K142244 · Infinium Medical · MWI · Feb 6, 2015 · Cardiovascular
Device Facts
Record ID
K142244
Device Name
CLEO PATIENT MONITOR
Applicant
Infinium Medical
Product Code
MWI · Cardiovascular
Decision Date
Feb 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The purpose and function of the CLEO patient monitor is to monitor basic physiological parameters including NIBP (systolic and diastolic) SpO2 ETCO2 The target population is for adults only. It can be used in all hospital areas and hospital-type facilities. It may be used as bedside or portable monitor and be used in all hospitals and hospital-type facilities such as clinics and emergency room facilities. It is to be used under the direct supervision of a licensed healthcare practitioner.
Device Story
CLEO Patient Monitor; compact, portable, battery-powered system for monitoring physiological parameters. Inputs: NIBP, SpO2, and ETCO2 via connected sensors/modules. Device processes, analyzes, and displays data on integrated screen; includes recorder. Used in hospitals, clinics, and emergency rooms by healthcare practitioners. Provides real-time vital sign data to support clinical decision-making; facilitates patient transport via internal battery. Output: visual display of physiological traces and numerical values.
Clinical Evidence
Bench testing only. Performance validated against standards: NIBP (ANSI/AAMI SP10:1992/2002, ANSI/AAMI/IEC 80601-2-30:2009), SpO2 (EN ISO 9919:2009), ETCO2 (80601-2-55:2011), Electrical Safety (IEC 60601-1:2005), and EMC (IEC 60601-1-2:2007). Biocompatibility confirmed via history of safe use and prior clearance of patient-contact components.
Technological Characteristics
Compact, portable monitor with touch screen. Powered by AC or internal battery (3-hour capacity). Modules: NIBP, SpO2, ETCO2. Compliance: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ANSI/AAMI SP10, EN ISO 9919, 80601-2-55. Software level of concern: Moderate.
Indications for Use
Indicated for adult patients requiring monitoring of NIBP, SpO2, and ETCO2 in hospital or hospital-type clinical settings under direct supervision of a licensed healthcare practitioner.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is a stylized image of three human profiles facing right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 6, 2015
Infinium Medical % John O'Brien Regulatory and Quality Systems Lead AJW Technology Consultants, Inc 445 Apollo Beach Blvd. Apollo Beach, Florida 33572
Re: K142244
> Trade/Device Name: Cleo Patient Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI, DXN, DQA, CCK Dated: January 6, 2015 Received: January 8, 2015
Dear John O'Brien.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known): K142244
Device Name: CLEO Patient Monitor
The purpose and function of the CLEO patient monitor is to monitor basic physiological parameters including
- NIBP (systolic and diastolic)
- SpO2
- ETCO2
The target population is for adults only.
It can be used in all hospital areas and hospital-type facilities. It may be used as bedside or portable monitor and be used in all hospitals and hospital-type facilities such as clinics and emergency room facilities. It is to be used under the direct supervision of a licensed healthcare practitioner.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
## 510(k) SUMMARY (as required by 807.92)
#### I. SUBMITTER
Infinium Medical 12151 62nd Street North #5 Largo, FL 33773
Phone: 727-531-8434 Fax: 727-531-8436
Contact Person: Oz Ozkaya Date Prepared: August 1, 2014
### REGULATORY CORRESPONDENT
AJW Technology Consultants, Inc 445 Apollo Beach, Blvd Apollo Beach, FL 33572
Phone: 813-645-2855 x101 Fax: 813-645-2856
Contact Person: John O'Brien Email: jobrien@ajwtech.com
#### II. DEVICE
Name of Device: CLEO Patient Monitor Common or Usual Name: Patient Physiological Monitor (without Arrhythmia Detection or Alarm) Classification Name: Monitor, physiological, patient (without Arrhythmia detection or alarms) Device Panel: Cardiovascular Regulatory Class: II Product Code: MWI
### Subsequent Product Codes:
| Regulation<br>Number | Description | Product Code |
|----------------------|-------------------------------------------|--------------|
| 870.1130 | Noninvasive Blood Pressure<br>Measurement | DXN |
| 870.2700 | Oximeter | DQA |
| 868.1400 | Carbon dioxide gas analyzer | CCK |
{4}------------------------------------------------
#### III. PREDICATE DEVICE
The CLEO patent monitor is substantially equivalent in intended use and similar technological characteristics of NIBP, ETC02, and Sp02 as the OMNI Express Patient Monitor which is one of the monitors cleared as part of K132229. The internal Infinium CAPNOTRAK CO2 sensor is substantially equivalent in intended use and technological characteristics to the Respironics LoFlo CO2 Sensor which was cleared under K053174.
This predicate has not been subject to a design-related recall. No reference devices were used in this submission
#### DEVICE DESCRIPTION IV.
The CLEO monitor is a comprehensive monitoring system with two or three traces compiling, and processing, analyzing and displaying data from up to three different patient parameters. It integrates parameter measuring modules, display and recorder in one device, featuring in compactness, lightweight and portability. Built-in battery facilitates transportation of patient.
The CLEO Patient monitor can be powered by an internal battery pack that provides 3 hours of monitoring from fully charged batteries. The batteries are continuously recharged when AC power is connected to the monitor.
The associated accessories include:
- Sp02 Finger Sensors ●
- NIBP Monitor ●
- Blood Pressure Cuff ●
- ETCO2 Module ●
#### V. INDICATIONS FOR USE
The purpose and function of the CLEO patient monitor is to monitor basic physiological parameters including
- NIBP (systolic and diastolic)
- SpO2 ●
- ETCO2 ●
The target population is for adults only.
It can be used in all hospital areas and hospital-type facilities. It may be used as bedside or portable monitor and be used in all hospitals and hospital-type facilities such as clinics and emergency room facilities. It is to be used under the direct supervision of a licensed healthcare practitioner.
{5}------------------------------------------------
#### VI. COMPARISON OF TECHNOLIGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The monitoring of basic physiological parameters is the technological principle for both the subject and predicate devices. It is based on the use of modules and accessories that are either connected internally or externally to the monitors in order to monitor the specific physiological parameter. At a high level, the subject and predicate devices are based on the following same technological characteristics:
- Energy Source ●
- The monitoring of Non-invasive blood pressure
- The monitoring of pulse oximetry
- The monitoring of End Tidal CO2 ●
- Touch Screen Technology ●
- . Software
Because CLEO Patient Monitor's NIBP subsystem is identical to Suntech Medical's NIBP subsystem cleared under K040799 substantial equivalence was not based on performance testing.
The following technological differences exist between the subject and predicate devices:
- . ECG waveform
- Heart Rate Monitoring ●
- Respiration ●
- Temperature
- Anesthetic agents
- Physical Dimensions
#### VII. PERFORMANCE DATA
The following performance date were provided in support of the substantial equivalence determination
### Biocompatibility Testing
The product contact materials utilized in the CLEO Patient Monitor have been well characterized chemically and physically and have a long history of safe use in predicate devices and are categorized as Generally Regarded as Safe (GRAS). In addition, all patient contact components have been FDA cleared through the 510(k) Premarket Notification process and have been tested for biocompatibility.
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Infinium device, consisting of the CLEO Monitor, NIBP Monitor and Cuff, SPO2 finger Sensors
{6}------------------------------------------------
and the ETCO> Module. The system complies with the IEC 60601-1 standards for safety and the IEC 60601-1-2 standard for EMC.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent flaw could directly result in minor injury to the patient or operator.
### Mechanical Testing
- CO2 Measurement Accuracy and Drift of Measurement Accuracy
- Gas Sampling Rate Accuracy ●
- System Response/Rise Time ●
- Respiratory Rate Accuracy ●
- Gas Mixture Accuracy ●
- Interfering Agents
- SPO2 Performance Test ●
- NIBP Performance Test ●
### Summary of Non-Clinical Data:
The Cleo Patient Monitor underwent testing according to several different performance standards. Below is a chart of the different testing that was completed.
| Devices | Performance Test | Standard of Compliance |
|--------------|-------------------|--------------------------------------------------------------------|
| CLEO Monitor | NIBP | ANSI/AAMI SP10:1992 and<br>2002, ANSI/AAMI/IEC 80601-<br>2-30:2009 |
| | SPO2 | EN ISO 9919:2009 |
| | Electrical Safety | IEC 60601-1:2005/ EN 60601-<br>1:2006 |
| | EMC | IEC 60601-1-2:2007 |
| | ETCO2 | 80601-2-55:2011 |
### VIII. CONCLUSIONS
{7}------------------------------------------------
The testing completed demonstrates that the CLEO Patient Monitor exhibits comparable mechanical and functional characteristics to the predicate device in addition to being biocompatible acceptable. Based on those characteristics, the CLEO Patient Monitor is substantially equivalent to the predicate device in safety and effectiveness in addition to being intended for the same uses.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.