Stabilynx PEEK Suture Anchor

K142643 · D3 Dc, LLC /Stabilynx, Inc. · MBI · Dec 15, 2014 · Orthopedic

Device Facts

Record IDK142643
Device NameStabilynx PEEK Suture Anchor
ApplicantD3 Dc, LLC /Stabilynx, Inc.
Product CodeMBI · Orthopedic
Decision DateDec 15, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Stabilynx Suture Anchor is a soft tissue anchor which will be used to secure soft tissue to bone during reconstructive surgery. The anchor is intended for use in such procedures as: • Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation Repair, Capsular Shift/Capsulolabral Reconstruction, Biceps Tendonesis, Deltoid Repair • Knee: Extra Capsular Repairs (Medial Collateral Ligament, Lateral Collateral Ligament, Posterior Oblique Ligament), Illiotibial Band Tendonesis, Patellar Tendon Repair • Elbow, Wrist, Hand: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Biceps Tendon Reattachment • Foot and Ankle: Medial Instability Repair/Reconstruction, Lateral Instability Repair/Reconstruction, Achilles Tendon Repair/Reconstruction, Midfoot Reconstruction, Hallux Valgus Reconstruction

Device Story

Stabilynx PEEK Suture Anchor is an implantable bone screw system for soft tissue-to-bone fixation. Device consists of PEEK bone screw, UHMWPE sutures, and disposable insertion tool (ABS handle, stainless steel shaft). Surgeon inserts anchor into bone; unique design allows creation of contiguous suture links between adjacent anchors; pre-loaded sutures can be exchanged between anchors. Used in orthopedic reconstructive procedures (shoulder, knee, elbow, wrist, hand, foot, ankle). Benefits include enhanced surgical flexibility for complex soft tissue repairs. Operated by orthopedic surgeons in clinical/OR settings.

Clinical Evidence

Bench testing only. Verification included dimensional/visual inspections, suture loop exchange force, fixation strength (with/without cycle loading), inserter/anchor torque strength, suture knot-pull tensile strength, and repair suture bifurcated loop strength. Biocompatibility testing per ISO 10993-1 (cytotoxicity, systemic toxicity). Sterilization validated per ISO 11135-1 and ISO 10993-7. Packaging/shipping validated per ASTM F88, ASTM 2096-11, and ASTM D4169-09. All tests met acceptance criteria.

Technological Characteristics

Materials: PEEK (bone screw), UHMWPE (suture), ABS (handle), stainless steel (shaft). Principle: Mechanical bone fixation fastener. Connectivity: None. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 10993-1, ISO 11135-1, ISO 10993-7, ASTM F88, ASTM 2096-11, ASTM D4169-09.

Indications for Use

Indicated for patients requiring soft tissue-to-bone fixation during reconstructive surgery of the shoulder, knee, elbow, wrist, hand, foot, and ankle.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other, with flowing lines extending from the bottom profile. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 15, 2014 Stabilynx, Incorporated Mr. Mike Rosenthal Vice President, Engineering Operations 3475-0 Edison Wav Menlo Park, California 94025 Re: K142643 Trade/Device Name: Stabilynx PEEK Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: September 15, 2014 Received: September 17, 2014 Dear Mr. Rosenthal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Lori A. Wiggins -S for Mark N. Melkerson Director Department of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K142643 Device Name Stabilynx PEEK Suture Anchor #### Indications for Use (Describe) The Stabilynx Suture Anchor is a soft tissue anchor which will be used to secure soft tissue to bone during reconstructive surgery. The anchor is intended for use in such procedures as: • Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation Repair, Capsular Shift/Capsulolabral Reconstruction, Biceps Tendonesis, Deltoid Repair · Knee: Extra Capsular Repairs (Medial Collateral Ligament, Lateral Ligament, Posterior Oblique Ligament), Illiotibial Band Tendonesis, Patellar Tendon Repair · Elbow, Wrist, Hand: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Biceps Tendon Reattachment · Foot and Ankle: Medial Instability Repair/Reconstruction, Lateral Instability Repair/Reconstruction, Achilles Tendon Repair/Reconstruction, Midfoot Reconstruction, Hallux Valgus Reconstruction Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Section 6: 510(k) Summary (21 CFR § 807.92(c)) # I: SUBMITTER INFORMATION | Submitter: | Stabilynx, Inc.<br>3475-0 Edison Way<br>Menlo Park, CA 94025 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Mike Rosenthal<br>Vice President, Engineering Operations, Stabilynx<br>Phone: 650.503.3329<br>Fax: 650.618.1440<br>Email: mrosenthal@d3dc.com | | Date Summary Prepared: | 15 September 2014 | #### II: SUBJECT DEVICE INFORMATION | Device Trade Name: | Stabilynx PEEK Suture Anchor | |----------------------|--------------------------------------------------------------------------| | Common Name: | Fastener, Fixation, Nondegradable, Soft Tissue | | Classification Name: | Smooth or threaded metallic bone fixation fastener<br>(21 CFR §888.3040) | | Product Code: | MBI | # III: PREDICATE DEVICE INFORMATION Stryker PEEK Zip Suture Anchor (K070758) Predicate Device: No recalls, market withdrawals or safety alerts were identified in FDA's database for the above referenced predicate device. No reference devices were used in this submission. #### IV: DEVICE DESCRIPTION: The Stabilynx PEEK Suture Anchor consists of an ethylene oxide sterilized, implantable, poly-ether-etherketone (PEEK) bone screw and ultra high molecular weight polyethylene (UHMWPE) suture anchor is supplied pre-mounted on a disposable Insertion Tool, which consists of a non-patient contact, acrylonitrile butadiene styrene (ABS) handle molded over an externally-communicating, stainless steel shaft. The anchor comes pre-loaded with two (2) high-strength, polyethylene, non-absorbable sutures. The Stabilynx PEEK Suture Anchor design allows a surgeon to create at least one contiguous suture link between one or more adjacent suture anchors after the suture anchors have been inserted into bone. The suture anchor is designed such that one or both of the pre-loaded sutures may be exchanged with one or more sutures connected to an adjacent suture anchor(s). {4}------------------------------------------------ # V. INDICATIONS FOR USE: The Stabilynx Suture Anchor is a soft tissue anchor which will be used to secure soft tissue to bone during reconstructive surgery. The anchor is intended for use in such procedures as: - . Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Acromio-Clavicular Separation Repair, Capsular Shift/Capsulolabral Reconstruction, Biceps Tendonesis, Deltoid Repair - . Knee: Extra Capsular Repairs (Medial Collateral Ligament, Lateral Collateral Ligament, Posterior Oblique Ligament), Illiotibial Band Tendonesis, Patellar Tendon Repair - . Elbow, Wrist, Hand: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Biceps Tendon Reattachment - . Foot and Ankle: Medial Instability Repair/Reconstruction, Lateral Instability Repair/Reconstruction, Achilles Tendon Repair/Reconstruction, Midfoot Reconstruction, Hallux Valgus Reconstruction # VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE: The primary technological principle for the subject and predicate devices is to anchor suture to bone so that the suture may be used to secure soft tissue to bone. The subject and predicate devices accomplish this function by similar means. At a high level, the subject and predicate devices are based on the following same technological elements: - . PEEK implantable bone screw and non-absorbable sutures - Ability to secure soft tissue to bone during reconstructive procedures The following technological differences exist between the subject and predicate devices: - Sutures in the subject device contain suture loops at either end - Suture loops of the subject device enable contiguous linking of adjacent suture anchors # VII. PERFORMANCE DATA: The following performance data were provided in support of the substantial equivalence determination. # Bench Testing Verification tests included: a) Dimensional and Visual Inspections; b) Suture Loop Exchange Force; c) Fixation Strength with and without Cycle Loading; d) Inserter and Anchor Torque Strength; e) Suture Knot-Pull Tensile Strength; and, f) Repair Suture Bifurcated Loop Strength. These tests were conducted on ethylene oxide sterilized units at baseline and on devices subjected to accelerated aging conditions equivalent to three (3) months. Passing results were obtained for all design verification tests. # Biocompatibility Testing The biocompatibility evaluation for the Stabilynx PEEK Suture Anchor was conducted in accordance with the requirements defined in ISO 10993-1 "Biological Evaluation of Medical Devices" and FDA's guidance ("Use of International Standard ISO 10993", draft document issued on 24 April 2013 and BlueBook Memorandum #G95-1 dated 01 May 1995). Testing included the following: 1) Cytotoxicity; and, 2) Systemic Toxicity. The device is comprised of: 1) an inserter shaft that is considered a tissue/bone {5}------------------------------------------------ contacting component for a duration of < 24 hours; and, 2) a bone anchor and suture that are considered permanent implants. ### Sterilization Data The sterilization parameters for the Stabilynx PEEK Suture Anchor device comply with the requirements prescribed in the applicable standards for ethylene oxide sterilization (ISO 11135-1:2007 "Sterilization of health care products -- Ethylene oxide -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices" and ISO 10993-7:2008 "Biological evaluation of medical devices -- Part 7: Ethylene oxide sterilization residuals"). The sterilization cycle ensures a SAL of 10 °. The EO and ECH residuals were below the limits specified in the standard. #### Packaging and Shipping Validation Packaging and shipping validation studies were successfully conducted on sterilized Stabilynx PEEK Suture Anchors pursuant to the applicable ASTM guidelines. These tests included seal peel and bubble emission tests at baseline and T= 3 months. Tests were conducted pursuant to ASTM F88 "Standard Test Method for Seal Strength of Flexible Barrier Materials"; and ASTM 2096-11 "Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)". Additionally, the ASTM D4169-09 standard was used to evaluate package performance under simulated transport conditions. All tests results met the acceptance criteria. #### VIII CONCLUSIONS: Based on the technological characteristics, indications for use, and bench performance data provided in this premarket notification, the Stabilynx PEEK Suture Anchor Device has a safety and effectiveness profile that is substantially equivalent to the predicate device. The information included in this 510(k) submission demonstrates the same intended use, similar indications for use and technological characteristics of the Stabilynx PEEK Suture Anchor Device as compared to the predicate device. The differences between the subject and predicate devices do not raise different types of safety or effectiveness questions and the performance data demonstrate substantial equivalence.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%