The Anyplex™ II HSV-1/2 Assay is a real-time polymerase chain reaction (PCR)-based in vitro diagnostic test intended for the qualitative detection and differentiation of Herpes Simplex Virus Type-1 (HSV-1) and Herpes Simplex Virus Type-2 (HSV-2) DNA from female skin lesions from anogenital sites. The test is intended for use as an aid in the diagnosis of anogenital HSV infection in symptomatic patients. Warning: Anyplex™ II HSV-1/2 Assay is not FDA cleared for use with cerebrospinal fluid (CSF). The assay is not intended to be used for prenatal screening.
Device Story
The Anyplex™ II HSV-1/2 Assay is an in vitro diagnostic test for qualitative detection and differentiation of HSV-1 and HSV-2 DNA. Input: female anogenital skin lesion specimens. Process: manual DNA extraction using QIAGEN QIAamp DNA Mini Kit; real-time PCR amplification on Cepheid SmartCycler® II Dx instrument; detection via fluorescently-labeled duplex Catcher probes. Output: qualitative detection of HSV-1 and HSV-2. Used in clinical laboratories by trained personnel. Results aid clinicians in diagnosing anogenital HSV infection in symptomatic patients. Includes positive, negative, and blank negative controls to ensure run validity and correct sample preparation.
Clinical Evidence
Clinical performance evaluated at three US sites using 656 prospective female anogenital specimens compared to ELVIS® HSV ID/Typing Test System. HSV-1 sensitivity: 98.9% (95% CI: 94.1-99.8%), specificity: 93.7% (95% CI: 91.0-95.6%). HSV-2 sensitivity: 97.2% (95% CI: 92.0-99.0%), specificity: 93.6% (95% CI: 91.3-95.4%). Discordant analysis performed via bidirectional sequencing.
Technological Characteristics
Real-time PCR-based assay. Uses fluorescently-labeled duplex Catcher probes for detection. Manual DNA extraction via QIAamp DNA Mini Kit. Instrumentation: Cepheid SmartCycler II Dx. Controls: HSV-1/2 plasmid positive control, internal control (IC) for extraction/process monitoring, and negative controls. Qualitative detection and differentiation of HSV-1 and HSV-2 DNA.
Indications for Use
Indicated for symptomatic female patients with anogenital skin lesions to qualitatively detect and differentiate HSV-1 and HSV-2 DNA. Not for use with CSF or prenatal screening.
Regulatory Classification
Identification
Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.
Special Controls
*Classification.* Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K142156
B. Purpose for Submission:
Clearance of New Device
C. Measurand:
Target DNA sequences from Herpes Simplex Virus type 1 (HSV-1) and Herpes Simplex Virus type 2 (HSV-2).
D. Type of Test:
An *in vitro* molecular diagnostic test for the qualitative detection and differentiation of HSV-1 and HSV-2 DNA from female skin lesions from anogenital sites.
E. Applicant:
Seegene
F. Proprietary and Established Names:
Anyplex™ HSV-1/2 Assay
G. Regulatory Information:
1. Regulation section: 21 CFR 866.3305
2. Classification: Class II
3. Product code: OQO
4. Panel: Microbiology (83)
H. Intended Use:
1. Intended use(s):
The Anyplex™ II HSV-1/2 Assay is a real-time polymerase chain reaction (PCR)-based *in vitro* diagnostic test intended for the qualitative detection and differentiation of Herpes Simplex Virus Type-1 (HSV-1) and Herpes Simplex Virus Type-2 (HSV-2) DNA from
{2}
female skin lesions from anogenital sites. The test is intended for use as an aid in the diagnosis of anogenital HSV infection in symptomatic patients.
Warning: Anyplex™ II HSV-1/2 Assay is not FDA cleared for use with cerebrospinal fluid (CSF). The assay is not intended to be used for prenatal screening.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Cepheid SmartCycler® II DX System (software version 3.0b).
I. Device Description:
The Anyplex™ II HSV-1/2 Assay uses PCR to generate amplified product from HSV-1 and HSV-2 DNA present in clinical specimens. The presence of HSV-1 and/or HSV-2 target DNA is indicated by the fluorescent signal generated through the use of fluorescently-labeled oligonucleotide probes (duplex Catcher) on the Cepheid SmartCycler® II Dx instrument. The probes do not generate a signal unless they are specifically bound to the amplified product. A preparation of HSV-1 and HSV-2 plasmids is included as the positive control in the Anyplex™ II HSV-1/2 Assay. The positive control serves to demonstrate that the HSV-1/2 PCR reagents are functional and to check the validity of the run. An internal control (IC) is also included in the assay kit. The IC is added to each sample specimen during sample preparation, and is also used to create a Blank Negative Control by adding a set amount to viral transport media to serve as an extraction control. In addition, RNase-free water is used to create the Negative Control by adding a set volume to the prepared master mix. Users are instructed to include all three controls, Positive, Negative, and Blank Negative Control with each test run.
The Anyplex™ II HSV-1/2 Assay detects and differentiates HSV-1 and HSV-2 genomic DNA in a single assay utilizing Seegene's TOCE™ technology. Clinical samples are spiked with an Internal Control and DNA is extracted utilizing a standard QIAGEN QIAamp® DNA Mini Kit. Extracted DNA is then mixed with the Seegene reagents and processed in a Cepheid SmartCycler® II Dx System. In approximately 2.5 hours, results can be available, including validity and acceptance criteria for the sample.
The one reagent kit for the Anyplex™ II HSV-1/2 Assay is sufficient for 50 reactions. The kit contains:
- 10X HSV-1/2 TOM - consisting of TOCE Oligo Mix (TOM), amplification and detection reagents
{3}
- 2X Anyplex™ II PCR Master Mix (with UDG) - consisting of DNA polymerase, Uracil-DNA glcosylase, and buffer containing dNTPs
- HSV-1/2 Positive Control (PC) - containing a mixture of Cloned DNA
- HSV-1/2 Internal Control (IC) - containing exogenous DNA
- RNase-free Water - Ultrapure PCR-grade quality
## J. Substantial Equivalence Information:
1. Predicate device name(s):
IMDx HSV-1/2 for Abbott m2000
Reference Method:
ELVIS® HSV ID/Typing Test System (Diagnostic Hybrids, Inc.) for clinical evaluation (K971662)
2. Predicate 510(k) number(s):
K140198
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Characteristic | Seegene Anyplex™ II HSV-1/2 Assay (New Device) | IMDx HSV-1/2 for Abbott m2000 (Predicate) |
| Intended use | The Anyplex™ II HSV-1/2 Assay is a real-time polymerase chain reaction (PCR)-based in vitro diagnostic test intended for the qualitative detection and differentiation of Herpes Simplex Virus Type-1 (HSV-1) and Herpes Simplex Virus Type-2 (HSV-2) DNA from female skin lesions from anogenital sites. The test is intended as an aid in the diagnosis of anogenital HSV infection in symptomatic patients.
Warning: The Anyplex™ II HSV-1/2 Assay is not indicated for use with cerebrospinal fluid (CSF). The assay is not intended to be used for pre-natal screening. | The IMDx HSV-1/2 for Abbott m2000 assay is an in vitro diagnostic test for the direct, qualitative detection and differentiation of Herpes Simplex Virus Type 1 (HSV-1) and Type 2 (HSV-2) DNA from male and female skin lesions from anogenital or oral sites. The test is intended for use as an aid in the diagnosis of HSV infection in symptomatic patients. The assay is intended to be run on the Abbott m2000 instrument system.
Warning: The IMDx HSV-1/2 for Abbott m2000 assay is not FDA cleared for use with cerebrospinal fluid (CSF). The assay is not intended for prenatal screening. |
{4}
| Similarities | | |
| --- | --- | --- |
| Characteristic | Seegene Anyplex™ II HSV-1/2 Assay (New Device) | IMDx HSV-1/2 for Abbott m2000 (Predicate) |
| Test Principle | Real-time PCR DNA amplification | Real-time PCR DNA amplification |
| Assay Results | Qualitative detection and differentiation of HSV-1 and HSV-2 DNA | Qualitative detection and differentiation of HSV-1 and HSV-2 DNA |
| Differences | | |
| Characteristic | Seegene Anyplex™ II HSV-1/2 Assay (New Device) | IMDx HSV-1/2 for Abbott m2000 (Predicate) |
| Instrumentation | Real-time PCR amplification/detection using the Cepheid SmartCycler II DX system. | Sample extraction and real-time PCR amplification/detection using the Abbott m2000 system. |
| Extraction Method | Manual extraction using the QIAGEN QIAamp® DNA Mini Kit | Automated on Abbott m2000 system |
| Sample Type | Female skin lesions from anogenital sites | Male and female skin lesions from anogenital or oral sites |
| Detection Method | Double-labeled (fluorophore and quencher) duplex Catcher. Measures increase in assay fluorescence with each PCR cycle. | Double-labeled (fluorophore and quencher) hydrolysis probes. Measures increase in assay fluorescence with each PCR cycle. |
# K. Standard/Guidance Document Referenced (if applicable): N/A
# L. Test Principle:
The Anyplex™ II HSV-1/2 Assay is based on real-time PCR technology combined with Seegene's TOCE™ (Tagging Oligonucleotide Cleavage and Extension) technology on the SmartCycler II Dx instrument system. A typical real-time PCR consists of denaturation of purified DNA, primer annealing and extension. The detection of target by the TOCE™ technology depends on two novel oligonucleotides, the Pitcher and the Catcher. When the PCR reaction initiates, Dual Priming Oligonucleotides (DPO™) specifically hybridize to the Glycoprotein D gene of HSV-1 and HSV-2 and the portion of the Pitcher at the 3'-end binds between the priming oligonucleotides. During PCR extension, the unbound 5' portion (Tag) of the Pitcher oligonucleotide is cleaved off by Taq DNA polymerase exonuclease. Subsequently, the released Tag anneals to a dual-labeled and single-stranded artificial template, called the Catcher. Extension of the Tag on the Catcher template results in a Duplex Catcher causing the physical distance of the fluorescent reporter from the quencher to increase, thereby generating signal. Thus, fluorescent signal is proportional to the concentration of Glycoprotein D gene of HSV-1 and HSV-2 in the original specimen, and is
{5}
displayed as the value of threshold cycle, Ct. The threshold cycle is the cycle number at which the fluorescence signal in the reaction crosses above the background of fluorescence.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Reproducibility:
Reproducibility studies were performed at three sites using three lots of reagent kits. Each site tested one lot of the reagent kit. The test panel of seven members was prepared randomized and blinded and tested for five (5) days non-consecutive days by two operators with each operator running the panel, in triplicate, once a day for a total of 90 results per panel member. The panel was formulated using intact virus diluted in M4 Viral Transport Media. Panel members were prepared with a single target present (HSV-1 or HSV-2) at three concentrations: ~3X LoD (High Positive), 1X LoD (Low Positive), and ~0.125X LoD (High Negative). A seventh panel member (HSV Negative) was prepared using M4 medium without HSV present. Reproducibility testing was carried out using panels that had been randomized and blinded prior to providing to the test site. The table below summarizes the results from each site as well as the combined results.
5
{6}
Reproducibility
| | | Site 1 | | Site 2 | | Site 3 | | All 3 sites Combined | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Panel Member | Concentration | % Agreement Agreement with expected result | Avg. Ct (%CV) | % Agreement Agreement with expected result | Avg. Ct (%CV) | % Agreement Agreement with expected result | Avg. Ct (%CV) | % Agreement Agreement with expected result | Avg. Ct (%CV) |
| HSV-1 High-Positive* | 3X LoD | 100.0%
30/30 | 36.4
(2.0%) | 100.0%
30/30 | 35.9
(2.6%) | 100.0%
30/30 | 36.1
(2.9%) | 100.0%
90/90 | 36.1
(2.6%) |
| HSV-1 Low-Positive* | 1X LoD | 100.0%
30/30 | 39.9
(2.2%) | 90.0%
27/30 | 39.8
(3.3%) | 100.0%
30/30 | 39.2
(1.7%) | 96.7%
87/90 | 39.6
(2.6%) |
| HSV-1 High-Negative** | <1X LoD | 6.7%
2/30 | 42.9
(3.2%) | 43.3%
13/30 | 42.6
(3.9%) | 3.3%
1/30 | 42.4
(2.4%) | 17.8%
16/90 | 42.6
(3.1%) |
| HSV-2 High-Positive* | 3X LoD | 100.0%
30/30 | 36.5
(1.6%) | 100.0%
30/30 | 35.7
(2.2%) | 100.0%
30/30 | 36.4
(1.7%) | 100.0%
90/90 | 36.2
(2.0%) |
| HSV-2 Low-Positive* | 1X LoD | 100.0%
30/30 | 39.8
(2.3%) | 93.3%
28/30 | 39.6
(2.4%) | 100.0
30/30 | 39.9
(2.4%) | 97.8%
88/90 | 39.8
(2.3%) |
| HSV-2 High-Negative** | <1X LoD | 30.0%
9/30 | 42.3
(2.8%) | 46.7%
14/30 | 42.1
(4.0%) | 16.7%
5/30 | 42.2
(2.5%) | 31.1%
28/90 | 42.2
(3.0%) |
| HSV Negative** | N/A | 100.0%
60/60 | 0.0 | 98.3%
59/60 | 42.7 | 100.0%
60/60 | 0.0 | 99.4%
179/180 | 42.7 |
*Expected result is positive. %CV calculated from results with non-zero Ct values.
** Expected result is negative. %CV calculated from results with non-zero Ct values.
b. Precision:
The precision studies were conducted in-house using a five membered panel tested twice a day by a single operator using one lot of reagents for twenty days. Panel members were tested in triplicate for each run (for a total of 600 data points for the 40 runs). The panel was prepared with a single target present (HSV-1 or HSV-2) at two concentrations:
HSV-1 Low Positive (1X LoD), HSV-1 High Positive (10X LoD), HSV-2 Low Positive (1X LoD), HSV-2 High Positive (10X LoD), and HSV-1/2 Negative. The HSV-1/2 Negative member was prepared using M4 medium without HSV present. The table below shows the Percent (%) Agreement for each panel member tested for the precision study.
{7}
Precision Study Average Ct Values
| Panel Member | Level | % Agreement with expected results | 95% Confidence Interval | Avg. Ct | SD Ct | %CV Ct |
| --- | --- | --- | --- | --- | --- | --- |
| HSV-1 Low Positive | 1X LoD | 100.00% (120/120) | 96.90% - 100.00% | 42.93 | 0.75 | 1.74% |
| HSV-1 High Positive | 10X LoD | 100.00% (120/120) | 96.90% - 100.00% | 40.32 | 0.71 | 1.76% |
| HSV-2 Low Positive | 1X LoD | 100.00% (120/120) | 96.90% - 100.00% | 41.31 | 1.06 | 2.56% |
| HSV-2 High Positive | 10X LoD | 100.00% (120/120) | 96.90% - 100.00% | 38.19 | 0.51 | 1.34% |
| Negative | N/A | 100.00% (120/120) | 96.90% - 100.00% | N/A | N/A | N/A |
c. Linearity/assay reportable range: N/A
d. Traceability, Stability, Expected values (controls, calibrators, or methods):
## Assay Controls
The Anyplex™ II HSV-1/2 Assay Positive Control, Negative Control, and Blank Negative Control are included in each run to evaluate run validity.
- The Anyplex™ II HSV-1/2 Positive Control is constructed from plasmids with gDNA extracts from HSV-1 HF (VR-260) and HSV-2 G (VR-734). The Positive Control serves to demonstrate that the HSV-1/2 PCR reagents are functional, and to check the validity of the run.
- The Anyplex™ II HSV-1/2 Negative Control is constructed of RNase-free water added to the Master Mix prior to PCR. The presence of HSV-1 and/or HSV-2 must not be detected in the Negative Control.
- The Anyplex™ II HSV-1/2 Blank Negative Control is comprised of Internal Control added to blank VTM. The Blank Negative Control is carried through sample preparation alongside the clinical specimens.
e. Detection limit:
The purpose of this study was to evaluate the Limit of Detection (LoD) of the Anyplex™ II HSV-1/2 Assay. The study was performed with two representative strains of HSV-1 (MacIntyre and HF) and two representative strains of HSV-2 (MS and G). A preliminary study was performed for each HSV-1 and HSV-2 strain. A series of seven (7) 10-fold dilutions of virus was made in pooled negative matrix, then tested with the Anyplex™ II HSV-1/2 Assay in replicates of three (3). Results from the preliminary LoD estimation study were used to determine the range for the
{8}
LoD analysis. The last dilution with a $100\%$ detection rate was used to prepare a series of five (5) 2-fold dilutions using negative matrix as a diluent. These five (5) dilutions were tested on three (3) Anyplex™ II HSV-1/2 Assay kit lots in replicates of 20. Results from the three (3) kit lots were combined and the detection rate was calculated for each dilution. From this study, the LoD was determined to be the lowest concentration of target that was detected in at least $95\%$ of the replicates.
The LoD of the Anyplex™ II HSV-1/2 Assay was determined to be $3.75 \times 10^{2}$ $\mathrm{TCID}_{50} / \mathrm{mL}$ for HSV-1 (MacIntyre), $1.875 \times 10^{2}$ $\mathrm{TCID}_{50} / \mathrm{mL}$ for HSV-1 (HF), and $3.75 \times 10^{1}$ $\mathrm{TCID}_{50} / \mathrm{mL}$ for both MS and G strains of HSV-2.
# Limit of Detection
HSV1-MacIntyre
| TCID50/mL | Positive/total | Positivity rate (%) | 95% CI | |
| --- | --- | --- | --- | --- |
| 1.5 X 103 | 60/60 | 100.00 | 93.98 | 100.00 |
| 7.5 X 102 | 60/60 | 100.00 | 93.98 | 100.00 |
| 3.75 X 102 | 57/60 | 95.00 | 86.30 | 98.29 |
| 1.875 X 102 | 49/60 | 81.67 | 70.08 | 89.44 |
| 9.375 X 101 | 17/60 | 28.33 | 18.51 | 40.77 |
HSV1-HF
| TCID50/mL | Positive/total | Positivity rate (%) | 95% CI | |
| --- | --- | --- | --- | --- |
| 1.5 X 103 | 60/60 | 100.00 | 93.98 | 100.00 |
| 7.5 X 102 | 60/60 | 100.00 | 93.98 | 100.00 |
| 3.75 X 102 | 60/60 | 100.00 | 93.98 | 100.00 |
| 1.875 X 102 | 57/60 | 95.00 | 86.30 | 98.29 |
| 9.375 X 101 | 46/60 | 76.66 | 64.56 | 85.56 |
HSV2-MS
| TCID50/mL | Positive/total | Positivity rate (%) | 95% CI | |
| --- | --- | --- | --- | --- |
| 1.5 X 102 | 60/60 | 100.00 | 93.98 | 100.00 |
| 7.5 X 101 | 60/60 | 100.00 | 93.98 | 100.00 |
| 3.75 X 101 | 59/60 | 98.33 | 91.15 | 99.71 |
| 1.875 X 101 | 54/60 | 90.00 | 79.85 | 95.34 |
| 9.375 X 100 | 40/60 | 66.67 | 54.06 | 77.27 |
{9}
9
# HSV2-G
| TCID_{50}/mL | Positive /total | Positivity rate (%) | 95% CI | |
| --- | --- | --- | --- | --- |
| V 1.5 X 10^{2} | 60/60 | 100.00 | 93.98 | 100.00 |
| 2 7.5 X 10^{1} | 60/60 | 100.00 | 93.98 | 100.00 |
| 3.75 X 10^{1} | 59/60 | 98.33 | 91.15 | 99.71 |
| 1.875 X 10^{1} | 40/60 | 66.67 | 54.06 | 77.27 |
| 9.375 X 10^{0} | 33/60 | 55.00 | 42.49 | 66.91 |
Analytical Reactivity/LoD Confirmation: In addition, an LoD confirmation study using titered stocks of cultured clinical isolates obtained from a supplier was conducted in house to confirm the observed LoD values for HSV-1 and HSV-2. A total of 42 clinical isolates (21 HSV-1 and 21 HSV-2) were diluted to the observed LoD in the negative matrix pool. Each HSV-1 and HSV-2 clinical isolate was extracted and tested in 3 replicates using the three different lots of kit reagents (9 replicates for each isolate). All strains were detected by the assay, demonstrating that the Anyplex™ II HSV-1/2 Assay can detect a broad range of HSV-1 and HSV-2 isolates.
## f. Analytical specificity:
A study was performed to evaluate cross reactivity of the Anyplex™ II HSV-1/2 Assay with fifty (50) microorganisms that might be found in anogenital lesion specimens. The panel members were obtained from suppliers as quantified cultures (CF), or prepared in house (IHC) by growing each organism and quantifying the culture. Bacteria were tested at $10^{6}$ cfu/mL or higher and viruses were tested at $10^{5}$ pfu/mL or higher. For bacteria that were difficult to obtain or grow, purified DNA was used in the place of the intact microorganism and tested at a concentration of $\geq 1\times 10^{6}$ genome copies/mL. Human DNA was tested at a concentration of $1\times 10^{5}$ genome copies/mL. All samples were prepared by diluting microorganisms or DNA into M4RT viral transport medium prior to testing for cross-reactivity. No microorganisms tested positive for HSV-1 or HSV-2 using the Anyplex™ II HSV-1/2 Assay indicating no cross reactivity with these microorganisms.
To assess microbial interference, each test microorganism from the cross reactivity panel was added to a sample tube with two representative strains of HSV-1 (MacIntyre and HF), and HSV-2 (MS and G) at 3X LoD in M4RT viral transport medium. Each potentially interfering microorganism was tested in three (3) replicates at the same levels used in the cross-reactivity study described above. No evidence of microbial interference was observed for any of the 50 test microorganisms included in the analysis.
{10}
Cross Reactivity & Microbial Interference Panel
| NO. | Organism | Test Concentration | Origin* |
| --- | --- | --- | --- |
| 1 | Atopobium vaginae | 1 X 10^{7} CFU/mL | IHC |
| 2 | Bacteroides fragilis | 1 X 10^{9} CFU/mL | CF |
| 3 | Candida albicans | 1 X 10^{7} CFU/mL | CF |
| 4 | Candida glabrata Z007 | 1 X 10^{8} CFU/mL | CF |
| 5 | Candida guilliemondii Z008 | 1 X 10^{7} CFU/mL | CF |
| 6 | Candida krusei Z009 | 1 X 10^{6} CFU/mL | CF |
| 7 | Candida lusitaniae Z010 | 1 X 10^{7} CFU/mL | CF |
| 8 | Candida parapsilosis Z011 | 1 X 10^{7} CFU/mL | CF |
| 9 | Candida tropicalis Z012 | 1 X 10^{6} CFU/mL | CF |
| 10 | Chlamydia trachomatis (D-UW3) | 1 X 10^{8} IFU/mL | CF |
| 11 | Chlamydia trachomatis (serotype E) | 1 X 10^{6} IFU/mL | OPZ |
| 12 | Chlamydia trachomatis (serotype F) | 1 X 10^{6} IFU/mL | OPZ |
| 13 | Chlamydia trachomatis (serotype G) | 1 X 10^{4} IFU/mL | OPZ |
| 14 | Chlamydia trachomatis (serotype H) | 1 X 10^{6} IFU/mL | OPZ |
| 15 | Chlamydia trachomatis (serotype I) | 1 X 10^{7} IFU/mL | OPZ |
| 16 | Chlamydia trachomatis (serotype J) | 5 X 10^{3} IFU/mL | OPZ |
| 17 | Chlamydia trachomatis (serotype K) | 1 X 10^{7} IFU/mL | OPZ |
| 18 | Cytomegalovirus (AD169) | 1 X 10^{6} IU/mL | CF |
| 19 | Enterococcus casseliflavus | 5 X 10^{8} CFU/mL | IHC |
| 20 | Enterococcus faecalis | 1 X 10^{9} CFU/mL | IHC |
| 21 | Enterococcus faecium | 1 X 10^{9} CFU/mL | IHC |
| 22 | Enterococcus gallinarum | 1 X 10^{9} CFU/mL | IHC |
| 23 | Enterovirus (Type 71) | 1 X 10^{6} TCID_{50}/mL | CF |
| 24 | Epstein-Barr virus (B95-8 strain) | 5 X 10^{8} copies/mL | CF |
| 25 | Escherichia coli | 1 X 10^{7} CFU/mL | IHC |
| 26 | Gardnerella vaginalis | 1 X 10^{7} CFU/mL | IHC |
| 27 | Human Herpes 6B virus (Z29 strain) | 1 X 10^{5} TCID_{50}/mL | CF |
| 28 | Human Herpes 7 virus (SB strain) | 5 X 10^{6} TCID_{50}/mL | CF |
* In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of * In the case of *
{11}
| 29 | Human Papilloma virus-16 (Caski) | 1,000 cell/mL | CF |
| --- | --- | --- | --- |
| 30 | Human Papilloma virus-18 (Hela) | 5,000 cell/mL | CF |
| 31 | Klebsiella pneumoniae Z026 | 1 X 107 CFU/mL | CF |
| 32 | Lactobacillus acidophilus | 1 X 108 CFU/mL | CF |
| 33 | Lactobacillus crispatus | 1 X 108 CFU/mL | IHC |
| 34 | Lactobacillus gasseri | 1 X 108 CFU/mL | IHC |
| 35 | Lactobacillus jensenii | 1 X 106 CFU/mL | IHC |
| 36 | Moraxella catarrhalis Ne 11 | 1 X 108 CFU/mL | CF |
| 37 | Mycoplasma hominis | 5.27 X 103CCU/mL(1.00 X 106copies/mL) | OPZ |
| 38 | Neisseria gonorrhoeae | 1 X 107 CFU/mL | CF |
| 39 | Rubella virus | 1 X 105TCID50/mL | CF |
| 40 | Serratia marcescens | 1 X 107 CFU/mL | IHC |
| 41 | Staphylococcus saprophyticus | 1 X 107 CFU/mL | IHC |
| 42 | Streptococcus mitis | 1 X 109 CFU/mL | CF |
| 43 | Streptococcus mutans Z072 | 1 X 109 CFU/mL | CF |
| 44 | Streptococcus oralis | 1 X 107 CFU/mL | IHC |
| 45 | Streptococcus pneumoniae 19F | 1 X 108 CFU/mL | CF |
| 46 | Streptococcus pyogenes Rosenbach | 1 X 107 CFU/mL | IHC |
| 47 | Toxoplasma gondii | 1 X 106tachyzoites/mL | CF |
| 48 | Trichomonas vaginalis | 5 X 105parasites/mL | OPZ |
| 49 | Ureaplasma urealyticum | 5 X 107 CCU/mL | OPZ |
| 50 | Varicella Zoster virus | 1 X 108 Copies/mL | CF |
* CF – Culture Fluid; OPZ – Organism Propagation from ZeptoMetrix; IHC – In-House Culture
{12}
# g. Interfering Studies
This study was performed to evaluate potential interference with the Anyplex™ II HSV-1/2 Assay with a panel of twenty-two (22) biological and chemical substances. All of the potentially interfering substances were tested at concentrations at or above physiological levels or typical usage levels with HSV-1 (MacIntyre) and HSV-2 (MS) at 3X LoD to evaluate potential interference with the detection of the HSV-1 and HSV-2 targets. The study was also carried out in the absence of HSV to evaluate potential interference with the detection of the internal control of the Anyplex™ II HSV-1/2 Assay. Each potentially interfering substance was tested in triplicate. No interference was observed with any of the substances tested.
Interfering Substance Panel
| Substance | Potential Inhibitor | Concentration |
| --- | --- | --- |
| Whole blood with EDTA | Heme, DNA, proteases, nucleases | 10% v/v |
| Female Urine | Non-specific PCR inhibitors | 10% v/v |
| Male Urine | Non-specific PCR inhibitors | 10% v/v |
| Acyclovir Cream 50mg/g | Acycloguanosine | 7% w/v |
| Albumin | Albumin | 10 mg/mL |
| Casein | Casein | 1 mg/mL |
| K-Y® Brand Jelly | Glycerin, Cellulose | 7% w/v |
| Douche | Decyl Glucoside; Octoxynol-9 | 7% w/v |
| Monistat® 1 | Miconazole nitrate cream 2% | 7% w/v |
| Monistat® 3 | Miconazole nitrate cream 4% | 7% w/v |
| Vagisil Cream | Benzocaine, Resorcinol | 7% w/v |
| Triconazole 1 | Tioconazole 6.5% | 7% w/v |
| Balneol® Hygienic Cleansing Lotion | Mineral oil, Fatty acids | 7% w/v |
| CVS Anti-Itch Cream | Benzocaine; Benzalkonium Chloride | 7% w/v |
| Listerine® Antiseptic Mouth Wash | Ethanol, Menthol, Thymol, Eucalyptol | 7% w/v |
| Abreva® | Docosanol 10% | 7% w/v |
| Carmex® Cold Sore Lip Balm | Menthol 0.7%, Camphor 1.7%, Phenol 0.4% | 7% w/v |
| Releev® cold sore treatment | Benzalkonium Chloride 0.13% | 7% w/v |
{13}
| Substance | Potential Inhibitor | Concentration |
| --- | --- | --- |
| Lip clear Lysine+® | Zinc Oxide 1.2% | 7% w/v |
| Toothpaste | Surfactants, fluorides, antibacterials | 7% w/v |
| Contraceptive Jelly | Non-oxynol-9 | 7% w/v |
| Vaginal Contraceptive Foam | Non-oxynol-9 | 7% w/v |
h. Specimen Stability in Viral Transport Media
The performance of the Anyplex™ II HSV-1/2 Assay was evaluated with respect to the specimen stability in viral transport media such as Remel M4, Remel M4RT, Remel M5, Remel M6, Copan Universal Transport Media (UTM), and BD Universal Viral Transport (UVT). Each transport medium was spiked with HSV-1 or HSV-2 positive clinical specimens at 3X LoD and stored and tested at the following temperatures and intervals: Refrigerated (2°C to 8°C)/Day 0, 3, 5, and 7; Frozen (-20°C)/Day 0, 15, 30, and 35. All samples were run in triplicate for all time points. Testing consisted of removing specimens from storage temperature, thawing (if frozen), extracting the DNA, and then running the Anyplex™ II HSV-1/2 Assay.
There was no interference observed with any of the viral transport media listed above for the detection of HSV-1 and HSV-2 target or the internal control. The study data supported the stability claims for HSV-1 and HSV-2 specimens collected in Remel M4, Remel M4RT, Remel M5, Remel M6, Copan Universal Transport Media (UTM), and BD Universal Viral Transport (UVT) refrigerated (2°C to 8°C) for 7 days and frozen (-20°C) for 30 days.
i. Competitive Inhibition
Competitive inhibition of the Anyplex™ II HSV-1/2 Assay was evaluated to assess the potential for interference in HSV-1/2 target detection when both HSV-1 and HSV-2 are present in a sample. The study was conducted by using simulated samples with equal or varying concentrations of HSV-1 and HSV-2 (3X, 10X, 100X and 1000X LoD). The results showed that competitive inhibition was not observed when the concentration of HSV1 and HSV2 strains are equal or varying. There was no influence of competitive inhibition between HSV-1 and HSV-2 on the performance of the Anyplex™ II HSV-1/2 Assay.
h. Carry-over/Cross Contamination
The analytical carry-over/cross contamination study was performed to assess the potential for carry-over contamination of HSV target from a high-positive sample into a negative sample during the Anyplex™ II HSV-1/2 Assay run. The carry-over/cross
{14}
contamination study was performed using HSV-1 MacIntyre and HSV-2 MS high-positive (~100,000X LoD) samples in M4 medium. Alternating high positive samples and negative samples (M4 medium) were extracted, placed in sequential wells of the SmartCycler® II Dx and run. The run was repeated five (5) times.
No carry-over (0/100) or cross contamination (0/100) events were observed between positive and negative samples.
j. Assay cut-off: Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The clinical performance evaluation was performed against a gold standard/reference method i.e., Cell Culture using an enzyme linked virus inducible system with HSV typing by fluorescently labeled antibodies.
b. Matrix comparison: N/A
3. Clinical studies:
a. Clinical Sensitivity: N/A
b. Clinical Specificity: N/A
c. Other clinical supportive data (when a. and b. are not applicable):
The performance of the Anyplex™ II HSV-1/2 Assay was compared with the ELVIS® HSV ID/Typing Test System (Diagnostic Hybrids, Inc.) which is a gold standard/reference method, i.e., Cell Culture using an enzyme linked virus inducible system with HSV typing by fluorescently labeled antibodies.
## Clinical Performance
The performance of the Anyplex™ II HSV-1/2 Assay was evaluated at three geographically diverse locations within the United States. A total of 656 prospective specimens were tested by the Anyplex™ II HSV-1/2 Assay and compared to results obtained with the ELVIS® (Enzyme Linked Virus Inducible System) HSV ID and D³ Typing Test System (Diagnostic Hybrids, Athens, OH). The reference ELVIS viral culture method used in this study is unable to detect co-infected specimens and cannot identify HSV-1 if HSV-2 is identified first. Consequently, if a specimen was positive for HSV-2, it was removed from the calculation of the HSV-1 clinical performance. Quality Controls (positive, negative, blank negative, and internal control) were run with each Anyplex™ II HSV-1/2 Assay run. The internal control was introduced into each specimen and the blank negative control prior extraction.
Results from the prospective studies are presented in the tables below.
14
{15}
HSV-1 Results for Female Anogenital Specimens
| HSV-1 | Reference Method | | | |
| --- | --- | --- | --- | --- |
| | | POS | NEG | Total |
| Seegene | POS | 91 | 29^{a} | 120 |
| | NEG | 1^{b} | 429 | 430 |
| | Total | 92 | 458 | 550 |
| Sensitivity; 95% CI | | 98.9% (91/92); 95% CI [94.1%-99.8%] | | |
| Specificity; 95% CI | | 93.7% (429/458); 95% CI [91.0%-95.6%] | | |
a Discordant analysis (bidirectional sequencing) was performed on all 29 samples identified as HSV-1 positive by the Anyplex™ II HSV-1/2 Assay. HSV-1 was detected in 18 of the 29 samples. The remaining 11 specimens remained discordant (HSV-1 was not detected).
b Discordant analysis (bidirectional sequencing) was performed on the one sample identified as negative by the Anyplex™ II HSV-1/2 Assay. HSV-1 was detected in this sample.
HSV-2 Results for Female Anogenital Specimens
| HSV-2 | ELVIS HSV ID and D^{3} Typing | | | |
| --- | --- | --- | --- | --- |
| | | POS | NEG | Total |
| IMDx | POS | 103 | 35^{c} | 138 |
| | NEG | 3^{d} | 515 | 518 |
| | Total | 106 | 550 | 656 |
| Sensitivity; 95% CI | | 97.2% (103/106); 95% CI [92.0%-99.0%] | | |
| Specificity; 95% CI | | 93.6% (515/550); 95% CI [91.3%-95.4%] | | |
e Discordant analysis (bidirectional sequencing) was performed on all 35 discordant specimens identified as HSV-2 positive by the Anyplex™ II HSV-1/2 Assay. HSV-2 was detected in 21 of the 35 specimens. The remaining 14 remained discordant (HSV-2 was not detected).
d Discordant analysis (bidirectional sequencing) was conducted for the three specimens identified as HSV-2 negative by the Anyplex™ II HSV-1/2 Assay. HSV-2 was not detected in any specimen.
4. Clinical cut-off: N/A
5. Expected values/Reference range:
Prevalence: The prevalence of HSV-1 and HSV-2 observed during the multi-center clinical trial was calculated for the Anyplex™ II HSV-1/2 Assay. The prevalence rate for HSV-1 was established as 21.8% (120/550) for anogenital samples. The prevalence rate for HSV-2 was established as 21.0% (138/656) for anogenital samples.
{16}
Age Distribution for Female Anogenital Specimens
| Age (Years) | HSV-1 Positive | HSV-2 Positive |
| --- | --- | --- |
| <1 to 17 | 5/34
(14.7%) | 2/35
(5.7%) |
| 18 to 30 | 67/237
(28.3%) | 65/287
(22.6%) |
| 31 to 40 | 21/110
(19.1%) | 25/131
(19.1%) |
| 41 to 50 | 10/73
(13.7%) | 18/88
(20.4%) |
| 51 to 60 | 8/55
(14.5%) | 15/68
(22.1%) |
| 61 to 70 | 9/29
(31.0%) | 12/35
(34.3%) |
| 71 to 80 | 0/9
(0.0%) | 1/9
(11.1%) |
| 81 to 90 | 0/3
(0.0%) | 0/3
(0.0%) |
| 91 to 95 | 0/0
(0.0%) | 0/0
(0.0%) |
| Total | 120/550
(21.8%) | 138/656
(21.0%) |
Positive and Negative Predictive Value: Hypothetical positive and negative predictive values (PPV & NPV) for the Anyplex™ II HSV-1/2 Assay are shown below. These calculations are based on hypothetical prevalence and overall sensitivity and specificity per specimen type as determined in the clinical trial.
For HSV-1, these calculations are based upon an overall sensitivity and specificity of 98.9% and 93.7%, respectively, for anogenital swabs.
For HSV-2, these calculations are based upon an overall sensitivity and specificity of 97.2% and 93.6%, respectively, for anogenital swabs.
PPV was calculated using: (Sensitivity x Prevalence)/(Sensitivity x Prevalence + [1 - Specificity] x [1 - Prevalence]).
NPV was calculated using: (Specificity x [1 - Prevalence])/( [1 - Sensitivity] x Prevalence + Specificity x [1 - Prevalence]).
{17}
Prevalence vs. Hypothetical Predictive Values
| Prevalence (%) | Female Anogenital Swabs | | | |
| --- | --- | --- | --- | --- |
| | HSV-1 | | HSV-2 | |
| | PPV (%) | NPV (%) | PPV (%) | NPV (%) |
| 2 | 23.6 | 100.0 | 25.2 | 100.0 |
| 5 | 44.3 | 99.9 | 46.5 | 99.9 |
| 10 | 62.6 | 99.8 | 64.8 | 99.7 |
| 20 | 79.1 | 99.5 | 80.5 | 99.4 |
| 30 | 86.6 | 99.1 | 87.6 | 98.9 |
| 40 | 91.0 | 98.7 | 91.7 | 98.3 |
| 50 | 93.8 | 98.0 | 94.3 | 97.5 |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.