The PolyCatch Retrieval Device is intended for capture and retrieval of excised polyps, foreign bodies, tissue samples and calculi during flexible or rigid endoscopy procedures.
Device Story
PolyCatch Retrieval Device is a single-use, disposable instrument for capturing polyps, foreign bodies, tissue samples, and calculi during endoscopy. Device consists of a handle, outer sheath, snare wire, and non-latex thermoplastic polyurethane (TPU) pouch. Operated by a clinician through the working channel of a flexible or rigid endoscope. Clinician advances the snare wire to deploy the pouch, maneuvers it to capture the target object, and uses the 3-ring handle to retract the pouch against the sheath, securing the object. The endoscope and device are then removed together. The device facilitates safe removal of excised tissue or foreign objects from the gastrointestinal or respiratory tracts, reducing the risk of loss during withdrawal.
Clinical Evidence
Bench testing only. Performance testing was conducted using a test fixture simulating actual clinical use. Biocompatibility testing was performed. No clinical data was required or provided.
Technological Characteristics
Single-piece, disposable endoscopic retrieval device. Components: handle, outer sheath, snare wire, and non-latex thermoplastic polyurethane (TPU) pouch. Pouch attached to snare wire via medical-grade adhesive. Mechanical operation: 3-ring handle controls pouch deployment and retraction through endoscope working channel. Non-electric.
Indications for Use
Indicated for patients undergoing flexible or rigid endoscopy requiring capture and retrieval of excised polyps, foreign bodies, tissue samples, or calculi.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines above them, possibly representing hair or movement. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 30, 2014
Medi-Globe Corporation Scott Karler Regulatory Affairs Coordinator 110 West Orion Street. Suite 136 Tempe, AZ 85283
Re: K142258
> Trade/Device Name: PolyCatch Retrieval Device Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscope electrosurgical unit and accessories Regulatory Class: II Product Code: FDI Dated: August 8, 2014 Received: August 14, 2014
Dear Scott Karler,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known): TBD
Device Name: PolyCatch Retrieval Device
Indications for Use: The PolyCatch Retrieval Device is intended for capture and retrieval of excised polyps, foreign bodies, tissue samples and calculi during flexible or rigid endoscopy procedures.
| Prescription Use | <span style="font-size:20px">✔</span> |
|-----------------------------|---------------------------------------|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
lof 1
Page 1
{3}------------------------------------------------
## 510(k) SUMMARY
### Date of Preparation: 08/06/2014
- A. Submitter Information:
Submitter's Name: Medi-Globe Corporation
Submitter's Address: 110 West Orion Street #136 Tempe, Arizona 85283
Contact Person: Scott Karler
Contact Person's Telephone Number: (480) 897-2772 ext. 208
Contact Person's FAX Number: (480) 897-2878
Contact Person's Email Address: skarler@mediglobe.com
- B. Proposed Device Information
- Trade Name: PolyCatch Retrieval Device
- Common Name: Snare, Flexible ●
- Classification Name: Endoscopic electrosurgical unit and accessories 21 CFR ● 876.4300
- Product Code: FDI, GCJ ●
- C. Predicate Device:
- Trade Name: Roth Net Retriever ●
- Manufacturer: United States Endoscopy Group, Inc. ●
- . Clearance Number: K122462
- Common Name: Snare, Flexible ●
- Classification Name: Endoscopic electrosurgical unit and accessories 21 CFR ● 876.4300
- . Product Code: FDI, GCJ
{4}------------------------------------------------
#### Device Description: D.
The PolyCatch Retrieval Device is a complete one-piece device for retrieval of polyps and foreign bodies from the gastrointestinal and respiratory tracts and is a disposable instrument intended for single patient use only. The PolyCatch Retrieval Device is used through the working channel of a legally marketed endoscope. The PolyCatch Retrieval Device consists of a handle, outer sheath and a snare wire/loop to which a non-Latex pouch is attached. As the snare wire is advanced from of the outer sheath of the device, it expands to its full circumference thereby deploying the non-Latex pouch. The pouch can then be manipulated by the clinician to engage a polyp or foreign body. After the polyp or foreign body is captured within the pouch the handle is used to draw the snare wire and pouch back against the device sheath. The endoscope along with the PolyCatch is then removed together from the patient to complete the retrieval procedure.
- E. Intended Use:
The PolyCatch Retrieval Device is intended for capture and retrieval of excised polyps, foreign bodies, tissue samples and calculi during flexible or rigid endoscopy procedures.
- F. Technological Characteristics Summary:
The PolyCatch Retrieval Device is very similar in design to a non-electric polypectomy snare, with the exception that it features a pre-formed pouch manufactured from a Thermoplastic Polyurethane. (TPU). The pouch is fixed to the snare wire using a medical grade adhesive. The device handle allows the pouch to be advanced out of the device catheter in order to capture the intended tissue or foreign body. Once the tissue/foreign body has been captured within the pouch, the 3-ring handle is used to draw the pouch back towards the device catheter, which then closes around the tissue/foreign body, allowing the device along with the endoscope to be removed from the patient.
{5}------------------------------------------------
- G. Non-Clinical Performance Data:
Design verification data and biocompatibility testing has demonstrated that the proposed PolyCatch Retrieval Device is safe and effective. Performance testing was conducted in accordance with the applicable Guidance Document titled "Guidance for the content of Premarket Notifications for Biopsy Devices Used in Gastroenterology and Urology. Additionally, bench performance testing was conducted using a specially designed test fixture intended to simulate actual device use of the PolyCatch Retrieval Device.
#### H. Conclusions of Non-Clinical Performance Data:
Medi-Globe Corporation has demonstrated that the proposed PolyCatch Retrieval Device is as safe and effective as the predicate device, (Roth Net Retriever, K122462). The PolyCatch Retrieval Device is considered to have the same intended use, diagnostic/therapeutic effect, method of introduction/use, technical characteristics and general range of descriptive features as the predicate device, (Roth Net Retriever, K122462).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.