K142679 · Datex-Ohmeda, Inc. · CBK · Jun 12, 2015 · Anesthesiology
Device Facts
Record ID
K142679
Device Name
CARESCAPE R860
Applicant
Datex-Ohmeda, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Jun 12, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The CARESCAPE R860 ventilator is designed to provide mechanical ventilation or support to neonatal, pediatric, and adult patients weighing 0.25 kg and above. The CARESCAPE R860 ventilator is a microprocessor based, electronically controlled, pneumatically driven ventilator that includes integrated monitoring of FiO2, airway pressure, flow, and volume. Additional respiratory gas monitoring capabilities are supported through the use of optional GE patient monitoring modules. Not all features are available for all patient types or product configurations. The CARESCAPE R860 ventilator is not a pulmonary function calculation device. The system is designed for facility use, including within-facility transport, and should only be used under the orders of a clinician.
Device Story
CARESCAPE R860 is a microprocessor-based, pneumatically driven critical care ventilator for facility use. It accepts inputs from integrated sensors (FiO2, airway pressure, flow, volume) and optional GE patient monitoring modules. The device features a touchscreen interface for clinicians to configure ventilation modes (e.g., A/C VC, SIMV, CPAP/PS, APRV) and view patient status, historical trends, and clinical decision support data. It supports nebulization and interfaces with central information management systems. The ventilator provides mechanical ventilation to patients weighing ≥0.25 kg. Clinicians use the output to manage respiratory support, potentially improving patient outcomes by providing tailored ventilation strategies and real-time monitoring of respiratory parameters.
Clinical Evidence
No clinical data. Substantial equivalence was established through non-clinical testing, including risk analysis, design reviews, unit-level testing, integration testing, performance verification, safety testing, and simulated use/user requirements validation.
Technological Characteristics
Microprocessor-based, pneumatically driven ventilator. Features 15-inch LCD resistive touchscreen and Trim Knob. Includes integrated monitoring for FiO2, airway pressure, flow, and volume. Supports optional GE patient monitoring modules and Aerogen nebulizer. Compliant with ANSI/AAMI ES60601-1:2005 (R 2012), IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, ISO 80601-2-12, IEC 62366, ISO 5356-1, and IEC 62304. Connectivity includes interface with central information management systems.
Indications for Use
Indicated for mechanical ventilation or support of neonatal, pediatric, and adult patients weighing 0.25 kg and above with pulmonary impairment ranging from minor to severe.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
GE Datex-Ohmeda Engström Carestation, Engström Pro (K111116)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 12, 2015
Datex-Ohmeda, Inc. Trishia Mercier Regulatory Affairs Leader PO Box 7550 Madison, WI 53707
Re: K142679
Trade/Device Name: CARESCAPE R860 Regulation Number: 21 CFR 868.5895 Regulation Name: Ventilator, continuous, facility use Class: II Product Code: CBK Dated: May 12, 2015 Received: May 13, 2015
Dear Ms. Mercier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Mercier
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142679
Device Name CARESCAPE R860
#### Indications for Use (Describe)
The CARESCAPE R860 ventilator is designed to provide mechanical ventilation or support to neonatal, pediatric, and adult patients weighing 0.25 kg and above. The CARESCAPE R860 ventilator is a microprocessor based, electronically controlled, pneumatically driven ventilator that includes integrated monitoring of FiO2, airway pressure, flow, and volume.
Additional respiratory gas monitoring capabilities are supported through the use of optional GE patient monitoring modules.
Not all features are available for all patient types or product configurations. The CARESCAPE R860 ventilator is not a pulmonary function calculation device.
The system is designed for facility use, including within-facility transport, and should only be used under of a clinician.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | September 17, 2014 |
|-----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare<br>Datex-Ohmeda, Inc.<br>3030 Ohmeda Drive<br>P.O. Box 7550<br>Madison, WI 53707-7550<br>USA |
| Primary Contact Person: | Trishia Mercier<br>Regulatory Affairs Leader<br>Telephone: (608) 709-3260<br>Fax: (608) 646-6488<br>Email: Trishia.Mercier@ge.com |
| Secondary Contact Person: | Monica Morrison<br>Regulatory Affairs Director<br>Telephone: (608) 709-3439<br>Fax: (608) 646-7464<br>Email: Monica.Morrison@ge.com |
| Device Trade Name: | CARESCAPE R860 |
| Common/Usual Name: | Ventilator, Continuous |
| Classification Names: | Ventilator, continuous, facility use |
| Product Code: | CBK |
| Regulation Number: | 21 CFR 868.5895 |
| Predicate Device(s): | GE Datex-Ohmeda Engström Carestation, Engström Pro K111116<br>Maquet Servo-I Ventilator K123149 |
| Intended Use: | The CARESCAPE R860 ventilator is designed to provide<br>mechanical ventilation or support to neonatal, pediatric, and<br>adult patients weighing 0.25 kg and above. The<br>CARESCAPE R860 ventilator is a microprocessor based,<br>electronically controlled, pneumatically driven ventilator that<br>includes integrated monitoring of FiO2, airway pressure,<br>flow, and volume.<br>Additional respiratory gas monitoring capabilities are<br>supported through the use of optional GE patient monitoring |
| modules. | |
| Not all features are available for all patient types or product<br>configurations. | |
| The CARESCAPE R860 ventilator is not a pulmonary<br>function calculation device. | |
| The system is designed for facility use, including within-<br>facility transport, and should only be used under the orders of<br>a clinician. | |
{4}------------------------------------------------
# Device Description:
The CARESCAPE R860 is a flexible, adaptable, intuitive critical care ventilator. Touchscreen capability allows the user to quickly and easily access patient information and procedures. A wide selection of performance options gives the user full control of the system configuration. The CARESCAPE R860 features patient monitoring, patient ventilation, and the capability of interfacing with central information management systems.
The CARESCAPE R860 is designed to provide mechanical ventilation for adult, pediatric and neonatal patient types weighing 0.25 kg and above, and having degrees of pulmonary impairment varying from minor to severe.
The CARESCAPE R860 introduces a new user interface with touch screen capabilities. Icons represent configurable views of past (historical trends), present (patient status), and possible future patient needs through clinical decision support, including Spontaneous Breathing Trial to evaluate a patient's ability to breath spontaneously for a limited, specified duration of time.
This ventilator comes with standard ventilation modes as well as purchasable ventilation modes and clinical decision support features.
Standard ventilation modes:
- A/C VC (Assist Control Volume Control) .
- A/C PC (Assist Control Pressure Control) .
- A/C PRVC (Assist Control Pressure Regulated Volume Control)
- SIMV VC (Synchronized Intermittent Mandatory Ventilation Volume Control) ●
- SIMV PC (Synchronized Intermittent Mandatory Ventilation Pressure Control)
- CPAP/PS (Continuous Positive Airway Pressure/Pressure Support)
- SBT (Spontaneous Breathing Trial) .
Purchasable ventilation modes:
- . nCPAP (nasal Continuous Positive Airway Pressure)
- . SIMV PRVC (Synchronized Intermittent Mandatory Ventilation Pressure Regulated Volume Control)
{5}------------------------------------------------
- . BiLevel
- . BiLevel VG (BiLevel airway pressure ventilation Volume Guaranteed)
- VS (Volume Support) ●
- NIV (Non-Invasive Ventilation) .
- . APRV (Airway Pressure Release Ventilation)
Additional features:
- . FRC (Functional Residual Capacity)
- . SpiroDynamics
The CARESCAPE R860 is based on the Engström Carestation feature set and contains similar performance characteristics to the Engström family of ventilators.
The CARESCAPE R860 is a microprocessor-based, pneumatically controlled, data driven ventilator which includes integrated FiO2, airway pressure, spirometry and volume monitoring and an Aerogen Aeroneb nebulizer control board. The ventilator consists of two main components: the display and the ventilator unit. The display allows the user to interface with the system through a resistive touch screen and Trim Knob with keys. The CARESCAPE R860 also includes an optional module bay which allows the integration of various Datex-Ohmeda patient monitoring modules with the ventilator.
The user interface for control of nebulization is provided via the ventilator display unit. The standard nebulizer board is provided with the CARESCAPE R860. Users have the option to configure the system to use an external pneumatic nebulizer in place of the standard nebulizer.
Optional accessories common to the CARESCAPE R860 and the predicate Engström family of ventilators include a trolley/cart, integrated air compressor, support arm, humidifier and water trap mounting brackets. Additional optional accessories include airway modules, intratracheal pressure sensor, auxiliary electrical outlets, adjustable mounting rail, nebulizer and components, and module bay.
The optional medical air compressor is intended for use as an accessory to provide a dry, filtered, breathable compressed air supply. The compressor is installed in the base of the ventilator cart. The compressor is powered from AC mains only. A source of compressed oxygen is required to be connected to ventilator equipped with the optional compressor. The use of an integrated air compressor was first cleared on the predicate Engström Carestation and Engström Pro in K050597.
Optional functionality includes integrated respiratory gas monitoring, capabilities to measure SpiroDynamics via a GE supplied intratracheal pressure sensor in patients using sized 6.5 tracheal tubes and larger, and calculation of functional residual capacity of mechanically ventilated patients using Nitrogen Wash In/Wash Out method. The integrated respiratory gas monitoring is provided via the Datex-Ohmeda Gas Modules, E-CO, E-COV, E-COVX, E-CAiO, E-CAiOV, E-CAiOVX (K051092), E-MiniC module (K052582), or E-sCO, E-sCOV, E-sCAiO, E-sCAiOV (K123195) which are physically
{6}------------------------------------------------
integrated into the CARESCAPE R860, receive electronic power from the CARESCAPE R860 and communicate measured values to the CARESCAPE R860 for display on the system display unit.
### Summary of the Technological Characteristics of the Device:
The CARESCAPE R860 is based on the Engström Carestation feature set and contains similar performance characteristics to the Engström family of ventilators. Changes include the addition of an upgraded display, new graphical user interface, added accessories and upgraded software. The CARESCAPE R860 is designed to be compliant with ANSI/AAMI ES60601-1:2005 (R 2012), Medical electrical equipment, Part 1: General requirements for basic safety and essential performance and the relevant collateral standards. There are no changes to the intended use or fundamental scientific technology of the ventilator.
### Summary of Non-Clinical Testing for the Device:
The CARESCAPE R860 ventilator has been thoroughly tested through verification of specifications and validation, including software validation, to ensure the product is substantially equivalent to the predicate Engstrom Carestation. Verification of compliance with applicable standards has also been completed. The following quality assurance measures were applied during the development of the CARESCAPE R860 system:
- · Risk Analysis
- · Requirements/Specification Reviews
- · Design Reviews
- · Testing on unit level
- · Integration testing
- · Performance Testing (Verification)
- · Safety Testing (Verification)
- · Simulated Use/User Requirements Testing (Validation)
- · Standards Compliance the list of standards to which the CARESCAPE R860 complies is listed below:
- ANSI/AAMI ES60601-1:2005 (R 2012 O
- IEC 60601-1-2:2007 + 2010 Interpretation O
- . Includes additional testing applicable to RFID frequency ranges of 125 kHz/134 kHz, 13.56 MHz, 902-915 MHz and 2.4 GHz
- IEC 60601-1-6: 2010 O
- IEC 60601-1-8: 2006 O
- ISO 80601-2-12:2011 + Technical Corrigendum 1 O
{7}------------------------------------------------
- IEC 62366:2008 O
- ISO 5356-1 O
- IEC 62304 O
Extensive non-clinical testing was performed to establish substantial equivalence of the CARESCAPE R860. Verification and validation testing was performed according to predetermined acceptance criteria, which concluded that the CARESCAPE R860 is substantially equivalent to the predicate Engström Carestation.
## Summary of Clinical Testing for the Device:
The CARESCAPE R860 ventilator incorporates modifications to the predicate Engstrom Carestation. These modifications did not require clinical testing. The changes made were completely evaluated by non-clinical tests to verify and validate the substantial equivalence of the ventilator.
### Summary of Changes:
The following is an overview of the differences between the proposed CARESCAPE R860 and the predicate Engström Carestation:
- Upgraded 15 inch LCD and Tough Screen
- Simplified graphic User Interface for the features that exist in 7.x, but with a simplified hierarchy, designed for ease of use
- 트 Updated the names of the ventilation modes
- 트 Added Volume Support for Adult and Pediatric patients
- 트 Updated accessories list, including addition of a new optional compressor, the EVair compressor, Inspiratory Safety Guard, Accessory Rail and updated gas monitoring modules from GE
- . Visual differentiation of the neonatal patient type from the adult and pediatric patient type. Labeling will more clearly designate the neonatal patient type from the adult and pediatric
- I Updated User Requirements Manual and Technical Reference Manual which reflect the new user interface
- Compliant with ANSI/AAMI ES60601-1:2005 (R 2012), Medical electrical I equipment, Part 1: General requirements for basic safety and essential performance
## Determination of Substantial Equivalence:
Datex-Ohmeda. Inc.. doing business as GE Healthcare, considers the CARESCAPE R860 to be as safe and as effective, and performance is substantially equivalent to as the predicate device, the Engström Carestation. The summary above demonstrates that there are no new questions of safety or effectiveness for the CARESCAPE R860.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.