HUMMINGBIRD(TM) TYMPANOSTOMY TUBE SYSTEM (TTS)

K142282 · Preceptis Medical · ETD · Apr 23, 2015 · Ear, Nose, Throat

Device Facts

Record IDK142282
Device NameHUMMINGBIRD(TM) TYMPANOSTOMY TUBE SYSTEM (TTS)
ApplicantPreceptis Medical
Product CodeETD · Ear, Nose, Throat
Decision DateApr 23, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3880
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The Hummingbird™ Tympanostomy Tube System is intended to deliver a tympanostomy tube in tympanostomy procedures in which the patient is receiving a tympanostomy tube.

Device Story

Hummingbird™ Tympanostomy Tube System (TTS) is a single-use, sterile manual surgical instrument for myringotomy and ventilation tube placement. Device consists of a handle and removable tip assemblies; each tip contains a positioning rod and a pre-loaded ventilation tube within a sharpened sheath. Operated by a surgeon under direct visualization; user manually advances the sheath to create a myringotomy and position the tube; manual actuator on the handle retracts the sheath to release the tube into the tympanic membrane. Designed for bilateral procedures via interchangeable tip assemblies. Used in clinical settings under moderate sedation with local anesthetic. Simplifies procedure by combining myringotomy, tube placement, and suctioning into one device; reduces need for general anesthesia; benefits patient by minimizing anesthesia-related risks.

Clinical Evidence

Clinical performance assessed in two studies (feasibility and multi-site) involving 69 children (136 ears, mean age 2.4 years) undergoing tympanostomy under moderate sedation. 100% of patients received ventilation tubes as planned. No intra-operative or unanticipated adverse events reported. Conversion rates from moderate sedation to general anesthesia were 30% in the first study and 7% in the second, attributed to increased surgical experience and design improvements.

Technological Characteristics

Manual surgical instrument; includes handle and removable tip assemblies with sharpened sheath, positioning rod, and pre-loaded ventilation tube. Combines myringotomy, tube placement, and suctioning functions. Single-use, sterile. No electronic components or software.

Indications for Use

Indicated for patients undergoing tympanostomy procedures requiring the placement of a ventilation tube.

Regulatory Classification

Identification

A tympanostomy tube is a device that is intended to be implanted for ventilation or drainage of the middle ear. The device is inserted through the tympanic membrane to permit a free exchange of air between the outer ear and middle ear. A type of tympanostomy tube known as the malleous clip tube attaches to the malleous to provide middle ear ventilation. The device is made of materials such as polytetrafluoroethylene, polyethylene, silicon elastomer, or porous polyethylene.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 23, 2015 Preceptis Medical, Inc. Mr. Keith Leland Vice President of Research and Development 505 Highway 169 North, #365 Plymouth, MN 55441 Re: K142282 Trade/Device Name: Hummingbird™ Tympanostomy Tube System (TTS) Regulation Number: 21 CFR 874.3880 Regulation Name: Tympanostomy Tube Regulatory Class: Class II Product Code: ETD Dated: November 13, 2014 Received: November 14, 2014 Dear Mr. Leland: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ Page 2 - Mr. Keith Leland forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. ## William H. Maisel -S William H. Maisel, MD, MPH Director, Office of Device Evaluation (Acting) Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K142282 Device Name Hummingbird™ Tympanostomy Tube System Indications for Use (Describe) The Hummingbird™ Tympanostomy Tube System is intended to deliver a tympanostomy procedures in which the patient is receiving a tympanostomy tube. Type of Use (Select one or both, as applicable) | | Prescription Use (Part 21 CFR 801 Subpart D) | |--|----------------------------------------------| | | Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | Submitter<br>Information: | Preceptis Medical, Inc.<br>505 Highway 169 North, #365<br>Plymouth, MN 55441<br>763.568.7819 | |------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Keith Leland, VP of R&D | | Date<br>Prepared: | 23 April 2015 | | Trade Name | Hummingbird™ Tympanostomy Tube System (TTS) | | Product Code | ETD (21 CFR Part 874.3880) | | Common<br>Name | Tympanostomy Tube Inserter with pre-loaded ventilation tube | | Predicate<br>Devices | Preceptis Tympanostomy Tube System, 510(k) K133921 | | Device<br>Description | The Hummingbird™ Tympanostomy Tube System (TTS) which includes<br>a tympanostomy tube inserter (TTI) with a preloaded ventilation tube, is a<br>single-use, sterile manual surgical instrument which is used to create a<br>myringotomy in the tympanic membrane and place a ventilation tube.<br>The TTS includes a handle with one or more tip assemblies which contain<br>a sterile tympanostomy tube.<br>Each tip assembly can be removably attached to the handle and includes<br>a positioning rod and a ventilation tube pre-loaded inside the distal end<br>of a sharpened sheath. Attaching the tip assembly to the handle also<br>connects the sheath and actuator, allowing the user to retract the sheath<br>by manually scrolling an actuator located on the handle.<br>The user manually advances the sharpened sheath to create a<br>myringotomy and simultaneously positions the ventilation tube within<br>the myringotomy, always under direct visualization. The user then<br>manually retracts the sharpened sheath away from the myringotomy<br>using the manual actuator located on the handle. The retraction of the<br>sheath releases the tube within the myringotomy.<br>A first tip assembly can then be removed from the handle and<br>replaced with a second preloaded tip assembly for bilateral<br> | | Indications<br>For Use | The Hummingbird™ Tympanostomy Tube System is intended to deliver<br>a tympanostomy tube in tympanostomy procedures in which the<br>patient is receiving a tympanostomy tube. | | Technical<br>Characteristics | The TTS is intended to deliver a tympanostomy tube (also referred to as<br>a ventilation tube) through the tympanic membrane (TM) of the patient.<br>It combines the separate functions of creating a myringotomy,<br>positioning and placing a ventilation tube across the TM, and suctioning.<br>A comparison between the TTS and predicate device shows that the<br>devices are identical. | | Performance<br>Data | Clinical performance data was submitted to assess the performance of the<br>device when placed under moderate sedation with local anesthetic. In two<br>studies (an initial feasibility study followed by a multi-site study), a total of<br>69 children (136 ears) underwent tympanostomy procedures under<br>moderate sedation using the Preceptis TTS. The mean age of the patients<br>was 2.4 years (range of 8 months to 8.9 years). Results: | | | 100% of the children received ventilation tubes as planned. There were no intra-operative adverse events, no unanticipated adverse events, and adverse event rates were well within peer-reviewed literature reported rates. Moderate sedation was per definition from the ASA guidelines, Continuum of Depth of Sedation: Definition of General Anesthesia and Levels of Sedation/Analgesia. The moderate sedation regimen was determined by the surgeon and anesthesiologist. In these two studies, the surgeon and anesthesiologist chose to use nitrous oxide (50-70%) in all cases. In the 1st study, there were 12 conversions (30%) from moderate sedation to general anesthesia due to surgical challenges. In the 2nd study, 2 cases (7%) were converted. The reduction in the conversion rate was likely due to increased surgical experience and design improvements to the TTS to improve visibility and repeatability. In most cases during the study, the surgeon chose to use a local anesthetic on the tympanic membrane to reduce discomfort for the patient. Under moderate sedation, local anesthesia of the tympanic membrane should be used to reduce the risk of head movement and increase patient comfort. | {4}------------------------------------------------ {5}------------------------------------------------
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%