K142207 · Eppendorf AG · MQJ · Jan 23, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K142207
Device Name
TransferMan 4m Micromanipulator
Applicant
Eppendorf AG
Product Code
MQJ · Obstetrics/Gynecology
Decision Date
Jan 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6150
Device Class
Class 2
Indications for Use
The TransferMan® 4m Micromanipulator is intended for use in assisted reproduction procedures requiring coarse and fine positioning of a microtool under the microscope. The CellTram® Microinjectors are intended for use in Intra-Cytoplasmic Spermatozoa Injection (ICSI) procedures to aspirate and inject sperm into oocytes.
Device Story
TransferMan 4m is a motorized micromanipulator; consists of motor module and control board with joystick; mounts to inverted microscope; provides coarse/fine tool positioning for IVF/ICSI. CellTram Air/Oil/vario are manual piston pumps; generate pressure differences via piston/cylinder system; transmit pressure to microcapillary via tube; used for aspiration/injection of sperm/oocytes. Operated by embryologists/technicians in clinical IVF labs. Output is physical movement of microtools or fluid pressure for cell manipulation. Benefits include precise, controlled cell handling during assisted reproduction.
Clinical Evidence
No clinical data. Safety and effectiveness demonstrated via non-clinical bench testing, including electrical safety (UL 61010-1), EMC (EN 61326-1, EN 61000 series), and endurance/packaging validation (ASTM D4169-09).
Technological Characteristics
TransferMan 4m: Motorized X,Y,Z-axis micromanipulator; joystick control; serial interface; 120-240V AC power. CellTram: Manual piston/cylinder pumps; air or oil transmission media; 2,640 µL (Air) or 960 µL (Oil/vario) total volume. Standards: EN 61326-1, EN 55011, EN 61000 series, UL 61010-1, ASTM D4169-09.
Indications for Use
Indicated for assisted reproduction procedures requiring microtool positioning (TransferMan 4m) and ICSI procedures for sperm aspiration and injection into oocytes (CellTram series).
Regulatory Classification
Identification
Assisted reproduction micromanipulators are devices intended to control the position of an assisted reproduction microtool. Assisted reproduction microinjectors are any device intended to control aspiration or expulsion of the contents of an assisted reproduction microtool.
Special Controls
*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center is a stylized caduceus symbol, which is often associated with healthcare. The caduceus is depicted with a modern, abstract design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 23, 2015
Eppendorf AG % JoAnne Bronikowski Regulatory Services Senior Project Manager Aptiv Solutions 62 Forest Street, Suite 300 Marlborough, MA 01752
Re: K142207
Trade/Device Name: TransferMan® 4m Micromanipulator CellTram® Air Microinjector CellTram® Oil Microinjector CellTram® vario Microinjector Regulation Number: 21 CFR 884.6150 Regulation Name: Assisted reproduction micromanipulators and microinjectors Regulatory Class: Class II Product Code: MQJ Dated: December 17, 2014 Received: January 23, 2015
Dear JoAnne Bronikowski,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | |
|-----------------------------------------|--|
| Food and Drug Administration | |
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
### Indications for Use
510(k) Number (if known) K142207
#### Device Name
TransferMan® 4m Micromanipulator, CellTram® Air Microinjector, CellTram® Vario Microinjector
Indications for Use (Describe)
TransferMan® 4m Micromanipulator Indications for Use:
The TransferMan® 4m Micromanipulator is intended for use in assisted reproduction procedures requiring coarse and fine positioning of a microtool under the microscope.
CellTram® Air Microinjector / CellTram® Oil Microinjector / CellTrans® vario Microinjector Indications for Use: The CellTram® Microinjectors are intended for use in Intra-Cytoplasmic Spermatozoa Injection (ICSI) procedures to aspirate and inject sperm into oocytes.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | lebrand |
|------------------------------------------------------------------------------|---------|
|------------------------------------------------------------------------------|---------|
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
PSC Publishing Services (101) 443-6740
{3}------------------------------------------------
# 510(k) Summary for the Eppendorf AG
TransferMan® 4m Micromanipulator and CellTram® Microinjectors (per 21CFR 807.92 and http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/default.htm)
### 1. SUBMITTER/510(K) HOLDER
Eppendorf AG Barkhausenweg 1 22339 Hamburg Germany Phone: +49 40 53801 - 188 Establishment Registration No: 801047
Contact: Dr. Cornelia Hildebrandt, Global Product Line Manager Cell Handling Contact's Phone: +49 40 53801 - 188 Contact's Email: hildebrandt.c@eppendorf.de Date Prepared: July 28, 2014
### 2. DEVICE NAME
| Trade Name: | TransferMan <sup>®</sup> 4m Micromanipulator<br>CellTram <sup>®</sup> Air Microinjector<br>CellTram <sup>®</sup> Oil Microinjector<br>CellTram <sup>®</sup> vario Microinjector |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Micromanipulator and microinjectors |
| Device Name: | Assisted reproduction micromanipulators and microinjectors |
| Classification Panel: | Obstetrics/Gynecology Devices |
| Classification Number: | 884.6150 |
| Product Code: | MQJ |
#### 3. PREDICATE DEVICES
The proposed TransferMan® 4m Micromanipulator and CellTram Microinjectors are substantially equivalent to other legally marketed assisted reproduction micromanipulators and microinjectors including the following:
- IM-11 Pneumatic Microinjector (Narishige Co., Ltd., K113712) ●
- NAI-5 Micromanipulator Set (Narishige Co., Ltd., K120877)* ●
* Eppendorf is claiming substantial equivalence of the proposed TransferMan® 4m Micromanipulator only to the Micromanipulator, not to the NAI-5 Micromanipulator Set as a whole.
A comparison of the intended use and features of the proposed TransferMan® 4m Micromanipulator and CellTram Microinjectors and the predicate devices described in K113712 and K120877 is provided in Table 5-1.
{4}------------------------------------------------
### 4. DEVICE DESCRIPTION
The TransferMan® 4m Micromanipulator is a motorized micromanipulator consisting of a Motor module unit and a Control board that allows the user to precisely control the movement of tools (i.e., microcapillaries) used for Intra-Cytoplasmic Spermatozoa Injection (ICSI) and other In Vitro Fertilization (IVF) procedures. The tool is mounted in a Universal Capillary Holder on the TransferMan® 4m Micromanipulator's Motor module unit that is, itself, mounted to an inverted microscope. The user controls the movements of the tool using a joystick on the Control board. The Motor module unit shifts the position of the tool in response to the joystick motions.
The CellTram Microinjectors are manual piston pumps for holding and transfer of suspension cells (e.g., oocytes and sperms for use in ICSI procedures), used in manual microinjection procedures. Using a movable piston in a cylinder system, the CellTram® Microinjectors generate differences in pressure that are transferred directly to a microcapillary via a connected pressure tube. Depending on the piston movement, material can be aspirated or dispensed. Three versions of the CellTram® Microinjectors will be marketed for use in ICSI procedures, the CellTram® Air, the CellTram® Oil, and the CellTram vario. The three versions differ in the medium used for pressure transmission (air versus oil).
## 5. INDICATIONS FOR USE
The TransferMan 4m Micromanipulator is intended for use in assisted reproduction procedures requiring coarse and fine positioning of a microtool under the microscope.
The CellTram Microinjectors are intended for use in Intra-Cytoplasmic Spermatozoa Injection (ICSI) procedures to aspirate and inject sperm into oocytes.
## 6. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICES
Both the proposed TransferMan® 4m Micromanipulator and the predicate NAI-5 Micromanipulator are motor-driven micromanipulators that allow the user to control the movement of a tool using the joystick on the Control unit. Motors in the Motor module unit shift the position of the tool in response to the joystick motions in X, Y, and Z directions. The manufacturers of both the proposed and predicate devices provide adapters for attaching the Motor module unit to inverted microscopes commonly used for assisted reproduction procedures. The proposed TransferMan® 4m Micromanipulator and predicate NAI-5 Micromanipulator are indicated for coarse and fine tool positioning.
The CellTram Microinjectors and the predicate IM-11 Pneumatic Microinjector are used
{5}------------------------------------------------
to control aspiration of the sperm into the microcapillary, and after perforation of the oocyte to release of the sperm into the ooplasm. Both the proposed and predicate devices consist of manually controlled pistons that slide within a cylinder system in and out to aspirate or dispense. The CellTram Air, the CellTram® Oil, and the CellTram® vario have different mediums used for pressure transmission (air versus oil). The predicate IM-11 Pneumatic Microinjector operates using air only.
The manufacturer believes that the technological characteristics of the TransferMan 4m Micromanipulator and the CellTram Microinjectors are substantially equivalent to those of the predicate devices.
The tables below provide the comparison of the characteristics between the proposed and predicated devices.
| | Feature | TransferMan® 4m<br>Micromanipulator Eppendorf AG<br>(Proposed) | NAI-5 Micromanipulator Set<br>Narishige Col. Ltd System<br>(K120877) |
|--------------|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| General | Intended Use | Intended for use in assisted<br>reproduction procedures requiring<br>coarse and fine positioning of a<br>microtool under the microscope. | : "… enables course and fine<br>positioning of a microtool under the<br>microscope and is used in assisted<br>reproduction procedures." |
| | Components | Motor Module<br>Control Board | Drive Unit<br>2 Control Units for coarse and fine<br>Power Supply |
| | Power Supply | Input: 120-240 V AC; 50-60 Hz | Input: 100-240 V; 50 Hz |
| | Range of movement | Coarse: $5 \mu m$ to $12,500 \mu m$ in $5 \mu m$<br>increments<br>Fine: $5 \mu m$ to $2,000 \mu m$ in $5 \mu m$<br>increments<br>X-Fine: $1 \mu m$ to $600 \mu m$ in $1 \mu m$<br>increments | Coarse: 22mm<br>Fine: 10mm<br>Increments not known |
| | Speed | 0 - 10,000 µm/s | Max. 1.4 mm/s |
| | Step Motors | X-,Y-,Z- Module | X-,Y-,Z- Module |
| | Step Size | < 20 nm (calculated resolution) | |
| Motor Module | Mechanical<br>Adjustability | >80 mm | Not known |
| | Dimensions | Module (X,Y,Z): 129 mm x 51 mm x<br>36 mm (W x D x H) | Drive Unit: 57 mm x 138 mm x<br>94 mm (W x D x H) |
| | Weight | Motor module (complete): 2.15 kg<br>Module (X,Y,Z): 570g | Not known |
| | Maximum Traveling<br>Distance | 20 mm | Coarse: 22mm<br>Fine: 10mm |
| | Swivel Joint<br>Direction of<br>Rotation | -45° - +90° | Not known |
| | | | |
Side-by-Side Comparison of the TransferMan® 4m Micromanipulator with the NAI-5 Micromanipulator
{6}------------------------------------------------
| | Feature | TransferMan® 4m<br>Micromanipulator Eppendorf AG<br>(Proposed) | NAI-5 Micromanipulator Set<br>Narishige Col. Ltd System<br>(K120877) |
|--------------|-------------------------------|----------------------------------------------------------------|----------------------------------------------------------------------|
| Control Unit | Capillary Exchange | Direction of rotation: forward (swivel out) | |
| | Sample Replacement | Direction of rotation: backward (swivel in) | |
| | Operating Angle of Angle Head | 0° - 90° | |
| | Microscope Compatibility | All major microscope brands: Leica, Nikon, Olympus, and Zeiss | All major microscope brands: Leica, Nikon, Olympus, and Zeiss |
| | Dimensions | 205 mm x 288 mm x 152 mm (W x D x H) | 70 mm x 100 mm x 120 mm (W x D x H) |
| | Weight | Control board: 1.8 kg (includes mains/power supply device) | Not known |
| | Control Mechanism | Dual Speed joystick | Joystick |
| | Working range | Coarse, fine, x-fine | Coarse, fine |
| | Programmable | Yes | No |
| | External device/PC Connection | Serial interface SubD9, male | Not known |
| | Speed Control | Proportional and dynamic kinetics | |
## Side-by-Side Comparison of the CellTram® Microinjectors with the IM-11 Pneumatic Microinjector
| | Feature | CellTram® Microinjectors<br>Eppendorf AG<br>(Proposed) | IM-11 Pneumatic Microinjector<br>Narishige Co., Ltd.<br>(K113712) |
|---------------------------|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General | Intended Use | Intended for use in Intra-Cytoplasmic<br>Spermatozoa Injection (ICSI)<br>procedures to aspirate and inject<br>sperm into oocytes. | Intended for use for Intra-<br>Cytoplasmic Sperm Injection (ICSI)<br>procedures to aspirate and inject<br>sperm into oocytes, and to hold<br>oocytes during the ICSI procedure |
| | Microinjector Type | CellTram® Air (air transfer medium)<br>CellTram® Oil (oil transfer medium)<br>CellTram® vario (oil transfer medium,<br>coarse/fine adjustment) | Air transfer medium |
| | Operation | Manual microinjection | Manual microinjection |
| | Minimum<br>Adjustment Volume | <0.2 µL | Not known |
| | Adjustment total<br>volume | 2,640 μL | Not known |
| | Applications | Gentle holding of larger cells, e.g.<br>oocytes<br>Manual microinjection, aspiration and<br>dispensing of cells (e.g. sperms) | Holding oocytes<br>Manual aspiration of sperm into<br>injection pipette<br>Microinjection of sperm into oocytes |
| | Air | Type | Manual pressure |
| | | Pressure Generation | Piston/cylinder system, filled with air,<br>maintenance-free |
| | | Max Pressure | 2,900 hPa |
| Minimum Setting<br>Volume | | <200 nL | Not known |
| | Feature | CellTram® Microinjectors<br>Eppendorf AG<br>(Proposed) | IM-11 Pneumatic Microinjector<br>Narishige Co., Ltd.<br>(K113712) |
| | Total Setting<br>Volume | 2,640 µL | Not known |
| | Change in Volume<br>(per dial revolution) | 88 µL | coarse: 1.0 mL<br>fine: 250 µL |
| | Max Range | 15 mm (air) | 40 mm (coarse and fine combined)<br>30 mm coarse<br>17 mm fine |
| | Applications | Gentle holding of larger cells, e.g.<br>oocytes<br>Manual microinjection, aspiration and<br>dispensing of cells (e.g. sperms) | |
| | Type | Manual hydraulic | |
| Oil | Pressure Generation | Piston/cylinder system, filled with oil | |
| | Max Pressure | 20,000 hPa | Not Applicable |
| | Minimum Setting<br>Volume | <20 nL | |
| | Total Setting<br>Volume | 960 µL | |
| | Change in Volume<br>(per dial revolution) | 9.6 µL | |
| | Max Range | 50 mm | |
| vario | Applications | Gentle holding of larger cells, e.g.<br>oocytes<br>Manual microinjection, aspiration and<br>dispensing of cells (e.g. sperms)<br>Manual aspiration, dispensing, and<br>removal of cells and organelles (e.g.<br>polar bodies). | |
| | Type | Manual hydraulic | |
| | Pressure Generation | Piston/cylinder system, with gear,<br>filled with oil | Not Applicable |
| | Max Pressure | 20,000 hPa | |
| | Minimum Setting<br>Volume | <20 nL/<2 nL (coarse/fine) | |
| | Total Setting<br>Volume | 960 µL | |
| | Change in Volume<br>(per dial revolution) | course: 9.6 µL<br>fine: 0.96 µL | |
| | Max Range | 50 mm | |
______________________________________________________________________________________________________________________________________________________________________________
{7}------------------------------------------------
## 7. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE
The safety and effectiveness of the proposed TransferMan® 4m Micromanipulator has been confirmed by hardware and software testing. The proposed micromanipulator complies with applicable requirements of the following standards:
- EN 61326- 1:2006, "Electrical Equipment for Measurement, Control and Laboratory .
{8}------------------------------------------------
Use - EMC Requirements"
- . EN 55011:2011, "Industrial, Scientific and Medical (ISM) Radio-Frequency Equipment - Radio Disturbance Characteristics - Limits And Methods Of Measurement"
- EN 61000-4-2:2009-12, Electromagnetic compatibility (EMC) Part 4: Testing and . measurement techniques - Section 2: Electrostatic discharge immunity test
- EN 61000-4-3:2011-04, "Electromagnetic compatibility (EMC) Part 4-3: Testing . and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test"
- EN 61000-4-4:2013-04, "Electromagnetic compatibility (EMC) Part 4: Testing and . measurement techniques - Section 4: Electrical fast transient/burst immunity test"
- EN 61000-4-5:2007-06, "Electromagnetic compatibility (EMC) Part 4: Testing and . measurement techniques – Section 5: Surge immunity test"
- EN 61000-4-6:2009-12, "Electromagnetic compatibility (EMC) Part 4: Testing and . measurement techniques – Section 6: Immunity to conducted disturbances, induced by radio-frequency fields"
- EN 61000-4-11:2005-02, "Electromagnetic compatibility (EMC) Part 4: Testing and . measuring techniques - Section 11: Voltage dips, short interruptions and voltage variations immunity tests"
- UL 61010-1 Edition 2:2008/10/28, "Electrical Equipment For Measurement, Control, . and Laboratory Use; Part 1: General Requirements''
The safety and effectiveness of the proposed CellTram® Microinjectors has been confirmed by hardware testing. The proposed CellTram® Microinjectors underwent endurance testing. Packaging validation testing has been performed according to ASTM D4169-09, "Standard Practice for Performance Testing of Shipping Containers and Systems".
## 8. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS
The indications for use, principles of operation, and technological characteristics of the TransferMan® 4m Micromanipulator and CellTram® Microinjectors are substantially equivalent to the predicate devices NAI-5 Micromanipulator (subject of K120877) and IM-11 Pneumatic Microinjector (subject of K113712), respectively. Differences between the proposed devices and the predicate devices are limited to minor differences in technological characteristics. These differences do not impact the safety and effectiveness of the micromanipulator or microinjectors for their intended use.
{9}------------------------------------------------
The safety and performance of the TransferMan® 4m Micromanipulator and CellTram® Microinjectors for their intended use is demonstrated by non-clinical testing. Based on the evidence provided, Eppendorf believes that the proposed TransferMan® 4m Micromanipulator and CellTram Microinjectors are substantially equivalent to the predicate devices NAI-5 Micromanipulator and IM-11 Pneumatic Microinjector and the differences between the products are minor, and raise no new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.