K142463 · Bio2 Technologies, Inc. · MQV · Dec 18, 2014 · Orthopedic
Device Facts
Record ID
K142463
Device Name
Bio2 CLM BG Bioactive Scaffold
Applicant
Bio2 Technologies, Inc.
Product Code
MQV · Orthopedic
Decision Date
Dec 18, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bio2 CLM-BG Bioactive Scaffold is intended for use as a bone void filler for bony voids or gaps that are not intrinsic to the stability of the bony structure. Bio2 CL.M+BG Bioacive Scaffold is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. extremitics and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
Device Story
Bio2 CLM-BG Bioactive Scaffold is a synthetic, non-collagenous, osteoconductive bone void filler composed of fiber-based bioactive glass. Available as cylinders, blocks, and wedges, the device is packed into non-structural bony voids or gaps in the extremities and pelvis. Upon implantation, the bioactive glass surface undergoes kinetic modification, forming a calcium phosphate layer similar to bone hydroxyapatite; this layer acts as a scaffold for new bone tissue infiltration. The fibers are slowly resorbed and replaced by host bone during the healing process. The device is provided sterile for single-use by surgeons in clinical settings. The radiopaque nature of the material allows for post-operative monitoring via X-ray. The device benefits patients by facilitating the repair of osseous defects through a resorbable, bioactive scaffolding mechanism.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing and animal studies. In vitro testing (ISO 23317:2012) confirmed formation of a hydroxyapatite layer in simulated body fluid. In vivo testing in a rabbit critical size defect model evaluated bone formation at 8 and 16 weeks using X-ray, histology, histomorphometry, SEM, and EDX. Results demonstrated the device is as safe and effective as the control material.
Technological Characteristics
Material: Fiber-based bioactive glass; synthetic and non-collagenous. Complies with ASTM F-1538. Form factor: Cylinders, blocks, and wedges. Principle: Osteoconductive, bioactive, resorbable scaffold. Radiopaque. Sterilization: Provided sterile for single use. Connectivity: None.
Indications for Use
Indicated for patients with bony voids or gaps in the skeletal system (extremities and pelvis) that are not intrinsic to the stability of the bony structure, including surgically created osseous defects or those resulting from traumatic injury.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure in profile, with three faces overlapping to symbolize health, human services, and the relationship between them. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18, 2014
Bio2 Technologies, Inc. Janet Krevolin, Ph.D. Chief Technical Officer 12 R Cabot Road Woburn, Massachusetts 01801
Re: K142463
Trade/Device Name: Bio2 CLM.BG Bioactive Scaffold Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: November 6, 2014 Received: November 7, 2014
Dear Dr. Krevolin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K142463
Device Name Bio2 CLM.BG Bioactive Scaffold
### Indications for Use (Describe)
The Bio2 CLM-BG Bioactive Scaffold is intended for use as a bone void filler for bony voids or gaps that are not intrinsic to the stability of the bony structure. Bio2 CL.M+BG Bioacive Scaffold is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. extremitics and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 5. 510(k) Summary
Summary of 510 (k) safety and effectiveness information upon which the substantial equivalence determination is based:
| I. Submitter | |
|-----------------------|-------------------------------------------------------------------|
| Date Prepared: | December 4, 2014 |
| Device Submitter: | Bio2 Technologies |
| | 12-R Cabot Road |
| | Woburn, MA 01801 |
| Phone: | 781-721-6309 |
| Contact Person: | Janet Krevolin, PhD |
| II. Device | |
| Device Name: | Bio2 CLM•BG Bioactive Scaffold |
| Common Name: | Bone void filler |
| Classification Name: | Resorbable calcium salt bone void filler device (21 CFR 888.3045) |
| Regulatory Class: | II |
| Product Code: | MQV |
| III. Predicate Device | |
| Predicate Device: | NovaBone K101860 |
| | Inion BioRestore K090177 |
| | OsteoBiologics Inc. PolyGraft K033707 |
| | Synthes chronOS K013072 |
IV. Device Description
The Bio2 Technologies implants are bone void fillers in the shape of cylinders, blocks and wedges. The devices are osteoconductive, bioactive, bone void fillers. The implants are made from a fiber based bioactive glass. The device structure allows tissue infiltration between the bioactive glass fibers. The fibers then are slowly absorbed and replaced by new bone tissue during the healing process. The cylinders, blocks and wedges are provided sterile and are intended for single use.
V. Indications for Use
The Bio2 CLM+BG Bioactive Scaffold is intended for use as a bone void filler for bony voids or gaps that are not intrinsic to the stability of the bony structure. Bio2 CLM-BG Bioactive Scaffold is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e.
{4}------------------------------------------------
extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.
The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
#### VI. Comparison of technological characteristics with the predicate device
Bio2 CLM-BG Bioactive Scaffold like the predicate devices is an osteoconductive, bioactive bone graft device. The Bio2 CLM.BG Bioactive Scaffold complies with the requirements of ASTM F-1538. Similar to the predicate devices In vitro testing confirms the formation of a hydroxyapatite layer on the surface of the implant when immersed in simulated body fluid. In vivo tests have demonstrated bone formation at each post-implantation time point. In the study critical size defects were filled with Bio2 CLM+BG Bioactive Scaffold and the control material. Similar to the predicate device the rabbit femurs were evaluated at 8 and 16 weeks using x-ray, histology, histomorphometry, SEM and EDX. The Bio2 CLM+BG Bioactive Scaffold was shown to be as safe and effective as the control material.
Like the predicate devices Bio2 CLM+BG Bioactive Scaffold consists of different size cylinders, blocks and wedges made of degradable and resorbable bioactive glass. When implanted, a kinetic modification of the surface occurs, resulting in the formation of a calcium phosphate layer that is essentially similar in composition and structure to the hydroxyapatite found in bone mineral. This apatite layer provides scaffolding onto which the patient's new bone will grow allowing repair of the defect. The material is radiopaque. Bio2 CLM+BG Bioactive Scaffold implants are intended for single use and are provided sterile to the user. They are completely synthetic and non-collagenous.
| Item | Bio2 | NovaBone | Synthes |
|------------------------------------------------------------------|---------------------------------|-------------------------------------|---------------------------------------------|
| Use | Single use/sterile | Single use/sterile | Single use/sterile |
| Intended use | Bone void filler | Bone void filler | Bone void filler |
| Material | Bioactive glass | 45S5 Bioglass/<br>synthetic binder. | $\beta$ -Tricalcium<br>Phosphate[Ca3(PO4)2] |
| | Synthetic and non-collagenous | Synthetic and non-collagenous | Synthetic and non-collagenous |
| Available Shapes | Cylinders, blocks<br>and wedges | Putty | Cylinders, blocks and<br>wedges |
| Biocompatible | Yes | Yes | Yes |
| Form HA in SBF<br>solution | Yes | Yes | - |
| Radiopaque | Yes | - | Yes |
| Resorb and form bone in<br>critical size defect,<br>rabbit model | 8 and 16 weeks | 8 and 16 weeks | - |
{5}------------------------------------------------
#### VII. Performance Data
Biocompatibility of the device has been established according to blue book memorandum #G95-1 entitled "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing," In vitro bioactivity testing (ISO 23317:2012) shows the material forms a surface apatite layer when submerged in simulated body fluid.
In vivo animal studies show the device achieves bony healing in a critical defect model, confirmed with radiographs, histology and histomorphometry.
#### Conclusions VIII.
The Bio2 CLM+BG Bioactive Scaffold when compared to the predicate have the same intended use and similar indications, technological characteristics, and principals of operation as its predicate devices. In vivo test data demonstrates that the Bio2 CLM-BG Bioactive Scaffold is as safe and effective as the predicate devices. Thus the Bio2 CLM•BG Bioactive Scaffold is substantially equivalent to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.