MAGNETOM Combi Suite Neurosurgery for the MAGNETOM Aera, MAGNETOM Combi Suite Neurosurgery for the MAGNETOM Skyra
K142515 · Siemens Medical Solutions USA, Inc. · LNH · Oct 8, 2014 · Radiology
Device Facts
Record ID
K142515
Device Name
MAGNETOM Combi Suite Neurosurgery for the MAGNETOM Aera, MAGNETOM Combi Suite Neurosurgery for the MAGNETOM Skyra
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Oct 8, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
Your MAGNETOM MR system [Aera/Skyra] is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis. Your MAGNETOM MR system [Aera/Skyra] may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Device Story
MAGNETOM Combi Suite Neurosurgery is an optional hardware package for MAGNETOM Aera and Skyra MR systems; enables patient transfer between OR table and MR scanner via Combi Dockable Table. Facilitates intraoperative imaging during neurosurgical procedures. System inputs include MR signals processed into transverse, sagittal, coronal, and oblique cross-sectional images/spectra. Operated by clinical staff in hospital/neurosurgical environments. Output consists of diagnostic images/spectra interpreted by physicians to assist in diagnosis and surgical guidance. Benefits include seamless patient transport and real-time intraoperative imaging capabilities without requiring patient repositioning between separate OR and MR tables.
Clinical Evidence
Bench testing only. Includes biocompatibility (EN ISO 10993-1: cytotoxicity, skin irritation, sensitization), electrical safety (IEC 60601-1), and software verification/validation. Performance testing covered MR imaging, spectroscopy, static magnetic field strength, acoustic noise, dB/dt, and RF heating. No clinical data presented.
Technological Characteristics
MR diagnostic device; hardware includes Combi Dockable Table, interface boxes, transfer boards, and NORAS2 head holder. Complies with IEC 60601-1 (electrical safety) and ISO 14971 (risk management). Software version syngo MR D13A/D13C. No active electrical circuits in dockable table requiring EMC testing per IEC 60601-2-33.
Indications for Use
Indicated for patients requiring diagnostic MR imaging of the head, body, or extremities, including intraoperative imaging during neurosurgical procedures. Contraindications include patients with non-MR-safe implants or devices.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
MAGNETOM Artis Combi Suite for MAGNETOM systems Aera and Skyra (K140253)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads facing to the right. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 8, 2014
Siemens Medical Solutions USA, Inc. % Cordell L. Fields, Esq. Regulatory Affairs Technical Specialist 51 Valley Stream Parkway, Mail Code D02 MALVERN PA 19355
Re: K142515
Trade/Device Name: MAGNETOM Combi Suite Neurosurgery for the MAGNETOM Systems Aera and Skyra Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: September 5, 2014 Received: September 8, 2014
Dear Mr. Fields:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K142515
#### Device Name
"MAGNETOM Combi Suite Neurosurgery" for the MAGNETOM systems Aera and Skyra
#### Indications for Use (Describe)
Your MAGNETOM MR system [Aera/Skyra] is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis.
Your MAGNETOM MR system [Aera/Skyra] may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Section 5 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
## l General Information
| Establishment | Siemens Medical Solutions USA, Inc.<br>51 Valley Stream Parkway<br>Mail Code D02<br>Malvern, PA 19355, USA<br>Registration Number: 2240869 |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Siemens AG<br>Henkestrasse 127<br>D-91052 Erlangen, Germany<br>Registration Number: 3002808157<br><br>SIEMENS SHENZHEN MAGNETIC RESONANCE<br>LTD.<br>Siemens MRI Center<br>Hi-Tech Industrial park (middle)<br>Gaoxin C. Ave., 2nd<br>Shenzhen 518057, P.R. CHINA<br>Registration Number: 3004754211 |
| Contact Person | Cordell L. Fields, Esq.<br>Regulatory Affairs Technical Specialist<br>Siemens Medical Solutions USA, Inc.<br>51 Valley Stream Parkway<br>Mail Code D02<br>Malvern, PA 19355, USA<br>Phone: (610) 219-8518<br>Fax: (610) 427-3465<br>E-mail: Cordell.Fields@siemens.com |
### Classification and Device Name
| Trade name: | MAGNETOM Combi Suite Neurosurgery for the<br>MAGNETOM systems Aera and Skyra. |
|-----------------------|-------------------------------------------------------------------------------|
| Classification Name: | Regulation Description:<br>- Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| Regulation number: | 21 CFR § 892.1000 |
MAGNETOM Combi Suite Neurosurgery for the MAGNETOM systems Aera/Skyra
{4}------------------------------------------------
Device Class:
Product Code:
## II Safety and Effectiveness Information Supporting Substantial Equivalence
ll
LNH
### Device Description
MAGNETOM Combi Suite Neurosurgery is the marketing name for an optional package for the MAGNETOM systems Aera and Skyra that includes the modified Combi Dockable Table for use with standalone OR tables (Operation Room) in a clinical workflow. The market name, MAGNETOM Combi Suite Neurosurgery, refers to a sales bundle of single items that have been combined to build up a patient transfer workflow. MAGNETOM Combi Suite Neurosurgery enables the user to move a patient via the Combi Dockable Table from the OR table to the MR and vice versa in order to acquire intraoperative images.
The intended use of the MAGNETOM Aera, MAGNETOM Skyra, and the supported OR-Tables remains unchanged.
The MAGNETOM Combi Suite Neurosurgery will only be available for MAGNETOM systems Aera and Skyra and will be available as an option to newly manufactured scanners; existing scanners can be upgraded with this sales bundle. The sales bundles include new hardware for the MAGNETOM systems Aera and Skyra.
### Summary of new features for MAGNETOM Combi Suite Neurosurgery:
### Software
No software changes have been made with respect to the predicate device MAGNETOM Artis Combi Suite with K140253 cleared on March 20, 2014.
### Hardware
Maquet Package
- . Combi Dockable Table
- Interface Box
- NORAS2 head holder ●
<sup>2</sup> The NORAS head holder is a medical device in its own right and was cleared under K133506 on August 6, 2014 and is not subject to this 510(k) submission
{5}------------------------------------------------
### Trumpf Package (TruSystem 7500)
- . Combi Dockable Table
- . Interface Box
- Transfer Board
- Support cushion for Transfer Board ●
- NORAS2 head holder
- Floor dock o
### Substantial Equivalence
The sales bundle MAGNETOM Combi Suite Neurosurgery for the MAGNETOM systems Aera and Skyra is substantially equivalent to the following current legally marketed device (Table 1):
| Table 1: | Predicate device for MAGNETOM Combi Suite Neurosurgery for the<br>MAGNETOM systems Aera and Skyra |
|----------|---------------------------------------------------------------------------------------------------|
|----------|---------------------------------------------------------------------------------------------------|
| Predicate Device<br>Name | FDA Clearance<br>Number / Date | Manufacturer | Manufacturer,<br>Product Code | Claim Substantial<br>Equivalence to |
|----------------------------------------------------------------------------|---------------------------------------|--------------|-------------------------------|-------------------------------------------------------------------------|
| MAGNETOM Artis<br>Combi Suite for<br>MAGNETOM<br>systems Aera and<br>Skyra | K140253,<br>cleared March<br>20, 2014 | Siemens AG | LNH | MAGNETOM Artis<br>Combi Suite for<br>MAGNETOM systems<br>Aera and Skyra |
This predicate has not been subject to any recalls.
### Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk Management is ensured via a Risk Analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions USA. Inc. and Siemens AG adhere to recognized and established industry standards, such as the IEC 60601-1 series, to minimize electrical and mechanical hazards.
The MAGNETOM systems Aera and Skyra with sales bundle MAGNETOM Combi Suite Neurosurgery conform to the applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as recommended by the respective MR FDA Guidance Document.
### Indications for Use
The MAGNETOM MR [Aera/Skyra] system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.
Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
{6}------------------------------------------------
The MAGNETOM MR system [Aera/Skyra] may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR-safe biopsy needles.
### Comparison of Technological Characteristics with the Predicate Device & Conclusion as to Substantial Equivalence
MAGNETOM Aera/Skyra with software syngo MR D13A as well as MAGNETOM Skyra with software syngo MR D13C combined with MAGNETOM Combi Suite Neurosurgery has the same intended use and the same technical characteristics as the predicate device (Table 1: Predicate device for MAGNETOM Combi Suite Neurosurgery for the MAGNETOM systems Aera and Skyra with respect to the magnetic resonance features and functionalities. The differences between the subject device and the predicate device. which include the aforementioned hardware, provide the MAGNETOM systems Aera and Skyra with additional capabilities by allowing the user to transfer a patient via the Combi Dockable Table from the OR table to the MR and vice versa in order to acquire intraoperative images.
There are no new issues of safety or effectiveness introduced with the MAGNETOM Combi Suite Neurosurgery. The basic functionality of MAGNETOM Combi Suite Neurosurgery remains the same to the predicate device (Table 1: Predicate Neurosurgery for the MAGNETOM systems Aera and Skyra). However with this package, the MAGNETOM CombiSuite can now also be used within a Neurosurgery workflow / environment.
Therefore, Siemens believes that the subject devices, MAGNETOM Combi Suite Neurosurgery, are substantially equivalent to the predicate device, MAGNETOM Artis Combi Suite (Table 1).
### III Performance Data
The following performance data is being provided in support of the substantial equivalence:
### Biocompatibility Testing
The biocompatibility evaluation for MAGNETOM Combi Suite Neurosurgery was conducted in accordance with EN ISO 10993-1. The testing included the following tests:
- · Cytotoxicity
- Skin Irritation .
- o Sensitization
### Electrical safety and electromagnetic compatibility (EMC)
The MAGNETOM Combi Suite Neurosurgery does not contain active electrical circuits that require electromagnetic compatibility testing when not connected to the MR System.
{7}------------------------------------------------
In addition, within particular standard IEC 60601-2-33, chapter 202.6.101 the components of the MR system located inside the controlled area, are excluded from the requirements given in IEC60601-1-2: 2007, thus the Combi Dockable Table, even when energized via the MR System, does not need to be tested for electromagnetic compatibility.
For electrical safety the system complies to the IEC 60601-1.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and Staff "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a "moderate" level of concern, since a failure or latent design flaw could directly result in minor injury to the patient or operator.
### Bench and Acoustic Testing
- . MR Imaqing
- . Spectroscopy
- Safetv
- Static Magnetic Field Strength O
- Acoustic noise level o
- dB/dt O
- RF Heating o
## IV Conclusions
The submitted data support the safety of the device. The hardware and software verification and validation demonstrate that the MAGNETOM CombiSuite Neurosurgery device performs as specified.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.