← Product Code [LPH](/productcode/LPH) · K142606

# Trident® Tritanium® PST® Acetabular Shells (K142606)

_Mako Surgical Corp. · LPH · Jan 26, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K142606

## Device Facts

- **Applicant:** Mako Surgical Corp.
- **Product Code:** [LPH](/productcode/LPH.md)
- **Decision Date:** Jan 26, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Trident® Tritanium® PST® Acetabular Shell System is indicated for use in skeletally mature individuals undergoing surgery for total hip replacement due to: . A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, or congenital hip dysplasia; . Acute traumatic fracture of the femoral head or neck; Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement. The Trident® Tritanium® PST® Acetabular Shell is intended for cemented fixation. The porous structured surface provides biological fixation when used in a cementless application.

## Device Story

Acetabular shell component for total hip replacement; implanted in surgical setting by orthopedic surgeon. Device features porous titanium alloy surface for biological fixation (cementless) or cemented fixation. Modified from predicate PST shell to incorporate bearing mating features compatible with existing Stryker Orthopaedics acetabular bearings (polyethylene and Modular Dual Mobility families). Available in solid or cluster screw hole configurations for optional supplemental bone screw fixation. Provides stable acetabular component for hip joint reconstruction; restores joint function; reduces pain in patients with degenerative or traumatic hip conditions.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including disassembly testing (push-out and torsional) per ASTM F1820, comparing subject shells with worst-case compatible Stryker Orthopaedics acetabular bearings.

## Technological Characteristics

Titanium alloy construction; porous structured surface for biological fixation. Available in solid or cluster screw hole configurations. Compatible with titanium alloy bone screws and CP titanium/titanium alloy dome hole occluders. Mechanical testing per ASTM F1820. Sterilization method same as predicate.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- PST® Acetabular Shells ([K112802](/device/K112802.md))
- Trident® Tritanium® ([K081171](/device/K081171.md))
- Trident® PSL® ([K983382](/device/K983382.md))
- Trident® Hemispherical Shells ([K013676](/device/K013676.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles incorporated into its design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 26, 2015

MAKO Surgical Corporation % Terry Sheridan Powell M Squad Associates, Incorporated 575 Eighth, Avenue, Suite 1212 New York City, New York 10018

Re: K142606 Trade/Device Name: Trident® Tritanium® PST® Acetabular Shells Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, JDI Dated: December 22, 2014 Received: December 29, 2014

Dear Terry Sheridan Powell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

#### Page 2 - Terry Sheridan Powell

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K142606

Device Name

Trident® Tritanium® PST® Acetabular Shells

#### Indications for Use (Describe)

The Trident® Tritanium® PST® Acetabular Shell System is in skeletally mature individuals undergoing surgery for total hip replacement due to:

· A severely painful and/or disabled joint from osteoarthritis, theumatoid arthritis, avascular necrosis, or congenital hip dysplasia;

· Acute traumatic fracture of the femoral head or neck;

· Failed previous hip surgery including joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement.

The Trident® Tritanium® PST® Acetabular Shell is intended for cemented fixation. The porous structured surface provides biological fixation when used in a cementless application.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) Summary is provided in accordance with the requirements of 21 CFR 807.92.

#### 1.1. Device Name and Classification

| Device Trade Name:                 | Trident® Tritanium® PST® Acetabular Shells                                                                                                                                              |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Common Name:                | Artificial Total Hip Replacement                                                                                                                                                        |
| Regulation Number and Description: | 888.3358 - Hip joint metal/polymer/metal semi-<br>constrained porous-coated uncemented<br>prosthesis; and 888.3350 - Hip joint<br>metal/polymer semi-constrained cemented<br>prosthesis |
| Device Class:                      | II                                                                                                                                                                                      |
| Product Codes:                     | LPH - prosthesis, hip, semi-constrained,<br>metal/polymer, porous uncemented<br>JDI - prosthesis, hip, semi-constrained,<br>metal/polymer, cemented                                     |
| Advisory Panel:                    | Orthopedic                                                                                                                                                                              |

### 1.2. Address and Registration

| Submitter's Name:                  | MAKO Surgical Corp.                           |
|------------------------------------|-----------------------------------------------|
| Address:                           | 3 Wing Drive Suite 102 Cedar Knolls, NJ 07927 |
| Establishment Registration Number: | 3009701876                                    |
| Contact Person:                    | Jonathan Reeves                               |
| Telephone Number:                  | 954-628-0665                                  |
| Fax Number:                        | 954-927-0446                                  |
| Date Summary Prepared:             | January 8, 2015                               |

### 1.3. Purpose of Submission

The Trident® Tritanium® PST® Acetabular Shell is a new device.

### 1.4. Identification of Legally Marketed Device to which Submitter Claims Equivalence

The subject device is substantially equivalent to the following predicate devices:

| Device Description                                                      | 510(k) Numbers            |
|-------------------------------------------------------------------------|---------------------------|
| PST® Acetabular Shells (and dome hole plug)                             | K112802                   |
| Trident® Tritanium®, Trident® PSL® and Trident®<br>Hemispherical Shells | K081171, K983382, K013676 |

{4}------------------------------------------------

### 1.5. Device Description

The subject Trident® Tritanium® PST® Acetabular Shell is the predicate PST® Acetabular Shell, which has been modified to incorporate the acetabular bearing mating features of the predicate Trident® Tritanium®, Trident® PSL®, and Trident® Hemispherical Shells, to allow compatibility with existing Stryker Orthopaedics acetabular bearings including the Trident polyethylene and Modular Dual Mobility (MDM) bearing families.

The subject Trident® Tritanium® PST® Acetabular Shells are manufactured from titanium alloy and feature a porous structured surface. The subject shells feature a dome hole, and are available either in a solid shell (no screw holes) configuration, or in a cluster screw hole configuration for optional supplemental bone screw fixation. The subject Tritanium® PST® Acetabular Shells are compatible with the optional predicate PST Acetabular Shell system bone screws (titanium alloy), with the optional predicate Stryker Orthopaedics Torx head acetabular dome hole occluder (CP titanium), and with an optional new hex head acetabular dome hole occluder (titanium alloy).

# 1.6. Intended Use

The Trident® Tritanium® PST® Acetabular Shell System is indicated for use in skeletally mature individuals undergoing surgery for total hip replacement due to:

- . A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, or congenital hip dysplasia;
- . Acute traumatic fracture of the femoral head or neck;
- Failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty or total hip replacement.

The Trident® Tritanium® PST® Acetabular Shell is intended for cemented fixation. The porous structured surface provides biological fixation when used in a cementless application.

# 1.7. Comparison of Technological Characteristics

The subject Trident® Tritanium® PST® Acetabular Shells have the same indications for use, same outer profile, same porous structured surface, and are made from the same materials using the same manufacturing, similar packaging, and same sterilization methods as the predicate PST® Acetabular Shells. The subject Trident® PST® Acetabular Shells incorporate the acetabular bearing mating features of the predicate Trident® Trident® PSL® and Trident® Hemispherical Shells, and are therefore compatible with existing Stryker acetabular bearings, including the Trident polyethylene and Modular Dual Mobility (MDM) bearing families.

# 1.8. Performance Testing

Comparisons of materials, manufacturing methods, and design features, as well as disassembly testing per ASTM F1820 (including push-out, and torsional testing) for the subject shells with the worst case style and size of compatible Stryker Orthopaedics acetabular bearings, were relied upon to demonstrate the substantial equivalence of the modified subject device to the predicate devices.

{5}------------------------------------------------

## 1.9. Clinical Testing

Clinical testing was not required to demonstrate substantial equivalence to the predicates.

### 1.10.Conclusions

The subject Trident® Tritanium® PST® Acetabular Shells share indications for use, same outer profile, same porous structured surface, and are made from the same materials using the same manufacturing, similar packaging, and same sterilization methods as the predicate PST® Acetabular Shells. The subject Trident® Tritanium® PST® Acetabular Shells share the acetabular bearing mating features of the predicate Trident® Trident® PSL® and Trident® Hemispherical Shells, for compatibility with the same predicate Trident polyethylene and Modular Dual Mobility (MDM) bearing families. Therefore, the subject Trident® Tritanium® PST® Acetabular Shells are substantially equivalent to the identified predicate acetabular shells.

---

**Source:** [https://fda.innolitics.com/device/K142606](https://fda.innolitics.com/device/K142606)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
