BD PHOENIX AUTOMATED MICROBIOLOGY SYSTESM - TIGECYCLINE (0.313-4UG/ML) GP

K142170 · Becton, Dickinson and Company · LON · Oct 28, 2014 · Microbiology

Device Facts

Record IDK142170
Device NameBD PHOENIX AUTOMATED MICROBIOLOGY SYSTESM - TIGECYCLINE (0.313-4UG/ML) GP
ApplicantBecton, Dickinson and Company
Product CodeLON · Microbiology
Decision DateOct 28, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.

Device Story

BD Phoenix Automated Microbiology System performs in vitro antimicrobial susceptibility testing (AST). System utilizes sealed, self-inoculating polystyrene trays containing dried reagents and cation-adjusted Mueller-Hinton broth. Input: pure culture bacterial colonies suspended in broth (standardized to 0.5 McFarland via nephelometer or automated Phoenix AP instrument). Process: panels incubated at 35°C; instrument continuously monitors growth via colorimetric oxidation-reduction indicator (blue to pink to colorless) and turbidity. Readings taken every 20 minutes for 4-16 hours. Output: MIC values and categorical interpretations (Susceptible/Non-susceptible). Software-driven 'EXPERT' system applies rules derived from CLSI standards and FDA labeling to interpret results. System eliminates manual reading; provides automated, quantitative susceptibility data to clinicians to guide antimicrobial therapy.

Clinical Evidence

Performance evaluated using clinical, stock, and challenge isolates across multiple US sites. Comparison against CLSI reference broth microdilution method. Study included 1021 isolates for Tigecycline. Results: 98.0% Essential Agreement (EA) and 100.0% Category Agreement (CA). Reproducibility study (intra- and inter-site) showed >95% agreement (+/- 1 dilution).

Technological Characteristics

Polystyrene trays with 136 micro-wells; cation-adjusted Mueller-Hinton broth with 0.01% Tween 80; redox indicator (colorimetric oxidation-reduction); automated incubation at 35°C; optical monitoring of turbidity and color change; software-driven rule-based interpretation (EXPERT system); connectivity via automated instrument platform.

Indications for Use

Indicated for in vitro quantitative determination of antimicrobial susceptibility (MIC) of pure culture bacterial isolates. Specifically for Tigecycline (0.0313-4 ug/mL) in Gram-positive ID/AST or AST only panels. Target populations include Enterococcus faecalis (vancomycin-susceptible) and Staphylococcus aureus (methicillin-susceptible and -resistant).

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM ASSAY ONLY TEMPLATE A. 510(k) Number: K142170 B. Purpose for Submission: Addition of Tigecycline to the BD Phoenix Gram positive ID/AST and AST only panels. C. Measurand: Tigecycline 0.0313 - 4 µg/mL D. Type of Test: Antimicrobial Susceptibility Test (Quantitative) colorimetric, oxidation-reduction, growth based E. Applicant: Becton, Dickinson and Company F. Proprietary and Established Names: BD Phoenix™ Automated Microbiology System G. Regulatory Information: 1. Regulation section: 21 CFR 866.1645 – Short-Term Antimicrobial Susceptibility Test System 2. Classification: Class II 3. Product code: LON – System, Test, Automated, Antimicrobial Susceptibility, Short Incubation {1} 4. Panel: 83 Microbiology H. Intended Use: 1. Intended use(s): The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus. 2. Indication(s) for use: The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacterial isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus. This premarket notification is for the addition of the antimicrobial agent tigecycline at concentrations of 0.0313-4μg/mL to Gram-positive ID/AST or AST only Phoenix panels. Tigecycline has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package inserts for this antimicrobial agent. **Active In Vitro and in Clinical Infections Against:** Enterococcus faecalis (vancomycin-susceptible isolates) Staphylococcus aureus (including methicillin-susceptible and -resistant isolates) 3. Special conditions for use statement(s): For prescription use only. The current absence of resistant isolates precludes defining any results other than "Susceptible". Isolates yielding MIC result suggestive of "Non-susceptible" category should be submitted to a reference laboratory for further testing. 2 {2} 4. Special instrument requirements: For use with the BD Phoenix™ Automated Microbiology System I. Device Description: The BD Phoenix™ Automated Microbiology System includes instrumentation and software, sealed and self-inoculating molded polystyrene trays with 136 micro-wells containing dried reagents, and specific inoculum broth formulations for identification (ID) and antimicrobial susceptibility testing (AST). The organism to be tested must be in pure culture and preliminarily identified as Gram-positive or Gram-negative. Colonies are then suspended in the ID broth, and, using one of the recommended BD nephelometer devices, brought to a concentration of 0.5 McFarland. A further dilution is made into the AST broth. Prior to inoculation of the panel, an AST broth indicator is added which changes the AST broth to a blue color. The AST broth is a cation-adjusted formulation of Mueller-Hinton broth containing 0.01% Tween 80, and has a final isolate concentration of approximately 5×10⁵ CFU/mL. After inoculation and incubation, the AST broth indicator in the AST broth changes from blue to pink to colorless as organism growth and reduction in the panel well occurs. Inoculated panels are barcode scanned and loaded into the BD Phoenix™ Automated Microbiology System instrument where the panels are incubated at 35°C and continuously monitored for changes in the indicator and bacterial turbidity to determine bacterial growth in the presence of an antimicrobial agent. Organisms killed or inhibited by a given antimicrobial do not cause reduction of the indicator and therefore do not produce a color change. Additional interpretation is done using the software driven "EXPERT" System triggered rules derived from the CLSI standards and/or FDA drug labeling. Readings are taken every 20 minutes and a final AST result in 4 to 16 hours. The AST result is determined via automated readings; no manual readings are possible with this system. J. Substantial Equivalence Information: 1. Predicate device name(s): Vitek® Antimicrobial Susceptibility Test System 2. Predicate 510(k) number(s): N50510 {3} # 3. Comparison with predicate: Table 1. Similarities and Differences of the BD Phoenix Tigecycline and the Predicate | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | BD Phoenix Automated Microbiology System (Tigecycline) | VITEK (N50510) | | Intended Use | Determination of in vitro antimicrobial susceptibility testing of aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria | Same | | Source of Microorganisms for Testing | Bacterial colonies isolated from culture | Same | | System | Automated instrumentation for in vitro antimicrobial susceptibility testing (AST) | Same | | Incubation Time | Short-term (<16 hours) | Same | | Test Card | Contains dried antimicrobials and substrates | Same | | Results | MIC and interpretive criteria (i.e., susceptible, intermediate, resistant and non-susceptible) | Same | | Technology | Automated growth-based detection | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Methodology | MIC determination based on serial two-fold dilution format | MIC determination based on computer assisted extrapolation of doubling dilutions | | Technology | Automated growth based, enhanced by use of a redox indicator (colorimetric oxidation-reduction) to detect organism growth | Automated growth based detection using attenuation of light measured by an optical scanner | # K. Standard/Guidance Document Referenced (if applicable): - CLSI M7-A8 "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically" {4} - CLSI M100-S22 “Performance Standards for Antimicrobial Susceptibility Testing” - Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test Systems; Guidance for Industry and FDA L. Test Principle: The BD Phoenix™ Automated Microbiology System is a broth based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in “growth control wells” that contain no antibiotic. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Two separate reproducibility studies were conducted for each inoculum preparation method. Reproducibility testing using inocula prepared manually (PhoenixSpec Nephelometer) was conducted at two external and one internal site. Fourteen isolates with on-scale MIC values as pre-determined by reference methods were provided to the testing sites by BD. Isolate identification and expected MIC result was blinded to those conducting the testing. Testing was performed in triplicate on three separate days. Reproducibility testing using inocula prepared by the automated (Phoenix AP) instrument was conducted at two external sites and one internal site. Testing was conducted using 14 isolates with on-scale MIC values. Isolate identification and expected MIC result was blinded to those conducting the testing. Testing was performed in triplicate on three separate days. Results of the inter-site and intra-site reproducibility studies were acceptable and demonstrated best case and worst case results of greater than 95%. A summary of the reproducibility study performance is illustrated in Table 2 below. 5 {5} Table 2. Summary of Reproducibility Studies – BD Phoenix Tigecycline - Gram Positive | BD Phoenix Instrument Platform | Inoculation Method | Best Case | Worst Case | | --- | --- | --- | --- | | BD Phoenix | Phoenix AP Instrument | 100% | 100% | | | Manual PhoenixSpec Nephelometer | 100% | 100% | Best case calculation for reproducibility assumes off-scale results are within one well of the mode MIC value. Worst case calculation for reproducibility assumes off-scale results are greater than one well from the mode MIC value. Regarding both the manual inoculation method, and the automated inoculation method (AP Instrument), no MIC results were off-scale. There were a total of 378 MIC values evaluated for the manual inoculation method and likewise a total of 378 MIC values evaluated for the automated inoculation method (AP Instrument). b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Testing of the FDA and CLSI recommended quality control (QC) strains, S. aureus and E. faecalis was performed each day of the challenge and clinical study testing with the reference method and with the BD Phoenix System. The inocula were standardized using both the automated (Phoenix AP) and manual (PhoenixSpec Nephelometer) methods. A sufficient number of tests were performed and all quality control results for the BD Phoenix fell within the acceptable ranges as per the FDA drug label, demonstrating that the BD Phoenix System can consistently produce quality control results in the recommended range for tigecycline. Quality control testing of the BD Phoenix Tigecycline was conducted using two recommended QC strains. A total of 275 results were obtained for S. aureus ATCC 29213 and 275 for E. faecalis ATCC 29212. Results demonstrate that acceptable results were achieved &gt;95% of the time by both auto-dilution and manual dilution inoculum preparation methods as shown in Table 3 and Table 4 below. {6} Table 3. Quality Control Results - BD Phoenix Tigecycline S. aureus ATCC 29213 | ORGANISM | Tigecycline (μg/mL) | Reference | Auto-Dilution (Phoenix AP) | Manual Dilution (Phoenix SpecTM) | | --- | --- | --- | --- | --- | | S. aureus 29213 Expected Range: 0.03 – 0.25 μg/mL | | | | | | | ≤ 0.0313 | | | | | | 0.0625 | 2 | 4 | 15 | | | 0.125 | 66 | 78 | 99 | | | 0.25 | 7 | 1 | 3 | | | Total | 75 | 83 | 117 | Table 4. Quality Control Results - BD Phoenix Tigecycline E. faecalis ATCC 29212 | ORGANISM | Tigecycline (μg/mL) | Reference | Auto-Dilution (Phoenix AP) | Manual Dilution (Phoenix SpecTM) | | --- | --- | --- | --- | --- | | E. faecalis 29212 Expected Range: 0.03 – 0.12 μg/mL | | | | | | | ≤0.03 | | 1 | | | | 0.0625 | 45 | 76 | 99 | | | 0.125 | 28 | 7 | 18 | | | 0.25 | 1 | | | | | Total | 74 | 84 | 117 | Growth Rate: The overall growth rate during the clinical studies was $99.7\%$ (1150/1153). Purity Check Plates: Purity check plates were inoculated from the standardized organism suspensions for both the Phoenix and reference methods. Any isolate that showed mixed growth on the purity check plate was considered noncompliant and not included in result analysis. Inoculum Density Control: The PhoenixSpec Nephelometer was used to prepare the inocula for testing of the clinical, challenge, reproducibility and QC isolates. The same inoculum suspension was used for both the Phoenix System and the reference method testing. The BD Phoenix AP instrument was used to standardize the inocula for challenge, QC, and reproducibility isolates. Validation data for both the PhoenixSpec and the Phoenix AP instrument was provided and found to be acceptable. d. Detection limit: Not applicable e. Analytical specificity: Not applicable {7} f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: The accuracy of results obtained with the Phoenix System was determined by comparison to the CLSI-recommended broth dilution method (reference method). Reference panels were prepared according to CLSI M07-A8 guidelines. In addition, specific to the drug tigecycline, fresh Mueller Hinton broth (&lt;12 hours old) was used in the preparation of the reference panels. Sites performed testing on Gram positive isolates of *E. faecalis* and *S. aureus* using Phoenix and reference panel formats appropriate for these organisms. Antimicrobial agents in the test and reference panels had identical dilution ranges which were appropriate for the interpretive breakpoints of the drug. Testing was performed using at least two different production lots of Phoenix panels, AST broth and AST indicator at each study site. A minimum of three different lots of the Phoenix panel were used across all sites for the entire study. Phoenix and reference panels were inoculated using the same organism suspension. Growth in the Phoenix panels was determined from data recorded by the instrument. Performance was analyzed using FDA breakpoints for tigecycline, and results were compared to results obtained by the broth microdilution reference method based on the guidelines provided in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems. A total of 977 clinical isolates were tested at the three study sites and included both fresh and stock isolates (783 *S. aureus*, 194 *E. faecalis*). Clinical isolates tested included representatives of species listed in the FDA pharmaceutical drug label. Clinical isolates were tested using inocula prepared using the PhoenixSpec nephelometer (manual method). A total of 44 challenge isolates were supplied to the testing sites by the sponsor (30 *S. aureus*, 14 *E. faecalis*). Challenge isolates were obtained from BD’s internal collection and from external laboratories. Results obtained for Challenge isolates using the Phoenix System were compared to expected MIC results; expected MIC values and categorical interpretations were derived from testing with multiple lots of reference broth dilution panels over a three-month period. The challenge set was divided into subsets and an individual subset was distributed to each of the three study sites. Identification and expected results were masked to the study sites. The inocula for the challenge isolates were prepared using both the PhoenixSpec nephelometer (manual method) and the Phoenix AP (automated method). The performance evaluation summary of essential and categorical agreement results 8 {8} for clinical and challenge isolates with inocula prepared using the PhoenixSpec nephelometer (manual method) is shown in Table 4 below. The performance evaluation summary of essential and categorical agreement results for challenge isolates with inocula prepared using the Phoenix AP (automated method) is shown in Table 5 below. Table 4. BD Phoenix Tigecycline (PhoenixSpec Nephelometer - Manual Inoculum Preparation) | | Tot | EA N | % EA | Total Eval | EA Eval N | %EA Eval | CA N | % CA | #NS | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | | | | | | | | | | | | | | S. aureus(methicillin-resistant and methicillin-susceptible isolates combined) | 783 | 769 | 98.2 | 778 | 764 | 98.2 | 783 | 100 | 2 | 0 | 0 | 0 | | E. faecalis(vancomycin-susceptible isolates only) | 193 | 187 | 96.9 | 182 | 182 | 100 | 193 | 100 | 0 | 0 | 0 | 0 | | Challenge | | | | | | | | | | | | | | S. aureus(methicillin-resistant and methicillin-susceptible isolates combined) | 30 | 30 | 100 | 30 | 30 | 100 | 30 | 100 | 0 | 0 | 0 | 0 | | E. faecalis(vancomycin-susceptible isolates only) | 14 | 14 | 100 | 14 | 14 | 100 | 14 | 100 | 0 | 0 | 0 | 0 | | | | | | | | | | | | | | | | Combined Clinical and Challenge * | 1021 | 1001 | 98.0 | 1005 | 991 | 98.6 | 1021 | 100 | 2 | 0 | 0 | 0 | *Combined results include one isolate of E. faecalis with unknown vancomycin characterization. Table 5. BD Phoenix Tigecycline (Phoenix AP - Auto Inoculum Preparation) | | Tot | EA N | % EA | Total Eval | EA Eval N | %EA Eval | CA N | % CA | #NS | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Challenge | | | | | | | | | | | | | | S. aureus(methicillin resistant and methicillin susceptible isolates combined) | 30 | 30 | 100 | 30 | 30 | 100 | 30 | 100 | 0 | 0 | 0 | 0 | | E. faecalis(vancomycin susceptible isolates only) | 14 | 14 | 100 | 14 | 14 | 100 | 14 | 100 | 0 | 0 | 0 | 0 | | | | | | | | | | | | | | | | Combined Challenge | 44 | 44 | 100 | 44 | 44 | 100 | 44 | 100 | 0 | 0 | 0 | 0 | EA = Essential Agreement NS = Non-susceptible maj = major discrepancies CA = Category Agreement min = minor discrepancies vmj = very major discrepancies {9} Essential Agreement (EA) occurs when there is agreement between the result of the reference method and that of BD Phoenix within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the BD Phoenix panel and the reference method. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the BD Phoenix result. For the clinical and challenge organism testing performed for tigecycline using the BD Phoenix, the overall % EA and % CA consistently met the acceptance criteria of greater than or equal to 90%. Overall, there were no major, very major, or minor categorical errors. The clinical data or Staphylococcus aureus demonstrated downward trending in the MIC of the BD Phoenix Tigecycline compared to the reference method. The following footnote to inform end users of the observed downward trending was added to the table in the package insert: "BD Phoenix MIC values for isolates of Staphylococcus aureus may be lower by one dilution compared to reference broth microdilution". For challenge isolates two methods of organism suspension standardization were used in the evaluation of tigecycline with the Phoenix System. Suspensions were prepared using both the PhoenixSpec nephelometer (manual method) and the Phoenix AP instrument (automated method). A comparison of the performance of the two standardization methods is illustrated in Table 6 below. Table 6. Comparison of Challenge Isolate Inoculation Standardization Methods | | Tot | EA N | % EA | Total Eval | EA Eval N | %EA Eval | CA N | % CA | #NS | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Inoculum Method | | | | | | | | | | | | | | PhoenixSpec (Manual) | 44 | 44 | 100 | 44 | 44 | 100 | 44 | 100 | 0 | 0 | 0 | 0 | | Phoenix AP (Auto) | 44 | 44 | 100 | 44 | 44 | 100 | 44 | 100 | 0 | 0 | 0 | 0 | | | | | | | | | | | | | | | For the challenge organisms tested using suspensions prepared with either the manual (PhoenixSpec) method or using the Phoenix AP instrument, the overall % EA and % CA consistently met the acceptance criteria of greater than or equal to 90%. There were no very major, major or minor categorical errors with either inoculation method. b. Matrix comparison: Not applicable {10} 11 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: The MIC interpretive criteria are illustrated in Table 7 below. Table 7. MIC Interpretive Criteria - Tigecycline | Antibiotic Concentration Reviewed | Tigecycline - Susceptibility Interpretive Criteria (MIC in μg/mL) | | | | --- | --- | --- | --- | | Tigecycline 0.0313-4 μg/mL | | | | | Organism | FDA (SIR) | BD Phoenix | CLSI | | Enterococcus faecalis (vancomycin susceptible isolates) | ≤ 0.25*, -, - | ≤ 0.25*, -, - | Not Applicable | | Staphylococcus aureus (including methicillin resistant isolates) | ≤0.5*, -, - | ≤0.5*, -, - | Not Applicable | | Quality Control Organisms | Expected Range | | | | Staphylococcus aureus ATCC 29213 | 0.03-0.25 | ≤0.0313-0.25 | 0.03-0.25 | | Enterococcus faecalis ATCC 29212 | 0.03-0.12 | ≤0.0313-0.125 | 0.03-0.12 | *The current absence of resistant isolates precludes defining any results other than "Susceptible". Isolates yielding MIC result suggestive of "Non-susceptible" category should be submitted to a reference laboratory for further testing. {11} N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 12
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...