K142389 · Ivoclar Vivadent, AG · EBF · Nov 25, 2014 · Dental
Device Facts
Record ID
K142389
Device Name
Variolink Esthetic
Applicant
Ivoclar Vivadent, AG
Product Code
EBF · Dental
Decision Date
Nov 25, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Variolink Esthetic LC: - Permanent adhesive luting of glass-ceramic, lithium disilicate glass-ceramic and composite restorations (inlays, onlays and veneers) - Only use Variolink Esthetic LC for restorations with a low thickness of <2mm that have sufficient translucency (e.g. restorations made of IPS e.max R HT). Variolink Esthetic DC: - Adhesive luting of glass-ceramic, lithium disilicate glass-ceramic and composite restorations (inlays, onlays, partial crowns, crowns, bridges). - Restorations made of opaque ceramics, e.g. oxide ceramics, can only be permanently cemented if an adhesive is additionally used that is separately light-cured. Variolink Esthetic Trv-In To evaluate the overall effect of the restoration in conjunction with the various Variolink Esthetic shades prior to permanent cementation.
Device Story
Variolink Esthetic is a color-stable adhesive luting system for permanent cementation of ceramic and composite resin restorations. Available in light-curing (LC) and dual-curing (DC) versions; includes water-soluble glycerine Try-In pastes for shade visualization prior to cementation. Used by dental professionals in clinical settings. LC version dispensed via syringe; DC version dispensed via double-push syringes with mixing tips for ready-mixed application. System provides five shade gradations (Light+, Light, Neutral, Warm, Warm+). Enables clinicians to achieve aesthetic, durable bonding of restorations; Try-In pastes allow pre-cementation shade verification to ensure patient satisfaction with final appearance.
Clinical Evidence
Bench testing only. Device tested per ISO 4049:2009 for film thickness, depth of cure, ambient light sensitivity, working time, flexural strength, water sorption, solubility, and radiopacity. Biocompatibility assessed per EN ISO 10993-1:2009 and EN ISO 7405:2008.
Technological Characteristics
Polymer-based resin luting material. Available in LC (syringe) and DC (double-push syringe with mixing tips) formats. Tested per ISO 4049:2009. Biocompatibility per EN ISO 10993-1 and EN ISO 7405. Storage 2-28 °C.
Indications for Use
Indicated for permanent adhesive luting of glass-ceramic, lithium disilicate glass-ceramic, and composite restorations (inlays, onlays, veneers, partial crowns, crowns, bridges). LC version restricted to translucent restorations <2mm thick. DC version requires additional light-cured adhesive for opaque/oxide ceramics.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other. The faces are stylized and appear to be moving forward.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 25,2014
Ivoclar Vivadent, Inc. Donna Marie Hartnett Director QA/Regulatory Affairs 175 Pineview Drive Amherst, New York 14228
Re: K142389
Trade/Device Name: Variolink® Esthetic Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: II Product Code: EBF Dated: August 8. 2014 Received: August 27, 2014
#### Dear Ms. Hartnett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): K142389
Device Name: Variolink® Esthetic
Indications For Use:
### Variolink Esthetic LC:
– Permanent adhesive luting of glass-ceramic, lithium disilicate glass-ceramic and composite restorations (inlays, onlays and veneers)
- Only use Variolink Esthetic LC for restorations with a low thickness of <2mm that have sufficient translucency (e.g. restorations made of IPS e.max R HT).
#### Variolink Esthetic DC:
- Adhesive luting of glass-ceramic, lithium disilicate glass-ceramic and composite restorations (inlays, onlays, partial crowns, crowns, bridges).
- Restorations made of opaque ceramics, e.g. oxide ceramics, can only be permanently cemented if an adhesive is additionally used that is separately light-cured.
#### Variolink Esthetic Trv-In
To evaluate the overall effect of the restoration in conjunction with the various Variolink Esthetic shades prior to permanent cementation.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1
{3}------------------------------------------------
## 510(K) SUMMARY Variolink Esthetic
Image /page/3/Picture/1 description: The image shows the logo for Ivoclar Vivadent. The logo consists of the company name in blue, with the words "Ivoclar" stacked on top of "Vivadent". Above the company name is a semi-circle of green and blue squares. Below the company name is the company slogan, "passion vision innovation" in black.
Contact: Donna Marie Hartnett
Company: Ivoclar Vivadent. AG Bendererstrasse 2, Schaan, FL-9494, Liechtenstein +423-235-3535
November 25, 2014 Date Prepared:
Variolink® Esthetic Proprietary Name:
Classification Name: Material, Tooth Shade, Resin (872.3690) (Classification Code EBF)
Predicate Device: Variolink II (K971372)
- Device Description: Variolink® Esthetic is a color-stable adhesive luting system for the permanent cementation of ceramic and composite resin restorations. Variolin Esthet6ic is offered in a purely light-curing (LC) and dual-curing (DC) version. Try-In pastes are also included for shade matching purposes. The materials are offered in five shade gradations (Light+, Light, Neutral, Warm and Warm+)
#### Intended Use: Variolink Esthetic LC:
- Permanent adhesive luting of glass-ceramic, lithium disilicate glass-ceramic and composite restorations (inlays, onlays and veneers)
- Only use Variolink Esthetic LC for restorations with a low thickness of <2mm that have sufficient translucency (e.g. restorations made of IPS e.max R HT).
#### Variolink Esthetic DC:
- Adhesive luting of glass-ceramic, lithium disilicate glass-ceramic and composite restorations (inlays, onlays,partial crowns, crowns, bridges).
- Restorations made of opaque ceramics, e.q. oxide ceramics, can only be permanently cemented if an adhesive is additionally used that is separately light-cured.
#### Variolink Esthetic Try-In
To evaluate the overall effect of the restoration in conjunction with the various Variolink Esthetic shades prior to permanent cementation.
{4}------------------------------------------------
### 510(K) SUMMARY Variolink Esthetic
Image /page/4/Picture/1 description: The image is the logo for Ivoclar Vivadent. The words "ivoclar vivadent" are in blue, with "ivoclar" on top and "vivadent" on the bottom. Above the words is a semi-circle of green and blue dots. The dots are larger in the center and smaller on the ends.
passion vision innovation
Comparison to Predicate: The predicate device to which Variolink® Esthetic has been compared is Variolink II (K971372). For this application, Variolink® Esthetic has been compared to its predicate with regard to chemical composition, physical properties, and indications for use. The comparison shows that Variolink® Esthetic is substantially equivalent to the predicate device.
| Predicate - Variolink II K971372 | Subject Device |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Variolink II is a dual-curing (light- and self-curing)<br>luting composite system for the adhesive luting of<br>ceramic and composite restorations. | Variolink Esthetic is a color-stable, adhesive luting<br>system for the permanent cementation of ceramic<br>and composite resin restorations. |
| Variolink II may also be applied in the light-curing<br>technique only (e.g. luting of veneers). For this<br>purpose, only the Variolink II Base is to be used. | Variolink Esthetic is offered in a light-curing version<br>(Variolink Esthetic LC) and a dual-curing version<br>(Variolink Esthetic DC). |
| Variolink II Try-in Pastes. Water-soluble glycerine<br>pastes shades to match the cement and used<br>temporarily before cementation to visualize the final<br>shade. | Variolink Esthetic also includes an accessory<br>product, Variolink Esthetic Try-in which are water-<br>soluble glycerine pastes shaded to match the<br>Variolink Esthetic cement and used temporarily<br>before cementation to visualize the final shade. |
| Variolink II is available in 6 base shades and 2<br>catalyst shades. The catalyst is available in different<br>viscosities. Variolink II is available in syringes. Base<br>and catalyst are hand-mixed. | Variolink Esthetic is available in 5 shades:<br>Light +, Light, Neutral, Warm, Warm +<br>Variolink Esthetic DC is available in one consistency.<br>Variolink Esthetic LC is available in syringes.<br>Variolink Esthetic DC is available in double-push<br>syringes with mixing tips so the material is dispensed<br>ready-mixed. |
| Cool storage (2-8 °C/36-46 °F) is indicated for the<br>catalyst paste. Syringes should be closed<br>immediately after use. Exposure to light causes<br>premature polymerization. | 2-28 °C/36-82 °F for all items. Close Variolink<br>Eshetic LC syringes immediately after use. Exposure<br>to light causes premature polymerization. |
Technological Characteristics and Testing Summary: The device has been tested and designed in accordance with ISO 4049:2009 Dentistry - Polymer-based restorative materials for film thickness, Depth of Cure, sensitivity to ambient light, working time, flexural strength, water sorption, solubility, and radiopacity. Biocompatibility testing has been assessed according EN ISO 10993-1:2009 and EN ISO 7405:2008. The device design, i.e. delivery form, and intended use of Variolink® Esthetic and the predicate device are the same.
{5}------------------------------------------------
## 510(K) SUMMARY Variolink Esthetic
Image /page/5/Picture/1 description: The image shows the logo for Ivoclar Vivadent. The logo is blue and green, with the word "Ivoclar" in a large, bold font. Above the word "Ivoclar" are a series of green and blue squares. Below the word "Ivoclar" is the word "Vivadent" in a smaller, less bold font. The logo is simple and modern, and it is likely used to represent a company that specializes in dental products.
Guidance Document: This submission complies with the requirement of "Guid Industry and FDA Staff: Dental Composite Resin Devices - Premarket Notification innovation Submissions" dated October 26, 2005.
CONCLUSION: The above data and analysis demonstrates that Variolink® Esthetic is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.