of lead in a whole blood sample | Same  |
|  Methodology | Anodic Stripping Voltammetry | Same  |
|  Sensor (test strip) | Screen printed sensors with conductive inks; plastic spacer and lid; capillary fill | Same  |
|  Active Test Electrode area | Thin layer of colloidal gold in an inert polymer matrix | Same  |
|  Calibration | Electronic calibration button | Same  |
|  Blood Collection | Fingerstick or venipuncture | Same  |
|  Sample Matrix | Whole blood; up to 72 hours from time of draw | Same  |
|  Unit of Measure | Results displayed in micrograms of lead per deciliter of whole blood (μg/dL) | Same  |
|  Reportable Range | 1.9 – 65 μg/dL | Same  |
|  Controls | 2 levels of external liquid controls | Same  |
|  Power Source | AC Adapter or batteries | Same  |
|  Test Time | 3 minutes | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device LeadCare® Plus™ | Predicate LeadCare Ultra® (K123563)  |
|  Throughput | 1 sample at a time | 6 samples at a time  |

# K. Standard/Guidance Document Referenced (if applicable):

- CLSI EP5-A2 Evaluation of Precision Predominance of Quantitative Measurement Methods
- CLSI EP6-A Evaluation of Linearity of Quantitative Measurement procedures: A Statistical Approach
- CLSI EP7-A2 Interference Testing in Clinical Chemistry
- CLSI EP9-A2 Method Comparison and Bias Estimation Using Patient Samples
- CLSI EP17-A2 Protocols for Determination of Limits of Detection and Limits of Quantitation
FCC 47CFR Part 15, Subpart B (2014)
- CEI/IEC 61010-1: 2001 Safety Requirements for electrical equipment for measurement, control and laboratory use.

{3}

- CEI/IEC 61326-2-6:2006 Electrical equipment for measurement control and laboratory use - EMC requirements.

# L. Test Principle:

The LeadCare® Plus™ Blood Lead Testing System is an in vitro diagnostic device that relies on electrochemistry (Anodic Stripping Voltammetry or ASV) with a sensor to detect lead in whole blood. When a sample of whole blood is mixed with Treatment Reagent (a dilute solution of hydrochloric acid), the red blood cells are lysed and the lead becomes available for electrochemical detection. When a test is run, the analyzer applies an electrical potential that causes the lead to collect on the Sensor. The analyzer measures the amount of lead on the Sensor and displays the result in micrograms per deciliter.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

The LeadCare® Plus™ precision was demonstrated using bovine blood standards at lead concentrations of 3.1, 5.1, 11.7, 24.7 and  $45.4~\mu \mathrm{g / dL}$ . An additional human blood sample at a lead concentration of  $59.1~\mu \mathrm{g / dL}$  was also included to challenge the high end of the range. Eighty (80) data points were collected per concentration level over a 20 day period using two Sensor/Reagent Lot Pairs. The combined data set is shown:

|  Mean, μg/dL | WR* SD, μg/dL | Total SD, μg/dL | WR CV | Total CV  |
| --- | --- | --- | --- | --- |
|  3.1 | 0.44 | 0.49 | 14.1% | 15.6%  |
|  5.1 | 0.44 | 0.50 | 8.5% | 9.6%  |
|  11.7 | 0.64 | 0.71 | 5.3% | 6.0%  |
|  24.7 | 0.80 | 1.00 | 3.2% | 4.0%  |
|  45.4 | 1.61 | 1.71 | 3.5% | 3.7%  |
|  59.1 | 1.89 | 2.42 | 3.2% | 4.0%  |

$\mathrm{WR} =$  within run

# b. Linearity/assay reportable range:

Linearity assessments were performed on two Sensor/Reagent Lot Pairs using nine human donor blood samples, spiked with lead to concentrations of  $0 - 2\mu \mathrm{g / dL}$ , 2-5  $\mu \mathrm{g / dL}$ , 5-10  $\mu \mathrm{g / dL}$ , 15-25  $\mu \mathrm{g / dL}$ , 25-35  $\mu \mathrm{g / dL}$ , 35-45  $\mu \mathrm{g / dL}$ , 45-55  $\mu \mathrm{g / dL}$ , 55-65, 65-70  $\mu \mathrm{g / dL}$ . One whole blood sample, unadulterated in  $\mathrm{K}_2\mathrm{EDTA}$  vacutainers, was also included.

{4}

The Linear Regression results for each of the two Sensor/Reagent Lot Pairs are as follows:

|  Lot Pair #1 | Lot Pair #2  |
| --- | --- |
|  Y = 1.11x - 1.49 | Y = 1.08x - 0.803  |
|  R² = 0.998 | R² = 0.997  |

The results of the study support the sponsor's claimed reportable range of 1.9 – 65 µg/dL

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability: The LeadCare® Plus™ system is traceable to the NIST Standard Reference Material 955c analyzed on Graphite Furnace Atomic Absorption Spectrometer (GFAAS).

Test Kit Stability: Protocols and acceptance criteria were previously reviewed and found to be acceptable (k123563) for the test kit. The sponsor's stability claims are for 18 months when stored at 59°F to 80°F (15°C - 27°C). The labeling instructs the user not to freeze or refrigerate the reagent. The sensors should be kept sealed until the sample is prepared and the treatment reagent should be used immediately after opening the tube.

Control Solutions: Two levels of Quality Controls material are provided with the test kit, Level 1 and Level 2. Theses control solutions were previously cleared k063398. The labeling instructs the user to store the control solutions at: 15°C - 27°C (59°F - 80°F) and not to refrigerate.

Sample stability study protocols and acceptance criteria were reviewed and found to be acceptable to support the claims in the labeling that samples collected in microcapillary K2EDTA tubes can be stored for up to 72 hours (3 days) from collection at 33°F - 77°F (1°C - 25°C) prior to being mixed with treatment reagent.

d. Detection limit:

The Limits of Blank, Detection and Quantification were established using the Total Error Analysis Method and by following the CLSI EP17-A2 guidelines.

The Limit of Blank (LoB) was determined by running 60 replicates of near blank bovine blood samples, over 5 days. Samples were analyzed using two Sensor/Reagent Lot Pairs. The average LoB was calculated to be 0.98 µg/dL.

The Limit of Detection (LoD) and Limit of Quantification (LoQ) were determined by running 60 replicates of 10 different whole blood samples, collected in K₂EDTA vacutainers. The replicates were run, using two Sensor/Reagent Lot Pairs on

5

{5}

LeadCare Plus analyzers over five days. The average LoD was calculated to be 1.2  $\mu \mathrm{g} / \mathrm{dL}$ .

The Limit of Quantification LoQ was calculated using the Total Error equation:

Total Error  $\mathrm{LoQ} =$  absolute (Bias)  $^+$  (2 x SD)

The average LoQ was calculated to be  $1.6\mu \mathrm{g / dL}$

The Total Error at the LoQ was  $18\%$ .

Obtained LoB, LoD, and LoQ values for the LeadCare® Plus™ differ slightly from those obtained for the predicate, LeadCare Ultra® (See k123563). For product family consistency between LeadCare® Plus™ and LeadCare Ultra®, the claimed values for LoB, LoD and LoQ of the LeadCare® Plus™ are provided in the table below and are identical to the values obtained for the LeadCare Ultra®:

|   | LeadCare® Plus™  |
| --- | --- |
|  LoB | 1.5 μg/dL  |
|  LoD | 1.9 μg/dL  |
|  LoQ | 1.9 μg/dL  |

# e. Analytical specificity:

The assay reagents and controls are unchanged from the predicate, therefore please refer to k123563 where analytical specificity was established.

# f. Assay cut-off:

Not applicable.

# 2. Comparison studies:

# a. Method comparison with predicate device:

A method comparison study was conducted at one site with two Sensor/Reagent Lot Pairs, two labs, two users and two stations of six LeadCare Plus analyzers. To assess system accuracy 169 whole blood samples collected in  $\mathrm{K}_2\mathrm{EDTA}$  Vacutainers with lead concentrations within the claimed analytical range of  $1.9 - 65~\mu \mathrm{g / dL}$  were run on the LeadCare® Plus™ system and the results were compared to the reference method, Graphite Furnace Atomic Absorption Spectroscopy (GFAAS). The LeadCare® Plus™ results from each sample were compared to the average of two GFAAS results. The linear regression results are as follows:

$$
y = 1. 0 2 9 x - 0. 9 8, R ^ {2} = 0. 9 9 4
$$

# b. Matrix comparison:

{6}

A total of 50 micro-capillary tubes with K₂EDTA fingerstick samples were spiked with lead to achieve concentrations of 1.9-63.1 μg/dL lead as measured by GFAAS. Samples tested included unaltered native patient samples and spiked samples. These samples were analyzed on the LeadCare® Plus™ system and compared to values obtained from GFAAS. The linear regression results are as follows:

$$
y = 1.018X - 0.023; \ R^2 = 0.983
$$

The assay reagents and controls are unchanged from the predicate, therefore please refer to k123563 for matrix comparison information for EDTA Vacutainer and Heparin Vacutainer samples.

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable.

b. Clinical specificity:

Not applicable.

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable.

4. Clinical cut-off:

Not applicable.

5. Expected values/Reference range:

The current labeling includes the following:

In 2012, based on the increased body of evidence demonstrating there is no safe level of lead in the blood, experts established a new reference value to identify children who have elevated blood lead levels (BLL). According to the Centers for Disease Control (CDC) website, this level is based on the U.S. population of children ages 1-5 years who are in the top 2.5% of children when tested for lead in their blood (when compared to children who are exposed to more lead than most children). Currently this reference value is 5 μg/dL, a change from the previously utilized "level of concern" set at 10 μg/dL.*

*CDC Response to Advisory Committee on Childhood Lead Poisoning Prevention Recommendations in "Low Level Lead Exposure Harms Children: A Renewed Call of Primary Prevention." June 7, 2012.

{7}

The labeling also includes a note that it is important to check the CDC website, as well as specific state regulations, for up-to-date information regarding blood lead testing and provides the following website link, (www.cdc.gov/nceh/lead/acclpp/blood_lead_levels.htm).

## N. Instrument Name:

LeadCare Plus Blood Lead Analyzer

## O. System Descriptions:

1. **Modes of Operation:**

Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?

☑ Yes ☐ X or No

Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?

☐ Yes ☐ or No ☑ X

2. **Software:**

FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:

☑ Yes ☐ X or No

3. **Specimen Identification:**

Specimen identification is based on sample ID, sensor lot number, and user ID.

4. **Specimen Sampling and Handling:**

The labeling provides instructions on specimen sampling and handling and states the following user instructions: Handle all products and objects containing human blood as if capable of transmitting diseases. Follow established recommendations for prevention of blood-borne transmissible diseases. For example, consult the "Universal Precautions" issued by the U.S. Public Health Service, Centers for Disease Control. Review your internal protocol for preventing transmission of blood-borne pathogens and your biohazardous waste disposal procedures prior to implementing the LeadCare® Plus™ Blood Lead Testing System.

5. **Calibration:**

8

{8}

Calibration is done using the calibration button which is supplied in each test kit. The calibration button is specific for the test kit in which it is supplied. It is used to download all calibration information, analytical test parameters, and expiration date code information for the sensor lot supplied within the test kit. This lot specific calibration button is held against the calibration reader on the analyzer to calibrate the analyzer for use with a particular test kit. The labeling states that the analyzer must be calibrated for each test kit lot and instructs the user to make sure the calibration code on the calibration button matches the lot number on the sensor container and on the controls.

6. Quality Control:

Two levels of liquid Quality Controls (Level 1 and Level2) are supplied with the Test Kit. The labeling states that controls should be tested, at a minimum, as follows: each day or shift before patient samples are tested.

P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

The sponsor provided the appropriate documentation certifying that acceptable electromagnetic testing (EMC) had been performed.

Q. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K142705](https://fda.innolitics.com/device/K142705)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
