SMR 40MM GLENOSPHERE

K142139 · Lima Corporate S.P.A. · PHX · Nov 7, 2014 · Orthopedic

Device Facts

Record IDK142139
Device NameSMR 40MM GLENOSPHERE
ApplicantLima Corporate S.P.A.
Product CodePHX · Orthopedic
Decision DateNov 7, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3660
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SMR Shoulder System is intended for partial or total, primary or revision shoulder joint replacement. The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to: - non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; - inflammatory degenerative joint disease such as rheumatoid arthritis; - treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation methods; - revision of a failed primary implant; - cuff tear arthropathy (CTA Heads only). The Large Resection Stems (not available in the US) are indicated for oncology applications. The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The Modular SMR Shoulder System allows the assembly of components in various humeral and glenoid constructs. The constructs are intended for cemented and uncemented use as specified in the following table. In the Anatomic shoulder the humeral construct consists of the humeral body, the adaptor taper and the humeral head. In the Reverse shoulder the humeral construct consists of the humeral stem, the reverse humeral body and the reverse liner. On the humeral side the fixation of the humeral stem determines if the construct is cemented or uncemented. The Anatomic glenoid construct consists of an all polyethylene glenoid or a metal back assembled with a liner while the Reverse glenoid construct consists of the metal back, the connector and the glenosphere. On the glenoid side, the fixation of the all polyethylene glenoid or the metal back determines if the construct is cemented or uncemented.

Device Story

SMR 40mm Glenosphere is a modular shoulder prosthesis component; part of SMR Reverse Shoulder System. Device consists of CoCrMo alloy glenosphere (40mm diameter) and UHMWPE liner. Glenosphere attaches to metal-back glenoid via double male taper and screw. System replaces damaged shoulder joint surfaces; restores joint function in rotator cuff deficient patients. Used in orthopedic surgery; implanted by surgeons. Provides articulation surface for humeral components. Benefits include pain relief and restored mobility in severe arthropathy cases.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, specifically fatigue resistance testing of the modular connection to the metal-back glenoid component using worst-case constructs.

Technological Characteristics

Materials: CoCrMo (ASTM F1537, ISO 5832-12), UHMWPE (ASTM F648, ISO 5834-2), Ti6Al4V (ASTM F1472, ISO 5832-3). Modular design: 40mm glenosphere, reverse liners (standard/retentive), metal-back glenoid, connector, bone screws. Fixation: Cemented or uncemented (press-fit).

Indications for Use

Indicated for patients requiring partial or total, primary or revision shoulder joint replacement due to non-inflammatory/inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, rheumatoid arthritis), acute humeral head fractures, failed primary implants, or cuff tear arthropathy. Reverse system indicated for grossly rotator cuff deficient joints with severe arthropathy; requires anatomically/structurally suited joint and functional deltoid muscle.

Regulatory Classification

Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,” (vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 7, 2014 Limacorporate S.p.A % Dr. Stephen J. Peoples, VMD, MS Principal Consultant Peoples & Associates Consulting, LLC 5010 Lodge Pole Lane Fort Wayne, Indiana 46814 Re: K142139 Trade/Device Name: SMR 40mm Glenosphere Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS Dated: October 21, 2014 Received: October 24, 2014 Dear Dr. Peoples: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ Page 2 - Dr. Stephen J. Peoples, VMD, MS the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number K142139 Device Name: SMR 40mm Glenosphere Indications for Use: ## SMR 40mm Glenosphere Indications for Use The SMR Shoulder System is intended for partial or total, primary or revision shoulder joint replacement. The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to: - non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; ● - inflammatory degenerative joint disease such as rheumatoid arthritis; ● - treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation ● methods: - . revision of a failed primary implant; - cuff tear arthropathy (CTA Heads only). ● The Large Resection Stems (not available in the US) are indicated for oncology applications. The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The Modular SMR Shoulder System allows the assembly of components in various humeral and glenoid constructs. The constructs are intended for cemented and uncemented use as specified in the following table. In the Anatomic shoulder the humeral construct consists of the humeral body, the adaptor taper and the humeral head. In the Reverse shoulder the humeral construct consists of the humeral stem, the reverse humeral body and the reverse liner. On the humeral side the fixation of the humeral stem determines if the construct is cemented or uncemented. The Anatomic glenoid construct consists of an all polyethylene glenoid or a metal back assembled with a liner while the Reverse glenoid construct consists of the metal back, the connector and the glenosphere. On the glenoid side, the fixation of the all polyethylene glenoid or the metal back determines if the construct is cemented or uncemented. {3}------------------------------------------------ | System | | Components | Material | Use | | Available<br>in US | Available<br>in Canada | | | |--------|--------------------|------------------------------------------------------------------------------------------------------------------------|---------------------------|-----|------------|--------------------|------------------------|--|--| | A | R | | | Cem | Not<br>Cem | | | | | | ● | ● | SMR Stems (Cemented, Cemented Revision) | Ti6Al4V | X | | ● | ● | | | | ● | ● | SMR Stems (Cementless Finned, Cementless Revision) | Ti6Al4V | | X | ● | ● | | | | ● | ● | SMR Large Resection stems | Ti6Al4V | X | | | ● | | | | ● | ● | SMR Modular Augments | Ti6Al4V | X | | | ● | | | | ● | | SMR Humeral Bodies (Trauma, Finned) | Ti6Al4V | X | X | ● | ● | | | | | ● | SMR Reverse Humeral Body | Ti6Al4V | X | X | ● | ● | | | | | ● | SMR Reverse HA Coated Humeral Body | Ti6Al4V+HA | | X | | ● | | | | ● | ● | Humeral Extension | Ti6Al4V | X | X | ● | ● | | | | | | SMR Humeral Heads (Standard, CTA) | CoCrMo | X | X | ● | ● | | | | ● | | | Ti6Al4V | X | X | | ● | | | | ● | | SMR Adaptor Tapers (Neutral, Eccentric) | Ti6Al4V | X | X | ● | ● | | | | ● | | SMR CTA Head Adaptor for Reverse Humeral Body | Ti6Al4V | X | X | ● | ● | | | | | | | CoCrMo | | X | ● | ● | | | | | | SMR Glenospheres | Ti6Al4V | | X | | ● | | | | | ● | | UHMWPE X-Lima<br>+Ti6Al4V | | X | | | | | | | ● | SMR Connectors | Ti6Al4V | | X | ● | ● | | | | | | Reverse Liners | UHMWPE | X | X | ● | ● | | | | | ● | | UHMWPE X-Lima | X | X | | | | | | | | | CoCrMo | X | X | | | | | | | | | Alumina | X | X | | | | | | ● | | SMR Cemented Glenoids | UHMWPE | X | | ● | ● | | | | ● | | SMR 3 Pegs Cemented Glenoids | UHMWPE | X | | ● | | | | | | ● | SMR Metal Back Glenoids | Ti6Al4V+PoroTi | X* | X* | ● | ● | | | | ● | | | Ti6Al4V+PoroTi+HA | | X | | ● | | | | | ● | SMR TT Metal Back Baseplate | Ti6Al4V | X* | X* | | | | | | ● | ● | SMR TT Metal Back Peg | Ti6Al4V | X | X | | | | | | ● | | SMR Metal Back Liner | UHMWPE | X* | X* | | | | | | ● | | SMR Metal Back Liner | UHMWPE X-Lima | | X | ● | ● | | | | • * | ● | SMR Bone screws | Ti6Al4V | | X | ● | ● | | | | | ● | SMR Glenoid Plates | Ti | | X | | | | | | | Material Standards | | | | | | | | | | | | Ti6A4V (ISO 5832-3 - ASTM F1472) - CoCrMo (ISO 5832-12 - ASTM F1537) - Ti (ASTM F67) - UHMWPE (ISO 5834-2 - ASTM F648) | | | | | | | | Alumina (ISO 6474) – PoroTi Titanium Coating (ASTM F1580) – HA Hyroxyapatite Coating (ISO 13779) #### A= Anatomic / R=Reverse #### *NOTE : - In the US, the SMR Metal Backed Glenoid/Liner construct, used as part of the SMR Anatomic ● Shoulder Replacement, is intended for use with bone cement and should be used without bone screws. - Outside the US, the SMR Metal Backed Glenoid/Liner construct, used as part of the SMR Anatomic Shoulder Replacement, is intended for uncemented use with the addition of screws for fixation. - The SMR Metal Backed Glenoid/Connector/Glenosphere construct, used as part of the SMR Reverse Shoulder replacement, is intended for uncemented use with the addition of screws for fixation. - The SMR TT Metal Back is indicated for primary implant in anatomic or reverse total shoulder arthroplasty with poor glenoid bone quality or in revision case with consistent bone loss on the {4}------------------------------------------------ glenoid side. The implant is suitable for implantation with or without bone graft. These indications are not yet approved in the US. In the US the SMR TT Metal Back Baseplate used as part of the SMR Anatomic Shoulder Replacement, is intended for use with bone cement and should be used without bone screws; while when used as part of the SMR Reverse Shoulder replacement, is intended for uncemented use with the addition of screws for fixation. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {5}------------------------------------------------ # 510(k) Summary Date: July 23, 2014 Manufacturer: Limacorporate S.p.A. Via Nazionale, 52 33038 – Villanova di San Daniele Udine - Italy U.S. Contact Person: Dr. Stephen J. Peoples Principal Consultant Phone: 260-645-0327 FAX: +39 0432945512 | Product | Common<br>Name | Product<br>Code | Regulation and Classification Name | |-------------------------|------------------------|-----------------|----------------------------------------------------------------------------------------------| | SMR 40mm<br>Glenosphere | Shoulder<br>Prosthesis | KWS | Shoulder joint metal/polymer semi-<br>constrained cemented prosthesis per 21<br>CFR 888.3660 | | | | PHX | Shoulder joint metal/polymer semi-<br>constrained cemented prosthesis per 21<br>CFR 888.3660 | ### Description: The SMR 40mm Glenosphere is a Co-Cr-Mo glenosphere intended to be used with the 40mm standard UHMWPE reverse liner (standard or retentive) as part of the SMR Reverse Shoulder System (K110598). The SMR Reverse Shoulder System consists of a humeral stem, a reverse humeral body, a reverse liner, a metal-back glenoid, a glenosphere and a connector with screw. Bone screws are used for the fixation of the metal-back glenoid to the scapula. Humeral stems are provided for both cemented and cementless (K100858, K101263, K111212, K113523) fixation as well as reverse humeral bodies (K110598). The SMR Reverse Shoulder System metal back glenoids (K110598. K113254 and K133349) and glenospheres (K110598) are intended for uncemented pressfit use only with the addition of screws for fixation. SMR 40mm Glenosphere is made of CoCrMo alloy that conforms to ASTM F1537 - ISO 5832-12 in two (2) different designs: a standard glenosphere that can be centered with respect to the glenoid component or eccentrical glenosphere that provides offset. Both designs have a spherical shape of 40mm diameter and articulate with the standard 40mm ultra-high molecular weight polyethylene (ASTM F648 - ISO 5834-2) liners that are coupled to the humeral body. Two designs for the 40mm liners are available: standard and retentive. The 40mm glenospheres are coupled to a SMR metal-back glenoid (K113254, K133349) through a double male taper, one side of which is connected to the glenosphere while the {6}------------------------------------------------ other is coupled with the glenoid component. To increase the solidity of the system, a screw is used to link the glenosphere to the glenoid component. The screw is inserted through a hole on the surface of the glenosphere, passing through the internal cavity of the connector and then screwed to the metal-back. The SMR 40mm Liners are made from standard ultra-high molecular weight polyethylene conforming to ASTM F648 – ISO 5834-2. They are coupled to the SMR Reverse Humeral Bodies (K110598) via a taper. The liners are designed with a bevel cut in their inferior aspect to reduce the possibility of accidental contact between the polyethylene liner and the scapular bone during adduction - abduction movements. The 40mm liners are available in standard and retentive designs in three (3) thicknesses (STD, +3mm and +6mm) with the retentive liners characterized by a deeper spherical concavity than is present in the standard liners. ### Intended Use / Indications for Use: The SMR Shoulder System is intended for partial or total, primary or revision shoulder joint replacement. The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to: - . non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; - inflammatory degenerative joint disease such as rheumatoid arthritis; . - treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation ● methods: - . revision of a failed primary implant; - . cuff tear arthropathy (CTA Heads only). The Large Resection Stems (not available in the US) are indicated for oncology applications. The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The Modular SMR Shoulder System allows the assembly of components in various humeral and glenoid constructs. The constructs are intended for cemented and uncemented use as specified in the following table. In the Anatomic shoulder the humeral construct consists of the humeral stem, the humeral body, the adaptor taper and the humeral head. In the Reverse shoulder the humeral construct consists of the humeral stem, the reverse humeral body and the reverse liner. On the humeral side the fixation of the humeral stem determines if the construct is cemented or uncemented. {7}------------------------------------------------ The Anatomic glenoid construct consists of an all polyethylene glenoid or a metal back assembled with a liner while the Reverse glenoid construct consists of the metal back, the connector and the glenosphere. On the glenoid side, the fixation of the all polyethylene glenoid or the metal back determines if the construct is cemented or uncemented. | | | System | Components | Material | Use | | Available<br>in US | Available<br>in Canada | |----|---|--------|----------------------------------------------------|---------------------------|-----|------------|--------------------|------------------------| | A | R | | | | Cem | Not<br>Cem | | | | ● | ● | | SMR Stems (Cemented, Cemented Revision) | Ti6Al4V | X | | ● | ● | | ● | ● | | SMR Stems (Cementless Finned, Cementless Revision) | Ti6Al4V | | X | ● | ● | | ● | ● | | SMR Large Resection stems | Ti6Al4V | X | | | ● | | ● | ● | | SMR Modular Augments | Ti6Al4V | X | | | ● | | ● | ● | | SMR Humeral Bodies (Trauma, Finned) | Ti6Al4V | X | X | ● | ● | | ● | ● | | SMR Reverse Humeral Body | Ti6Al4V | X | X | ● | ● | | ● | ● | | SMR Reverse HA Coated Humeral Body | Ti6Al4V+HA | | X | | ● | | ● | ● | | Humeral Extension | Ti6Al4V | X | X | ● | ● | | ● | ● | | SMR Humeral Heads (Standard, CTA) | CoCrMo | X | X | ● | ● | | ● | ● | | SMR Adaptor Tapers (Neutral, Eccentric) | Ti6Al4V | X | X | ● | ● | | ● | ● | | SMR CTA Head Adaptor for Reverse Humeral Body | Ti6Al4V | X | X | ● | ● | | ● | | | SMR Glenospheres | CoCrMo | | X | ● | ● | | ● | ● | | | Ti6Al4V | | X | ● | ● | | ● | | | | UHMWPE X-Lima<br>+Ti6Al4V | | X | ● | ● | | ● | ● | | SMR Connectors | Ti6Al4V | | X | ● | ● | | ● | | | Reverse Liners | UHMWPE | X | X | ● | ● | | ● | ● | | | UHMWPE X-Lima | X | X | ● | ● | | ● | | | | CoCrMo | X | X | ● | ● | | ● | | | | Alumina | X | X | ● | ● | | ● | | | SMR Cemented Glenoids | UHMWPE | X | | ● | ● | | ● | | | SMR 3 Pegs Cemented Glenoids | UHMWPE X-Lima | X | | ● | ● | | ● | | | | UHMWPE | X | | ● | ● | | ● | ● | | SMR Metal Back Glenoids | Ti6Al4V+PoroTi | X* | X* | ● | ● | | ● | | | | Ti6Al4V+PoroTi+HA | | X | ● | ● | | ● | ● | | SMR TT Metal Back Baseplate | Ti6Al4V | X* | X* | ● | ● | | ● | ● | | SMR TT Metal Back Peg | Ti6Al4V | X | X | ● | ● | | ● | | | SMR Metal Back Liner | UHMWPE | X* | X* | ● | ● | | ● | | | SMR Metal Back Liner | UHMWPE X-Lima | | X | ● | ● | | ●* | ● | | SMR Bone screws | Ti6Al4V | | X | ● | ● | | ● | ● | | SMR Glenoid Plates | Ti | | X | ● | ● | #### A= Anatomic / R=Reverse *NOTE : - . In the US, the SMR Metal Backed Glenoid/Liner construct, used as part of the SMR Anatomic Shoulder Replacement, is intended for use with bone cement and should be used without bone screws. - . Outside the US, the SMR Metal Backed Glenoid/Liner construct, used as part of the SMR Anatomic Shoulder Replacement, is intended for uncemented use with the addition of screws for fixation. {8}------------------------------------------------ - . The SMR Metal Backed Glenoid/Connector/Glenosphere construct, used as part of the SMR Reverse Shoulder replacement, is intended for uncemented use with the addition of screws for fixation. - The SMR TT Metal Back is indicated for primary implant in anatomic or reverse total ● shoulder arthroplasty with poor glenoid bone quality or in revision case with consistent bone loss on the glenoid side. The implant is suitable for implantation with or without bone graft. These indications are not yet approved in the US. - . In the US the SMR TT Metal Back Baseplate used as part of the SMR Anatomic Shoulder Replacement, is intended for use with bone cement and should be used without bone screws: while when used as part of the SMR Reverse Shoulder replacement, is intended for uncemented use with the addition of screws for fixation. ### Predicate Devices: - SMR Reverse Shoulder System (Limacorporate, K110598); - Promos Reverse Shoulder System (Plus Orthopedics, K081016); ● - Inverse/reverse Shoulder (Zimmer, K053274); ● - Delta Xtend Reverse Shoulder (DePuy, K091751). ## Comparable Features to Predicate Device(s): The SMR 40mm Glenosphere and Liner are similar to the predicate devices in terms of intended use, indications, design and materials. The SMR 40mm Glenosphere and the predicates are all intended for total primary or revision reverse shoulder joint replacement. All systems are provided with reverse liners and glenospheres of similar diameters that are intended to articulate together. In all the systems, the glenospheres are fixed to the glenoid bone through the use of a glenoid component and connector. The components of the SMR 40mm Glenosphere and Liner are manufactured from the same or similar materials as the predicate devices. ### Non-Clinical Testing: The SMR 40mm Glenosphere was tested for fatigue resistance of the modular connection to the metal-back glenoid component. Mechanical testing was performed on worst case components or constructs. The testing results demonstrated the device's ability to perform under expected clinical conditions. ## Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence of the SMR 40mm Glenosphere to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...