K142424 · North East Monitoring, Inc. · MWJ · Nov 25, 2014 · Cardiovascular
Device Facts
Record ID
K142424
Device Name
DR300 HOLTER MONITOR
Applicant
North East Monitoring, Inc.
Product Code
MWJ · Cardiovascular
Decision Date
Nov 25, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2800
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Holter Mode: Detection of Arrhythmias, Efficacy of Pharmacological Treatment and Pacemaker Evaluation. Event Recorder: The event recorder module is a patient activated device designed to record and for diagnostic evaluation of transient symptoms (such as dizziness, palpitations, syncope, and chest pain). Once data is recorded, the data is transmitted for evaluation.
Device Story
Ambulatory ECG recorder (Holter/Event monitor) for cardiac rhythm monitoring; records ECG signals via patient-worn electrodes and cables; stores data on removable SD card. Features Bluetooth transmitter for wireless data transmission to NorthEast Monitoring gateway or PC USB adapter. Operated by physician/technician for hookup; patient-activated for event recording. Data downloaded to TelePro analysis software for physician review. Assists in diagnosing arrhythmias and evaluating treatment efficacy. Not for real-time telemetry of life-threatening arrhythmias.
Clinical Evidence
Bench testing only. Verification included 11 Holter tests (24-hour to 14-day duration) to confirm functionality with added Bluetooth and shielding. Wireless communication of encrypted data was validated through simulated bench testing (24-hour to 48-hour duration). Compliance with IEC 60601-1, 60601-1-6, 60601-1-11, 60601-2-47, and 60601-1-2 standards demonstrated.
Technological Characteristics
Ambulatory ECG recorder; 2 or 3-channel recording; uses standard Ag/AgCl electrodes. Powered by 1.5V AA alkaline, NiMH, or Lithium L91 battery. Connectivity via Bluetooth wireless transmitter (encrypted data). Includes LCD for status/lead quality. Standards: IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-2-47, IEC 60601-1-2.
Indications for Use
Indicated for patients, including infants <10 kg, requiring ambulatory cardiac monitoring for arrhythmia detection, pharmacological treatment efficacy, pacemaker evaluation, or diagnostic evaluation of transient symptoms like dizziness, palpitations, syncope, and chest pain.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 25, 2014
North East Monitoring, Inc. c/o Mr. Joe Azary Official Correspondent 2 Clock Tower Place, Suite 555 Maynard, Massachusetts 01754
Re: K142424
> Trade/Device Name: DR300 Holter Monitor Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MWJ Dated: October 3, 2014 Received: October 6, 2014
Dear Mr. Joe Azary,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K142424
Page __1__ of__1_
510(k) Number (if known):
Device Name: __ DR300 Holter Monitor__________________________________________________________________________________________________________________________________________
Indications For Use:
Holter Mode: Detection of Arrhythmias, Efficacy of Pharmacological Treatment and Pacemaker Evaluation.
Event Recorder: The event recorder module is a patient activated device designed to record and for diagnostic evaluation of transient symptoms (such as dizziness, palpitations, syncope, and chest pain). Once data is recorded, the data is transmitted for evaluation.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109)
Over-The-Counter Use
OR
{3}------------------------------------------------
# K142424 Page 1 of 5
# 510(k) Summary for North East Monitoring DR300
### 1. SUBMITTER/510(K) HOLDER
North East Monitoring Two Clock Tower Place Suite 555 Maynard, MA 01754
Contact Name: Joseph Azary Email: jazary@erols.com Telephone: (203) 242-6670
Date Revised: November 4, 2014
### 2. DEVICE NAME
| Proprietary Name: | DR300 Holter Monitor |
|----------------------|-------------------------------------------------------------------------------------------|
| Common/Usual Name: | Ambulatory ECG recorder, Ambulatory Electrocardiograph (without analysis), Holter Monitor |
| Classification: | Class 2, 21 CFR 870.2800 |
| Classification Name: | Ambulatory ECG recorder |
| Product Code: | MWJ |
#### 3. PREDICATE DEVICES
- K061293 Telaheart Digital Recorder / DR200 .
#### 4. DEVICE DESCRIPTION
The DR300 is an ambulatory monitor, sometimes called a Holter, which is a painless method to monitor the heart beat for a period of time (such as 24 hours, 48 hours, or 72 hours). The Holter is a small recording device that records the heart beat while being worn by the patient.
The physician or technician places electrodes and wires on the patient. The wires are connected to the Holter or digital recorder.
{4}------------------------------------------------
The DR300 Recorder has two modes that allow it to be used either as a standard Holter monitor or a looping Event recorder. The device is designed to facilitate ambulatory cardiac monitoring on the order of a physician, of those patients (including infants weighing less than 10 kg) who may benefit from such monitoring.
The data obtained during monitoring is not analyzed at the time of recording. After the recording is complete, the data must later be downloaded to a compatible NorthEast Monitoring Holter or Event analysis system (TelePro Software) to be analyzed. The Holter Analysis Software was cleared by FDA under K930564.
The device is not intended to replace real time telemetry monitoring for patient suspected of having life threatening arrhythmias.
The DR300 Digital Recorder package includes:
- . DR300 Digital Recorder
- Operation Manual
- SD Card ●
- Patient Cable
The DR300 digital recorder is powered by one 1.5 volt AA alkaline battery (MN1500 or the equivalent), one AA rechargeable NiMH (nickel metal hydride) battery, or one AA Eveready Lithium L91 battery. Batteries should not be re-used for a second patient. The batteries are not included; users are instructed to purchase 2 AA batteries.
The device is compatible with standard silver / silver chloride ECG electrodes. Electrodes are not provided with the subject device. The user is instructed to purchase standard silver / silver chloride ECG electrodes.
The DR300 digital recorder uses NorthEast Monitoring NorthEast Monitoring shielded patient cables with either 5 or 7 patient leads for a 3 channel recording or three leads for 2 channel Holter Recording.
The DR300 has an LCD that is used to display the time of day (during the recording), prompts and error messages (during the hookup procedure or during recording), or lead quality (during the hookup procedure).
The data collected by the DR300 digital recorder is stored on a removable SD Card.
The DR300 is packaged in a plastic bag in a cardboard shipping carton. The shipping carton will also include a patient cable and an SD Card.
{5}------------------------------------------------
The difference between the DR300 and the DR200/Telaheart is the wireless capability. The DR300 digital recorder is equipped with wireless Bluetooth transmitter. A NorthEast Monitoring Bluetooth Gateway or USB Dongle is able to receive the encrypted Holter and Event data.
In order for the wireless transmission to occur, the DR300 will need to come in range of a North East Monitoring transceiver - either a gateway or a PC with a USB Adapter. If there is sufficient data to send, the recorder will attempt to locate a transceiver every 20 minutes. When the transceiver is found the data will be sent. Transmission may be started manually by pressing the enter button.
### 5. INTENDED USE
The DR300 Digital Recorder can be used in Holter mode and Event Recorder mode.
### Holter Mode
Detection of Arrhythmias, Efficacy of Pharmacological Treatment, and Pacemaker Evaluation.
### Event Recorder
The DR300 event recorder module is a patient activated device designed to record and for diagnostic evaluation of transient symptoms (such as dizziness, palpitations, syncope, and chest pain). Once data is recorded, the data is transmitted wirelessly for evaluation.
#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The DR300 substantially equivalent to the Telaheart / DR 200 cleared under 510(k) K061293. The DR300 is equipped with a wireless Bluetooth transmitter and has additional shielding to be immune to radiofrequency interference and in order to pass FCC and EMC test requirements.
When the wireless function is enabled by the physician/technician, once every preset interval (normally set at 20 minutes) the device will check for the availability of any new data since the last successful transmission. If there is any, the Bluetooth module will be enabled and an attempt made to establish a link. If a link is established, then all available new ECG data will be transmitted. After each "block" of data is transmitted (normally 16 K bytes or less) a verification will be made for a successful transmission by the detection of a reply from the receiving end. The data will be re-transmitted if needed.
{6}------------------------------------------------
### 7. PERFORMANCE TESTING
The subject device conforms to the following standards:
- . IEC 60601-1-6: 2010 Medical Electrical Equipment Part 1-6: General Requirements for Safety - Collateral Standard: Usability including IEC 62366: Application of usability engineering to medical devices.
- IEC 60601-1-11: 2010 Medical Electrical Equipment Part 1-11: General . Requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 60601-1: 2012 Medical Electrical Equipment Part 1: General Requirements . for Safety
- IEC 60601-1-2-47 Medical Electrical Equipment Part 2 47: Particular ● Requirements for the Safety including Essential Performance of Ambulatory Electrocardiographic Systems.
- IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements . for Basic Safety and Essential Performance Collateral Standard: Electromagnetic Compatibility.
- FDA Guidance Radio Frequency Wireless in Medical Devices (August 14, 2013) ●
## 8. SAFETY AND EFFICACY
The subject device was subjected to testing including:
- · IEC 60601-1-6: 2010 Medical Electrical Equipment Part 1-6: General Requirements for Safety - Collateral Standard: Usability including IEC 62366: Application of usability engineering to medical devices.
· IEC 60601-1-11: 2010 Medical Electrical Equipment Part 1-11: General Requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
· IEC 60601-1: 2012 Medical Electrical Equipment Part 1: General Requirements for Safety
· IEC 60601-1-2-47 Medical Electrical Equipment Part 2 – 47: Particular Requirements for the Safety including Essential Performance of Ambulatory Electrocardiographic Systems.
· IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance Collateral Standard: Electromagnetic Compatibility.
- Software Verification and Validation ●
- Simulated bench and wireless testing (see summary below)
{7}------------------------------------------------
The testing included 11 Holter Tests for the purposes of verifying that the DR300 functions as intended. Testing was conducted related to the functionality of the Holter and Event functions of the DR300 without the wireless function being utilized. These tests verified that the device functions as intended after the inclusion of the Bluetooth transmitter and additional shielding. The testing ranged from monitoring over a 24 hour to 14 day period. Additional testing was performed to evaluate the wireless Bluetooth technology and communication of encrypted data. The wireless testing ranged from monitoring over a 24 hour to 48 hour period.
All testing confirmed that the device using the wireless technology functioned as intended and found no issues of safety or effectiveness.
## 9. CONCLUSION
North East Monitoring believes that based on the indications for use, technological characteristics, and comparison to predicate devices the DR300 has been shown to be substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.