The Veenstra Instruments VDC-506 Dose Calibrator is designed to measure the amount of radioactive material in vials, syringes and capsules. It is used to measure the amount of activity used to prepare radiopharmaceutical kits, measure the activity in syringes of radiopharmaceuticals prior to injection and to quantify the activity remaining in the syringe following injection. It is indicated for use in the preparation of radiopharmaceuticals and verification of the activity prior to patient administration.
Device Story
VDC-506 Dose Calibrator measures radioactive material in vials, syringes, and capsules; utilizes VIK-202 ionization chamber (argon-filled, pressurized well chamber) to detect radiation; PC-based software (Windows XP or higher) acts as control and display unit; connected via RS-232 cable. Used by clinicians to prepare radiopharmaceutical kits, verify activity before injection, and quantify residual activity post-injection. Provides quantitative activity readouts; assists in ensuring accurate dosing for patient safety. System includes QA protocols for daily/quarterly checks and Mo-99 breakthrough protocol.
Clinical Evidence
Bench testing only. No clinical data. Performance verified against IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 61145 (calibration/usage), IEC 61303 (radionuclide calibrator performance), and IEC 62366 (usability). Software verified and validated per IEC 62304 and FDA guidance.
Technological Characteristics
Argon-filled, pressurized well ionization chamber (VIK-202); 150V lithium battery ionization voltage; PC-based control/display via RS-232; Windows XP or higher OS; measurement range 0.001 MBq–200 GBq; accuracy ±3%; linearity ±1%. Standards: IEC 60601-1, IEC 60601-1-2, IEC 61145, IEC 61303, IEC 62304, IEC 62366.
Indications for Use
Indicated for use in the preparation of radiopharmaceuticals and verification of radioactive activity prior to patient administration in clinical settings.
Regulatory Classification
Identification
A radionuclide dose calibrator is a radiation detection device intended to assay radionuclides before their administration to patients.
Predicate Devices
Nuclear Associates Cal/Rad Mark VI/VDC 505 Dose Calibrator, Model 34-165 (K030066)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized emblem with three abstract human profiles facing right, stacked on top of each other. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 18, 2014
Comecer Netherlands b.v. % Mr. Thomas Kroenke Principal Correspondent Speed To Market, Inc. P.O. Box 3018 NEDERLAND CO 80466
Re: K142325
Trade/Device Name: Veenstra Instruments Regulation Number: 21 CFR 892.1360 Regulation Name: Radionuclide dose calibrator Regulatory Class: II Product Code: KPT Dated: August 22, 2014 Received: August 25, 2014
Dear Mr. Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K142325
Device Name
Veenstra Instruments VDC-506 Dose Calibrator
#### Indications for Use (Describe)
The Veenstra Instruments VDC-506 Dose Calibrator is designed to measure the amount of radioactive material in vials, syringes and capsules. It is used to measure the amount of activity used to prepare radiopharmaceutical kits, measure the activity in syringes of radiopharmaceuticals prior to injection and to quantify the activity remaining in the syringe following injection. It is indicated for use in the preparation of radiopharmaceuticals and verification of the activity prior to patient administration.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submission Date: | 14 August 2014 | | |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|-----------------------------------------------------------------------------------|
| Submitter: | Comecer Netherlands b.v.<br>Madam Curieweg 1<br>Joure, Fryslan 8501 XC<br>The Netherlands | | |
| Submitter Contact: | Mr. Gert Jan Klok<br>Phone: +011 031-513-416964<br>Fax: +011 031-513-416919<br>Email: gklok@comecer.com | | |
| Application<br>Correspondent: | Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 USA<br>Mr. Thomas Kroenke<br>Phone: +1 (303) 956-4232<br>Fax: +1 (303) 642-3141<br>Email: tkroenke@speedtomarket.net | | |
| Manufacturing Site: | Comecer Netherlands b.v.<br>Madam Curieweg 1<br>Joure, Fryslan 8501 XC<br>The Netherlands | | |
| Trade Name: | Veenstra Instruments VDC-506 Dose Calibrator | | |
| Common Name: | Radionuclide dose calibrator | | |
| Classification Name: | Radionuclide dose calibrator | | |
| Classification<br>Regulation: | 21 CFR §892.1360 | | |
| Product Code: | KPT | | |
| Substantially<br>Equivalent Devices: | New Model | Predicate<br>510(k) Number | Predicate<br>Manufacturer / Model |
| | Veenstra Instruments<br>VDC-506 Dose<br>Calibrator | K030066 | Nuclear Associates<br>Cal/Rad Mark VI/VDC<br>505 Dose Calibrator,<br>Model 34-165 |
{4}------------------------------------------------
| Device Description: | The Veenstra Instruments VDC-506 Dose Calibrator (VDC-506) is a<br>software application installed on a personal computer (PC) running<br>Windows XP, and is used in conjunction with the Veenstra Instruments<br>VIK-202 ionization chamber (VIK-202) to measure the radioactive<br>strength of radioactive material in vials, syringes and ampoules placed<br>inside the ionization chamber.<br><br>The VDC-506 software, in conjunction with the PC, serves as the<br>control mechanism and display for the VIK-202 ionization chamber.<br>The PC on which the VDC-506 software is installed is connected to the<br>VIK-202 ionization chamber using a RS-232 cable.<br><br>It is recommended that the PC includes a touchscreen display for<br>convenient navigation within the software; however, a touchscreen is<br>not necessary. |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | The Veenstra Instruments VDC-506 Dose Calibrator is designed to<br>measure the amount of radioactive material in vials, syringes and<br>capsules. It is used to measure the amount of activity used to prepare<br>radiopharmaceutical kits, measure the activity in syringes of<br>radiopharmaceuticals prior to injection and to quantify the activity<br>remaining in the syringe following injection. It is indicated for use in<br>the preparation of radiopharmaceuticals and verification of the activity<br>prior to patient administration. |
{5}------------------------------------------------
### Technology Comparison:
The VDC-506 employs the same technological characteristics as the predicate device.
| Characteristic | Predicate Device | Proposed Device |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Controller/<br>Display Unit | Control Unit which contains a<br>single board computer. | Personal computer (PC) |
| Controller/<br>Display Unit<br>Operating System | Microsoft® Windows XP<br>Embedded | Microsoft® Windows XP for<br>PC or higher |
| VDC-506<br>Readout | 0.001 MBq – 200 GBq<br>0.01 µCi – 6000 mCi | Same |
| VDC-506<br>Software<br>Features | Future Dose Calculator<br>QA Protocols for Daily and<br>Quarterly Checks<br>Mo-99 Breakthrough Protocol | Same |
| Ionization<br>Chamber Model | Veenstra Instruments VIK-202 | Same |
| Chamber Type | Argon-filled, pressurized, well<br>chamber | Same |
| Ionization<br>Voltage | 150 V lithium battery | Same |
| Overall Accuracy | ± 3 % dependant of specific<br>calibration source and<br>geometric variations | Same |
| Electrometer<br>Accuracy | ±1% | Same |
| Temperature<br>Coefficient | 0.1 % / ° C between 10° C and<br>40° C at 5 MBq and up | Same |
| Reproducibility | ± 1% over 24 hours, stable<br>conditions | Same |
| Reproducibility | ± 1% over 24 hours, stable<br>conditions | Same |
| Geometry of the<br>Sensitive Volume<br>(Well Size) | Height: 250 mm (~9.84 in)<br>Diameter: 60 mm (~2.36 in) | Same |
| Linearity | ± 1 % between 1 MBq and<br>200 GBq (Tc-99m) | Same |
{6}------------------------------------------------
### Summary of Performance Testing:
| Electrical Safety | The VIK-202 ionization chamber was tested for performance in accordance with the following Standards:<br>• IEC 60601-1: 2005, Medical electrical equipment – Part 1. General requirements for basic safety and essential performance |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Test results indicated that the VIK-202 complies with the Standards. |
| Electromagnetic<br>Compatibility (EMC)<br>Testing | The VIK-202 ionization chamber was tested for performance in accordance with the following Standard:<br>• IEC 60601-1-2: 2007, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests. |
| | Test results indicated that the VIK-202 complies with the Standards. |
| Software Testing | Software device modifications made to the VDC-506 system were designed and developed according to a robust software development process, and were rigorously verified and validated.<br>Software information is provided in accordance with internal documentation and the following Standards and guidance documents:<br>• FDA guidance: The content of premarket submissions for software contained in medical devices, 11 May 05;<br>• FDA guidance: Off-the-shelf software use in medical devices, 09 Sep 99; and<br>• IEC 62304: 2006, Medical device software – Software life cycle processes. |
Test results indicate that the VDC-506 complies with its predetermined
specification and the Standards and guidance documents.
{7}------------------------------------------------
| <b>Performance Testing</b> | The VDC-506 system and VIK-202 ionization chamber were verified<br>for performance in accordance with internal documentation and the<br>following Standards: |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
- o IEC 61145: 1992, Calibration and usage of ionization chamber systems for assay of radionuclides;
- o IEC 61303: 1994, Medical electrical equipment – Radionuclide calibrators – Particular methods for describing performance; and
- IEC 62366: 2007, Medical devices Application of usability o engineering to medical devices.
Test results indicated that the VDC-506 system and VIK-202 comply with predetermined specification and with the applicable Standards.
### Conclusion Verification and validation activities were conducted to establish the performance and safety characteristics of the software device modifications made to the VDC-506 system. The results of these activities demonstrate that the VDC-506 system is safe and effective when used in accordance with its intended use and labeling.
Therefore, the VDC-506 system is considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.