K142619 · Asante Solutions, Inc. · LZG · Dec 18, 2014 · General Hospital
Device Facts
Record ID
K142619
Device Name
Snap Insulin Pump System
Applicant
Asante Solutions, Inc.
Product Code
LZG · General Hospital
Decision Date
Dec 18, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Snap Insulin Pump System is intended for continuous, subcutaneous delivery of insulin at programmable basal and bolus rates for the management of diabetes mellitus in adult patients requiring insulin.
Device Story
Wearable, battery-powered insulin delivery system; consists of reusable software-controlled Controller module and single-use disposable Pump Body. User inserts pre-filled Eli Lilly Humalog U100 glass cartridge into Pump Body; Controller programs basal/bolus delivery parameters. Pump Body houses drive mechanism applying force to cartridge plunger to deliver insulin via connected infusion set (Asante Comfort/Conset or Luer-compatible via adapter). Used by adult patients for diabetes management; provides continuous subcutaneous infusion. Controller interface allows user to set delivery rates; system includes download cradle for data management. Benefits include precise, programmable insulin administration.
Clinical Evidence
Bench testing only. Verification and validation performed per FDA infusion pump guidance. Testing included risk analysis, software validation (IEC 62304), electrical safety (IEC 60601-1, IEC 60601-2-24), electromagnetic compatibility (IEC 60601-1-2, IEC 61000 series), and water ingress (IEC 60529). All testing met acceptance criteria, confirming performance reliability and substantial equivalence to the predicate.
Indicated for continuous subcutaneous insulin delivery at programmable basal and bolus rates for management of diabetes mellitus in adult patients requiring insulin.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Unomedical Asante Comfort or Asante Conset Infusion Set (K120872)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name, "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA," arranged in a circular pattern around the edge. At the center of the seal is a stylized emblem consisting of three human profiles facing to the right, stacked on top of each other, representing the department's focus on health and human well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18,2014
Asante Solutions, Inc. Edward J. Sinclair Vice President, Regulatory and Quality Affairs 352 East Java Drive Sunnyvale, California 94089
Re: K142619
Trade/Device Name: Snap Insulin Pump System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LZG Dated: November 20, 2014 Received: November 21, 2014
Dear Mr. Sinclair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno, DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K142619
Device Name Asante SnapTM Insulin Pump System
Indications for Use (Describe)
The Asante Snap Insulin Pump System is indicated for continuous delivery of insulin at programmable basal and bolus rates for the management of diabetes mellitus in adult patients requiring insulin.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
## 510(k) Summary
## Submitter Information
| Company: | Asante Solutions<br>352 East Java Drive<br>Sunnyvale, CA 94089<br>(408) 716-5600<br>Edward J. Sinclair<br>Vice President, Regulatory and Quality Affairs |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | December 18, 2014 |
## Device Name and Classification
| Common Name: | Insulin Infusion Pump |
|----------------------|---------------------------|
| Proprietary Name: | Snap™ Insulin Pump System |
| Classification Name: | Pump, Infusion, Insulin |
| Product Code: | LZG |
| Regulation Number: | 880.5725 |
| Class: | II |
## Predicate Device
Asante Pearl Diabetes Management System (K122483)
## Intended Use
The Snap Insulin Pump System is intended for continuous, subcutaneous delivery of insulin at programmable basal and bolus rates for the management of diabetes mellitus in adult patients requiring insulin.
{4}------------------------------------------------
## Device Description
The Snap™ Insulin Pump System is a wearable, continuous, programmable insulin delivery system. It is a portable, battery-powered system capable of delivering insulin at fixed and variable rates. The Snap system consists of the following components and accessories:
- 1. a software-controlled, programmable, reusable Controller module;
- 2. a single-use, disposable Pump Body;
- 3. an Unomedical Asante Comfort™ or Asante Conset™ Infusion Set (K120872), or an equivalently FDA-cleared infusion set with Luer fitting using the Asante sterile, single-use disposable Infusion Set Adapter; and
- 4. accessories, including the AsanteSync™ download cradle with USB cable, and a fabric Case and plastic Clip that provide a means of attaching the Snap system to the user's belt or other clothing.
The Controller module has a user interface to program delivery parameters for basal and bolus insulin delivery and the Pump Body provides the drive mechanism and battery power and holds the pre-filled insulin cartridges. The user obtains Eli Lilly Humalog fast-acting U100 insulin from their pharmacy and inserts the pre-filled glass cartridge into the Asante Pump Body. The Pump Body is connected to the Controller module and either (a) an Unomedical Asante Comfort or Asante Conset infusion set with proprietary connector, or (b) an Asante Infusion Set Adapter with Luer fitting that connects to commercially available Luered infusion sets to allow delivery of insulin for the management of diabetes.
## Summary of Technological Characteristics
The Snap Insulin Pump System has the same technological characteristics as the predicate device. The system uses existing established and unchanged technology and materials, relative to the previously cleared device. The system employs the same drive mechanism as that of the cleared Pearl System (#K122483) to apply controlled and accurate force to move the plunger in the insulin cartridge.
## Summary of Changes
The following changes have been made:
- Product trade name change from Pearl Diabetes Management System to Snap Insulin Pump System
- . Component substitutions in the Controller module to address discontinued parts from the respective manufacturer
{5}------------------------------------------------
- . Component additions to the Controller module for added electrical protection during assembly
- Increase in the water ingress rating of the Pump Body from IPX5 to IPX7
- Addition of a fabric case, plastic clip and download accessories ●
- Minor user interface improvements
- . Other minor software changes to accommodate componement substitutions. software anomaly resolution and improvements
# Performance Testing
Asante completed the appropriate verification and validation activities required by the Guidance of Industry and FDA Staff - Total Product Life Cycle: Infusion Pump -Premarket Notification [510(k)] Submissions Draft Guidance and other guidance, as applicable. The system has been verified for performance and functionality to provide assurance that the device modifications have been designed and tested to assure conformance to the requirements for its intended use, based on an evaluation of risks for the type of change involved. All the changes made to the device were assessed to make sure all the risks were sufficiently mitigated and that no new risks were introduced as a result of these modifications.
The following performance and safety testing has confirmed that the proposed device is substantially equivalent to the predicate device:
- Risk Analysis; Identification of the hazards associated with the modified device . in order to evaluate, estimate and control the associated risks was performed in accordance with ANSI/AAMI/ISO 14971 Application of risk analysis to medical devices.
- . Physical characteristics.
- . Software; documentation was prepared and submitted for a "major" Level of Concern device in accordance with FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Device software was verified and validated according to IEC 62304:2006 - Medical device software -Software life cycle processes and FDA's General Principles of Software Validation; Final Guidance for Industry and FDA Staff.
- Electrical Safety; the modified Snap Insulin Pump System has been tested and . successfully passed all of the relevant sections of the following standards:
- o IEC 60601-1 Medical electrical equipment -- Part 1: General requirements for safety.
- o IEC 60601-1-2 General Requirements for basic safety and essential performance - Collateral standard: Electromagnetic Compatibility -Requirements and Tests.
{6}------------------------------------------------
- o IEC 60601-2-24 Medical electrical equipment Part 2-24: Particular requirements for the basic safety and essential performance of infusion pumps and controllers.
- Water Ingress: the modified Snap Insulin Pump System was tested for the ● protection of the equipment inside the Controller and Pump Body enclosures against harmful effects due to the ingress of water, in accordance with IEC 60529 Degrees of protection provided by enclosures (IP Code).
- Electromagnetic Interference; the modified device has been tested and . successfully passed all of the relevant sections of the following standards:
- IEC 61000-4-2. Electromagnetic Compatibility (EMC) Part 4-2: Testing o and measurement techniques – Electrostatic discharge immunity test.
- o IEC 61000-4-3, Electromagnetic Compatibility (EMC) Part 4-3: Testing and measurement techniques - - Radiated, radio-frequency, electromagnetic field immunity test.
- IEC 61000-4-8, Electromagnetic Compatibility (EMC) Part 4-8: Testing o and measurement techniques - Power frequency magnetic field test.
All testing met the acceptance criteria.
## Conclusion From Data
Performance testing on the modified device demonstrated that the Snap Insulin Pump System performs reliably and delivers insulin as intended through the infusion sets, and is as safe and as effective as the predicate. Design verification testing confirmed that no new questions of safety or effectiveness were identified during device testing of the modified device.
Based upon the successful safety and performance tests and the similarities to predicate device, the subject Snap Insulin Pump System is substantially equivalent to the predicate Pearl Diabetes Management System in design, features, performance, fundamental scientific technology, and indications for use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.