The Neo Cosmo (Polymacon) Soft (hydrophilic) Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic and non-aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or colored and may be used to enhance or alter the apparent color of the eye. Eye care practitioners may prescribe the lens frequent replacement wear with cleaning, disinfecting and schedule replacement. The lens may be disinfected using a chemical (not heat) lens care system only.
Device Story
Neo Cosmo is a hemispherical, flexible, hydrophilic contact lens made of polymacon (HEMA cross-linked with ethyleneglycol dimethacrylate). It covers the cornea and adjacent sclera to correct visual acuity in patients with myopia or hyperopia. The lens is available in clear or colored versions, utilizing FDA-approved color additives (e.g., titanium dioxide, iron oxides). It is prescribed by eye care practitioners for daily wear with a frequent replacement schedule. Patients must use chemical (non-heat) disinfection systems. The device functions as a refractive optical element; it does not involve electronic processing or clinical decision-making software. It provides vision correction and cosmetic enhancement.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including tensile testing, solution compatibility, physical property evaluation, and biocompatibility testing (ocular irritation, skin sensitization, and cytotoxicity).
Technological Characteristics
Material: Polymacon (HEMA 99.3%, ethyleneglycol dimethacrylate 0.7%). Water content: 38%. Refractive index: 1.428. Oxygen permeability: 9.77 x 10^-11 (cm2/s)(ml O2/ml x mmHg) at 35°C. Manufacturing: Lathe-cut (semi-molded). Sterilization: Steam validated autoclave. Packaging: Blister or vial with NaCl solution.
Indications for Use
Indicated for daily wear for correction of visual acuity in aphakic and non-aphakic persons with non-diseased eyes with myopia or hyperopia; includes patients with refractive astigmatism of 0.50 diopters or less.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol representing the human form, composed of three overlapping profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 7. 2014
Neo Vision Co., Ltd. % Albert Rego Ph.D. Regulatory Consultant Albert Rego, Ph.D., Inc. 27001 La Paz Road, Suite 312 Mission Viejo, CA 92691
Re: K142275
Trade/Device Name: Neo Cosmo (polymacon) Soft (hydrophilic) Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL Dated: July 31, 2014 Received: August 15, 2014
Dear Dr. Rego:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
Page 2 -
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation
(21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142275
#### Device Name
Neo Cosmo (polymacon) Soft (hydrophilic) Contact Lens
#### Indications for Use (Describe)
The Neo Cosmo (Polymacon) Soft (hydrophilic) Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic and non-aphakic persons with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or colored and may be used to enhance or alter the apparent color of the eye. Eye care practitioners may prescribe the lens frequent repacement wear with cleaning, disinfecting and schedule replacement. The lens may be disinfected using a chemical (not heat) lens care system only.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="unicode-bidi:embed;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="unicode-bidi:embed;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
(As required by 21 CFR 807)
| SUBMITTED BY | Neo Vision Co., Ltd.<br>17th Floor, U-Tower, 1029 Yeongdeok-<br>Dong, Giheung-gu, Yongin-si<br>Gyeonggi-do<br>Korea |
|---------------------|---------------------------------------------------------------------------------------------------------------------|
| CONTACT | Kyung Hwa Kim 82-31-889-2040<br>export@neolens.co.kr |
| SUBMISSION DATE | April 7, 2014 |
| TRADE NAME | Neo Cosmo (Polymacon) Soft (Hydrophilic<br>Contact Lens - Neo Cosmo |
| COMMON NAME | Lenses, Soft Contact, Daily Wear |
| CLASSIFICATION NAME | Class II (21 CFR 886.5925) |
| PRODUCT CODE | LPL |
| PREDICATE DEVICE | K051477 Migwang<br>Comfort 38<br>Decision Date: 03/10/2006 |
### DEVICE DESCRIPTION
Neo Vision Neo Cosmo (Polymacon) Soft (Hydrophilic) Contact Lens for Daily Wear is available as a single vision lens. The lens material, polymacon is hydrophilic polymer of 2-hydroximethyl methacrylate (HEMA) cross-linked with ethyleneglycol dimethacrylate(62%) and water(38%).
Neo Vision Neo Cosmo (Polymacon) Soft (Hydrophilic) Contact Lens for Daily Wear is colored with one or a combination of one or more of the following color additives: Titanium dioxide (white), C.I Pigment blue 36, Iron oxides, C.I. Pigment green 7, C.I. Pigment violet 23, D&C Yellow No.10.
Neo Vision Neo Cosmo is a hemispherical flexible shell which covers the cornea and a portion of the adjacent sclera with the following dimensions:
- Diameter: 13.8mm to 14.5mm
- Base Curve: 8.0mm to 9.0mm ●
- Center Thickness: 0.05mm to 0.135 (varies with power) ●
- Powers: 0.00D to -10.00D ●
{4}------------------------------------------------
The physical properties of the lens are:
- I Refractive Index: 1.43
- I Light Transmittance: >95% .
- I Water Contents: 38% .
- I Oxygen Permeability: 9.5 X 10-11(cm2/s) [ml X mmHg) at 35°C .
Neo Vision Neo Cosmo (Polymacon) Soft (Hydrophilic) Contact lens consists of:
| No. | Part<br>Name | Raw Material | Composition<br>(%) | Remark | Function |
|-----|-------------------|-----------------------------------|--------------------|------------------|------------------|
| 1 | Base | 2-hydroxyethylmethacrylate | 99.3 | CAS No. 868-77-9 | Main<br>Element |
| | | Ethylene glycol<br>dimethacrylate | 0.7 | CAS No. 97-90-5 | Cross-<br>linker |
| 2 | Color<br>Additive | Titanium dioxide (white) | $\leq$ 0.04 | CFR No. 73.3126 | Color |
| | | Iron oxides | | CFR No. 73.3125 | Color |
| | | C.I. Pigment violet -23 | | CFR No. 73.3107 | Color |
| | | D & C Yellow No. 10 | | CFR No. 73.3110a | Color |
| | | C.I. Pigment green -7 | | CFR No. 73.3124 | Color |
Neo Vision Neo Cosmo (Polymacon) Soft (Hydrophilic) Contact lens is packaged in either in a vial package or a blister package.
The vial package is a 5mL Borosilicate glass vial filled with NaCI 0.85 ~0.95 w/v% solution. The cap is aluminum with a locking rubber (Chlorobutyl rubber for medical use).
The blister package consists of 3mL Polypropylene filled with NaCI 0.85 ~0.95 w/v% solution with an aluminum sealing film sheet.
### INDICATIONS FOR USE
The Neo Cosmo (Polymacon) Soft (hydrophilic) Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic and non-aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or colored and may be used to enhance or alter the apparent color of the eye. Eye care practitioners may prescribe the lens frequent replacement wear with cleaning, disinfecting and schedule replacement. The lens may be disinfected using a chemical (not heat) lens care system only.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE
This device is substantially equivalent to the predicate devices in its intended use and technological characteristics, manufacturing process specified in following table;
{5}------------------------------------------------
| Substantial Equivalency | Soft (hydrophilic) contact lens | | |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Manufacture | MiGwang Contact Lens Co., Ltd. | Neo Vision | |
| Article name | Migwang Comfort 38 | Neo Cosmo | |
| Material<br>USAN name | Polymacon | Polymacon | |
| Type | Spherical, Toric | Spherical, Toric | |
| FDA Classification | Lenses, Soft Contact, Daily Wear<br>(Class II) | Lenses, Soft Contact, Daily<br>Wear (Class II) | |
| Water Content | 38±2% | 38±2% | |
| Light Transmittance | >90% | >90% | |
| Index of Refraction | 1.43 | 1.428 | |
| Oxygen permeability | 9.77 * 10-11 (cm²/sec)(ml O2/ml *<br>mm Hg @ 35℃) | 9.77 * 10-11 (cm²/sec)(ml<br>02/ml * mm Hq @ 35℃) | |
| Manufacturing Method | Lathe-cut<br>(semi molded) | Lathe-cut<br>(semi molded) | |
| Sterilization | Steam Validated Autoclave | Steam Validated Autoclave | |
| Packaging | Blister Pack & vial | Blister Pack & vial | |
| Visibility tint | Titanium Dioxide;<br>21CFR 73.3126 | Titanium Dioxide(White);<br>21CFR 73.3126 | |
| Visibility tint | Iron Oxides;<br>21CFR 73.3125 | Iron Oxides(Red);<br>21CFR 73.3125 | |
| Visibility tint | C.I Pigment Green 7;<br>21CFR 73.3124 | C.I Pigment Green 7;<br>21CFR 73.3124 | |
| Visibility tint | (Phthalocyaninato(2-)) Copper;<br>21CFR 74.3045 | (Phthalocyaninato(2-))<br>Copper;<br>21CFR 74.3045 | |
| Visibility tint | C.I. Reactive Black 5<br>21CFR 73.3121 | Reactive Black 5<br>21CFR 73.3121 | |
| Tint Process | Entrapment | Entrapment | |
| Indication | Soft Contact Lenses for daily<br>wear are indicated for the<br>correction of visual in aphakic and Contact Lens is indicated for<br>non-aphakic persons with non-<br>diseased eyes with myopia or<br>hyperopia and/or possesses<br>refractive astigmatism not<br>exceeding 5.00 diopter, and/or<br>are presbyopia.<br>NOTE: Refractive astigmatism<br>and presbyopia N/A for spherical | The Neo Vision Neo Cosmo<br>(Polymacon) Soft (hydrophilic)<br>daily wear for the correction of<br>visual acuity in aphakic and<br>non-aphakic persons with<br>non-diseased eyes with<br>myopia or hyperopia. The<br>lens may be worn by persons<br>who exhibit refractive<br>astigmatism of 0.50 diopters<br>or less where the astigmatism | |
| | lenses. | does not interfere with visual | |
| | | acuity. | |
| | | The lens is available clear or<br>colored and may be sued to<br>enhance or alter the apparent<br>color of the eye. | |
| | | Eye care practitioners may<br>prescribe the lens frequent<br>replacement wear with<br>cleaning, disinfecting and<br>schedule replacement. not<br>exceeding 5.00 diopter,<br>and/or are presbyopia. | |
| Powers | +25.00 ~ -25.00 Diopter | +25.00 ~ -25.00 Diopter | |
| Total diameter | 12.8 ~ 14.8 | 13.5 ~ 14.5 | |
| Geometrical center<br>thickness | 0.03 ~ 0.30 | 0.03 ~ 0.50 | |
| Curvature | 8.0 ~ 9.5 | 8.3 ~ 9.0 | |
| Optic zone | 6.0 ~ 12.5 | 6.0 ~ 12.5 | |
| Radial edge thickness | 0.03 ~ 0.12 | 0.03 ~ 0.05 | |
| Color | brown, gray, aqua, blue, violet,<br>green | brown, gray, aqua, blue | |
| Toxicity(Safety) | Non-Toxic | Non-Toxic | |
| Tensile Strength (MPa) | 0.427 | 0.425 | |
| % Elongation to Break | 128% | 127% | |
| Module of Elasticity<br>(MPa) | 0.338 | 0.350 | |
| Breaking Force (N) | 0.449 | 0.439 | |
{6}------------------------------------------------
## SUMMARY NON-CLINICAL TESTING
The Neo Vision Neo Cosmo (Polymacon) Soft (Hydrophilic) Contact Lens has been subjected to property evaluation, moist heat (steam) sterilization evaluation and biocompatibility evaluation, and has successfully passed all testing.
### PERFORMANCE TESTING
| TEST TITLE | FINAL REPORT<br>NUMBER | TEST<br>RESULT |
|----------------------------------------------------|------------------------------------------------------------|----------------|
| Tensile Testing | TP-001/TR-001 | Pass |
| Solution Compatibility (Cleaning Cycle<br>Testing) | Protocol per FDA<br>Guidance – Attached/<br>Report 13-0400 | Pass |
| Physical Properties of the Neo Cosmo | MTK-6310 | Pass |
| Physical Properties of the Neo Vision Neo<br>Cosmo | MSK-1241 | Pass |
| Physical Properties of the NEO Toric | MSK-1242 | Pass |
{7}------------------------------------------------
| Physical Properties of the NEXT | MSK-1243 | Pass |
|--------------------------------------|----------|------|
| Physical Properties of the NEW GREEN | MSK-1244 | Pass |
| Physical Properties of the NEO CLEAN | MSK-1245 | Pass |
#### BIOCOMPATIBILITY TESTING
| TEST TITLE | FINAL REPORT<br>NUMBER | TEST<br>RESULT |
|---------------------------------------------------------------------------------------------------------|---------------------------|----------------|
| Ocular Irritation Test | MTK-6310 | Pass |
| Skin Sensitization Test | MTK-6310 | Pass |
| Cytotoxicity Test<br>(Proliferation Inhabition Test) | MTK-6310 | Pass |
| Cytotoxicity Test<br>(Agar Diffusion Test) | MTK-6310 | Pass |
| Ocular Irritation Study of Neo Cosmo Soft<br>Contact Lens in NZW Rabbits | TBH-0375<br>(KG-2013-121) | Pass |
| Skin Sensitization Study of Neo Cosmo soft<br>contact lens using Guinea-Pig Maximization<br>Test (GPMT) | TBH-0376<br>(KG-2013-122) | Pass |
## SUMMARY CLINICAL TESTING
The Neo Vision Neo Cosmo (Polymacon) Soft (Hydrophilic) Contact Lens was not evaluated with clinical testing.
# CONCLUSION
Based on analyzing the device characteristics, bench performance and sterilization testing, and biocompatibility testing, it is the conclusion of Neo Vision Co., Ltd., that the Neo Vision Neo Cosmo (Polymacon) Soft (Hydrophilic) Contact lens is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.