Sensititre HP MIC Susceptibility Plate with Tedizolid (0.002-4 mcg/ml)
K142338 · Thermo Fisher Scientific · JWY · Mar 2, 2015 · Microbiology
Device Facts
| Record ID | K142338 |
| Device Name | Sensititre HP MIC Susceptibility Plate with Tedizolid (0.002-4 mcg/ml) |
| Applicant | Thermo Fisher Scientific |
| Product Code | JWY · Microbiology |
| Decision Date | Mar 2, 2015 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.1640 |
| Device Class | Class 2 |
Intended Use
The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae and Streptococcus species.
Device Story
Sensititre HP MIC Susceptibility plate is an in vitro diagnostic device used in clinical laboratories for antimicrobial susceptibility testing. It utilizes microbroth dilution to determine the Minimum Inhibitory Concentration (MIC) of Tedizolid against specific Streptococcus species. The device consists of multi-well plates containing pre-dispensed, dried antimicrobial agents. Clinical isolates are inoculated into the wells; after incubation, growth is assessed to determine the MIC. Results assist clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections. The device is intended for professional use in clinical settings.
Clinical Evidence
Performance evaluated via method comparison study at three sites using 225 clinical and 60 challenge Streptococcus isolates (S. pyogenes, S. agalactiae, S. anginosus). Overall EA was 100% (Manual) and 98.9% (Automated). CA was 100% (Manual) and 99.6% (Automated). Reproducibility >95% for both methods. No clinical data (patient outcomes) provided; performance based on bench-top comparison to CLSI reference broth microdilution.
Technological Characteristics
Multi-well plastic microtitre plates containing dried antimicrobial agents. Sensing principle: fluorescence detection of bacterial surface enzyme activity (cleavage of fluorogenic substrate). Energy source: electrical (for Autoinoculator/AIM, OptiRead, Vizion systems). Connectivity: OptiRead/ARIS systems. Sterilization: not specified. Software: automated reading algorithm for MIC determination.
Indications for Use
Indicated for clinical susceptibility testing of fastidious isolates, specifically Streptococcus pyogenes, Streptococcus agalactiae, and Streptococcus anginosus. Prescription use only.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Related Devices
- K062681 — SENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE (HP) MIC PLATE · Trek Diagnostic Systems, Ltd. · Oct 13, 2006
- K041534 — SENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE (HP) MIC SUSCEPTIBILITY PLATE AND SUSCEPTIBILITY TEST PANEL · Trek Diagnostic Systems, Ltd. · Jul 27, 2004
- K062722 — SENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE (HP) MIC PLATE · Trek Diagnostic Systems, Inc. · Oct 20, 2006
- K143526 — The Sensititre HP MIC Susceptibility plate with Oritavancin (0.0005-8mcg/ml) · Thermo Fisher Scientific · Mar 5, 2015
- K142151 — SENSITITRE 18-24 HOUR SUSCEPTIBILITY PLATES · Thermo Fisher Scientific · Dec 23, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 27, 2015
THERMO FISHER SCIENTIFIC CYNTHIA KNAPP DIRECTOR OF CLINICAL OPERATIONS 1 THERMO FISHER WAY OAKWOOD VILLAGE OH 44014-6
Re: K142338
Trade/Device Name: Sensititre Haemophilus/Streptococcus pneumoniae (HP) MIC Plates with Tedizolid in the dilution range of 0.002-4 ug/ml Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: JWY. LRG, LTT Dated: February 11, 2015 Received: February 12, 2015
Dear Ms. Knapp:
This letter corrects our substantially equivalent letter of March 2, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Ribhi Shawar -S
For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142338
#### Device Name
Sensititre Haemophilus / Streptococcus pneumoniae (HP) MIC plates with Tedizolid in the dilution range of 0.002-4ug/ml
Indications for Use (Describe)
The Sensititre HP MIC Susceptibility plate is an in vitro diagnostic product for clinical susceptibility testing of fastidious isolates.
This 510(k) is for the newly approved Tedizolid for the dilution range of 0.002-4ug/ml to the Sensititre HP MIC Susceptibility plate for testing Streptococcus spp.
The approved primary "Indications for Use" and clinical significance for Streptococcus spp. is for the following species: Streptococcus pyogenes Streptococcus agalactiae Streptococcus anginosus
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."