Surgical Gut Suture

K142656 · Covidien · GAL · Jan 29, 2015 · General, Plastic Surgery

Device Facts

Record IDK142656
Device NameSurgical Gut Suture
ApplicantCovidien
Product CodeGAL · General, Plastic Surgery
Decision DateJan 29, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4830
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Plain, Chromic and Mild Chromic gut absorbable sutures are in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in microsurgery, cardiovascular or neurological surgery.

Device Story

Absorbable sterile surgical sutures composed of purified bovine intestinal collagen; packaged in isopropanol/water/triethanolamine solution. Used by surgeons for soft tissue approximation and ligation. Manufacturing process includes sodium hydroxide bath for viral inactivation. Device provides mechanical wound closure; absorbed by body over time. Benefits include secure tissue approximation in general and ophthalmic procedures.

Clinical Evidence

Bench testing only. In vitro testing included visual/tactile inspection, suture removal, diameter, knot pull, and needle attachment. In vivo testing included strength loss and mass loss. Biocompatibility testing performed per ISO 10993-1.

Technological Characteristics

Absorbable suture material derived from bovine intestinal serosal collagen. Packaged in 87% isopropanol, 12% water, 1% triethanolamine. Tested per ISO 10993-1 for biocompatibility. Manufacturing includes 1N NaOH bath for viral inactivation.

Indications for Use

Indicated for general soft tissue approximation and/or ligation, including ophthalmic surgery. Contraindicated for microsurgery, cardiovascular, or neurological surgery.

Regulatory Classification

Identification

An absorbable surgical gut suture, both plain and chromic, is an absorbable, sterile, flexible thread prepared from either the serosal connective tissue layer of beef (bovine) or the submucosal fibrous tissue of sheep (ovine) intestine, and is intended for use in soft tissue approximation.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol of three human profiles facing to the right, representing health and human services. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 29, 2015 Covidien Ms. Mary Mellows Senior Specialist, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06473 Re: K142656 Trade/Device Name: Absorbable Surgical Gut Suture Regulation Number: 21 CFR 878.4830 Regulation Name: Absorbable surgical gut suture Regulatory Class: Class II Product Code: GAL Dated: December 23, 2014 Received: December 30, 2014 Dear Ms. Mellows: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {2}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. for Sincerely yours, # David Krause -S Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K142656 Device Name Absorbable Surgical Gut Suture Indications for Use (Describe) Plain, Chromic and Mild Chromic gut absorbable sutures are in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in microsurgery, cardiovascular or neurological surgery. Type of Use (Select one or both, as applicable) | <span style="font-size:10pt"> <span style="font-family:Arial"> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CEB 801 Subpart D) </span> </span> </span> | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size:10pt"> <span style="font-family:Arial"> <span> <input type="checkbox"/> Over-The-Counter Use (21 CER 801 Subpart C) </span> </span> </span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary This 510(k) summary information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.98. | SUBMITTER: | Covidien<br>60 Middletown Avenue<br>North Haven, CT 06473<br>(203) 492-5284 (T) | |--------------------------------------------------------------------|-------------------------------------------------------------------------------------| | CONTACT PERSON: | Mary Mellows<br>Senior Specialist, Regulatory Affairs | | DATE PREPARED: | September 17, 2014 | | TRADE/PROPRIETRY NAME: | Surgical Gut Suture | | COMMON/USUAL NAME: | Absorbable Surgical Gut Suture | | CLASSIFICATION NAME: | Suture, Absorbable, Natural | | FDA PANEL NUMBER: | 79 | | PRODUCT CODE: | GAL | | CLASS CODE: | Pursuant to 21 CFR 878.4830, absorbable surgical gut<br>suture is a Class II device | | LEGALLY MARKETED<br>DEVICES TO WHICH<br>EQUIVALENCY IS<br>CLAIMED: | Surgical Gut Suture (K885018) | {5}------------------------------------------------ REASON FOR 510(K) TECHNICAL SUBMISSION: Obtain clearance for Covidien's surgical gut suture (plain, mild, and chromic) with a manufacturing process modification to include an additional sodium hydroxide (1N NaOH) bath or soaking step inactivate any potential viruses. to In addition, the Contraindications and Warning sections of the Instructions for Use are being modified to make them clearer for the user. - DEVICE DESCRIPTION: Covidien's surgical gut sutures are absorbable sterile surgical sutures composed of purified connective tissue (mostly collagen) derived from the serosal layer of bovine intestines. They are packaged in a solution of 87% isopropanol, 12% water and 1% triethanolamine. - INTENDED USE: Plain, Chromic and Mild Chromic qut absorbable sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in microsurgery, cardiovascular or neurological surgery. - CHARACTERISTICS: The proposed surgical gut suture is substantially equivalent and its fundamental scientific technology has not been altered as compared to the predicate devices. - MATERIALS: All components of the surgical gut suture are similar to the predicate surgical gut suture. All materials are similar and have been tested in accordance with ISO Standard 10993-1. - PERFORMANCE DATA: Design verification and pre-clinical validation studies were conducted to demonstrate that the proposed surgical gut sutures are safe and effective and perform as intended. In vitro and in vivo testing to support the intended use of this device includes: - . In Vitro - Visual/Tactile o - Suture Removal O - Diameter O - Knot Pull O - Needle Attachment O - In Vivo ● - Strenath Loss O - Mass Loss o - Biocompatibility . CONCLUSION: The result of these tests demonstrates that the proposed surgical gut sutures are substantially equivalent to the predicate device and does not introduce additional risk to the patient.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...